CCD diagnostics and 04/12/2003 data - PowerPoint PPT Presentation

1 / 13
About This Presentation
Title:

CCD diagnostics and 04/12/2003 data

Description:

Image IP laser spot. Basler A302f. Camera specifications. Monochrome. 8bit ADC. 780-582 resolution. 8.3. m 30 fps. Shutter/Gain/Brightness. Triggered operation ... – PowerPoint PPT presentation

Number of Views:25
Avg rating:3.0/5.0
Slides: 14
Provided by: bes84
Category:

less

Transcript and Presenter's Notes

Title: CCD diagnostics and 04/12/2003 data


1
CCD diagnostics and 04/12/2003 data
  • Talk outline
  • Introduction to CCD use in PETRA laserwire
  • Readout and software
  • Readout problems
  • Analysis of images
  • Results for scans of 04/12/2003
  • Summary of results
  • Near term future work
  • Summary

2
CCD Diagnostics
  • System
  • 2 cameras
  • Pre/post IP
  • Image IP laser spot
  • Basler A302f
  • Camera specifications
  • Monochrome
  • 8bit ADC
  • 780-582 resolution
  • 8.3 ?m
  • 30 fps
  • Shutter/Gain/Brightness
  • Triggered operation
  • Firewire
  • External line
  • Small footprint

3
Readout software
  • Firewire
  • Digital
  • No readout card required or grabber
  • BCAM driver simple API
  • Cable lengths (lt10m)
  • Readout
  • C library
  • Simple to use
  • Tied to Microsoft
  • No problem now (Labview)
  • Multiple file formats (ascii, bmp)
  • MFC based windows executable
  • Fast (25-30 fps)
  • See images directly as DAQ runs
  • Upgrade for next run
  • Triggered image acquisition
  • Multiple cameras
  • Analysis included

4
Readout problems
  • Trigger
  • Laser 30 Hz CCD 30 fps
  • Empty frames
  • Partial images
  • Recently found
  • Only occurs in 1st scan
  • Cause unknown
  • Laser damage
  • Care must be taken not to damage devices
  • Filtering
  • Currently use OD 2-5 ND filters
  • Reflection reduction in intensity

5
CCD image analysis
  • Analysis procedure
  • Delete burn marks
  • Projections x and y
  • Fit background and subtract
  • Fit Signal
  • Mean/RMS calculation
  • Fit signal and background
  • Fit in 2D?
  • Average over all images for a scan point
  • Important to remember
  • Area of interest (AoI)
  • Fits and mean/RMS calculation
  • ?70 pixels
  • Errors on pixel values
  • Constant?
  • Dependent on signal?
  • Propagation through to background subtraction

Raw projection
Background subtracted
6
CCD Background
  • Problematic when
  • Laser mode non-Gaussian
  • Shutter
  • Ambient light noise should be reduced
  • Source of background
  • Ambient light
  • Diffuse laser reflections?
  • Electronics noise
  • Thermal?
  • Accelerator induced electronics noise?
  • Define AoI!

7
CCD analysis Low Current Scan
  • Data taking
  • 10-30 frames/camera/scan point before calorimeter
    measurement

8
CCD analysis High Current Scan
  • Data taking
  • 50 frames/camera/scan point before calorimeter
    measurement
  • 30 frames too low.

9
CCD calibration summary
  • Summary of CCD/scanner results
  • Mean mean position / scanner voltage
  • Fit-? Gaussian mean position / scanner voltage
  • RMS average RMS
  • Fit-? average Gaussian width
  • Calibration
  • Fit and simple calculation consistent
  • Low and high current scans are not consistent!

Camera 1 Camera 1 Camera 2 Camera 2
Scan1 Scan 2 Scan 1 Scan 2
Mean ?m/volt 74.8?0.6 70.1?0.5 71.2?00.6 65.9?0.6
Fit-? ?m/volt 75.4?0.6 70.4?0.5 71.3?0.7 66.1?0.7
RMS ?m 163.1?0.5 164.0?0.6 141.2?0.5 143.8?0.6
Fit-? ?m 160.5?0.8 164.6?0.9 138.9?0.8 140.4?0.9
10
Future work
  • Analysis
  • Gain/Brightness tuning loop
  • Online analysis fitting etc
  • Trigger
  • 100 filled images
  • Camera from RHUL, kitted out PC _at_ UCL
  • Cooling/Radiation
  • Temperature monitoring
  • Investigate cooling options
  • Background
  • Background levels with/without
  • PETRA beam
  • Laser
  • Area of Interest
  • Calibration
  • CCD to chamber centre

11
Short term developments
  • Trigger
  • Signal generator
  • Pulse generator with delay (trigger signal)
  • LED (for fake signal)
  • Stephen Malton
  • Software (triggered version is quite complicated)
  • Testing
  • Large increase in good CCD data factor 2-3
  • Environment monitoring
  • Damaged 2 cameras last year!
  • Temperature
  • Humidity
  • Radiation
  • Ambient light (background)
  • Insert during next visit?

12
Longer term
  • High speed scanning
  • CMOS technology
  • Basler 00
  • 500 fps _at_ 1Mpixel
  • 7000 fps maximum
  • More radiation hard
  • Even faster scanning with multiple laser pulses
    per
  • Larger dynamic range
  • Basler A302f
  • 8 bit
  • How about 12-16 bit output?
  • Better signal to noise
  • Basler A600
  • Effective 16bit output _at_ VGA resolution

13
Summary
  • First use of CCDs during scan
  • Fit and simple calculation consistent
  • Low and high current scans are not consistent!
  • Much to be done
  • Calibration
  • Analysis
  • Software
  • Background
  • Reliability