Title: GAS ANALYSIS OF NITRITE LABORATORY DISCUSSION
1GAS ANALYSIS OF NITRITE LABORATORY
DISCUSSION NO2-(aq) H (aq) NH2SO3- (aq) ?
N2(g) HSO4- (aq) H2O(l) Ptot P of N2 P
of H2O SCHEDULE WEEK 1 (Learn
technique) Experiment set-up (leak tests,
determining amount of sample) First runs with
unknown nitrite sample Estimate precision of
results (5 successive runs agree ?V(N2)/msmpl
1) WEEK 2 (Apply technique) Do runs until
precision achieved. Boil 1L of deionized water
for 10 minutes, allow to cool without agitation,
then transfer to 1 L bottle. CO2 free water for
use in preparing NaOH solution for next
experiment. Report due next day.
2- SAMPLE PRECAUTIONS
- Unknown from stock desiccator
- Transfer to transport desiccator
- Write down sample number
- Weigh only 1 sample at a time
- Use gelatin capsules
- Return unknown vials to stock desiccator at end
of period - Estimate size of unknown
- 70 NaNO2
- 730 torr, 295K, 35 mL
3Keys to Procedure!!!
- Before connecting the Erlenmeyer with the sample,
raise leveling bulb until levels equalize with
the buret reading near 0 mL and only 2-3 cm of
water showing in the leveling bulb. - Connect the Erlenmeyer carefully but firmly.
- Adjust buret and bulb water levels until they are
the same. Read buret volume. - Measure the temperature of the bath and the
temperature of air near the top of the buret (for
PV nRT). Make sure thermometer is dry for the
air measurement. - Lower leveling bulb before starting reaction.
4- THINGS TO REMEMBER
- Measure barometric pressure one time each week.
Barometer located near ovens. - Before connecting Erlenmeyer, adjust leveling
bulb so that buret is near 0 mL -
and leveling bulb is only 2-3 cm full - Only read volume when buret level even with
leveling bulb level - If bath ?T gt 0.1ºC at end of run, use ice chips
to equalize temperature before -
reading
final volume - Adjust mass sample to achieve approximately 35
mL change in volume - At days end, turn in "kit", replace transfer
desiccator, put unknown vial storage
-
desiccator - In write-up, use v. p. of water at your T to
subtract water P from total P
5CHANGES FROM LAB MANUAL Step 4 Add sulfamic
acid and then the capsule. Shake periodically.
The capsule does not dissolve immediately. Step
5 Measure T of water in the bath and
T of air near the buret (with dry
thermometer!!!) Use bath T to determine v. p. of
water. Use air T in PVnRT to determine
moles N2 from PN2
6WEEK 2 Do runs until precision achieved.
Boil 1L of deionized water for 10 minutes w/
boiling chips Allow to cool without
agitation, then transfer to 1 L bottle. CO2
free water for use in preparing NaOH solution for
next experiment.
7Cooling StorageLab assistant add 4.05 mL of
50 NaOH
- Water has been dyed for viewing purposes.