CTD AND XBT TEMPERATURE MEASUREMENT COMPARISON. LT Scott Boedeker. CTD 10/XBT 4 ... Manufacturing Defects. Method of deployment -CTD casts are made nearly ... – PowerPoint PPT presentation

Number of Views:34
Avg rating:3.0/5.0
Slides: 11
Provided by: sabo5
Category:

less

Transcript and Presenter's Notes

Title: CTD AND XBT TEMPERATURE MEASUREMENT COMPARISON


1
CTD AND XBT TEMPERATURE MEASUREMENT COMPARISON
  • LT Scott Boedeker

2
(No Transcript)
3
(No Transcript)
4
CTD 10/XBT 4
5
REPRESENTATIVE PAIRS
6
(No Transcript)
7
  • For the 27 comparisons from this cruise, the XBT
    temperature has a warm bias of .0882 degrees C
    over the depth range
  • This is similar to results obtained from previous
    OC3570 cruises
  • The greatest bias was in the upper levels

8
CTD-XBT Temp Difference
R/V Point Sur
08/2001 02/2001
07-08/2000 Roth (2001)
Schmeiser (2000)  
9
Why the Bias?
  • Manufacturing Defects
  • Method of deployment

-CTD casts are made nearly vertical with the
vessel stopped. -XBT casts are made with the
vessel in motion and are therefore not as
vertical. This results in the XBT being
shallower than it is recorded.