Title: External Second Gate, Fourier Transform Ion Mobility Spectrometry:
1External Second Gate, Fourier Transform Ion
Mobility Spectrometry FT-IMS Next
Generation Ion Mobility Spectrometer
Edward E. Tarver, Ph.D. Analytical Material
Sciences Department Sandia National
Laboratories-Livermore, California
2Ion Mobility Spectrometry
- Real-time response few seconds analysis time.
- Sensitivity low part-per-billion detection
without pre-concentration. - Versatility simultaneous/universal response.
- Simplicity of electronics no vacuum
pumps/chromatographs. - Field portability low power, size and weight.
- Battery powered military and commercial units
available. - Unattended monitoring perimeter and network
defense. -
3Drift Gas Exhaust
Air Drift Gas Inlet
Faraday Collector Signal Out
-
Ion Drift Region
Sample Inlet
-
63Ni Ionization Region
Drift Gas Flow
-
-
High Voltage Repeller
Entrance Gate
Focusing Rings
Aperture Grid
Commercial/Military IMS Drift Tube
4(No Transcript)
5The observed peak tailing is due to ion-molecule
reactions occurring during time-of-flight and
further compounded by the signal averaging
process.
6Fourier Transform Ion Mobility Spectrometry
- Increased Sensitivity, Lower Detection Limits
Sensitivity depends on the duty cycle. - FT-IMS operates with 50 ion gating efficiency
compared to 1 with conventional IMS. - Fifty times more ions transmitted and detected
than conventional IMS. -
- Improved Resolution, Fewer False Alarms FT-IMS
dual-gate design eliminates broadening due to
ion-molecule reactions and averaging process. - Conventional IMS sums all variations in ion
velocity, broadening peaks and reducing
resolution. No need to average with FT-IMS. - Suited for Miniaturization FT-IMS performance
allows miniaturization of detectors. - Adaptable to Current IMS Systems No hardware
modifications to drift tube.
7Ion Gating in FT-IMS
CYCLE REPEATED (IF DESIRED)
LOW FREQUENCY
HIGH FREQUENCY
open
Entrance gate pulse
closed
open
Exit gate pulse
closed
1. Gates are open and closed for equal amounts of
time no matter how frequently they are pulsed. 2.
Ion collection during half of the analytical
cycle time, i.e., 50 duty cycle. 3. Low
frequency greater Signal/Noise, High frequency
better Resolution.
8Fourier Transform of the Ion Mobility
Interferogram
Fourier Transform
Ion Mobility Interferogram
Ion Mobility Spectrum
8
9Conventional IMS vs. FT-IMS
10FT-IMS Allows Tunable Resolution
11(No Transcript)
12TNT Response as a Function of Scanning Time
13PETN Response as a Function of Scanning Frequency
14HNS Response at 10kHz and 20kHz Scanning Frequency
15HMX Response Frequency Range and Scan Time
20
16RDX Response as a Function of Frequency Range
Scanned
17Resolution vs. Aspect Ratio as Indicator of Peak
Quality
- RESOLUTION (R) R Drift Time (ms) / Peak Width
at Half Height (ms) - Resolution calculation ignores peak broadening
below Half Height - where peak tailing and overlap limits
ability to separate adjacent peaks. - Drift time dependent broad, low intensity peaks
with long drift times can - give higher Resolution (R) than strong, sharp
peaks with short drift times. - Misleading indicator of instrumental resolving
power. - ASPECT RATIO AR Peak Height (h) / Peak Width
at Base (w) - Unbiased indicator of peak quality, includes peak
width below Half Height. - Aspect Ratio is Independent of drift time and
describes actual peak shape.
18Resolution Number vs. Aspect Ratio
(Drift Time/w1/2) (Peak
Height/wb)
R 5/2 2.5
R 20/2 10
R 32/2 16 R 40/2.5
16 AR 3.25/.375 8.6
AR 8.6
AR 8.6 AR 0.235
0 5 10 15 20 25 30 35 40 45 Drift Time (ms)
19Resolution in IMS
- Selected Bench-top IMS Instruments
- IMS 5000 UVIMS-MCC Itemiser
AirSentry IonScan 400B - Draeger G.A.S.
G.E./Ion Track SAES/Molecular Smiths
Detection - Safety Co. Technol.
Analytics - Germany Germany U.S.A.
Italy U.K. - Tritium 63Ni or UV 63Ni
63Ni
63Ni - 30-60 NA
25 44 - Selected Handheld IMS Instruments
- RAID-M IMS Mobile µIMS VaporTracer
Quantum Sniffer LCD3.2 - Bruker Draeger G.A.S.
G.E./Ion Track Implant Sciences Smiths
Detection - Daltonics Safety Co. Technol.
Corporation - Germany Germany Germany U.S.A.
U.S.A. U.K.
20Peak Quality Determines False Alarm Rate
Peak Resolution R td/w1/2 Aspect Ratio AR
h/wb PEAK IMS X2G-FT-IMS IMS
X2G-FT-IMS Ko 1.84 SA 10K 20K 40K
SA 10K 20K 40K TNT 40.97 30.27
36.59 10.74 156.8 101.6 PETN 41.23 28.74
39.56 13.68 209.8 18.88
HNS 41.94 28.74 34.31 5.98 188.4 130.2
HMX 41.35 28.57 40.98 3.02 185.6 36.56
RDX ------ 28.84 50.92 ------ 113.4
31.89 Averages 41.37 29.03 37.72
8.35 170.8 63.82 Ko 1.54
TNT 45.59 30.41 30.75 42.47 9.12 156.8
134.0 56.87 PETN 38.20 37.42 41.40
------ 5.68 47.14 75.90 ------
HNS 45.70 26.86 40.67 ------ 12.8 51.70
77.13 ------ HMX 42.04 31.76 41.49
65.99 7.52 147.4 56.84 29.81
RDX 46.33 ------ 34.11 75.27 9.32 ------
17.86 ------ Averages 43.57 31.61 37.68
61.24 8.88 100.8 72.34 ------
21acetone
RDX
reactant ion peak 8.5 ms
228.5 ms
RDX
Note the comparative resolution of the peak a 8.5
ms. FT-IMS is able to resolve both species
Present whereas signal averaging cannot. The
peak at 12 ms is residual acetone.
23Handheld FT-IMS
24FT-IMS Rear View
25FT-IMS 9-Volt Batteries in Parallel
26FT-IMS Interior View
27FT-IMS Vertical Battery Arrangement
28Acknowledgements Sandia National Laboratories,
Research Foundations Laboratory Directed
Research and Development Grants Sandia National
Laboratories, Livermore CA Analytical Material
Sciences Department Dr. Jim Wang, Mr. Anh Phan,
Dr. Kent Pfeiffer, Mr. John Warmouth Professor
Herbert Hill, Washington State University,
Pullman WA Professor David Harris, Harvey Mudd
College, Claremont CA United States Department
of the Navy Contract N4175603GO14803