Title: ScreenFilm Radiography II
1Screen-Film Radiography II
- Characteristics of film
- Screen-film system
- Scattered radiation
2Composition and function
- Unexposed film consists of one or two layers of
film emulsion coated onto a flexible Mylar sheet - Tabular grain emulsions are used in modern
radiographic film laser cameras and older
radiographic films use cubic grain emulsions - Grains of silver halide (AgBr and AgI) are bound
in a gelatin base and together comprise the film
emulsion
3Scanning electron micrographs of film emulsions
4Optical density
- X-ray film is a negative recorder increased
light (or x-ray) exposure causes the developed
film to become darker - Degree of darkness is quantified by the OD,
measured with a densitometer - Transmittance and OD defined as
5OD examples
6Hurter and Driffield curve
- The response of film as a function of x-ray
exposure is nonlinear - Curve describing OD versus the logarithm (base
10) of exposure is called the HD curve - This curve has a sigmoid shape
7A Hurter and Driffield (HD) curve
8Contrast
- Contrast of a radiographic film is related to the
slope of the HD curve - Regions of higher slope have higher contrast
- Regions of reduced slope (e.g., the toe and
shoulder) have lower contrast - A single number, which defines the overall
contrast of a given type of radiographic film, is
the average gradient
9Average gradient from HD curve
10Average gradient
- Always measured at the same OD points
- OD1 0.25 base fog
- OD2 2.0 base fog
- Average gradients for radiographic film range
from 2.5 to 3.5
11Contrast as a function of exposure or optical
density
12Speed
- Sensitivity or speed of a screen-film system can
be seen from HD curves - As the speed of a screen-film combination
increases, the amount of x-ray exposure required
to achieve the same OD decreases - Absolute speed is the inverse of the exposure (in
roentgens) required to achieve an OD of 1.0
base fog - Commercial system for defining speed is a
relative measure. Par speed systems have an
arbitrary speed rating of 100.
13HD curves for two different screen-film detectors
14Latitude
- Contrast is desirable in screen-film radiography
- Latitude is the range of x-ray exposures that
deliver ODs in the usable range - High contrast systems have reduced latitude
- It is more difficult to consistently achieve
proper exposures with low-latitude screen-film
systems
15HD curves for screen-film systems which differ
in contrast
16Screen-film system
- Film emulsion should be sensitive to the
wavelengths of light emitted by the screen - Calcium tungstate emits blue light to which
silver halide is sensitive - Rare earth screens emit green light sensitizers
are added to the film emulsion to increase the
sensitivity of silver halide to these wavelengths - Screens and films are usually purchased as a
combination
17Reciprocity law
- The relationship between exposure and OD should
remain constant regardless of the exposure rate - For very long or very short exposures, and
exposure rate dependency between exposure and OD
is observed, and this is called reciprocity law
failure
18(No Transcript)
19Contrast and dose
- Screen-film system governs the overall detector
contrast - Total x-ray exposure time (motion artifacts) and
radiation dose to the patient are considerations - Lower kVp settings yield higher subject contrast,
especially for bone imaging - Appropriate kVp for each specific type of
examination is dogmatic, and these values have
been optimized over a century of experience - Adjustments may be required due to patient size
20Entrance skin exposure (ESE) vs peak kilovoltage
(kVp)
21Contrast and dose as a function of peak
kilovoltage (kVp) 1-mm bone chip within a 23-cm
patient
22Scattered radiation
- For virtually all radiographic procedures except
mammography, most photon interactions in soft
tissue produce scattered x-ray photons - Detection of scattered photons causes film
darkening but does not add information content to
the image
23Scattered radiation and the scatter-to-primary
ratio
24Effect of collimation
- As the field of view is reduced, the scatter is
reduced - An easy way to reduce the amount of x-ray scatter
is by collimating the x-ray field to include only
the anatomy of interest and no more
25Scatter-to-primary ratio vs x-ray field size for
different thickness
26Antiscatter grid
- An antiscatter grid is placed between the patient
and the screen-film cassette - The grid uses geometry to reduce the amount of
scattered reaching the detector
27Geometry of an antiscatter grid used in
radiography
28Antiscatter grid (cont.)
- Antiscatter grid is composed of a series of small
slits, aligned with the focal spot, that are
separated by highly attenuating septa - Primary x-rays have a higher chance of passing
through the slits unattenuated by the adjacent
septa - Septa (grid bars) are usually made of lead
openings (interspaces) between the bars can be
made of carbon fiber, aluminum, or even paper
29Details of grid construction
30Grid ratio
- Single most important parameter that influences
the performance of a grid - Grid ratio is the ratio of the height to the
width of the interspaces (not the grid bars) in
the grid - Grid ratios of 81, 101, and 121 are common in
general radiography grid ratio of 51 is common
in mammography
31Distribution of scattered x-rays vs scatter
incidence angle
32Other grid parameters
- Focal length determines the amount of slant of
the slits in the grid - Grid frequency refers to the number of grid bars
per unit length. Grids with 40 and 60 grid lines
per centimeter commonly available - Interspace material influences dose efficiency.
Air or carbon fiber required for mammography - The Bucky factor is the ratio of the entrance
exposure to the patient when the grid is used to
the entrance exposure without the grid
33Bucky factor for four different grid ratios
34Grid orientation errors
35Artifacts caused by grids
- Most artifacts associated with the use of a grid
have to do with mispositioning - Generally occurs only during installation or
servicing of the system - Portable systems may attach the grid to the
cassette, increasing the chance of mispositioning - Radiologist should be able to identify the
presence and cause of various grid artifacts when
present
36Air gap geometry for scatter reduction
37Air gaps
- Clinical utility of this approach is compromised
by several factors - Additional object magnification, often causing
loss of spatial resolution unless a very small
focal spot is used - Reduced field of view that is imaged by a
detector of fixed dimensions - Grids still routinely used in all of radiography
other than extremity studies, some pediatric
studies, and magnification mammography views