Title: Dosimetry Fundamentals II
1Dosimetry Fundamentals II
- General Characteristics of Dosimeters
2Absoluteness
- An absolute dosimeter is one that can be
assembled and used to measure the absorbed dose
deposited in its own sensitive volume without
requiring calibration in a known field of
radiation - It may, however, need some kind of calibration
not involving radiation, such as
electrical-heating calibration of a calorimetric
dosimeter
3Absoluteness (cont.)
- Three types of dosimeters are now generally
regarded as being capable of absoluteness - Calorimetric dosimeters
- Ionization chambers
- Fricke ferrous sulfate dosimeters
4Absoluteness (cont.)
- These are not always employed as absolute
dosimeters, however, because calibration offers
certain advantages - A calibration can be stated in terms of some
quantity of interest other than the absorbed dose
in the sensitive volume, e.g., tissue dose or
exposure - It can also provide traceability to an
authoritative standardization laboratory such as
the NRCC
5Absoluteness (cont.)
- When an absolute dosimeter is used independently,
it relies on its own accuracy instead of
referring to a standard dosimeter in common with
other radiation users - Thus an error may go undetected in an absolute
dosimeter unless comparisons with others are
carried out, or a calibration is obtained at a
standardization laboratory
6Absoluteness (cont.)
- The calorimetric dosimeter has the fundamental
advantage of directly measuring the heat to which
the absorbed dose degrades, without dependence on
any coefficient of conversion such as to
ionization (W) or to chemical yield (G) - Thus if there is a hierarchy of dosimeter
absoluteness, the calorimeter ranks at the top - Note that the absoluteness of a dosimeter is
independent of its precision or its accuracy
7Precision and Accuracy
- The concept of the precision or reproducibility
of dosimeter measurements was discussed earlier
it has to do with random errors due to
fluctuations in instrumental characteristics,
ambient conditions, and so on, and the stochastic
nature of radiation fields - Precision can be estimated from the data obtained
in repeated measurements, and is usually stated
in terms of the standard deviation - High precision is associated with a small
standard deviation
8Precision and Accuracy (cont.)
- The accuracy of dosimeter measurements expresses
the proximity of their expectation value to the
true value of the quantity being measured - Thus it is impossible to evaluate the accuracy of
data from the data itself, as is done to assess
their precision - Accuracy is a measure of the collective effect of
the errors in all the parameters that influence
the measurements - In experiments that are limited to relative
measurements, only the precision, not the
accuracy, is important
9Precision and Accuracy (cont.)
- Clearly precision and accuracy are separate
characteristics - Measurements may be highly precise but
inaccurate, or vice versa, or may be strong in
both or neither of these virtues - If one speaks of a dosimeter as being a
high-precision instrument, one means that it is
capable of excellent measurement reproducibility
if properly employed
10Precision and Accuracy (cont.)
- Poor technique, a hostile environment (e.g., high
atmospheric humidity) or faulty peripheral
equipment (e.g., ion-chamber cables or
electrometer) may cause poor reproducibility - Accuracy depends on the type of radiation being
measured, and dosimeter calibrations are more or
less specific in that respect - A dosimeter that is accurately calibrated to
measure the exposure at one x-ray quality may be
significantly in error at another
11Precision and Accuracy (cont.)
- The quantity that a dosimeter is inherently the
most capable of measuring accurately, and that is
the least influenced by changing the type or
quality of the radiation, is the absorbed dose
deposited in the dosimeters own sensitive volume
12Dose Range Dose Sensitivity
- To be useful, a dosimeter must have adequate dose
sensitivity (dr/dDg) throughout the dose range
to be measured - A constant dose sensitivity throughout the range
provides a linear response, that is desirable for
ease of calibration and interpretation - However, single-valuedness of the function
r(Dg), even if nonlinear, may be acceptable,
though it requires that the calibration be
carried out at a multiplicity of doses to provide
a calibration curve
13Dose Range Background Readings and Lower Range
Limit
- The lower limit of the useful dose range may be
imposed by the instrumental background or
zero-dose reading - The is the value of r r0 observed when Dg 0
sometimes it is referred to as spurious
response, since it is not caused by radiation - Examples of r0 include charge readings due to
ion-chamber insulator leakage, and
thermo-luminescence dosimeter readings resulting
from response of the reader to infrared light
emission by the dosimeter heater
14Background Readings (cont.)
- The instrumental background should be subtracted
from any dosimeter reading - The usual procedure for determining this
correction is to make measurements with the same
dosimeter treated in the same way (including
duration of the time) except for the absence of
the applied radiation field - In this way the quantity one measures is r0 plus
the radiation background reading rb
15Background Readings (cont.)
- If a background reading is very reproducible from
run to run, subtracting it from a dosimeter
reading may have little effect on the precision
of the instrument - In many cases, however, the background reading
exhibits significant nonreproducibility - The lower limit of the practical dose range of a
dosimeter is usually estimated to be the dose
necessary to double the instrumental background
reading
16Background Readings (cont.)
- Evaluation of the precision of the measurements
from repeated readings of both the radiation and
the background will of course provide more
quantitative information - If ?? is the S.D. of the average of a group of
radiation readings r, and ??0 is the S.D. of the
average of the background readings r 0, then the
S.D. of the net radiation reading r r0 is given
by - (Note that these are not percentage S.D.s)
17Background Readings (cont.)
- If the background reading is negligibly small,
then the lower dose limit is imposed by the
capability of the dosimeter readout instrument to
provide a readable value of r for the dose to be
measured, Dg - If r is less than 10 of full scale on analogue
instruments, or contains fewer than three
significant figures on digital readouts, the
precision and accuracy may both become
unsatisfactory - A more sensitive scale should then be used
18Dose Range Upper Limit of the Dose Range
- The upper limit of the useful dose range of a
dosimeter may be imposed simply by external
instrumental limitations, such as reading off
scale on the least sensitive range of an
electrometer - Alternatively some kind of inherent limit may be
imposed by the dosimeter itself
19Upper Limit (cont.)
- Causes of this type include
- Exhaustion of the supply of atoms, molecules, or
solid-state entities (traps) being acted upon
by the radiation to produce the reading - Competing reactions by radiation products, for
example in chemical dosimeters - Radiation damage to the dosimeter (e.g.,
discoloration of light-emitting dosimeters, or
damage to electrical insulators)
20Upper Limit (cont.)
- Usually the upper limit of the dose range is
manifested by a decrease in the dose sensitivity
(dr/dDg) to an unacceptable value - It may be reduced to zero, or to a negative
value, as in the following diagram, which causes
the dose-response function to become double-valued
21Illustrating a double-valued dose-response
function resulting from a decrease in the
dosimeter sensitivity at high doses
22Upper Limit (cont.)
- In such a case other information is needed to
decide which dose is represented by a large
r-value, as shown in the figure - It is of course possible in principle to make use
of the negative-slope part of a dose-response
curve such as that in the figure for dosimetry
purposes if it is sufficiently reproducible
23Dose-Rate Range For Integrating Dosimeters
- If a dosimeter is to be used for measuring the
time-integrated dose (not the dose rate), then it
is necessary that its reading not depend on the
rate at which the dose is delivered, at least
within the range of dose rates to be encountered - Usually there will not be any low-dose-rate
limitation except that imposed by the lower dose
limits already discussed
24Integrating Dosimeters (cont.)
- One case of a genuine low-dose-rate limitation is
reciprocity-law failure in photographic film
dosimeters - It occurs only with low-LET radiation (e.g., x
rays or electrons) and is due to the necessity
for several ionizing events to occur in a single
grain of silver bromide to make it developable - Low-LET radiation can only create one ion pair at
a time in a small volume like a AgBr grain in
photographic emulsion, and after a time the ions
can recombine
25Integrating Dosimeters (cont.)
- Thus the grain repairs itself at low enough dose
rates, and never produces a latent image, that
is, reaches a condition of developability - Consequently it never contributes to the opacity
of the film, which is the parameter used to
measure the dose - Biological damage by low-LET radiation exhibits
similar time-repair characteristics
26Integrating Dosimeters (cont.)
- The upper limit of dose-rate independence usually
occurs when charged-particle tracks are created
close enough together in space and time to allow
the ions, electron-hole pairs, or active chemical
products such as free radicals to interact
between tracks - In an ion chamber this is called general or
volume ionic recombination - Similar back reactions also occur in solid or
liquid dosimeters, resulting in a loss of
contribution to the reading r
27Dose-Rate Range For Dose-Rate Meters
- It is desirable in dose-rate-measuring dosimeters
that the reading r be proportional to the dose
rate dDg/dt, or at least to be a single-valued
function of it - Jamming or paralysis of an instrument, causing
it to read zero or a small response at high dose
rates, as can occur in Geiger-Müller counters
when the dead time overlaps and becomes
continuous, is intolerable, especially in
personnel monitoring meters
28Dose-Rate Meters (cont.)
- The upper limit on the usable dose-rate range
more usually takes the form of some kind of
saturation of the reading vs. dose rate, due to
ionic recombination or other results of track
proximity - The counting of two or more events as one when
they occur temporally too close together in
pulse-counting dosimeters also may cause
saturation - Other modes of saturation may also occur in
various kinds of dosimeters
29Dose-Rate Meters (cont.)
- In dose-rate measurements the response time
constant, while not a limitation on the dose-rate
range, is also an important consideration - It is defined as the time it takes for the
reading in a constant field to rise to within 1/e
of its steady-state value, or to decay to 1/e of
that value upon removal from the field - A long time constant will cause a dose-rate meter
to seek a mean reading value in a repetitively
pulsed radiation field
30Stability Before Irradiation
- The characteristics of a dosimeter should be
stable with time until it is used - That includes shelf life and time spent in situ
until irradiated (e.g., worn by personnel if a
health-physics monitoring dosimeter) - Effects of temperature, atmospheric oxygen or
humidity, light, and so on can cause a gradual
change in the dose sensitivity or the
instrumental background - Photographic, chemical, or solid-state dosimeters
are generally more susceptible to these
influences than ion chambers or counters
31Stability After Irradiation
- The latent reading in some types of integrating
dosimeters (e.g., photographic, chemical,
solid-state) may be unstable to some extent,
suffering fading losses during the time
interval between irradiation and readout - Again, harsh ambient conditions of elevated
temperature or humidity, direct sunlight or
bright fluorescent lighting, and so on, may
aggravate this effect
32After Irradiation (cont.)
- If such time-dependent fading losses are
unavoidable, it is advantageous to make them as
reproducible as possible through standardization
of laboratory technique so that a fading
correction can be applied to the readings - The following diagram outlines a protocol for
measuring both the pre- and postirradiation
instabilities of a group of identical dosimeters
33Protocol for measuring pre- and postirradiation
instability effects in integrating dosimeters,
where a common dosimeter preparation time tp is
used
34Energy Dependence Specification
- Generally speaking, the energy dependence of a
dosimeter is the dependence of its reading r, per
unit of the quantity it is supposed to measure,
upon the quantum or kinetic energy of the
radiation, as illustrated in the following
diagram - Pane A shows the reading r obtained from an
imaginary dosimeter vs. some dosimetric quantity
J (such as exposure, absorbed dose in water under
CPE conditions, etc.)
35Illustration of the general concept of energy
dependence
36Energy Dependence (cont.)
- Let us suppose that the calibration curves shown
have been obtained at the three different
radiation energies (or qualities) E1, E2, and E3,
as shown - In this example the dosimeter response is assumed
to be linear at energy E1, but becomes
progressively more nonlinear at E2 and E3 - The corresponding plots of r/J vs. J are shown in
B - The energy-dependence curves for the two J-values
J1 and J2 are given in C, and are seen to differ
in this case for E gt E1
37Energy Dependence (cont.)
- If only a single curve of r vs. J, for instance
the E3 curve in A, were obtained at all energies,
then the dosimeter would be energy-dependent at
all J-levels - For each value of J a horizontal line would
result as in D, producing a family of such
energy-dependent r/J curves for different
J-values - If the single energy-independent calibration
curve were linear, then a common horizontal line
would result in D, providing a single r/J value
that would be applicable to all J-values and all
radiation energies
38Energy Dependence ? Dependence of the Dosimeter
Reading, per Unit of X- or ?-ray Exposure, on the
Mean Quantum Energy or Quality of the Beam, r/X
vs. E
- This usage is commonly found in health-physics
personnel monitoring or any application in which
exposure is commonly referred to - 60Co ?-rays are frequently used as the reference
energy for normalization, producing
energy-dependence curves looking typically like
the following figure for dosimeters made of
materials higher than, equal to, and lower than
air in atomic number
39Typical energy-dependence curves in terms of the
response per unit exposure of x- or ?-rays
40r/X vs. E (cont.)
- The rise in the top curve below about 0.1 MeV is
caused by the onset of photoelectric effect in
the sensitive volume of the dosimeter - The flat maximum usually occurs at about 30-50
keV, below which the curve may slowly descend due
to attenuation in the dosimeter, onset of
photoelectric effect in the reference material
(air), and LET dependence of the dosimeter
41r/X vs. E (cont.)
- The shape of the curves can be estimated by
- where the subscript g refers to the material
in the dosimeters sensitive volume
42r/X vs. E (cont.)
- This equation is based on the assumptions that
- The dosimeters sensitive volume is in
charged-particle equilibrium, and the wall medium
w g - Attenuation is negligible in the dosimeter, both
for incident photons and for fluorescence photons
generated in the dosimeter - A given absorbed dose to the sensitive volume
produces the same reading, irrespective of photon
energy (i.e., the dosimeter is LET-independent)
43r/X vs. E (cont.)
- These assumptions are all questionable and may
require suitable corrections for their failure in
specific cases - Matching the wall medium to the material in the
dosimeters sensitive volume can satisfy
assumption 1 - Substituting (?/?)g for (?tr/?)g in (?en/?)g has
the effect of assuming the local reabsorption of
all fluorescence photons generated in the
sensitive volume, thus providing an upper limit
for the influence of that effect
44r/X vs. E (cont.)
- Attenuation of photons entering the dosimeter can
be simply estimated by the straight-ahead
approximation - Failure of assumption 3 is referred to as LET
dependence of a dosimeter - The total effect of assumptions 2 and 3 may cause
a perfectly air-equivalent dosimeter to decrease
its reading at low photon energies, as indicated
by the dashed curve
45Energy Dependence ? Dependence of the Dosimeter
Reading per Unit of Absorbed Dose in Water on the
Photon or Electron-Beam Energy
- This usage is commonly found in radiotherapy
literature, where absorbed dose always refers
to water (or muscle tissue) unless otherwise
specified - Inasmuch as water and tissue are not identical,
one should say which is meant, but since the
differences are small (1) in the megavolt
region, this choice frequently remains unspecified
46Absorbed Dose in Water (cont.)
- For x rays the equation by which a homogeneous
dosimeters energy dependence can be estimated is - which depends on water as a reference material
and 60Co ? rays for normalization - The following figure illustrates this equation
over the energy range from 1.25 to 50 MeV for LiF
and bone-equivalent dosimeters
47X-ray energy dependence estimated for a LiF and a
bone-equivalent dosimeter, in terms of response
per unit absorbed dose in water, normalized to
60Co ? rays
48Absorbed Dose in Water (cont.)
- Because of the large secondary-electron ranges at
these energies, this equation is only satisfied
to the extent that TCPE is present, g wall w,
and ? is the same in water as in the dosimeter - Also, considerable x-ray attenuation occurs in
the thick walls, and the size of the resulting
dosimeter may be impractical anyway - In radiotherapy dosimetry these problems are
usually avoided by doing the measurements in a
phantom, letting it comprise most of the
dosimeters wall thickness
49Absorbed Dose in Water (cont.)
- For electron beams of kinetic energy T (MeV), the
corresponding equation for estimating energy
dependence in terms of the dose to water,
normalized to T 1 MeV, is
50Absorbed Dose in Water (cont.)
- This approximation assumes that
- CPE exists for ?-rays entering and leaving the
sensitive volume - The incident electrons lose only a very small
fraction of their energy in traversing the
dosimeter - Electron scattering is the same in g as in water
- The reading per unit dose to the dosimeters
sensitive volume remains energy-independent
(LET-independent)
51Absorbed Dose in Water (cont.)
- Items 1 and 3 are suspect, while 2 and 4 are
easily satisfied in the energy region above 1 MeV - The following figure illustrates this equation
for an air-cavity chamber, LiF, and
bone-equivalent dosimeters - Clearly the lack of polarization effect in the
gaseous air relative to water causes a large
energy dependence in that case - Neither LiF nor a bone-equivalent dosimeter shows
much dependence - This illustrates the fact that collision
stopping-power ratios are insensitive to electron
energy unless the polarization effect is involved
52Electron-energy dependence estimated for LiF, a
bone-equivalent dosimeter, and an air-filled ion
chamber, in terms of response per unit absorbed
dose in water, normalized to T 1 MeV
53Energy Dependence ? Dependence of the Dosimeter
Reading per Unit of Absorbed Dose to the Material
in the Sensitive Volume Itself, on the Radiation
Energy or Beam Quality
- This kind of energy dependence is the most
fundamental, inasmuch as it reflects the
dosimeters energy efficiency, i.e., the ability
of the dosimeter to give the same reading for the
same amount of absorbed energy in its own
sensitive volume, regardless of radiation type or
quality
54Dose to Sensitive Volume (cont.)
- It is often called LET dependence because it
usually manifests itself as a change in the
reading per unit dose as a function of
charged-particle track density, due to ionic
recombination or other second-order effects that
depend on the proximity of radiation products to
the dosimeter - For example, ion chambers show such LET
dependence only at radiation energies low enough
(?10 keV) so the value of W for the gas is no
longer constant but begins to rise
55Energy Dependence Modification
- The energy dependence of a dosimeter can be
changed to some extent, especially when the
photoelectric effect is causing an overresponse - In that case a medium-Z (e.g., tin), high-Z
(e.g., lead), or composite filter can be
incorporated into the design of the dosimeter
capsule - The thickness t can be chosen to correct the
overresponse at about 100 keV, using e-?t as a
guide
56Energy Dependence Modification (cont.)
- Having determined the thickness needed to correct
the response at 100 keV, it will be found to have
been overcorrected at lower energies, the reading
being essentially zero at 50 keV - This can be rectified by perforating the filter
using the unfiltered height of the maximum
overresponse as a guide - Experimental testing is of course required to
verify and finally adjust the design
57Energy Dependence Modification (cont.)
- This approach to the modification of energy
dependence adds weight and size, and introduces
directional dependence to the dosimeter reading,
influenced by the geometrical design - A sophisticated example of a design that
minimizes the directional dependence is shown in
the following diagram
58A perforated filter to reduce the photoelectric
overresponse of a dosimeter while retaining
response to photoelectrons below ?80 keV. The
spherical design minimizes directional dependence.
59Miscellany
- The configuration of a dosimeter sometimes is
crucial to its use - It may be necessary to simulate as closely as
possible the geometry of the test object - A thin plastic-film dosimeter might best measure
the dose in a layer of biological cells - Finally, small size of a dosimeter is of primary
importance in its application in vivo in patients
or test animals
60Miscellany (cont.)
- A dosimeter needs a relevant calibration that is
appropriate to the radiation type and quality, as
well as to the quantity to be measured - A calibration in terms of the dose to the
dosimeters own sensitive volume is more
generally applicable than other types of
calibrations
61Miscellany (cont.)
- If different types of radiation coexist in the
field to be measured, attention must be paid to
the relative sensitivity of the dosimeter to the
different components - It may be possible in specific cases to
discriminate against one type of radiation (e.g.,
by attenuation) so that another may be measured
without competition
62Miscellany (cont.)
- There are many dosimeters in the literature that
may be useful but have never become commercially
available - Given a choice, a commercial system is usually
easier to apply, so long as it satisfies other
requirements - Chemical dosimetry must still be done on the
basis of local preparation, primarily because of
shelf-life instability - Calorimetric dosimetry has also eluded commercial
manufacture, so far
63Miscellany (cont.)
- The reusability of a dosimeter has several
important implications - Reusable dosimeters such as TLDs can be
individually calibrated single-use dosimeters
such as film badges cannot - The latter can only be batch-calibrated by
irradiating and measuring representative samples - The precision of the measurements refers to the
reproducibility of readings obtained from
different members of the dosimeter batch after
they have been given identical irradiations
64Miscellany (cont.)
- The advantage of reusability of a dosimeter
depends on how difficult or convenient it is to
restore it to its original condition - If it cannot be fully purged of the effects of an
earlier dose, some of the advantage is lost - A shift in sensitivity, for example, means that
dosimeters must be segregated on the basis of
prior history and recalibrated before reusing - Economies that may be realized through reuse of
dosimeters may thus be limited