Title: Measurement of cardiac output
1Measurement of cardiac output
University College of Medical Sciences GTB
Hospital, Delhi
2Methods used for measurement of Cardiac Output
- Invasive
- PA catheter
- - Ficks cardiac output measurement
- - Thermodilution Technique
- - Mixed venous oximetry pulmonary catheter
- Minimally invasive
- Doppler Ultrasound
- Lithium dilution cardiac output monitoring
- Pulse contour cardiac output monitoring
- Transpulmonary thermodilution
- Non-invasive
- Bio impedence cardiac output monitoring
- Partial carbon dioxide re breathing cardiac
output monitoring
3INVASIVE METHODS OF CARDIAC OUTPUT MEASUREMENT
(PAC)
4- PULMONARY ARTERY CATHETERISATION
- First used by Swan, Ganz for hemodymamic
monitoring of patients - PAC can be placed from any central venous
cannulation sites, but right internal jugular
vein is most commonly used. - Standard PAC is 7.0 to 9.0 Fr in circumference,
110 cm long, has 4 internal lumens
5(No Transcript)
6PHYSIOLOGICAL CONCEPTS OF CARDIAC OUTPUT
MEASUREMENT
7- INDICATOR DILUTION TECHNIQUE
- Tracer substance is injected into the bloodstream
concentration change measured at a
downstream site - Indocyanin green most commonly used dye
- Stewart Hamilton Equation
- I
- Q
- ? CI dt
- Where
- Q CO
- I Amount of indicator
- ? CI dt Integral of indicator concentration
over time - Drawbacks of indicator dilution method
- Limited to cardiac catheterization laboratories
- Continuous withdrawal of arterial blood to plot
the dye concentration curve - Dye needs regular injections (can accumulate)
8- Other guidelines for placement waveforms
were already discussed
9- Complication of PA catheterisation-
- Infection, endocarditis
- Thrombo embolism
- Endocardial damage, valve injury
- PA infarction
- PA rupture
- Catheter knotting
- Ventricular fibrillation, arrhythmia, RBBB
10Bolus - Thermodilution Cardiac Output Monitoring
- Variant of indicator dilution technique
- Iced indicator/ room temperature indicator
(bolus)- 10 ml or 0.15ml/kg in children - Advantages
- Performed quickly, repeatedly
- Does not require advanced diagnostic or technical
skills - Uses non-toxic, non-accumulative indicator
11Stewart Hamilton equation is modified
- (TB TI) x K
- Q
- ? ? TB (t) dt
- Where
- Q CO
- TB Blood temp.
- TI Injectate temp.
- K Computational constant
- ? ? TB (t) dt Integral of temp. change over
time
12- Method
- Volume of ice cold or room temperature fluid is
injected as bolus - ?
- Change in pulmonary artery blood temperature is
recorded - Source of error
- Intra or extra-cardiac shunt
- Tricuspid or pulmonary valve regurgitation
- Inadequate delivery of indicator
- Thermister malfunction
- Unrecognised blood temperature fluctuation
- Respiratory cycle influence
13(No Transcript)
14Continuous - Thermodilution CO monitoring
- Warm or cold thermal indicator
- Methods
- Release of small quantity of heat from a 10 cm
thermal filament incorporated into right
ventricular portion of a PAC approx. 15-20 cm
from catheter tip - ?
- Heating filament is cycled on off
- ?
- Thermal signal measured
- ?
- CO derived from cross-correlation of measured
pulmonary artery temp. - ?
- Displayed value of CO is updated every 30-60 sec
represents the average value for cardiac output
measured over 3-6 min
15- Advantages
- External system for cold fluid injection is not
required - Fewer measurement error
- Less risk of fluid overload and infection
- Measures average CO value - derived over several
mins - Beat-to-beat variation in SV that occur during
single respiratory cycle are equally represented - In contrast bolus technique measures cardiac
output values depending on phase of respiration
16MINIMALLY INVASIVE METHODS OF CARDIAC OUTPUT
MEASUREMENT
17Doppler Ultrasound
- Doppler principle
- When USG waves strike moving objects, these waves
are reflected back to their source at a different
frequency, termed the Doppler shift frequency
that is directly related to the velocity of
moving object and the angle at which the USG beam
strikes these objects - Red blood cells serve as moving object target
182
- Where
- f Doppler shift frequency
- v Velocity of red blood cell targets
- f0 Transmitted USG beam frequency
- 0 Angle b/w the USG beam and the vector of
RBC flow - C Velocity of USG in blood (approx. 1570
m/sec) - Cosine 0 1 as long as angle of insonation
is small
19SV v x ET x CSA
- Where
- SV Stroke volume
- v Spatial average velocity of blood
flow (cm/sec) - ET Systolic ejection time
- CSA Cross-sectional area of vessel
- -Estimated CSA close to the mean value during
systole obtained from a nomogram stored in the
computer - -Measured CSA using an M mode echo transducer
incorporated in the probe
20- Types of probe-Suprasternal(Ascending aorta)
-
Esophageal(Descending aorta) - Suprasternal probe position instability limited
their use for extended period of time - Esophageal probes have 2 advantage over the
suprasternal probe - Smooth muscle tone of the oesophagus maintains
the probe position - Its in close proximity to the aorta thereby
minimizing signal interference
21(No Transcript)
22The shape of the waveform allows Assessment of
the venticular preload, afterload and
contractility
23PULSE CONTOUR CARDIAC OUTPUT MONITORING
- Cardiac output is determined through analysis of
arterial pressure wave form obtained from an
arterial catheter or from a non invasive finger
blood pressure waveform - CO is measured on a beat to beat basis
- Wesseling and colleagues devised an algorithm
for the calculation of SV from aortic impedence
and changes in arterial pressure during systole - SV ? dP/dt
- Z
24- Advantage-
- It has the potential for continuous, beat to beat
monitoring of cardiac output - Disadvantage-
- Baseline calibration with known cardiac output is
required - Recalibration is required every 8 to 12 hrs
Require calibration to compensate for the
algorithms inability to independently assess the
ever changing effects of vascular tone - A well defined arterial pressure waveform is
needed
25- Pulse contour cardiac output estimation without
external calibration(Flo Trac) - Doesnt require external calibration
- The algorithm works on the principle that SV is
directly proportional to pulse pressure and
inversely proportional to aortic compliance - The aortic pressure is sampled at 100Hz analysed
and updated every 20 sec - SV K(SdAP)
- The standard deviation-SdAP is proportional to
the pulse pressure, which is proportional to SV.
K is the constant derived from patient
characteristics as described by Langewouler and
co workers
26NON INVASIVE CARDIAC OUTPUT MEASUREMENT
27BIOIMPEDENCE CARDIAC OUTPUT MONITORING
- Developed by Kubiceck and NASA researchers in
1960s - Based on changes in electrical resistance of the
thoracic cavity occurring with change in aortic
blood volume during systole diastole - 4Pairs -Each pair of electrode consists of a
transmitting and a sensing electrode - Two pairs are applied to the base of the neck on
opposite sides, two pairs are applied to the
lateral aspect of the thorax at the level of the
xiphoid process on opposite sides
28(No Transcript)
29- The electrodes mark the upper and lower
boundaries of the thorax - An alternating current of low amplitude and high
frequency is applied which is sensed by
electrodes placed over the neck lateral aspect
of the chest. - Volume of thorax is calculated according to the
height, weight and gender
30- Advantage-
- Non invasive, continuous monitoring
- Measures thoracic fluid content, left ventricular
ejection time, cardiac index - Disadvantage-
- Susceptibility to electrical interference
- Relies on correct placement of the electrodes
31(No Transcript)
32References
- Lailu M, Kalyan RK. Cardiac output monitoring.
Annals of Cardiac monitoring.2008 1156-61 - Jhanji S, Dawson J and Pearse R M. Cardiac output
monitoringbasic science and clinical
application. Anaesthesia .2008 172-78 - Rebecca A, Schroeder, Atilio B, Shahar B and
Jonathan B. Cardiovascular Monitoring. Millers
Anaesthesia 7th edition 1314-21 - William F Ganong, Review of medical physiology
22nd edition 819 - Kaplan JA. Hemodynamic monitoring. Kaplans
Cardiac Anaesthesia 5th edition 283-86 - Edward Morgan.Patient monitors. Clinical
Anaesthesiology 4th edition 137-139
33