Defibrillator Implants: - PowerPoint PPT Presentation

1 / 25
About This Presentation
Title:

Defibrillator Implants:

Description:

... RF System Power Control Systems Pulser Power Transformers Pulser Power Transformers RF Distribution Waveguides Klystron Gallery Linac Klystrons Case ... – PowerPoint PPT presentation

Number of Views:131
Avg rating:3.0/5.0
Slides: 26
Provided by: temp255
Learn more at: https://orau.org
Category:

less

Transcript and Presenter's Notes

Title: Defibrillator Implants:


1
Defibrillator Implants
Occupational Exposure to Electromagnetic Radiation
  • Bruce Murdoch PhD CHP CLSO
  • Argonne National Laboratory
  • Argonne, IL USA
  • btmurdoch_at_anl.gov
  • AIHce 2011, Portland OR Roundtable
    Session RT 201

2
Typical ICD Device
  • Dimensions 2 X 2.5 X 0.2

3
Implant Placement
  • EM Vulnerabilities
  • Leads
  • Control Unit
  • Failure Modes
  • False Trigger
  • Stopping

Source www.defibrillator-help.com
4
CASE STUDY Where? APS
(Advanced Photon Source) Who?
Accelerator Operator w/ICD Why?
Work Clearance
5
(No Transcript)
6
(No Transcript)
7
APS Control Room
8
Manufacturer Test Standard
  • ANSI/AAMA PC69 - 2007

9
Typical Manufacturer (Medtronic) Exposure Limit
Recommendations
  • Static B-Field B 0.5 mT (5G)
  • 50/60 Hz B 0.1 mT (1G)

  • E 6000 V/m
  • 10 kHz B 1.25 mT (12.5 G)
  • R-F E-Field ( 150 kHz) E 100 V/m

10
ACGIH 2010 TLV Book Exposure Limit
Recommendations
  • Static B-Field B 0.5 mT (5G)
  • ELF (5 Hz2kHz) B 0.1 mT (1G)
  • VLF (2kHz400 kHz) No recommendation
  • R-F (100 kHz3GHZ) E 3-10 V/m

11
Static B-Fields
  • MetroLab
  • Model ETM-1
  • 3-Axis Magnetometer

12
ELF
  • Holaday HI-3627
  • ELF Magnetic Field Meter
  • w/ HI-3602 Probe
  • 30 Hz 2 KHz

13
VLF
  • Holaday HI-3637
  • VLF Magnetic Field Meter
  • w/ HI-3091 Probe
  • 2kHz 400 KHz

14
R-F
  • Narda NBM-520
  • RF Survey Meter
  • w/ EF3091 E-Field Probe
  • 100 kHz 3 GHz

15
Spectrum Coverage
RF
16
The Instrument Ensemble
17
Survey Plan
  • Daily Control Room work area
  • Weekly maintenance walkthroughs
  • ESH Coordinator guidance
  • 84 specific survey locations plus area sweeps
  • Full spectrum survey at each location

18
Control Room Displays and Instruments
19
Control Room Panels
20
RF Distribution System
21
Electrical Panels
  • ELF, VLF B 0.1 G
  • RF E 4 V/m

22
Electrical Distribution Room
  • ELF B 0.4 G _at_ 6 inches from transformer
  • RF E 4 V/m

23
Mezzanine RF System
  • Freq 325 MHz
  • E(RF) up to 3.5 V/m _at_ 6 in.

24
Power Control Systems
25
Pulser Power Transformers
26
Pulser Power Transformers
27
RF Distribution Waveguides
  • Freq 325 mHz

28
Klystron Gallery
  • Freq 2.865 Mhz

29
Linac Klystrons
  • Freq 2.8565 GHz
  • E(RF) 2 V/m _at_ 1 ft

30
Case Study Results
  • Employee barred from
  • Fundamental RF Room 7 V/m RF
  • Electrical Distribution Room 4 V/m RF
  • Cautions
  • Some instrument racks 4 V/m RF _at_
    12
  • Some waveguides 3.5 V/m
    RF _at_ 12

31
Credits
  • Technical Assistance
  • Mary Malik
  • Adam Sauter
  • Photography
  • George Joch
  • Edmund Chang
Write a Comment
User Comments (0)
About PowerShow.com