Fasteners / Joint Design - PowerPoint PPT Presentation

About This Presentation
Title:

Fasteners / Joint Design

Description:

Using a long narrow bolt results in a much higher bolt elasticity than that of the Flag (10X) ... With a strain as high as .008' washers will prevent preload ... – PowerPoint PPT presentation

Number of Views:120
Avg rating:3.0/5.0
Slides: 28
Provided by: michaelk170
Category:

less

Transcript and Presenter's Notes

Title: Fasteners / Joint Design


1
Fasteners / Joint Design
NSTX TF FLAG JOINT REVIEW 4/10/03
Michael Kalish
2
Stud Preload
  • Maintaining the preload on the stud is critical
    for maintaining contact pressure and contact
    resistance
  • Using a long narrow bolt results in a much higher
    bolt elasticity than that of the Flag (10X).
  • Applied cyclical loading adds relatively small
    additional loading to the stud.
  • With higher elasticity, loss in preload due to
    deflection is minimized.

3
Preload Continued
  • Belleville washers are used to account for any
    unexpected yielding of bolt or copper
  • While the bolt length provides adequate
    elasticity to accommodate design load scenarios
    the addition of Belleville washers prevents
    relief of the preload in the event of
    unanticipated strain
  • Washer has ½ the stiffness of the bolt, for every
    .001 inch strain only 125 lbf preload is lost
    (total washer deflection .032)
  • With a strain as high as .008 washers will
    prevent preload from dropping below 3,560 lbf.
    (the first .0045 drops the bolt pre-load to
    4,000 lbf then the washer takes over)
  • Testing of prototype will verify that preload is
    maintained.
  • A washer plate is added to spread out the
    compressive forces under the nut and minimize
    local yielding of the copper.
  • Bolts to be pre-tensioned to eliminate stored
    torque

4
Bolt Characteristics
  • Bolt is a a 3/8-16 stud threaded at both ends
  • To increase elasticity the bolt shank diameter
    will be just slightly larger than the root
    diameter of the threads resulting in a relatively
    elastic bolt (a creep of .001 results in a loss
    of 225 lbf of preload)
  • Loading
  • A preload of 5,000 lbf is applied with an
    equivalent root diameter stress of 64,700 psi
  • Thermal loading after ratcheting applies an
    enforced deflection of .0043 inches corresponding
    to a stress adder of 12,700 psi
  • As a result of stiffening the hub structure
    additional mechanical loading is minimal so that
    almost all fatigue loading is the result of
    thermal stress
  • With the 5,000 lbf preload and the thermal
    stresses applied the bolt sees a mean tensile
    stress of 71.2 ksi and a mean amplitude of 6.5
    ksi
  • The ultimate tensile strength for the A286 is 145
    ksi and the yield is 100 ksi

5
Modified Goodman Diagram For A286 Stud
6
Threaded Insert
  • A TapLok 3/8-16 Medium Length insert is used
    (OD into copper is .50)
  • Loading
  • The bolt preload of 5,000 lbf results in 11,800
    psi in shear at the outer threads of the insert
    into the copper.
  • Thermal loading adds a cyclical load of 2,300 psi
  • Additional mechanical loading is minimal so that
    almost all fatigue loading is the result of
    thermal stress
  • Per the inspection certification the Tensile
    strength 38 kpsi equivalent to a Shear Strength
    22 kpsi
  • With the 5,000 lbf preload and the thermal
    stresses applied the bolt sees a mean shear
    stress of 12.9 kpsi and a mean cyclical amplitude
    of 1.2 ksi

7
Modified Goodman Diagram for Insert in Copper
Conductor
8
Threaded Insert (cont.)
  • Testing shows margins may be greater then the
    numbers indicate. The lowest pull out force
    measured 11,500 lbf equivalent to 27 kpsi shear
    strength in the copper (as compared to 22 kpsi)
  • The insert was tested for both pull out strength
    and pull out strength after cycling
  • Cycling (one sample) did not indicate significant
    degradation to pull out strength (fatigue sample
    pulled at 12,380 lbf)
  • Further testing is planned. A mechanical
    prototype will test for maintenance of preload
    after application of a cyclical load.

9
Testing
NSTX TF FLAG JOINT REVIEW 4/10/03
  • Test Setup
  • Pull Tests
  • Cyclical Pull Testing
  • Friction Tests
  • Collar Shear Tests
  • E-Beam Weld Tests
  • Mechanical Prototype

Michael Kalish
10
Test Setup
  • MTS Hydraulic Test Stand
  • Plots Load vs Deflection up to 100,000 lbf
  • Provides Cyclical Testing Capabilities

11
Pull Test Setup
  • Designed to test pull out strength of inserts
  • Inserts installed in spare lengths of conductor
  • Keenserts and two lengths of Taplok inserts tested

12
Pull Test Setup (cont.)
  • Specimen in Test Fixture

13
Pull Test Results KeenSerts
  • KEEN SERT 1     11,120lbs
  • KEEN SERT 2     12,000lbs
  • KEEN SERT 3     11,880lbs
  • KEEN SERT 4     11,620lbs
  • KEEN SERT 5     11,500lbs
  • KEEN SERT 6     11,260lbs
  • KEEN SERT 7     11,500lbs
  • KEEN SERT 8     11,380lbs

14
Pull Test Results Tap-Lok
  • 3/8-16 H Series Tap-Lok Inserts
  • 1st Sample SUMMARY (peak force) Regular Length
    0.687, Tap Drill
  • TAP LOK 1 15,260lbs
  • TAP LOK 2 15,500lbs (bolt broke)
  • TAP LOK 3 15,260lbs (bolt broke)
  • 2nd Sample SUMMARY (peak force) Regular Length
    0.687Tap Drill Plug Tapped Only
  • TAP LOK 1 16,000lbs
  • TAP LOK 2 15,380lbs
  • TAP LOK 3 15,760lbs
  • TAP LOK 4 15,500lbs
  • TAP LOK 5 15,620lbs
  • 3rd Sample SUMMARY (peak force) Medium Length
    0.562, Insert Tool Only
  • TAP LOK 1 12,500lbs
  • TAP LOK 2 12,500lbs
  • TAP LOK 3 11,500lbs
  • TAP LOK 4 12,500lbs
  • TAP LOK 5 11,760lbs

Displacement vs Force Curve for Tap Lok 4, 3rd
Sample Peak Force 12,500 lb
15
Cyclical Testing, Pull Out
  • Using same test setup medium length Tap Lok
    insert was cycled then pulled
  • 5000 cycles at 4600 lbf to 6000 lbf
  • 45000 cycles at 5400 lbf to 6000 lbf
  • Cycled at 1 Hz Sine Wave
  • Ranges represent thermal operational loads for
    25 degC and 5 degC cyclical thermal loading
  • Sample pulled after subjected to Fatigue along
    with insert installed in same conductor piece.

16
Cyclical Testing Results
  • Shaded area may indicate drift or creep (first
    run exhibited a negative creep drift)
  • Worst case creep of .0025(if real)
  • Top Plot5,000 cycles at 4,600lbf to 6,000lbf
  • Bottom Plot45,195 cycles at 5,400lbf to 6,000lbf

17
Cyclical Testing Pull Out Result
  • Fatigued Insert Pullout (bottom trace) vs
    Unfatigued Insert Pullout(top trace)
  • Fatigued 12,380 lbf
  • Unfatigued 13,380 lbf
  • Previous samples showed scatter of 11,500 lbf
    to 12,500 lbf
  • Pull out strength relatively unchanged
  • More testing to follow

18
Friction Test Setup
  • Two horrizontal load cells measure compressive
    force provided by eight 3/8th inch bolts
  • Specimens are machined and plated after each run
  • Vertical load is applied to offset middle block

19
Friction Test Calibrations
  • The two compression loads cells were checked
    against the MTS vertical load cell and found to
    be to within .30

20
Friction Tests Results
  • Force is recorded at point where inelastic
    behavior begins
  • Initial testing of unplated copper resulted in
    COF .12
  • Testing of plated copper increased COF values to
    .41

21
Friction Test Results (cont.)
  • Mean COF .41
  • Min. Value .39
  • Results consistent for varying compressive loads
  • Testing to continue
  • Will explore improvement of COF by changing
    surface conditions

22
Collar Shear Test
  • Representative sample of collar arrangement
    fabricated to characterize elasticity and shear
    strength

23
Collar Shear Test Results
  • No compressive load was applied
  • Shear area approx. 11.8 sq in.
  • For lower values split occurred on one side. For
    higher values on both sides simultaneously
  • Separation occurred at epoxy to SS interface.
  • Further testing planned for improved samples
  • SUMMARY
  • 1 60,000lbs, 2,540 psi
  • 2 58,590lbs, 3,000 psi
  • 3 43,100lbs, 1,820 psi
  • 4 40,000lbs, 2,040 psi
  • 5 61,900lbs, 2,610 psi
  • 6 65,000lbs, 2,720 psi

24
E-Beam Weld Test Specimens
  • Tensile Testing performed on 1 x .50 EBeam
    welded copper bars

25
E-Beam Weld Test Specimen Results
  • SUMMARY
  • 3L Peak Force 13,760lbs, 27,520 psi
  • 3R Peak Force 13,260lbs
  • 26,520 psi
  • 4L Peak Force 13,120lbs
  • 26,420 psi
  • 4R Peak Force 14,380lbs
  • 28,760 psi

Copper Bar Weld 4R Peak Force 14,380lbs, Peak
Displacement .199
26
E-Beam Weld Test Specimen Results
  • Bars separated at weld although weld area looked
    entirely homogeneous

27
Mechanical Prototype Testing
  • A prototype of the flag bolted to a section of
    conductor is in fabrication.
  • Vertical loading will be applied directly to the
    flag
  • This mockup will test the bolted joint for
    maintenance of preload after cycling.
  • Contact resistance will be monitored in real time.
Write a Comment
User Comments (0)
About PowerShow.com