Title: Diagnostic Radiology II


1
Diagnostic Radiology II
  • X-ray Tubes

2
Anode angle
  • Anode angle defined as the angle of the target
    surface with respect to the central ray in the
    x-ray field
  • Anode angles in diagnostic x-ray tubes range from
    7 to 20 degrees, with 12- to 15-degree angles
    most common

3
Focal spot size
  • Actual focal spot size is the area on the anode
    that is struck by electrons
  • Primarily determined by length of cathode
    filament and width of focusing cup slot
  • Effective focal spot size is the length and width
    of the focal spot projected down the central ray
    in the x-ray field

4
(No Transcript)
5
Focal spot size (cont.)
  • Effective focal spot width is equal to the actual
    focal spot width
  • Effective focal length actual focal length ?
    sin ?
  • Foreshortening of the focal spot length, as
    viewed down the central ray, is called the line
    focus principle

6
(No Transcript)
7
Anode angles
  • Optimal anode angle depends on the clinical
    imaging application
  • Small anode angle desirable for small
    field-of-view image receptors (cineangiographic
    and neuroangiographic equipment, where field
    coverage is limited by the image intensifier
    diameter)
  • Large anode angles necessary for general
    radiographic work to achieve large field area
    coverage at short focal spot-to-image distances

8
Variation with position
  • Effective focal spot length varies with the
    position in the image plane, in the anode-cathode
    direction
  • In the width dimension, the focal spot size does
    not change appreciably with position in the image
    plane
  • Nominal size specified at the central ray of the
    beam

9
(No Transcript)
10
Measuring focal spot size
  • Pinhole camera
  • Slit camera
  • Star pattern
  • Resolution bar pattern

11
Pinhole camera
  • Very small circular aperture (10 to 30 ?m
    diameter) in a disk made of a thin, highly
    attenuating metal such as lead, tungsten, or gold
  • Image of focal spot recorded with the pinhole
    camera positioned on the central axis between the
    x-ray source and the detector

12
(No Transcript)
13
Slit camera
  • The slit camera consists of a plate made of a
    highly attenuating metal (often tungsten) with a
    thin slit (typically 10 ?m wide)
  • Measuring the width of the distribution on the
    image yields one dimension of the focal spot
  • Second radiograph, taken with the slit
    perpendicular to the first, yields the other
    dimension of the focal spot

14
(No Transcript)
15
Star pattern
  • Star pattern test tool contains a radial pattern
    of lead spokes of diminishing width and spacing
    on a thin plastic disk
  • Imaging the star pattern at a known magnification
    and measuring the distance between the outermost
    blur patterns on the image provides an estimate
    of the resolving power of the focal spot in the
    directions perpendicular to and parallel to the
    anode-cathode axis

16
(No Transcript)
17
Resolution bar pattern
  • Similar to the star pattern tool, but containing
    radioopaque bars of varying widths and spacings
  • Bar pattern images demonstrate the effective
    resolution parallel and perpendicular to the
    anode-cathode axis for a given magnification
    geometry

18
(No Transcript)
19
Heel effect
  • The heel effect refers to a reduction in the
    x-ray beam intensity toward the anode side of the
    x-ray field
  • For a given field size, the heel effect is less
    prominent with a longer source-to-image distance
    (SID)
  • X-ray tube best positioned with the cathode over
    the thicker parts of the patient to balance the
    transmitted x-ray photons incident on the image
    receptor

20
(No Transcript)
21
Tube housing
  • X-ray tube housing supports, insulates and
    protects the x-ray tube insert
  • Oil in the housing provides heat conduction and
    electrical insulation
  • Lead shielding inside the housing attenuates the
    x-rays that are emitted in all directions
  • Regulations limit the maximum leakage exposure
    rate (0.1 _at_ 1m of the exposure rate at the same
    distance on the central axis of the useful beam)

22
(No Transcript)
23
Grid-biased tubes
  • Focusing cup is electrically insulated from the
    cathode filament
  • Approximately 2,000 V applied to focusing cup
    shuts off the tube current turning off the grid
    bias allows tube current to flow and x-rays to be
    produced
  • Used in applications such as pulsed fluoroscopy
    and cineangiocardiography

24
Filtration
  • In general diagnostic radiology, added filters
    attenuate the low-energy x-rays in the spectrum
    that have virtually no chance of penetrating the
    patient and reaching the x-ray detector
  • Patient radiation dose is reduced
  • Aluminum is the most commonly added filter
    material
  • Regulations prescribe minimum filtration (e.g.,
    2.5 mm Al for machines designed to operate at
    potentials above 70 kVp)

25
Compensators
  • Compensators are added filtration with a shape
    intended to change the spatial pattern of the
    x-ray intensity incident on the patient, so as to
    deliver a more uniform x-ray exposure to the
    detector
  • Placed close to the x-ray tube port or just
    external to the collimator assembly

26
Collimators
  • Collimators adjust the size and shape of the
    x-ray field emerging from the tube port
  • Adjustable parallel-opposed lead shutters define
    the x-ray field
  • Positive beam limitation collimators
    automatically limit the field size to the useful
    area of the detector

27
(No Transcript)