Title: bmd Crossvent 3
1bmd Crossvent 3
2Modes Touch Screen
- Adult to pediatric range
- Assist/Control Volume Control
- Assist/Control Pressure Control
- Sighs
- SIMV
- Pressure Support
- CPAP
3Turn knobs for setting
- Maximum pressure
- PEEP
- Flow
4LCD Touch ScreenSelect Mode then press MAIN menu
- Then press Rate TV
- adjust with up/down arrows
- IT, ET IE Ratio displayed
- Flow, PEEP Max Press limit
- adjust with dials
5On the left side
External Power Supply
O2 Sensor
ON/OFF
Alarm Reset
LED Charging Light
Pt Circuit
Pneumotach/Flow Sensor Input
Airway Pressure Connector
Exhalation Valve Connector
6bmd CrossVent 3
- powered by a rechargeable NiMH battery
- lasts 6 hours
- BioMed 16vDC 3A charger/converter recharges
internal battery and allows for AC operation - LCD screen has LCD charging indicator and
visually large LOW BATTERY alarm
7Flow Sensor at proximal airway
8Flow Sensor
9On the right side
Gas Entrainment ON/OFF
Bleed Exhaust
Gas Supply Inlet (31 - 75 psig or 45 - 75
when using blender)
10Gas Entrainment
- When connected to 100 oxygen source gas with the
Gas Entrainment ON, the oxygen concentration is
diluted to 50. - There will be some decrease in delivered flow
rate with Entrainment ON under high load
conditions
11The rear panel
- Maximum pressure relief valve (120 cm H2O)
- Negative pressure relief valve opens at - 4 cm
H2O with machine shutdown
12SIMV settings
13Press arrows to increase or decrease Rate or
Tidal Volume
14Touch screen to highlight TV Adjust with
up/down arrows
15CPAP Mode
16ALARMS
- ALARM 1
- Peak Pressure
- Respiratory Rate
- Exhaled Tidal Volume (can be switched off)
- Exhaled Minute Volume (can be switched off)
- ALARM 2
- PEEP
- Mean Pressure (can be switched off)
- Oxygen (can be switched off)
17To set Alarms Press ALARM 1
- Press High or Low Limits to Change
- Adjust with up/down arrows
18Repeat for ALARM 2 settings
193rd Level Alarms
Low Gas Supply Pressure Low Battery Ventilator
Failure
20With Alarm Condition
the value for the exceeded limit flashes red
-1
21When power off, press and hold ALARM RESET