Title: Electron detectors for HAPPEX2 and He
1Electron detectors for HAPPEX2 and He
4
- Installation in HRS
- Linearity tests
2Detectors
3Installation
Installation procedure to be tested next week in
the Hall
Coordinate with installation of scanners
4Linearity and Asymmetry
- Output signal of the PMT
- Asymmetry measured
The nonlinear term is proportional to the signal
amplitude.
- HAPPEX requirement
5Mode of operation of the tube
- Integrating mode
- High beam intensity (80 100 mA)
- Rate in HRS t 50 MHz in the whole detector
- Low gain region
- Continuous signal less than 6 mA to match ADC
range - Pulse mode
- - Background study and Q2 measurement
- Low beam intensity (factor 200 less counting
rate to fire VDCs) - Rate in HRS t 10 KHz/cm2
- High gain region
- pulse signal of a few mA
- need large dynamic range and good
linearity capability -
33 ms
6PULSE mode linearity setup
Method
Blue LED (430 nm)
-HV
Att dB
Att dB
8854 BURLE PMT
Deviation from linearity
ADC
Remote Filter wheels
Pulse generator
La
Lf
Normalization (1st order)
S(La)
S(Lf)
Ped
Light flux
PMT response
7PULSE mode linearity
Active base
Resistive base
2
2
Operating range
250
250
Nge-
Nge-
- Non linearity at high gain due to saturation in
latter dynodes - operating range is in a good
linearity region
8Linearity in continuous mode setup
Shutter 1
Method Use of additivity
Shutter 2
-HV
Micro Amperemetre
DC voltage
Plans to implement this method online for the
HAPPEX-2 run
9Linearity in continuous mode
Resistive base
Active base
ADC range
2
Results for average continuous current Is the
pulse structure still relevant ? Evaluation of
the real output current is done with a duty
factor of 2 or 3
10Gain drift
Measurement of Drift in gain with a pulse signal
and a load of continuous light background of 1,
5, 15 mA
Load ON
Load OFF
Gain do depends on output current but timescale
very long. no systematics from rate-gain
correlation at 30 Hz
11conclusion
- Complete tests of linearity for the BURLE 8854
base - Resistive base H gain G
1E6-3E6 output 2-8 mA L gain G
5E3-1E4 output 10-20 mA filters
(0.5,0.2,0.1) - Active base H gain G
1E6-1E7 output 2-20 mA L gain G
1E4-2E4 output 20-40 mA filter (0.1)