Title: Diagnostics for intense ecooled ion beams
1Diagnostics for intense e-cooled ion beams
by Vsevolod Kamerdzhiev Forschungszentrum Jülich,
IKP, COSY
ICFA-HB2004, Bensheim, October 19, 2004
2Content
- Objects of diagnostics
- What is an electron-cooled ion beam from the
diagnostics point of view? - Parameters to be measured and corresponding
diagnostic methods. - Diagnostics for electron-cooled beams,
difficulties and advantages.
3Objects of diagnostics
- Electron beam
- High beam power
- Beam pipe is inside the solenoid
- Electron-cooled ion beam
- Intensities differ in orders of magnitude
- High beam density
- Small transverse dimensions
- Small momentum spread
4Measured parameters (e-beam)
- E-beam position
- Space charge field of the e-beam
- Current
- Temperature
- Neutralization
5Measured parameters (ions)
- Beam current
- Position along the orbit
- Momentum
- Momentum spread
- Profile
- Emittance
- Tune
- BTF
6Interceptive methods or not?
- Interceptive methods
- Not suitable for a circulating beam (operation)
- Any probe will melt when inserted in the dc
electron beam - Not interceptive methods
- Often indirect measurements
- Suitable for (high current) rings
7Cooler Synchrotron COSY
- COSY accelerates (polarized) protons and
deuterons between 300 and 3700 MeV/c for p 535 to
3700 MeV/c for d - Kicker extraction, stochastic extraction
- 4 internal and 3 external experimental areas
- Electron cooling at low energy
- Stochastic cooling at high energies
8COSY-Cooler
9COSY-Cooler
Electron energy Up to 100 kV Electron current
0.2 - 3 A Operating at 24,5kV 100-250 mA
10LEPTA
Septum
e trap
Collector
e-gun
e source
Quadrupole
Cooling section
Detector
B?
11LEPTA
Electron gun
12Parameters of e-cooled ion beam
Longitudinal Schottky spectra, uncooled and
cooled proton beam
- Small transverse size/emittance
- High density
- Small momentum spread
- During e-cooling the ion beam is dc
- Often unstable
13Diagnostics in the cooler section
- Pick-ups (at least two) are needed inside the
cooling section to measure the position of both
beams. - To measure the position of the e-beam
longitudinal modulation must be applied - Large dynamic range of preamplifiers (variable
gain) - Difficulties in mechanical design, bad service
possibilities (COSY, LEPTA)
14COSY BPMs
15 Diagnostics in the cooler section
- Count rate of the particles recombinating in the
cooler section can be used to find optimum
alignment of the electron and ion beams and for
fine tuning the energy of the electron beam. - Measurement of the profile of recombination
particles (e.g. MWPC) is the easiest way to
determine the ion beam profile (only during
cooling process)
16Example of H0-profile measurement at COSY
Emittance ?m rad
Calculated from the measured H0- Profiles
Beam radius mm
Proton beam current
horizontal vertical
17Diagnostics in the cooler section
- Looking at the ? signals of the pick-up located
in the cooling section in frequency domain gives
useful information about residual gas ions
oscillating in the cooler section. - Such a pick-up can be used also as a clearing
electrode (experience at COSY, I.Meshkov,
A.Sidorin). Applying ac-voltage to the clearing
electrodes makes it possible to kick out the
trapped ions, provided the frequency corresponds
to resonant the frequency of a particular ion
species.
18Space charge field
- To measure the space charge parabola of the
electron beam a low intensity cold ion beam can
be used. - In this case the ion beam is used as a probe
which scans the e-beam. - Procedure
- Inject ion beam in the machine, cool it, measure
the revolution frequency of the ion beam, make a
parallel shift of the e-beam using the cooler
magnetic system, measure the frev again, repeat
the procedure several times shifting the e-beam
in both directions from the initial position.
19Temperature of the e-beam
- Longitudinal temperature can be derived from the
Schottky spectrum of the cooled ion beam - Beam heating effects should be taken into
account - Transverse temperature can be measured by the
pepper pot method - Only in the pulsed mode
- Requires complex mechanical design
20The idea of T?-measurement
The electrons move in the longitudinal magnetic
field. Method based on the measurement of
transverse Larmor radius Pulse duration 20-50
?s
The optical analysis of the electron beam
temperature, V. Golubev et all., Proceedings of
the Workshop on Beam Cooling and Related Topics,
1993.
21Profile of the ion beam
- Can be based on
- Ionization of residual gas
- Laser induced luminescence
- Laser induced photo-neutralization
- Light radiation of residual gas, exited by the
beam particles - Wire scanner
22Ionization profile monitor
If collecting the electrons additional magnetic
field is required. Position sensitive detectors
are usually based on the MCPs. For dense beams
MCP life time is a crucial issue.
IPMs are installed in TSR, SPS, COSY,RHIC
23IPM at COSY
24Beam Profiles measured in COSY
Profile measurement
Electron cooled proton beam
The proton beam is not cooled
1,3109 particles in the ring, 45 MeV.
25Experience with IPM at COSY
- For the WS anode high amplification factor is
necessary - Use of two MCPs in chevron geometry
- High electron density in the second MCP
- Short life time of the MCPs
- Limitations on beam current
- Protection screen is installed
- Triggering of the MCP power supply is applied
- Using an MCP with a phosphor screen is probably
the best way to build a position sensitive
detector for IPM
26Laser profile monitor
- Laser induced luminescence (for ions) in
connection with laser cooling (ASTRID) - Watching the light using a camera
- Photo-neutralization for H- beam (LANL, BNL,
ORNL) - A tightly focused laser beam is directed
transversely through the beam, causing
photo-neutralization. - Scanning the ion beam with the laser and
simultaneously measure the beam current
27PM based on light radiation of residual gas,
exited by the beam particles
28Spectral analysis of the beam signals
- ?-signals of a pick-up in frequency domain give a
lot of information - Exiting the beam und measuring the betatron
frequencies gives the tune - Stability information can be obtained using the
Beam Transfer Function (BTF) method - Electron cooling improves S/N ratio
29Example of the beam spectrum at COSY
Vertical delta signal
30Vertical BTF
31Transverse stability diagram
32Longitudinal BTF at COSY
For different proton beam currents 0,8 mA 2,7
mA 4,5 mA
33Transverse BTF at COSY
Beam current 3,2 mA 2,5 mA
34Summary
- Electron cooling gives much better S/N ratios
- Schottky diagnostics is a very powerful method
- Schottky spectra of an e-cooled ion beam might
be strongly distorted - BTF, longitudinal and transverse
- Online BTF measurement should be further
developed - Profiles of an e-cooled ion beam are difficult to
measure - Better resolution is needed
- Life time of MCPs
- For new machines diagnostic must be planed
together with the machine design
35