Title: CCDD Rules and Regulations Training Seminar
1CCDD Rules and RegulationsTraining Seminar
- PID/FID Equipment - Use and Calibration
2PID/FID EquipmentUse and Calibration
- Equipment
- Pat Maloney
- J M Instruments
- Field Use at CCDD Facilities
- Matt Vondra
- Bluff City Materials
3PID/FID Equipment
- AGENDA
- Presentation on PID and FID Technology
- Demonstration of TVA (FID-PID)
- Demonstration of RAE PIDs
- Questions And Answers
4PID/FID EquipmentUse and Calibration
- Section 1100.205 Load Checking
- The owner or operator must institute and conduct
a load checking program designed to detect
attempts to dispose of waste at the facility. At
a minimum, the load checking program must consist
of the following components - a) Routine Inspections
- 1) An inspector designated by the facility must
inspect every load before its acceptance at the
facility utilizing an elevated structure, a
designated ground level inspection area, or
another acceptable method as specified in the
Agency permit. In addition to a visual
inspection, the inspector must use an instrument
with a photo ionization detector utilizing a lamp
of 10.6eV or greater or an instrument with a
flame ionization detector, or other monitoring
devices approved by the Agency, to inspect each
load. All instruments shall be interpreted based
on the manufacturers margin of error. Any
reading in excess of background levels using any
of these instruments must result in the rejection
of the inspected load. In addition, any reading
in excess of background levels on any monitoring
device used by the Agency during an Agency
inspection must result in the rejection of the
inspected load.
5PID/FID Equipment
- Introduction to FID and PID gas monitor equipment
and practical considerations for field use - Pat Maloney
- J M Instrument Co.
- (219) 924-4545 x104
6PID/FID Equipment
- MSA
- Permanent Portable Gas Monitors for Toxics
Combustible Gases - Refrigerant Monitors, Confined Space Meters
Self-Contained Breathing Apparatus (SCBA)
- SERVOMEX
- Gas Analyzers for Process
- Control, Combustion, Environmental
- Features Paramagnetic 02 Insitu Combustion CO
/or O2 Photometric IR/UV
- THERMO (including the former Foxboro TVA)
- Industrial Mass Spec for Fast On Line Analysis
of Process, Safety and Environmental Applications - On Line Sulfur for Gas Refining
7.
PID/FID Equipment
- YOKOGAWA
- Process Analyzers for Combustion O2, Gas
Chromatography, NDIR and Gas Density
- RAE
- Portable Gas Monitors for Toxic and Explosive
Gasses - Specialists in PID Technology, Including
Benzene Butadiene Specific Portables
8Photo Ionization Detection (PID)
PID/FID Equipment
- Basic Theory Sample is introduced into an
ionization chamber and exposed to an ultraviolet
lamp of a specific energy. The photons of UV
energy excites the sample and ions (less than or
equal IP to lamp) are attracted to a collecting
electrode. The collection of the ions result in
an increase current which is proportional to the
concentration of the compound (compared to a
known calibration standard). The sample exits
the system unaltered.
9 Diagram of PID
PID/FID Equipment
10Flame Ionization Detection (FID)
PID/FID Equipment
- Basic Theory Sample is introduced into an
ionization chamber and burned in a hydrogen
flame. This process separates free ions (from
hydrogen and carbon bonded hydrocarbons)
which in turn are attracted to a collecting
electrode. The collection of the ions result in
an increase current which is proportional to the
concentration of the compound (compared to a
known calibration standard). The by-products of
the process are H2O and CO2.
11 Diagram of FID
PID/FID Equipment
12Advantages FID
PID/FID Equipment
- Wide dynamic and linear range
- (0-50,000 ppm or 5 which is 100 LEL as Methane)
- Highly sensitive to hydrocarbon vapors
- Can See Methane
- Very stable and repeatable
- (with generally tighter correction factors vs
PID) - Unaffected by ambient levels of water vapor
13 Disadvantages FID
PID/FID Equipment
- Requires H2 Fuel Source
- Requires 16 02 for reading (so has Flameout
Issues) - Size and Weight
- Sees Methane
- (not really toxic can interfere with seeing
gas of interest) - More Complex Unit
- (Calibration Operation - Cost and
Maintenance) - Can NOT See Inorganics
14 Advantages PID
PID/FID Equipment
- Size and Weight (Newer Models) and Simplicity of
Use - No support gases required (works in inert
conditions no O2) - Better low end sensitivity (PPB newer models)
- Non-destructive detector (allows sample
collection) - Can measure many inorganic compounds (NH3 for
exam) - Sensitive to aromatic, chlorinated unsaturated
hydrocarbons - Immune to (does not see) Methane
- Easier and Less Costly to Maintain (newer models
have less - expensive and more accessible lamps and detectors)
15 Disadvantages PID
PID/FID Equipment
- Typically sufficient range but lower range than
FID - (0-2000/10,000 ppm and looses linearity at higher
ranges) - Limited to reading gasses at/below Ionization
Potential of the - UV lamp installed (typically 10.6 though 11.7
and 9.8 exist) - Affected by Moisture 95 RH may reduce reading
25-30 - (New models do allow for dryer tubes 15/30
minute run time) - Based on my field experience, beware the claim
for moisture - immunity/compensation of some PID manufacturers.
16IP and Lamps
PID/FID Equipment
- IEPA requires 10.6 minimum (best lamp)
- 9.8 reduced survey capability
- 11.7 expanded range
- Quickly degrades less stable
- short life
17CENTURY TVA1000B
PID/FID Equipment
18Internal Packaging
PID/FID Equipment
19 Diagram of Dual System
PID/FID Equipment
20CCDD Rules and Regulations
- PID/FID Field Use and Calibration
- Concerns Under CCDD Rules and Regulations
21PID/FID Field Use
- Filed Use of FID and PID gas monitor equipment at
CCDD Facilities - Concerns Under CCDD Rules and Regulations
- Matt Vondra
- Bluff City Materials
22PID/FID
PID/FID Field Use
- PID - Photoionization Detector
- FID - Flame Ionization Detector
- Both are sensitive air monitoring devices
- Air stream pumped through detector so gases can
be analyzed - Unit ionizes hydrocarbon compounds carried in the
sample air stream - Ionized compounds are converted by unit into an
approximate concentration when compared to a
known standard - Does not distinguish one type of compound from
another
23PID/FID Operation
PID/FID Field Use
- Calibration
- Calibrate to zero using hydrocarbon free air
- Calibrate using 100ppm isobutylene gas
- Screen CCDD materials by sampling vapors
emanating off the material - Hydrocarbons found in the material will
volatilize and be detected through the sample air
stream - Detector will give approximate concentration in
parts per million (ppm)
24PID/FID Field Use
- Place the tip of the PID near the edge of the
material collected with the sampling device - Record the peak measurement shown on the PID
- If the PID shows a reading greater than the
margin of error, the material could be
contaminated and that load should be rejected - Results must be documented
25PID/FID Field Use
26PID/FID Field Use
27PID/FID Field Use
28PID/FID Field Use
29Important Considerations
PID/FID Field Use
- Thoroughly review and understand manual
- Unit response time is usually within a few
seconds - Manufacturer provides a statement of the
sensitivity/accuracy of the device - Routine maintenance is needed and should be
documented - Charging
- Inlet filter change out
- Lamp change out
30Important Considerations
PID/FID Field Use
- Environmental Factors which may affect reading
accuracy - Temperature and humidity extremes
- Rain and sunlight
- Moisture and solids introduced into detector
through air sampling pump - Can purchase an outside filter to minimize the
environmental factors that could affect accuracy
of readings
31PID/FID Alternatives
PID/FID Field Use
- IEPA allows for the approval of alternate devices
- Field GC/MS gas chromatograph/mass spectrometer
- X-ray fluorescence (XRF) analyzer - screens for
the presence of metals - estimated 35,000 per XRF analyzer
32Calibrating PID/FID for Background Levels
PID/FID Field Use
- PID/FID devices should be calibrated daily using
the manufacturers suggested gas - May calibrate several times throughout the day to
account for any fluctuations in readings due to
weather conditions - Calibration should occur at the CCDD inspection
site - The device has a margin of error of /- 0.1 ppm
33Calibration Steps
PID/FID Field Use
- Each morning the inspector should
- Calibrate the device
- Establish background levels
- Apply the margin of error to determine the
rejection threshold and record this number - Utilize this number when determining which loads
are accepted/rejected -
- See Sample Calibration Log
34PID/FID Calibration Log
PID/FID Field Use
35Issues Worth Additional Discussion
PID/FID Equipment
- Sample preparation
- How to deal with Back Ground measurements
- PID Calibration Demonstration
-
36Sample Conditioning Considerations
PID/FID Equipment
- Sample Conditioning Concepts for improved
results not required procedures (to my
knowledge) - You want a warm representative sample(s) to
test to get better results - Cold Weather seal off sample and raise
temperature if possible (less of an issue in warm
weather but could be standardized as a
procedure. - Get a sample from the center not just a pass at
the top - As easy as a Zip lock bag by floor board heater 5
min or similar SAFE temperature increase
leave sufficient head space to test.
37How to deal with background
PID/FID Equipment
- I have long proposed to industrial users (LDAR)
to always use a real baseline ZERO and to
include the background as part of their readings.
They may have more room for inclusion than will
work for CCDD applications. However, I remain a
big fan of starting from a baseline ZERO and if
needed log the background and report that
against the reading.
38Zeroing out the background
PID/FID Equipment
- If you ever need to explain your numbers people
will understand a logged offset that was recorded
against the reading. (The instrument reported
ZERO on the Zero gas we noted 3 to 4 ppm
background and the unit reported 12 ppm sample
headspace (9 ppm actual difference after
accounting for lowest average background). - Now imagine explaining that same 9 ppm after site
personnel ZEROs to the back ground (unrecorded). - Baseline ZERO is simpler to maintain and explain
39PID Cal Demo
PID/FID Equipment
- Use of Demand Flow Reg (bags used FIDs)
- USE ZERO GAS (not background)
- Span to Isobutylene 100 ppm
- Marker Cap field check not marker
40MiniRAE 2000 PID Sensor
PID/FID Equipment
- Inspect sensor for damage and replace to rectify
the following conditions - Bent electrode fingers
- Teflon mask warped so that sensing electrode
shows - Electrodes not in the same plane
- If problems persist, replace PID sensor
Bent Electrode
Straight Electrodes with mask removed for clarity
Teflon Mask Warped
41Maintenance
PID/FID Equipment
- Clean PID Lamp Sensor
- When display creeps upwards after good zero
- When PID responds to moisture
- When movement of PID results in response on
display
42Maintenance
PID/FID Equipment
- Humidity Check
- Cup hand over inlet or breathe into inlet for
10-20 seconds - Do not block flow
- If M2K reads gt2 ppm or ppbRAE reads gt500 ppb,
then the sensor needs cleaning
43Maintenance
PID/FID Equipment
- How to Clean PID Sensor
- Always clean sample probe and replace or clean
filters FIRST! If PID holds a stable zero after
this step then further cleaning may not be
necessary - Use anhydrous methanol (Lamp cleaning solution),
never use water - Clean lamp face with lens tissue do not touch
with your bare hands
44Maintaining PID Sensor
PID/FID Equipment
- Cleaning the PID Sensor
- Clean sensor by immersion in cleaning solution
- Do not loosen or remove screws on PID sensor
- Remove the o-ring from the PID sensor
- If Ultrasonic Cleaner is not available
- Immerse in cleaning solution and agitate by hand
- Drying the PID Sensor
- Let air dry overnight
- Warm air (not hot) will speed drying
45Humidity Filtering II Tubes
PID/FID Equipment
- Temporary relief for a dirty sensor
- Dries sample gas for about ½ hour
- Measure VOCs multiple sample use OK
- Useful for gasoline and chlorinated solvents
- CAUTION May cause low response for some
compounds or at low temperature or concentration
46General Calibration Procedure
PID/FID Equipment
- Prior to daily use, it is good practice to
perform a Fresh Air/Zero Calibration - After Calibration, test for moisture response
- If readings increase more than 5ppm, clean lamp
sensor - If readings do not increase higher than 5ppm,
continue with pump stall test - If unit passes, test pump to verify it will
stall, if pump does not stall, service the pump - Check Stall Threshold
- Check tubing
- Clean/rebuild pump or replace
47MiniRAE 2000 Top Assembly
PID/FID Equipment
- Filter history
- Green dust filter was originally used, but proved
not to work effectively - C filter was used next but proved to absorb too
many VOCs - Now only porous metal filter is used
48Inside the MiniRAE 2000
PID/FID Equipment
49Questions?