Title: Geant4 Monte Carlo Characterization of Novel Electronic Brachytherapy System
1Geant4 Monte Carlo Characterization of Novel
Electronic Brachytherapy System
- Derek Liu
- E. Poon, M. Bazalova, B. Reniers, M. Evans
- J. Seuntjens, T. Rusch, F. Verhaegen
- McGill University
- Xoft, Inc.
2Axxent Miniature X-ray Tube
3Project Objectives
- Characterize x-ray tube
- Monte Carlo with experimental verification
- Beam quality
- HVL
- Attenuation curve
- Photon Spectrum
- Spectrum
- 2-D Dose distribution
- Gafchromic film in Lucite Phantom
4Attenuation Curve
- NE 2571 farmer
- 2 mm Pb collimator
- 0.5 cm aperture
- Src-collimator 1
- Src-Al foil 2
- Src-detector 10
Output monitor
Chamber response
Subtract scatter
5Spectrum
- Amp-Tek CdTe
- 1 mm Crystal
- 4 mil Be Window
- 2 mm W collimator
- 2 mm aperture
Attenuation correction
Stripping procedure
31.0 keV Ba-133 decay 14.9 keV Yttrium K-alpha
6Geant 4.8.1
- Why use Geant?
- Complex geometry
- Low energy physics
- C
- Coding flexibility
- Geometry
- Tip thickness 0.2 mm
- Transmission Tungsten target thickness 0.7 um
- 2 mm water
7Geant 4 Bremsstrahlung Production
Bremsstrahlung mean free path
Bremsstrahlung angular distribution
8Geant 4 Bremsstrahlung Production
9Source to Source Variation
10Comparison between Measured Spectra and Geant
11Source to Source Variation
12Aging Effects
132-D Dose Distribution
- Gafchromic EBT films
- Lucite Phantom
- Src parallel to film
- Src-film 1.5 cm
- Exposure time 5 min
- Max dose 5 Gy
Ir-192 calibrated in water
Document scanner
Matlab correlation routine
14Gaf EBT in Lucite Relative Dose distribution at
1.5cm
15Conclusion
- Geant4 Bremsstrahlung production comparable with
EGS - Angular distribution
- Bremsstrahlung mean free path
- Bremsstrahlung spectrum
- Spectral measurements
- Spectra in agreement with Geant4
- Attenuation curve measurements
- HVL in agreement with spectral HVL and Geant4
- Observed factors affecting beam quality
- Source variation
- Aging effect
- Dose distribution measurements
- General agreement with Geant4
- Deviation in forward and aft directions