Terry Kotrla, MS, MT(ASCP)BB - PowerPoint PPT Presentation

1 / 27
About This Presentation
Title:

Terry Kotrla, MS, MT(ASCP)BB

Description:

In 1984 Judd recommended deleting the autocontrol as part of routine pretransfusion testing. By 1986 the minor crossmatch was of historic interest only. – PowerPoint PPT presentation

Number of Views:249
Avg rating:3.0/5.0
Slides: 28
Provided by: TerryK2
Category:

less

Transcript and Presenter's Notes

Title: Terry Kotrla, MS, MT(ASCP)BB


1
Unit 8 Pretransfusion TestingPart 2
  • Terry Kotrla, MS, MT(ASCP)BB

2
Compatibility Testing - History
  • Early 1980s started to question utility of
  • Routine use of anti-A,B and A2 cells in ABO
    grouping
  • Repeat D typing of D positive donor units
  • Weak D testing
  • Repeat antibody screen on donor units
  • DAT testing
  • Performance of elutions
  • Significance of antibodies reactive at RT or
    below.
  • Usefulness of albumin in antibody detection tests
  • Use of AHG in both antibody screen AND crossmatch

3
Compatibility Testing - History
  • During 1984-85 FDA and AABB allowed the AHG phase
    of the CROSSMATCH to be deleted if the patients
    antibody screen was negative.
  • In 1984 Judd recommended deleting the autocontrol
    as part of routine pretransfusion testing.
  • By 1986 the minor crossmatch was of historic
    interest only.

4
Compatibility Testing Coombs Crossmatch
  • Who needs a crossmatch? Patients who are
  • experiencing clinical signs and symptoms of
    anemia.
  • actively bleeding.
  • having a surgical procedure where possibility of
    excessive bleeding is high.
  • What is the major crossmatch
  • Patient serum/plasma added to donor cells
  • Read at three phases IS, 37C and AHG.
  • Set up and read as part of antibody screen
    procedure
  • Agglutination and/or hemolysis are positive.
  • Donor cells reacting with patient sample at 37C,
    AHG or causing hemolysis are incompatible
    CANNOT be transfused.

5
Compatibility Testing - IS
  • When NO CLINICALLY SIGNIFICANT antibodies are
    detected in the antibody screen AND there is no
    history of antibodies, the AHG phase of testing
    is NOT required.
  • Rarely are AHG incompatible crossmatches obtained
    when antibody screen is negative.
  • MUST demonstrate ABO incompatibility by
    performing IS crossmatch.
  • Policy change requires medical director approval.

6
Compatibility Testing - IS
  • Decision to omit AHG phase based on the
    following
  • Incidence of incompatible crossmatches when
    antibody screen is negative and the reason.
  • Sensitivity of antibody detection procedure used.
  • Benefits of omitting AHG phase in the laboratory.
  • Expertise of individuals working in the
    transfusion service.
  • If clinically significant antibodies are present
    the AHG phase of the crossmatch is required.

7
Compatibility Testing - Computer
  • Antibody screen negative and computer validated
    on site to prevent release of ABO-incompatible
    blood it may be used to detect ABO compatibility
    instead of serologic testing.
  • Following conditions MUST be met
  • TWO determinations of patients ABO group by
    second type on same sample OR second current
    sample, or comparing to previous records.
  • Donor ABO/D type, unit number, component name and
    confirmatory type.
  • Patient ABO/D type, antibody screen result and
    two unique patient identifiers.
  • Method to ensure correct data entry.
  • Computer logic to alert to ABO/D discrepancies on
    unit label and testing and ABO incompatibility
    between recipient and donor.

8
Optional Pretransfusion Testing
  • ABO Grouping
  • Testing RBCs with anti-A,B
  • Serum/plasma tested against A2 cells
  • D typing
  • Weak D test
  • Rh control with chemically modified reagents
    unless AB pos
  • Antibody Screen (IAT)
  • RT incubation
  • Additives such as albumin or LISS
  • Enzymes
  • Polyspecific AHG in IAT

9
Optional Pretransfusion Testing
  • Autocontrol or DAT
  • Published data indicate performance is of limited
    value even in recently transfused patients.
  • Standards does not require autocontrol or DAT
  • Microscopic reading of tests, magnifier viewing
    lamp adequate.
  • Crossmatch
  • 37C and AHG when antibody screen and history is
    negative.
  • RT incubation
  • Enzyme tests
  • Polyspecific AHG
  • Minor crossmatch

10
Selection of Donor Group
  • When possible ABO identical.
  • D positive should be selected for D pos, although
    D neg is acceptable but should be reserved for D
    neg except
  • D neg short date unit can be given to D pos sure
    give.
  • Multiple antibodies present and D neg more likely
    to lack.
  • D negative should be selected for D neg to avoid
    immunization to D antigen.
  • Consult with medical director and patients
    physician if need is urgent.
  • Use D negative first.
  • Weigh risk of patient death versus immunization
    to D.
  • May be appropriate to administer Rh immune
    globulin especially after platelet transfusion.

11
Selection of Donor Group
Recipient ABO Compatible Donor ABO In Order of Selection
O O
B B, O
A A, O
AB AB, A, O, B why O before B??
12
Blood Administered after Non-Type Specific
Transfusion
  • Determine presence of anti-A and/or anti-B in
    patient.
  • When serum from freshly drawn sample is
    compatible at AHG with recipients own blood
    group may return to group specific.
  • If AHG incompatible must continue with
    alternative ABO group.
  • If change involved D only return to D type
    specific.

13
Other Blood Groups
  • Unnecessary to select units based on other blood
    groups UNLESS patient has clinically significant
    unexpected antibody.
  • If antibody strongly reactive use patient serum
    to screen then confirm with specific typing sera.
  • Weakly reactive screen units with specific typing
    sera.
  • If commercially prepared typing sera is not
    available use patient sample or plasma from donor
    with antibody.

14
Antibody Detection Techniques
  • Use 2 to 3 commercially prepared group O cells.
  • Relatively short shelf life, two weeks.
  • Antigen profile (antigram) provided with analysis
    of antigens present on each cell.
  • MAKE SURE lot number of screen cell matches
    antigram.
  • Add patient serum/plasma to screen cells and
    observe at
  • RT/IS
  • After incubation at 37C with enhancement media.
  • After washing and addition of AHG reagent.
  • Agglutination and/or hemolysis is POSITIVE.
  • Phase of reactivity of positive reaction
    extremely helpful.

15
Antibody Detection Techniques
  • Antibody detection procedure used determined by
    what is considered significant antibody.
  • Carefully considered if AHG crossmatch not
    performed.
  • Once adopted method written in SOP and must be
    followed exactly by all staff members.
  • Detection method chosen should
  • Detect as many clinically significant antibodies
    as possible.
  • NOT detect clinically insignificant antibodies.
  • Allow prompt delivery of blood to the patient.

16
Antibody Detection Techniques
  • Method should be sufficiently sensitive to detect
    low level of antibody in patient serum or plasma.
  • Undetected low levels of patient antibody may
    result in rapid production of antibody if antigen
    positive cells transfused.
  • Antibody present in donor plasma will not harm
    recipient.
  • Methods for antibody screen and crossmatch may be
    the same or different.
  • RT tests such as IS crossmatch required to detect
    ABO incompatibilities, may not be required for
    antibody screen.
  • Antibody screen MUST include AHG to detect
    clinically significant antibodies, crossmatch may
    be IS only.

17
Antibody Detection Techniques
  • Lab personnel should use same interpretations,
    notations and consistency in grading reactions.
  • Consistency in grading reactions crucial.
  • Hemolysis and/or agglutination constitutes
    visible endpoint of antigen-antibody reaction and
    must be observed accurately and consistently.
  • Use light source or optical aid to enhance
    sensitivity and consistency.
  • Microscopic observation is not required but is
    useful to
  • Distinguish rouleaux from true agglutination
  • Detect mixed field agglutination seen in anti-Sda.

18
Antibody Detection Techniques
  • Reactions must be observed for hemolysis then
    agglutination IMMEDIATELY after centrifugation.
  • Manner in which RBCs are dislodged is crucial.
  • Hold tube at angle so fluid cuts across cell
    button as tube is tilted.
  • Reaction is not interpreted until ALL cells
    resuspended.
  • Over shaking will result in weak or negative
    reactions.
  • Reactions are recorded IMMEDIATELY with tube held
    in hand in front of column to record in.

19
General Considerations
  • Labeling tubes
  • Each tube labeled properly BEFORE use.
  • Recipients initials (or other identifying
    information) and donor unit number or reagent RBC
    identification.
  • System must allow for accurate, rapid labeling.
  • NEVER rely on the position of a tube in a rack or
    centrifuge head to identify the contents of the
    tube.
  • ALWAYS place tubes in the serofuge head in the
    order they will be read.
  • Use the SAME organizational techniques when
    labeling and arranging tubes in rack to improve
    organization and speed.

20
General Considerations
  • Volume of serum or plasma.
  • Most procedures call for 2 drops.
  • Research has shown 2 drops provide optimal
    antibody to antigen ratio.
  • Some alloantibodies detected only when 3 to 4
    drops used.
  • High variability in delivery in transfer
    pipettes.
  • Standardize volumes based on equipment used in
    your lab.
  • Low ionic reagents require ration of 2 drops
    serum/plasma to 2 drops LISS, cannot vary.

21
General Considerations
  • Cell suspension
  • RBCs used for crossmatching obtained from sealed
    segment of original tubing attached to blood
    container.
  • Wash once and prepare 2-4 suspension, some
    workers prefer 2 as it increases sensitivity of
    the test system.
  • Best to use weakest suspension that can be
    observed for agglutination.
  • Too heavy of a cell suspension can cause weak
    antibodies to be missed.

22
Testing Techniques Saline Tube
  • Simplest to perform.
  • Mix serum or plasma with saline suspended RBCs,
    centrifuge and read, incubate at RT or 37C.
  • Used in crossmatching to detect ABO
    incompatibility.
  • In antibody tests used to detect IgM antibodies
    which react preferentially at RT anti-M, -N,
    -P1, -Le and I.
  • Rare examples of antibodies of other
    specificities may be observed at RT but more
    often will be reactive at 37C and/or AHG as well.

23
Testing Techniques Bovine Albumin Tube
  • Utilized to enhance agglutination of IgG
    antibodies since 1945.
  • Decreases amount of time required for incubation.
  • Controversy Decrease zeta potential (affects
    second stage of agglutination) or due to function
    of ionic strength of albumin diluent does it
    increase uptake of antibody onto cells?
  • Many antibodies have enhanced reactivity when
    albumin is added to test system.

24
Testing Techniques LISS Tube
  • Low Ionic Strength Saline shortens incubation
    time.
  • Increases antibody uptake onto cell, enhancing
    agglutination.
  • Several important factors to consider
  • Incubation time and sensitivity subsequent to AHG
    depends upon desired ionic conditions.
  • Adding additional serum will increase ionic
    strength, must not be done.
  • MUST adhere to manufacturers instructions.

25
Testing Techniques PEG Tube
  • Polyethylene Glycol (PEG) is a water soluble,
    neutral polymer which is an effective potentiator
    of antigen-antibody reactions.
  • Advantages over albumin include
  • Increases rate of detection of clinically
    significant antibodies.
  • Decreases detection of clinically insignificant
    antibodies.
  • May decrease need for other enhancement
    techniques.
  • Procedure
  • Serum or plasma added to RBCs, perform IS.
  • Add PEG and incubate at 37C IS NOT READ AFTER
    37C
  • Wash and add AHG.

26
Testing Techniques Enzymes Tube
  • More appropriate for antibody ID than routine
    testing.
  • GREATLY enhance reactivity of Rh antibodies.
  • CANNOT be only method used as M, N, S, Fy and
    other antigens are destroyed, those antibody
    specificities would not be detected.
  • Enzymes used include
  • Papain
  • Bromelain
  • Trypsin
  • Ficin MOST POPULAR

27
References
  • AABB Technical Manual, 16th edition, 2008
  • CAP Today http//tinyurl.com/4cd4qgd
  • Basic Applied Concepts in Immunohematology, 2nd
    edition, 2009
  • Ortho WIRE, http//www.ortho-wire.com
Write a Comment
User Comments (0)
About PowerShow.com