Predicting visual performance from wavefront quality metrics in cataract

About This Presentation
Title:

Predicting visual performance from wavefront quality metrics in cataract

Description:

Predicting visual performance from wavefront quality metrics in cataract Konrad Pesudovs Katja Ullrich NH&MRC Centre for Clinical Eye Research, Flinders Medical ... –

Number of Views:71
Avg rating:3.0/5.0
Slides: 9
Provided by: KatjaU
Category:

less

Transcript and Presenter's Notes

Title: Predicting visual performance from wavefront quality metrics in cataract


1
Predicting visual performance from wavefront
quality metrics in cataract
  • Konrad Pesudovs
  • Katja Ullrich
  • NHMRC Centre for Clinical Eye Research, Flinders
    Medical Centre Flinders University, Adelaide,
    South Australia

Financial disclosure The authors have no
financial interest
2
Background and Purpose
  • Cataract affects visual performance via higher
    order aberrations and light scatter
  • Wavefront aberrations occurring in cataract have
    been described in terms of the Zernike polynomial
    decomposition but neither Zernike terms nor RMS
    predict visual performance
  • Other methods for organising wavefront data exist
    wavefront quality metrics
  • Attempts to connect wavefront quality metrics to
    visual performance in cataract are lacking
  • PURPOSE To determine which wavefront quality
    metrics are most predictive of visual performance
    in patients with cataract

3
Population and Visual Performance
  • Prospective, cross-sectional study of consecutive
    patients attending the Cataract Assessment Clinic
    at Flinders Medical Centre
  • Inclusions all types of cataract
  • Exclusions ocular comorbidity, unable to
    measure whole eye wavefront
  • 206 eyes, age 73 years, 58 female
  • The clinical assessment was conducted by one
    clinician-KP
  • Refraction and best corrected
  • High contrast visual acuity(VA)
  • Pelli-Robson contrast sensitivity (PRCS)
  • Pelli-Robson contrast sensitivity under glare
    (PRCSglare)

4
Wavefront quality metrics
  • Whole eye wavefront sensing with Wavefront
    Sciences COAS-HD
  • Wavefront data exported to VOLPro software v7.25
    (Sarver and Associates) and 10th order Zernike
    expansion derived
  • Zernike data exported to GetMetrics v.2.02.006
    (University of Houston, College of Optometry) by
    Thibos and Applegate for calculation of wavefront
    quality metrics
  • 31 metrics of wavefront quality designed to be
    predictive of visual performance were calculated
    for the pupil plane and the image plane as per
    Thibos LN, Hong X, Bradley A, Applegate RA.
    Accuracy and precision of objective refraction
    from wavefront aberrations. J Vis
    20044(4)329-51.
  • Linear Regression with SPSS Software V15.0 (SPSS
    Inc)

5
Results - visual acuity and wavefront quality
metrics
  • The strongest correlate of all three measures of
    visual performance was the pupil fraction metric
    PFWc
  • Visual acuity and logPFWc, r2-0.37, plt0.001

6
Results contrast sensitivity and wavefront
quality metrics
  • The strongest correlate of each measure of
    contrast sensitivity was the pupil fraction
    metric PFWc
  • Pelli-Robson contrast sensitivity and logPFWc,
    r20.39, plt0.001
  • Pelli-Robson contrast sensitivity glare
    logPFWc, r20.32, plt0.001

7
Pupil fraction metrics
  • Pupil fraction is defined as the fraction of the
    pupil area for which the optical quality of the
    eye is good
  • The critical pupil method uses an area of good
    pupil which is a concentric zone
  • The red circle indicate the largest concentric
    zone for which the wavefront has reasonably good
    quality
  • PFWc which is a critical pupil defined as the
    concentric area for which RMSwltcriterion (?/4)

8
Conclusion
  • Pupil fraction metrics are the best correlates
    with visual performance in cataract, and also
    have performed well in normal eyes

Pupil fraction metrics should be used to organise
wavefront aberration data so as to be predictive
of visual performance