Intraoperative Mechanical (Structural) Failure of a Phacoemulsification Probe - PowerPoint PPT Presentation

About This Presentation
Title:

Intraoperative Mechanical (Structural) Failure of a Phacoemulsification Probe

Description:

Intraoperative Mechanical (Structural) Failure of a Phacoemulsification Probe Rachel E. Peck, MD* Herbert J. Ingraham, MD* Geisinger Medical Center, Department of ... – PowerPoint PPT presentation

Number of Views:46
Avg rating:3.0/5.0
Slides: 8
Provided by: C507
Category:

less

Transcript and Presenter's Notes

Title: Intraoperative Mechanical (Structural) Failure of a Phacoemulsification Probe


1
Intraoperative Mechanical (Structural) Failure of
a Phacoemulsification Probe
  • Rachel E. Peck, MD
  • Herbert J. Ingraham, MD
  • Geisinger Medical Center, Department of
    Ophthalmology
  • Danville, PA

No financial interest
2
PURPOSE
  • To report a case in which the distal 4 mm tip of
    a reprocessed flared 30 degree 1.1mm
    phacoemulsification probe broke off from its base
    intraoperatively.

3
CASE PRESENTATION
  • Our case was a routine cataract surgery, in which
    a superior scleral tunnel approach was used.
  • Within seconds of the initiation of
    phacoemulsification, the tip spontaneously broke
    off.
  • Phacoemulsification was immediately stopped and
    the phaco probe and broken tip were both safely
    removed from the eye.

4
The reprocessed phacoemulsification probe that
broke intraoperatively
A similar flared 30 degree 1.1mm
phacoemulsification probe
5
Enlarged photo showing the jagged edges of the
complete detachment of the tip of the probe from
its base.
6
REPROCESSED SINGLE USE DEVICES
  • Phacoemulsification probes are typically marketed
    as single use devices (SUDs) that the original
    equipment manufacturers (OEMs) have not validated
    for repeated use.
  • However, due to both economic and environmental
    advantages, many hospitals and ambulatory surgery
    centers have adopted policies that allow the use
    of SUDs that have been reprocessed by third
    parties or hospitals.
  • The FDA monitors the use of reprocessed SUDs, and
    believes that reprocessed SUDs that meet the
    FDAs regulatory requirements are as safe and
    effective as a new device. 1

1 Statement of Daniel Schultz, MD, Director CDRH,
Before the Committee on Governmental Reform-Sept
26 2006. www.fda.gov
7
CONCLUSION
  • This case raises the question of the frequency of
    mechanical failure of reprocessed
    phacoemulsification instruments and whether
    repeated use of the probe can result in
    structural weakness due to reprocessing and/or
    oscillatory power.
  • Reprocessing surgical instruments has economic
    and environmental benefits, however this must be
    balanced with safety and the ability of the
    phacoemulsification probe to undergo repeated
    use.
  • This is the only incidence of a reprocessed
    phacoemulsification instrument breaking
    intraoperatively that we are aware of. Perhaps
    it is a rare, random occurrence, or it could be
    an under-reported phenomenon. Further discussion
    is warranted.
Write a Comment
User Comments (0)
About PowerShow.com