Title: Medical Image Analysis
1Medical Image Analysis
- Medical Imaging Modalities
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
2- Anatomical or structural
- X-ray radiology, X-ray mammography, X-ray CT,
ultrasound, Magnetic Resonance Imaging - Functional or metabolic
- Functional MRI, (Single Photon Emission Computed
Tomography) SPECT, (Positron Emission Tomography)
PET, fluorescence imaging
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
3X-ray Imaging
Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
4Figure 4.1. Atomic structure of a tungsten atom.
An incident electron with energy greater than
K-shell binding energy is shown interacting with
a K-shell electron for the emission of an X-ray
photon.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
5X-ray Imaging
- Tungsten
- K-shell binding energy level 69.5 keV
- L-shell binding energy level 10.2 keV
- An emision of X-ray photon of 59.3 keV
- X-ray generation
- Electrons are released by the source cathode and
are accelerated toward the target anode in a
vacuum under the potential difference ranging
from 20,000 to 150,000 volts
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
6Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
7X-ray 2-D Projection Imaging
- Diagnostic radiology
- 2-D projection of the three-dimensional
anatomical structure of the human body - Localized sum of attenuation coefficients of
material air, blood, tissue, bone - Film or 2-D array of detectors
- Digital radiographic system
- Use scintillation crystals optically coupled with
photomultiplier
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
8Figure 4.2. (a). A schematic diagram of a 2-D
X-ray film-screen radiography system. A 2-D
projection image of the 3-D object is shown at
the bottom. (b). X-ray radiographic image of a
normal male chest.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
9Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
10X-ray 2-D Projection Imaging
- Scattering
- Create artifacts and artificial structures
- Reduce scattering
- Anti-scattered grids and collimators
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
11X-ray Mammography
- Target material
- Molybdenum K-, L-, M-shell binding energies
levels are 20, 2.8, 0.5 keV. The characteristic
X-ray radiation is around 17 keV. - Phodium K-, L-, M-shell binding energies levels
are 23, 3.4, 0.6 keV. The characteristic X-ray
radiation is around 20 keV. - A small focal spot of the order of 0.1mm
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
12Figure 4.3. A film-screen X-ray mammography
imaging system.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
13Figure 4.4. X-ray film-screen mammography image
of a normal breast.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
14X-ray Computed Tomography
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
15Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
16Figure 4.5. 3-D object representation as a stack
of 2-D x-y slices.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
17Figure 4.6. Source-Detector pair based
translation method to scan a selected 2-D slice
of a 3-D object to give a projection along the
y-direction.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
18Figure 4.7 The translate-rotate parallel-beam
geometry of first generation CT scanners.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
19X-ray Computed Tomography
- Generations
- First an X-ray source-detector pair that was
translated in parallel-beam geometry - Second a fan-beam geometry with a divergent
X-ray source and a linear array of detectors.
Use translation to cover the object and rotation
to obtain additional views
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
20- Generations
- Third a fan-beam geometry with a divergent
X-ray source and an arc of detectors. Without
translation. Additional views are obtained by
simultaneous rotation of the X-ray source and
detector assembly. Rotate only - Fourth use a detector ring around the object.
The X-ray source provides a divergent fan-beam of
radiation to cover the object
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
21Figure 4.8. The first generation X-ray CT scanner
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
22Figure 4.9. The fourth generation X-ray CT
scanner geometry.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
23Figure 4.10. X-ray CT image of a selected slice
of cardiac cavity of a cadaver.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
24Figure 4.11. The pathological image of the
selected slice shown with the X-ray CT image in
Figure 4.10
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
25Magnetic Resonance Imaging
- Nuclear magnetic resonance
- The selected nuclei of the matter of the object
- Blood flow and oxygenation
- Different parameters weighted,
weighted, Spin-density - Advance MR Spectroscopy and Functional MRI
- Fast signal acquisition of the order of a
fraction of a second
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
26Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
27Figure 4.12. MR images of a selected
cross-section that are obtained simultaneously
using a specific imaging technique. The images
show (from left to right), respectively, the
T1-weighted, T-2 weighted and the Spin-Density
property of the hydrogen protons present in the
brain.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
28Magnetic Resonance Imaging
- 1H high sensitivity and vast occurrence in
organic compounds - 13C the key component of all organic
- 15N a key component of proteins and DNA
- 19F high relative sensitivity
- 31P frequent occurrence in organic compounds
and moderate relative sensitivity
Adapted from the Wikipedia, www.wikipedia.org.
29MR Spectroscopy
Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
30MR Spectroscopy
Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
31Functional MRI
Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
32MRI Principles
- spin-lattice relaxation time
- spin-spin relaxation time
- the spin density
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
33MRI Principles
- Great web sites
- Simulations from BIGS - Lernhilfe für Physik und
Technik - http//www.cis.rit.edu/class/schp730/bmri/bmri.htm
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
34MRI Principles
- Spin
- A fundamental property of nuclei with odd atomic
numbers is the possession of angular moment - Magnetic moment
- The charged protons create a magnetic field
around them and thus act like tiny magnets
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
35MRI Principles
- the spin angular moment
- the magnetic moment
- a gyromagnetic ratio, MHz/T
- A hydrogen atom
- 42.58 MHz/T
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
36Figure 4.13. Left A tiny magnet representation
of a charged proton with angular moment, J.
Right A symbolic representation of a charged
proton with angular moment, J and a magnetic
moment, µ.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
37MRI Principles
- Precession of a spinning proton
- The interaction between the magnetic moment of
nuclei with the external magnetic field - Spin quantum number of a spinning proton ½
- The energy level of nuclei aligning themselves
along the external magnetic field is lower than
the energy level of nuclei aligned against the
external magnetic field
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
38Figure 4.14 (a) A symbolic representation of a
proton with precession that is experienced by the
spinning proton when it is subjected to an
external magnetic field. (b) The random
orientation of protons in matter with the net
zero vector in both longitudinal and transverse
directions.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
39MRI Principles
- Equation of motion for isolated spin
- Solution
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
40Longitudinal Vector OX at the transverse
position X
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
41Figure 4.15 (a). Nuclei aligned under thermal
equilibrium in the presence of an external
magnetic field. (b). A non-zero net longitudinal
vector and a zero transverse vector provided by
the nuclei precessing in the presence of an
external magnetic field.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
42Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
43MRI Principles
- The precession frequency
- Depends on the type of nuclei with a specific
gyromagnetic ratio and the intensity of the
external magnetic field - This is the frequency on which the nuclei can
receive the Radio Frequency (RF) energy to change
their states for exhibiting nuclear magnetic
resonance - The excited nuclei return to the thermal
equilibrium through a process of relaxation
emitting energy at the same precession frequency
44MRI Principles
- 90-degree pulse
- Upon receiving the energy at the Larmor
frequency, the transverse vector also changes as
nuclei start to precess in phase - Form a net non-zero transverse vector that
rotates in the x-y plane perpendicular to the
direction of the external magnetic field
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
45Figure 4.16. The 90-degree pulse causing nuclei
to precess in phase with the longitudinal vector
shifted clockwise by 90-degrees as a result of
the absorption of RF energy at the Larmor
frequency.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
46MRI Principles
- 180-degree pulse
- If enough energy is supplied, the longitudinal
vector can be completely flipped over with a
180-degree clockwise shidf in the direction
against the external magnetic field
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
47Figure 4.17. The 180-degree pulse causing nuclei
to precess in phase with the longitudinal vector
shifted clockwise by 180-degrees as a result of
the absorption of RF energy at the Larmor
frequency.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
48MRI Principles
- Relaxation
- The energy emitted during the relaxation process
induces an electrical signal in a RF coil tuned
at the Larmor frequency - The free induction decay of the electromagnetic
signal in the PF coil is the basic signal that is
used to create MR images - The nuclear excitation forces the net
longitudinal and transverse magnetization vectors
to move
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
49MRI Principles
- A stationary magnetization vector
- The total response of the spin system
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
50Figure 4.18. The transverse relaxation process of
spinning nuclei.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
51MRI Principles
- The longitudinal and transverse magnetization
vectors with respect to the relaxation times - where
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
52 Figure 4.19. (a) Transverse and (b) longitudinal
magnetization relaxation after the RF pulse.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
53MRI Principles
- The RF pulse causes nuclear excitation changing
the longitudinal and transverse magnetization
vectors - After the RF pulse is turned off, the excited
nuclei go through the relaxation phase emitting
the absorbed energy at the same Larmor frequency
that can be detected as an electrical signal,
called the Free Induction Decay (FID)
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
54MRI Principles
- The NMR spin-echo signal (FID signal)
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
55MR Instrumentation
- The stationary external magnetic field
- Provided by a large superconducting magnet with a
typical strength of 0.5 T to 1.5 T - Housing of gradient coils
- Good field homogeneity, typically on the order of
10-50 parts per million - A set of shim coils to compensate for the field
inhomogeneity
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
56Figure 4.20. A general schematic diagram of a MR
imaging system.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
57Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
58Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
59MR Instrumentation
- An RF coil
- To transmit time-varying RF pulses
- To receive the radio frequency emissions during
the nuclear relaxation phase - Free Induction Decay (FID) in the RF coil
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
60MR Pulse Sequences
- NMR signal
- The frequency and the phase
- Spatial encoding in MR imaging
- Frequency encoding and phase encoding
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
61Figure 4.21 (a). Three-dimensional object
coordinate system with axial, sagittal and
coronal image views. (b) From top left to
bottom right Axial, coronal and sagittal MR
images of a human brain.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
62Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
63MR Pulse Sequences
Figure 4.22. (a) Three-dimensional spatial
encoding for spin-echo MR pulse sequence. (b) A
linear gradient field for frequency encoding.
(c). A step function based gradient field for
phase encoding.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
64Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
65MR Pulse Sequences
- The phase-encoding gradient
- Applied in steps with repeated cycles
- If 256 steps are to be applied in the
phase-encoding gradient, the readout cycle is
repeated 256 times, each time with a specific
amount of phase-encoding gradient
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
66Spin Echo Imaging
-
- Between the application of the 90 degree pulse
and the formation of echo (rephasing of nuclei -
- Between the 90 degree pulse and 180 degree pulse
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
67Figure 4.23. The transverse relaxation and echo
formation of the spin echo MR pulse sequence.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
68Spin Echo Imaging
- K-space
- The placement of raw frequency data collected
through the pulse sequences in a
multi-dimensional space - By taking the inverse Fourier transform of the
k-space data, an image about the object can be
reconstructed in the spatial domain - The NMR signals collected as frequency-encoded
echoes can be placed as horizontal lines in the
corresponding 2-D k-space
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
69Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
70Spin Echo Imaging
- the cycle repetition time
- weighted
- A long and a long
- weighted
- A short and a short
- Spin-density
- A long and a short
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
71Figure 4.24. A spin echo pulse sequence for MR
imaging.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
72Spin Echo Imaging
- The effective transverse relaxation time from the
field inhomogeneities
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
73Spin Echo Imaging
- The effective transverse relaxation time from a
spatial encoding gradient
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
74Echo Planar Imaging
- A single-shot fast-scanning method
- Spiral Echo Planar Imaging (SEPI)
- where
75Figure 4.25. A single shot EPI pulse sequence.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
76Figure 4.26. The k-space representation of the
EPI scan trajectory.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
77Figure 4.27. The spiral scan trajectory of SEPI
pulse sequence in the k-space.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
78Figure 4.28. The SEPI pulse sequence
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
79Figure 4.29. MR images of a human brain acquired
through SEPI pulse sequence.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
80Gradient Echo Imaging
- Fast low angle shot (FLASH) imaging
- Utilize low-flip angle RF pulses to create
multiple echoes in repeated cycles to collect the
data required for image reconstruction - A low-flip angle (as low as 20 degrees)
- The readout gradient is inverted to re-phase
nuclei leading to the gradient echo during the
data acquisition - The entire pulse sequence time is much shorter
than the spin echo pulse sequence
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
81Figure 4.30. The FLASH pulse sequence for fast MR
imaging.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
82Flow Imaging
- Tracking flow
- Diffusion (incoherent flow) and perfusion
(partially coherent flow) - The FID signal generated in the RF receiver coil
by the moving nuclei and velocity-dependent
factors - MR angiography
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
83Figure 4.31. A flow imaging pulse sequence with
spin echo.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
84Figure 4.32 Left A proton density image of a
human brain. Right The corresponding perfusion
image.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
85Figure 4.33. Gradient echo based MR pulse
sequence for 3-D MR volume angiography.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
86Figure 4.34. An MR angiography image.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
87Nuclear Medicine Imaging Modalities
- Radioactivity decay
- Half-life of a radionuclide decay
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
88Nuclear Medicine Imaging Modalities
- The radioactivity of a radionuclide
- The average decay rate
- Curie (CI)
- disintegrations per second
(dps) - Becquerel (Bq)
- One dps
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
89Single Photon Emission Computed Tomography
- Radioisotope
- The radioisotopes are injected in the body
through administration of radiopharmaceutical
drugs that metabolize with the tissue - Gamma rays
- The gamma rays from the tissue pass through the
body and are captured by the detectors
surrounding the body to acquire raw data for
defining projections
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
90Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
91Single Photon Emission Computed Tomography
- Radionuclides
- Thallium
- Technetium
- Iodine
- Gallium
- Gamma ray
- Decay by emitting gamma rays with photon energy
ranging from 135 keV to 511 keV - Attenuation
92Figure 4.35. A schematic diagram of detector
arrays of SPECT scanner surrounding the patient
area.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
93Single Photon Emission Computed Tomography
- Scintillation detector
- Barium fluoride
- Cesium iodide
- Bismuth germinate BGO
- Photomultiplier tube
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
94Figure 4.36. A 99Tc SPECT image of a human brain
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
95Single Photon Emission Computed Tomography
- Attenuation and scattering
- Photoelectric absorption and Compton scattering
- Poor in structural information
- Attenuation and scattering
- Assessment of metastases or characterization of a
tumor - Lower cost than PET
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
96Positron Emission Tomography
- Concept
- Simultaneous detection of two 511keV energy
photons traveling in the opposite direction - Radionuclides
- Decay by emitting positive charged particles
called positrons - Fluorine 18-F
- Oxygen 15-O
- Nitrogen 13-N
- Carbon 11-C
97Figure 4.37. A schemtaic diaggram of PET scanner.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
98Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
99Positron Emission Tomography
- After emission
- Travel typically for 1-3 mm, losing some of its
kinetic energy - The annihilation of the positron with the
electron - Cause the formation of two gamma photons with
511keV traveling in opposite directions - Coincidence detection
- The point of emission of a positron is different
from the point of annihilation with an electron
100Positron Emission Tomography
- Radiopharmaceutical
- Fluorodeoxyglucose (FDG)
- Resolution and sensitivity of PET imaging is
significantly better than SPECT
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
101Figure 4.38 Serial images of a human brain with
FDG PET imaging.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
102Ultrasound Imaging
- Diagnostic imaging
- Anatomical structures, blood flow measurements
and tissue characterization - Safety, portability, low-cost
Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
103Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
104Ultrasound Imaging
- Velocity
- Relative intensity in dB
- Shorter waves
- Better imaging resolution
- Frequencies 2 MHz to 5 MHz are common
105Reflection and Transmission
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
106Figure 4.39. A path of a reflected sound wave in
a multilayered structure.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
107Refraction
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
108Figure 4.40. A schematic diagram of a
conventional ultrasound imaging system.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
109Figure comes from the Wikipedia,
www.wikipedia.org.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
110Ultrasound Imaging
- A-mode
- Records the amplitude of returning echoes from
the tissue boundaries with respect to time - Perpendicular incident angle
- Basic method
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
111Ultrasound Imaging
- M-mode
- Variations in signal amplitude due to object
motion - X-axis represents the time, while the y-axis
indicates the distance of the echo from the
transducer
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
112Figure 4.41. M-Mode display of mitral valve
leaflet of a beating heart.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
113Ultrasound Imaging
- B-mode
- Two-dimensional images representing the changes
in acoustic impedance of the tissue
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
114Figure 4.42. The B-Mode image of a beating
heart with mitral stenosis.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
115Ultrasound Imaging
- Doppler ultrasound imaging
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.
116Figure 4.43. A Doppler image of the mitral valve
area of a beating heart. Figures 4.4.3-5 are
taken from the website http//www2.umdnj.edu/shin
dler/ms.html.
Figures come from the textbook Medical Image
Analysis, by Atam P. Dhawan, IEEE Press, 2003.