Title: Update to Absolute QE


1
Update to Absolute QE Photon Flux
  • Need to know illumination power at CCD
  • Measure illumination power in integrating sphere
    during QE measurements
  • Use K ( power at CCD/Power at sphere) to
    determine light level at CCD
  • What if K is not constant but K(l)?
  • Measure K for newly design integrating sphere
  • Better uniformity at CCD?
  • No need to 2-sphere illumination

2
K Factor from Monochromator
  • K appears to depend on l
  • Is this dependence in illumination system or
    photodiode calibration?
  • Could have a 10 effect on QE measurement in
    near IR

3
K from Monochromator and Filter Wheel
4
Conclusion
  • Is K a function of l or this in the photodiode
    calibration?
  • Is there some l dependent absorber in system?
  • Either way, this will change the shape of the
    absolute QE curve