ScreenFilm Radiography II - PowerPoint PPT Presentation

1 / 37
About This Presentation
Title:

ScreenFilm Radiography II

Description:

Maximum OD used in medical radiography. 3.6. 0.00025. Very dark; requires hot lamp. 3 ... Contrast is desirable in screen-film radiography ... – PowerPoint PPT presentation

Number of Views:69
Avg rating:3.0/5.0
Slides: 38
Provided by: michae354
Category:

less

Transcript and Presenter's Notes

Title: ScreenFilm Radiography II


1
Screen-Film Radiography II
  • Characteristics of film
  • Screen-film system
  • Scattered radiation

2
Composition and function
  • Unexposed film consists of one or two layers of
    film emulsion coated onto a flexible Mylar sheet
  • Tabular grain emulsions are used in modern
    radiographic film laser cameras and older
    radiographic films use cubic grain emulsions
  • Grains of silver halide (AgBr and AgI) are bound
    in a gelatin base and together comprise the film
    emulsion

3
Scanning electron micrographs of film emulsions
4
Optical density
  • X-ray film is a negative recorder increased
    light (or x-ray) exposure causes the developed
    film to become darker
  • Degree of darkness is quantified by the OD,
    measured with a densitometer
  • Transmittance and OD defined as

5
OD examples
6
Hurter and Driffield curve
  • The response of film as a function of x-ray
    exposure is nonlinear
  • Curve describing OD versus the logarithm (base
    10) of exposure is called the HD curve
  • This curve has a sigmoid shape

7
A Hurter and Driffield (HD) curve
8
Contrast
  • Contrast of a radiographic film is related to the
    slope of the HD curve
  • Regions of higher slope have higher contrast
  • Regions of reduced slope (e.g., the toe and
    shoulder) have lower contrast
  • A single number, which defines the overall
    contrast of a given type of radiographic film, is
    the average gradient

9
Average gradient from HD curve
10
Average gradient
  • Always measured at the same OD points
  • OD1 0.25 base fog
  • OD2 2.0 base fog
  • Average gradients for radiographic film range
    from 2.5 to 3.5

11
Contrast as a function of exposure or optical
density
12
Speed
  • Sensitivity or speed of a screen-film system can
    be seen from HD curves
  • As the speed of a screen-film combination
    increases, the amount of x-ray exposure required
    to achieve the same OD decreases
  • Absolute speed is the inverse of the exposure (in
    roentgens) required to achieve an OD of 1.0
    base fog
  • Commercial system for defining speed is a
    relative measure. Par speed systems have an
    arbitrary speed rating of 100.

13
HD curves for two different screen-film detectors
14
Latitude
  • Contrast is desirable in screen-film radiography
  • Latitude is the range of x-ray exposures that
    deliver ODs in the usable range
  • High contrast systems have reduced latitude
  • It is more difficult to consistently achieve
    proper exposures with low-latitude screen-film
    systems

15
HD curves for screen-film systems which differ
in contrast
16
Screen-film system
  • Film emulsion should be sensitive to the
    wavelengths of light emitted by the screen
  • Calcium tungstate emits blue light to which
    silver halide is sensitive
  • Rare earth screens emit green light sensitizers
    are added to the film emulsion to increase the
    sensitivity of silver halide to these wavelengths
  • Screens and films are usually purchased as a
    combination

17
Reciprocity law
  • The relationship between exposure and OD should
    remain constant regardless of the exposure rate
  • For very long or very short exposures, and
    exposure rate dependency between exposure and OD
    is observed, and this is called reciprocity law
    failure

18
(No Transcript)
19
Contrast and dose
  • Screen-film system governs the overall detector
    contrast
  • Total x-ray exposure time (motion artifacts) and
    radiation dose to the patient are considerations
  • Lower kVp settings yield higher subject contrast,
    especially for bone imaging
  • Appropriate kVp for each specific type of
    examination is dogmatic, and these values have
    been optimized over a century of experience
  • Adjustments may be required due to patient size

20
Entrance skin exposure (ESE) vs peak kilovoltage
(kVp)
21
Contrast and dose as a function of peak
kilovoltage (kVp) 1-mm bone chip within a 23-cm
patient
22
Scattered radiation
  • For virtually all radiographic procedures except
    mammography, most photon interactions in soft
    tissue produce scattered x-ray photons
  • Detection of scattered photons causes film
    darkening but does not add information content to
    the image

23
Scattered radiation and the scatter-to-primary
ratio
24
Effect of collimation
  • As the field of view is reduced, the scatter is
    reduced
  • An easy way to reduce the amount of x-ray scatter
    is by collimating the x-ray field to include only
    the anatomy of interest and no more

25
Scatter-to-primary ratio vs x-ray field size for
different thickness
26
Antiscatter grid
  • An antiscatter grid is placed between the patient
    and the screen-film cassette
  • The grid uses geometry to reduce the amount of
    scattered reaching the detector

27
Geometry of an antiscatter grid used in
radiography
28
Antiscatter grid (cont.)
  • Antiscatter grid is composed of a series of small
    slits, aligned with the focal spot, that are
    separated by highly attenuating septa
  • Primary x-rays have a higher chance of passing
    through the slits unattenuated by the adjacent
    septa
  • Septa (grid bars) are usually made of lead
    openings (interspaces) between the bars can be
    made of carbon fiber, aluminum, or even paper

29
Details of grid construction
30
Grid ratio
  • Single most important parameter that influences
    the performance of a grid
  • Grid ratio is the ratio of the height to the
    width of the interspaces (not the grid bars) in
    the grid
  • Grid ratios of 81, 101, and 121 are common in
    general radiography grid ratio of 51 is common
    in mammography

31
Distribution of scattered x-rays vs scatter
incidence angle
32
Other grid parameters
  • Focal length determines the amount of slant of
    the slits in the grid
  • Grid frequency refers to the number of grid bars
    per unit length. Grids with 40 and 60 grid lines
    per centimeter commonly available
  • Interspace material influences dose efficiency.
    Air or carbon fiber required for mammography
  • The Bucky factor is the ratio of the entrance
    exposure to the patient when the grid is used to
    the entrance exposure without the grid

33
Bucky factor for four different grid ratios
34
Grid orientation errors
35
Artifacts caused by grids
  • Most artifacts associated with the use of a grid
    have to do with mispositioning
  • Generally occurs only during installation or
    servicing of the system
  • Portable systems may attach the grid to the
    cassette, increasing the chance of mispositioning
  • Radiologist should be able to identify the
    presence and cause of various grid artifacts when
    present

36
Air gap geometry for scatter reduction
37
Air gaps
  • Clinical utility of this approach is compromised
    by several factors
  • Additional object magnification, often causing
    loss of spatial resolution unless a very small
    focal spot is used
  • Reduced field of view that is imaged by a
    detector of fixed dimensions
  • Grids still routinely used in all of radiography
    other than extremity studies, some pediatric
    studies, and magnification mammography views
Write a Comment
User Comments (0)
About PowerShow.com