Title: PROGRAMMING
1PROGRAMMING
2MAPPING
- Individual patients have different perceptions
resulting from electrical stimulation to the
cochlea - ie the strength of the stimulation required to
elicit auditory perceptions of the same loudness
is different from patient to patient, and from
channel to channel for the same patient - Patients differ in their ability to perceive
pitch changes resulting from stimulation of
different channels
3MAPPING
- The CI can be programmed to fit the
psychophysical requirements of individual implant
recipients - The combination of all these individualised
parameters is called a MAP - MAPping is a PROCESS by which the optimal
programming parameters are determined for each
individual patient
4Speech Processing Strategies
5SPEAK
- Continuously analyzes the energy in the incoming
acoustic signal and determines the spectral
maxima. - Spectral maxima frequency regions in the
waveform that contain the greatest amounts of
energy. - Selection of only the highest amplitude spectral
maxima increases the probability that speech
rather than noise (static) is presented.
6SPEAK
Analysis bandwidth 116 to 7787 Hz Select the
spectral maxima Sequentially stimulate the
channels representing those frequency bands
7SPEAK
- Amount of current applied to each channel is
within the dynamic range (the difference between
C-level and T-level). - Acoustic input varies continuously over time, so
the spectral maxima will change and the channels
selected for stimulation will vary across the
array. - Average no. of maxima 6 8 (max 10)
- Actual no. depends on signal spectral
composition, signal intensity and the MAP.
8SPEAK
- The channels selected for stimulation represent
the largest spectral peaks. - They are stimulated in tonotopic order from high
to low frequency. - Stimulation rate 250 pps (overall rate 2500
pps) - For broadband signals more maxima, slower
stimulation rate - For sounds with limited spectral content less
maxima, higher stimulation rate
9SPEAK
10SPEAK
- SPEAK in SPrint
- Stimulation rate
- 250 Hz per channel (fixed)
- Not affected by high T/C levels
- Not affected by wide P.W. (lt 200 usec)
- SPEAK in ESPrit
- Selects 8 maxima (default)
- Max. PW 37 usec
- Requires at least 12 channels
- Alter apical gains if sound different to SPRint
using the gain shaper
11CIS
- Continuous interleaved sampling
- Uses smaller no. of channels at a higher
stimulation rate per channel to represent the
temporal changes. - Channel selection is fixed.
- You may choose to use 4, 6, 8 or 12 channels.
- Due to the power demands of this strategy, it was
used only with the SPRint speech processor. It
is currently available in ESPrit using lower
stimulation rate. - Monopolar modes only
12CIS
- You choose the no. of electrodes being used.
- A bandwidth is assigned automatically to each
one, following the tonotopic order of the
cochlea. - During stimulation, the output amplitudes from
each band are measured and represented as
variations in the amount of current sent to the
channels. - The current delivered is directly proportional to
the energy in the band and is within the dynamic
range. - Permissible stimulation rates per channel
- 900, 1200, 1800 and 2400 Hz
- (max overall rate lt 14400 Hz)
13CIS
14CIS
15CIS Secondary Parameters
- Jitter
- Recommended 10-20 jitter.
- May reduce pinging or bursty in onset at
higher rates and reduce buzz percepts at lower
rates. - Psychophysics
- Must be re-measured for different stimulation
rates. - Increasing stimulation rate will reduce T-level,
but C-levels less affected. - Increasing number of channels may require a
global reduction in C-levels.
16ACE
- Advanced Combination Encoder
- ACE SPEAK CIS
- Encodes important spectral and temporal speech
cues - Flexibility to meet individual needs
- No. of maxima up to 20
- Stimulation rate per channel 250, 500, 720,
900, 1200, 1800 or 2400 (overall up to 14400 Hz) - 6 maxima stimulation rate per channel 2.4 kHz
- 20 maxima stimulation rate per channel 720 Hz
- Monopolar modes only
17ACE
18COMPARISON OF STRATEGIES
19Stimulation Modes
20STIMULATION MODES
Electrical stimulation produces current flow
between an active and an indifferent
electrode Stimulation Mode describes the
location of the indifferent electrode relative to
the active electrode The physical separation of
the active and reference electrodes determines
the spread of electrical current, hence
controlling the area over which nerve endings are
stimulated.
21STIMULATION MODES
22Monopolar (MP) Stimulation
- Active electrode is intracochlear but the
indifferent electrode(s) is extracochlear. - MP1 current flows between designated active
intracochlear electrode and the ball electrode
(R1) - MP2 current flows between the active
intracochlear electrode and the plate electrode
(R2) - MP12 current flows between the active
intracochlear electrode and both extracochlear
electrodes (R1 and R2)
23Monopolar (MP) Stimulation
- Why MP modes?
- Wide spread current flow
- Much lower current levels for T and C-levels.
- Lower energy consumption to produce auditory
sensations - Longer battery life
- Auditory nerve can be stimulated at a higher
rates (good for both CIS and ACE)
24Bipolar (BP) Stimulation
- Current flows between two intracochlear
electrodes - Active indifference pair
- Examples
- Active Indifferent Total
no. of channels - BP 13 14 21
- BP1 13 15 20
- BP2 13 16 19
25Bipolar (BP) Stimulation
- The physical separation of these electrodes
determines the spread of current and the area
over which the spiral ganglion cells are
stimulated. - If T and C-levels are too high, a wider bipolar
mode should be selected. - As the bipolar configuration becomes wider, the
number of available pairs or channels for
stimulation decreases.
26Common Ground (CG) Stimulation
- Current flows between the designated active
electrode and all other electrodes on the array
inside the cochlea. - Allows all 22 electrodes to be used as active.
- T and C-levels are typically lower than in BP
mode. - It is used diagnostically to detect electrode
anomalies.
27Setting Stimulation Mode
- Try MP12 mode first.
- It provides the greatest flexibility.
- BP modes can use only the SPEAK strategy.
- CG modes can use only the SPEAK strategy.
- CG should not be used in cases of partial
insertions of electrode array because there may
be a higher probability of non-auditory
sensations. - Whenever you choose a different mode, T and
C-levels must be re-measured in the new mode
because the amount of current necessary to reach
T and C-levels will be different in each
stimulation mode.
28Frequency-to-Channel Allocation
29Frequency-to-Channel Allocation
- Electrodes are numbered sequentially.
- Tonotopic organization of the cochlea
- Electrode 1 Basal Highest pitch percept
- Electrode 22 Apical Lowest pitch percept
- Frequency-to-channel allocation depends on
- Speech processing strategy
- No. of channels available for stimulation in the
MAP
30SPEAK
- Maximum no. of channels 20
- 16 frequency-to-channel allocation tables
- Table 1-6 divide overall bandwidth into 20
channels - Table 7-14 divide overall bandwidth into fewer
channels when less than 18 active channels
are available - Default Table 6 (widest frequency range)
- 116 Hz to 7871 Hz
- There is less frequency selectivity in MAPs with
limited numbers of channels.
31CIS and ACE
- Fixed default frequency range 187 7937 Hz
- 27 frequency tables
- Automatically allocate a frequency table
according to no. of channels (412) available. - If necessary, the high frequency cut off point
can be reduced by lowering the frequency table
number.
32Electrical Stimulation Levels
33Electrical Stimulation
- Charge-balanced biphasic current pulses
Pulse Width
Stimulus Amplitude
0 mA
Increasing the amplitude of the pulse and/or
widening the pulse width results in more
electrical charge being delivered
34Current Level and Pulse Width
- Current Level is expressed in arbitrary units.
- Current Level 1 (10 mA) - 255 (1750 mA)
- Default pulse width (PW) 25 ms/phase
- Narrow pulses take less time to deliver
- Therefore, higher rates.
- Pulse width 25 400 ms/phase
- Maximum pulse width depends on the coding
strategy, the stimulation mode, the stimulation
rate and the number of maxima or channels
selected. - Dont widen the pulse width unless the user does
not receive sufficient loudness when using a
narrow pulse at maximum current amplitude.
35Stimulus Amplitude-to-Current Level MAPping
36Automatic Gain Control
- To control the gain so peaks in the acoustic
signals are limited to values that translate into
stimulation at C-level. - It rapidly decreases the gain so that the peak
value of the signal is set to a fixed value. - AGC slowly increases the gain after the
high-level peak has passed.
37Microphone Sensitivity Control
- It determines the minimum input signal level
required for stimulation. - Higher sensitivity settings lower SPL is
required to cause stimulation. - Lower sensitivity settings louder sound is
required to cause stimulation. - For normal listening situations,
- Sensitivity 8 on Sprint
- Sensitivity 3 on ESPrit
38Microphone Sensitivity Control
- If microphone sensitivity is always set above or
below the optimal settings - Overall C-levels in MAP are not set appropriately
- If sensitivity is set too low
- Gain applied to incoming signal is reduced
- Soft sounds will not be processed and cannot be
heard. - If sensitivity is set too high
- Gain applied to low-level inputs is increased
- Low level noise will be heard and signal-to-noise
level will decrease.
39Base Level
- Default 4
- Range 4 15 in SPEAK
- 0 15 in ACE and CIS
- Increasing the Base Level
- Increases the level of sound required to initiate
electrical stimulation, and then reduces the
background noise. - Reduces the dynamic range
- Dont raise the Base Level unless it is
necessary because it reduces the operating range
of the processor.
40Q-value
- Controls the steepness of the amplitude growth
- Determines the percentage of the electrical
dynamic range that is allocated to the top 10 dB
of the speech processors channel amplitude
range. - The lower the Q-value, the steeper the amplitude
growth function at the lower end. - Default 20
- 80 of electrical dynamic range is MAPped into
the top 10 dB of the processors range. - The lower 20 of the individuals dynamic range
is MAPped into the lower 20 dB of the processors
range.
41Q-value
- Lowering Q-value
- Making soft sounds, including background noise,
seem louder. - Raising Q-value
- Soft voices may not be heard. May reduce a
persons overall speech understanding ability. - 20 50 for SPEAK
- 10 50 for ACE and CIS
42Autosensitivity Control (ASC)
- Reduce background noise.
- Looks at the troughs (noise floor) in the
spectral enveolope. - Gain decreases slowly
43Channel Gains
- To alter the perceived sound quality of the MAP
- Default setting 0 dB across all channels
- Gain is applied to each channel output and
affects the stimulation levels sent to the
implant.
44How do we modify the loudness percept?
45How to modify the loudness percept?
- Use the T/C-level modifiers to increase/decrease
C-levels for all channels by a fixed percentage
of the dynamic range during live-voice testing of
the MAP. - Use the T/C-level modifiers to increase/decrease
T-levels for all channels by a fixed percentage
of the dynamic range, making the low-level
signals louder or softer. (Note You lower
T-levels only when you think T-levels has
over-estimated.) - Lowing Q-value results in a louder perception of
low-level signals, including background noise.
46How to modify the loudness percept?
- Raising the Base Level increases the minimum
acoustic signal that results in stimulation. - Use the volume control on the speech processor.
- 0 9 on SPRint
- 0 5 on ESPrit
- The volume control can only reduce the C-levels
(total reduction is 25)
47Magnet
48Magnet
- The magnet strength can be adjusted by turning
the magnet insert in the external coil. - Magnet strength should be assessed and if
necessary adjusted at first fitting. - This is critical especially for children
- Magnet inserts strength ½, 1, 2, 3, 4 are
available - Magnet strength required will depend on thickness
of skin flap, hair, swelling, fatty tissue etc
49Magnet
- Magnetic pull of magnet in headset should be as
loose as possible but still firm enough to stay
in place with movement of the head ie not to fall
off head
50Magnet
- Skin Flap Thickness
- Receiver-stimulator will operate for separation
of receiving and transmitting coils up to 6 to
9mm - If flap is thicker than 9mm
- the strongest magnet may not be strong enough
- the internal and external coil coupling will not
be as effective as it should be - the signal can be degraded
51Telemetry
52Objective Measurements
- electrode impedance
- compliance voltage
- neural response telemetry (NRT)
53Telemetry - purpose/benefits
- Impedance Telemetry
- Compliance Telemetry
- NRT Telemetry
- Measurement of electrode impedance and for
integrity testing (Implant Test) - Detection of out-of-compliance conditions which
result in abnormal loudness growth - Measurement of evoked auditory (neural) response
54Electrode Impedance
- electrode impedance is a measure of the
opposition to electrical current flow through the
lead wires, electrodes and tissue - electrode impedance calculated by dividing the
voltage at the electrode by the current flow
through the electrode. - impedance voltage/current
55Electrical Impedance
- can be used to detect conditions under which an
electrode should not be used -
- HIGH IMPEDANCE either the electrode lead wire
is broken and open circuited, or the electrode is
not in contact with body fluids - SHORT CIRCUIT low impedance (in common ground
mode) indicates that the electrode, or its lead
wire, is in contact with another electrode or
lead wire
56Compliance Telemetry
- compliance is the ability of the implant to
deliver sufficient voltage to generate the
desired current level - out-of-compliance refers to the condition where
the maximum voltage available from the implant is
not sufficient to generate the desired current
level
57Compliance Telemetry
- if implant is out-of-compliance
- the actual current delivered is less than the
current level requested - the patient does not perceive loudness growth
when current level is increased - the MAP will not provide appropriate loudness
information
58Compliance Telemetry
- causes of out-of-compliance
- high electrode impedance
- high current level
- WinDPS will provide warning message if implant is
out-of-compliance during T/C-level measurements - resolving out-of-compliance problems
- goal is to reduce T/C levels
- use wide stimulation mode eg monopolar
- increase pulse width
59Neural Reverse Telemetry NRT
- The stimulation of the implant causes nerve
fibers in the spiral ganglion to fire, which
generates an action potential.This is the NEURAL
RESPONSE. - The implant records the action potential by
amplifying signals from intracochlear electrodes.
The action potential is encoded and transmitted
back to the speech processor by radio frequency.
This is the TELEMETRY.