Chapter 16 - PowerPoint PPT Presentation

1 / 14
About This Presentation
Title:

Chapter 16

Description:

Infrared used to determine the major functional groups present. ... thermistor or bolometer- change in resistance is measured and related to amount ... – PowerPoint PPT presentation

Number of Views:64
Avg rating:3.0/5.0
Slides: 15
Provided by: maso85
Learn more at: http://mason.gmu.edu
Category:

less

Transcript and Presenter's Notes

Title: Chapter 16


1
Chapter 16 Infrared Spectroscopy
2
Introduction
  • Useful range from about 2.5 mm to 50 mm.
  • Infrared used to determine the major functional
    groups present.
  • Quantitative measurements possible but subject to
    large amount of error.
  • Atoms or groups of atoms in molecules are in
    continuous motion with different modes of
    vibration relative to each other.
  • Absorption of radiation changes amplitude of
    vibration but not frequency.

3
Vibrational Modes
  • Increased amplitude produces a change in dipole
    moment.
  • ? qr. where ? dipole moment, q change
    displacement, and r displacement from
    equilibrium.
  • Only vibrations that cause this change in
    electric dipole moment will be associated with an
    absorption of infrared radiation.
  • E.g. Symmetric and antisymmetric modes of
    vibration are possible with CO2
  • symmetric mode of vibration has no net change in
    its dipole moment, while antisymmetic mode has
    net change in dipole moment. The antisymmetric
    mode would be infrared active and the symmetric
    mode would not.

4
VIBRATIONAL MODELS
  • Mechanical model spring connects one or two
    moving bodies. Restoring force, F, pulling on
    atoms to return to initial positions.
  • Force related to force constant stiffness of
    spring (bond).
  • F ?ky. Negative sign means restoring force.
    (Hooks law)
  • dE ?Fdy or upon integrating between equilibrium
    position and y gives E ½ky2.
  • Potential energy curve parabolic maximum when
    spring is stretched or compressed and minimum at
    equilibrium position.

5
Vibrational frequency
  • F ma where a acceleration and m mass of the
    substance moving.
  • Acceleration is written as
  • Solution y A cos 2pumt where
  • .um vibrational frequency and
  • A maximum amplitude of the motion.
  • Leads to
  • Two moving particles frequency uses reduced mass
    .

6
Quantum Mechanical Vibrations
  • Harmonic oscillator is used to obtain wave
    equation for potential energy of oscillator.
  • where v vibrational quantum ( integer).
  • Quantum Mechanical Model of Atomic Movement only
    certain energies are allowed.
  • near equilibrium position, molecular vibrations
    similar to mechanical model..
  • Frequency of the vibration from the mechanical
    model as a reasonable estimate of true
    vibrational frequency .
  • But quantum mechanical energy is .
  • Quantum mechanical selection rule which states
    that Dv 1 (due to conservation of momentum of
    the combined photon and molecular system).
  • and
    where
  • k force constant (N/m) and m reduced mass
    (kg).

7
GROUP FREQUENCIES
  • Estimation of frequencies of vibration for
    various groups possible when force constant
    known.
  • E.g.1 force constant of CO bond is 1.23x103 N/m,
    determine vibrational frequency of this CO
    group.
  • mO
  • reduced mass of
  • Substituting
  • Units

8
Sample Problem
  • E.g.2 C-H stretch of alkane occurs at ? 2900
    cm?1 determine frequency of deuterated analog
    using mechanical equation
  • ratio of two equations for two forms of compound
  • or
  • reduced mass for each of two bonds will be mC
    1.99x10?26 kg,
  • mD 3.32x10?27 kg, mH 1.66x10?27 kg and
  • .mC-H 1.53x10?27 kg and mC-D 2.84x10?27 kg
  • Substituting
  • Deuterating convenient way to confirm presence
    of particular type of bond, since frequency shift
    is relatively large and predictable.

9
INSTRUMENTATION
  • Instrumentation is same as for other absorption
    instruments except that the sources, detector and
    the optical material are designed specifically
    for this spectral region.
  • Sources We have mentioned these detectors in
    our discussion of general features of the
    absorption spectrophotometer.
  • Nernst Glower cylinder composed of rare-earth
    oxides which is heated to some temperature before
    current can be passed directly through it.
  • Since it has a negative temperature coefficient,
    it is necessary to regulate the current (T kept
    at about 1800K)..
  • Globar SiC rod (T kept at about 1600K) kept at
    desired temperature by passing current through
    it positive temperature coefficient..
  • Both of these sources suffer from having low
    intensities ( 10?7 - 10?9 W) and has led
    authors to claim that source is energy limited.

10
Detectors
  • Low flux of photons and low energy of infrared
    radiation make it more difficult to detect.
  • photon detector- based upon photoconductive
    effect that occurs in certain semiconductor
    materials. Absorption of light causes resistance
    to decrease. Voltage drop across load resistor
    measured HgCdTe must be cooled to 77K and PbS2
    can be operated at room temperature.
  • Photon detectors (also called quantum detectors)
    have rapid response and thus are used with FTIR .
  • Material deposited on surface of a nonconducting
    material and is sealed in an evacuated tube to
    protect the semiconductor from reaction with the
    atmosphere.

11
Thermal Detectors
  • Absorption of IR radiation produces a heating
    effect which alters a physical property of the
    detector. Thermal detectors are usable over a
    wide wavelength range.
  • Thermocouple or therompile.Instrumental Methods
    of Analysis, Willard, Merritt, Dean Settle, p.
    193.thermocouples-piece of blackened foil placed
    over 2 wires made of dissimilar metals.
  • detector is placed inside vacuum to improve its
    response (minimize heat loss due to conduction).
  • Several thermocouples in series thermopile has
    higher sensitivity since voltages of individual
    thermocouples are additive.
  • Temperature differences of about 10?6 K can be
    determined this way.
  • Thermocouple is low impedance circuit so that
    high impedance preamplifier is necessary to avoid
    signal modification by the amplifier circuitry.
    The response time of these detectors is on order
    of 100 ms. 2s
  • Null Point Wheatsone Bridg
  • thermistor or bolometer- change in resistance is
    measured and related to amount of photons hitting
    detector.
  • Resistance measured with Wheatstone bridge
  • Radiation falls on R1R2 adjusted until circuit
    balanced (no voltage drop between C and D. Then
    R1/R2 R3/R4
  • Knowledge of all other resistances makes it
    possible to determine R1

12
INSTRUMENTS
  • Dispersive instruments have been the traditional
    instrument design for IRs high resolution is
    possible, FT-IR is essentially the same in modern
    instruments. The FT-IR has a much higher
    sensitivity which is very important due to the
    low signal levels in the IR region. This extra
    sensitivity makes it possible to use the FT-IR
    for quantitative work.
  • Sample Handling Solids, liquids and gases can be
    analyzed with IR.
  • Gases are generally are often constructed with a
    cylindrical glass body and are usually about 10
    cm in length.
  • The pressure can be from a few mm Hg up to
    several atm. depending upon the absorption
    characteristics of sample.
  • Liquids and solutions
  • Only a short pathlength (1mm) is required for
    liquids.
  • Some solvents absorb in IR region interfering
    with signal from sample.
  • Pathlength of thin IR sample holder can be
    determined by observation of the interference
    patterns associated with constructive
    interference from reflection of light from the
    two internal surfaces of cell .
  • E.g. Determine the pathlength from the fringes
    below

13
FT-IR
  • Michaelson Interferometer.
  • Laser-fringe reference system provides
    sampling-interval. Signal averaging can only be
    accomplished if the positioins of the mirror are
    precisely know. This is achieved by using a
    helium-neon laser as a reference. Radiation at
    exactly 632.8 nm traverses the same optical path
    as the IR beam.
  • A separate detector measures the interferogram
    produced, giving a sinusoidal signal with maxima
    separated by the laser frequency at 15,803 cm-1.
  • This signal is used to trigger the sampling of
    the IR signal very reproducibly
  • IR Source
  • Detector

14
Quantitative Analysis
  • IR has traditionally been used for qualitative
    analysis.
  • Difficult to use quantitatively due to chemical
    or instrumental effects.
  • Large sloping background often interferes with
    normal spectrum.
  • The base line method corrects involves selection
    of absorption band of the substance under
    analysis which is sufficiently separated from
    other matrix peaks and corrected as shown below.
  • Mixtures can be determined by same methods
    described earlier. Need to set up the correct
    number of simultaneous equations.
  • Convenient for measuring concentrations of gases
Write a Comment
User Comments (0)
About PowerShow.com