Head - PowerPoint PPT Presentation

About This Presentation
Title:

Head

Description:

Wu Chang-Chin , MD, Visiting Staff, Department of Orthopaedic Surgery, En ... Ebramzadeh E, Sangiorgio SN, Lattuada F, Kang JS, Chiesa R, McKellop HA, Dorr LD. ... – PowerPoint PPT presentation

Number of Views:35
Avg rating:3.0/5.0
Slides: 17
Provided by: liaw
Category:
Tags: changkang | head

less

Transcript and Presenter's Notes

Title: Head


1
Head Cup Simulator
  • Why do we use simulator?
  • To verify the accuracy of new methods.
  • When we invent a new measuring method,
    it is important to verify its accuracy and
    reliability. Many methods have been published,
    including using a special machine (see left
    picture, Fig. 1) 1, in which anteversion,
    inclination, and three directions wearing can be
    adjusted manually. Another method is to fabricate
    a cup inside plastic material with fixed
    orientation and wearing parameters. All these
    published methods are good but expensive. We
    propose an automatic, cheaper, and more accurate
    computer method to do the same thing.
  • To simulate every control condition.
  • For mechanical devices, to accurately
    adjust the orientation and wearing parameters is
    difficult. It is also hard to prove it. With
    computer, every condition is easy to simulate.
  • To reduce noise in early development of new
    methods.
  • In reality, noise is inevitable, but
    in simulated situation, noise is removable. In
    the early stage of developing a method, noise
    will always bother us. With the use of simulator,
    noise is removed, and we can focus more on the
    real measurement, thus facilitating the
    development.

Fig. 1. A mechanical device for
simulating wearing of every anteversion and
abduction from JBJS 1.
2
  • How does the simulator work?
  • Line tracing method
  • We simulate every X-ray beam
    traveling from source to the film. The simulated
    prosthesis projection will be calculated. The
    total thickness of metal is estimated and then we
    translate to gray level of the simulated X-ray.
  • Why not using OpenGL or DirectX?
  • DirectX is the 3D tool of Windows. The OpenGL
    is the open source version of 3D tool. They are
    useful and effective in virtual reality. They
    are good in drawing non-transparent object. In
    our case, the X-ray beam is a highly penetrating
    radiation. We are drawing a semi-transparent
    object. It is hard to use OpenGL or DirectX,
    although they are quite good in performance.
  • Absorption law
  • The X-ray is a kind of radiation beam with
    high energy. It follows the general rule of
    absorption.
  • Beer-Lambert law
  •                          -kbcPenetration
    ek molar absorbability
    (a)
  • b path length
  • c concentration
    (b)
  • It is hard for us to use it on our simulator.
  • Absorption example (Fig. 2)
  • Lookup table method
  • The simulated X-ray was analyzed. We build a
    table with known thickness, in which we can
    deduce its photo-density on X-ray. We then look
    up the table, interpolate the data and then get
    the result of photo-density.

Fig. 2. (a) A metal model made of titanium with
1, 2, , 5mm thickness. (b) Its X-ray image.
3
PE Wear Meter
  • What did we do before?
  • In the past, we measured PE wear manually. A
    well accepted method is Livermores method (Fig.
    3) 2. He finds the center of femoral head,
    measures the thinnest PE, and then reads the
    wear.
  • Whats new?
  • Fig. 4a. Hardinges method
    Fig. 4b. Shavers method (need to point out
    p1, p2, p3)

Fig. 3c. Then measure it.
Fig. 3b. Find the thinnest polyethylene.
Fig. 3a. Livermores method, find head center.
4
  • Our method
  • Find head-neck junction by cross-correlation
  • Fig. 5. Make a template and mark corresponding
    head-neck junction and the indicated place.
  • At first, we make an ordinary X-ray with
    template. Then we mark the important landmark and
    pointout the indication from it. The program will
    use cross-correlation method, find the
    corresponding landmark on the other X-rays, and
    then find the indicated place. This procedure is
    the most time-consuming procedure and costs about
    60 seconds (Pentium 500MHz computer). This is
    also our original contribution to this method.
  • Project 200 lines from indicated place to upper
    right side with about 135 degrees span.
  • Detect edges with vector edge detector.
  • Using vector edge detector subroutine, we
    detect the shell edges. These edges are rechecked
    again and again in order to remove noise such as
    screw or others. We consider currently published
    edge detectors unsuitable, so we invent a new
    one. It detects vectory edges, so we called it
    vector edge detector.

5
  • Our method (cont.)
  • Find center from the 200 detected edges.
  • The center of the 40 edges are found by
    calculus. Then the program will exclude the
    extremes and find center again and again, until
    there are no extremes.
  • Project 240 lines from the new center with 270
    degrees span. The lower left part is excluded.
  • Find head edges with the same vector edge
    detector.
  • Find the head center
  • Again, the center of femoral head is found with
    exclusion of extremes.
  • Check centers by Hill-Climbcing Search 7
  • Compare the two centers, and find the
    displacement vector.
  • .

6
Materials Methods
  • Simulated data
  • 64 simulated X-rays with 2 different anteversion
    angles, 2 different abduction angles, 4 different
    superior wears, and 4 different medial wears.
  • We measure these simulated X-rays automatically
    by our Auto PE Wear Meter program.
  • The results are compared.
  • Real X-rays
  • We digitized 88 total hip arthroplasty X-rays
    using Sony S70 digital camera size (20481680).
  • The camera was set with the same distance and
    same zoom.

7
  • Real X-rays (Cont.)
  • There is no exclusion criteria of digitization.
  • Results classification
  • Wrong detection on cross-correlation
  • Wrong detection of acetabulum
  • Wrong detection of head
  • Minor error on acetabular center estimation
  • Minor error on head center estimation
  • Unmeasurable X-ray by Livermores method
  • Partially measurable X-ray, which can be measured
    by other method
  • Measurable X-ray by Livermores method

8
Results
  • Simulated data
  • -Our program can detect 61 of 64 simulated X-rays
    without misdetection (Fig. 6).
  • -The three mis-detected X-rays images were due to
    misdetection of acetabulum (Fig. 7).
  • The error on measuring medial directional wear
    using our program is -0.0230.176mm.
  • The error on measuring upper wear using our
    program is
  • -0.1520.315mm.

Fig. 6. Four examples of successful detection
Fig. 7. The three misdetections.
9
  • Real X-rays
  • 47 excellent detections
  • 2 unknown errors
  • 3 wrong detections on cross-correlation (Fig. 8)
  • 23 wrong detections of acetabulum (Fig. 9)
  • 3 wrong detections of head (Fig.10)
  • 10 minor errors on acetabular center estimation
    (Fig. 11)
  • 0 error on head center estimation
  • 19 unmeasurable X-rays by Livermores method
    (Fig. 12)
  • 49 partially measurable X-rays, which can be
    measured with instrument (Fig. 13)
  • 21 measurable X-rays by Livermores method (Fig.
    14)

Fig. 8. Wrong detection due to cross correlation.
Fig. 9. Wrong detection on finding acetabulum.
Fig. 10. Wrong detection on finding head.
Fig. 11. Minor error on fitting the circle with
acetabulum edges.
Fig. 12. Head-acetabulum edges are too unclear
to use Livermores method.
Fig. 13. With instrument, the thinnest edges can
be detected by other edges.
Fig. 14. Measurable by Livermores method.
10
  • 2 X-rays measurable by program but unmeasurable
    manually
  • 27 X-rays measurable by program but unmeasurable
    by Livermores method
  • 24 X-rays measurable manually but unmeasurable by
    program
  • 8 X-rays measurable by Livermores method but
    unmeasurable by program

Fig. 15a. Head edges are unclear to be detect
manually.
Fig. 15b. Our program can detect the edges and
find the best fitting circle.
Fig. 16a. Unmeasurable by Livermores method.
Fig. 16b. The program can measure it.
Fig. 17a. Measurable manually.
Fig. 17b. Misdetection by our program.
Fig. 18b Misdetection by our program.
Fig. 18a. Measurable by Livermores method.
11
Discussion
  • In the simulated data, error in medial wear is
    -0.0230.176mm and error in upper wear is
    -0.1520.315mm.
  • Compared with literature 16, in which the
    error listed was 0.1mm, our method is as good as
    other computer-assisted methods.
  • Our program is the first fully automatic program
    in which no human intervention is needed for
    detection.
  • Detection rate
  • Our program 54.5 (40/88).If we accept minor
    error, the rate becomes 65.9 (58/88).
  • Livermores method 23.9 (21/88)
  • Others 78.4 (69/88)

12
  • Reasons of misdetection
  • Low kv
  • As we know, the kv of X-rays will change the
    penetration power. When the kv is low, the
    penetration power is low. On the other hand, when
    the kv is high, the penetration power is high. If
    the kv is too low, the bone and prosthesis will
    be saturated as white. If the kv is adjusted
    high, the prosthesis will be whiter. That makes
    detection easy.
  • Screw noise
  • The screws of acetabulum are located just above
    it. That is the place where our program detects
    edges for estimating acetabulum center.
    Sometimes, the screws cause noises and
    misdetections.
  • Bone noise
  • The edges of bone will cause white lines on
    X-rays. If the white lines intersect the
    prosthesis edges, they may cause misdetection.
  • Distortion
  • X-ray filming
  • When the X-rays pass through body to film,
    there are grids between body and film to filtrate
    scattering. The grids may cause torsions and make
    circles became ellipses.
  • Digitizing
  • There are several devices for digitizing
    X-rays. The digital cameras take pictures through
    lens. The lens will cause distortion.
  • Validation
  • Currently, we have not tested our program on
    revision hip arthroplasty, which is to measure
    wears after removal of polyethylene and compare
    it with preoperative X-rays. This should be our
    future work.

13
Conclusions
  • We designed a program which can automatically
    measure polyethylene wear. We have tested it on
    simulated X-rays and the result is encouraging.
    Clinical validation is needed in future work.

14
Authors
  • Liaw Chen-Kun, MD, Visiting Staff, Department of
    Orthopaedic Surgery, En Chu Kong Hospital
  • Wu Tai-Yin, MD, Fellow, Department of Family
    Medicine, Jen-Ai General Hospital
  • Hou Sheng-Mou, MD, Professor, Department of
    Orthopaedic Surgery, National Taiwan University
    Hospital
  • Yang Rong-Sen, MD, Professor, Department of
    Orthopaedic Surgery, National Taiwan University
    Hospital
  • Fuh Chiou-Shann, PhD., Professor, Department of
    Computer Science and Information Engineering,
    National Taiwan University
  • Liaw Shong-Hon, MD, Resident, Department of
    Emergency, Sin-Lau Christial Hospital
  • Wu Chang-Chin , MD, Visiting Staff, Department of
    Orthopaedic Surgery, En Chu Kong Hospital
  • Tai Han-Cheng , MD, Visiting Staff, Department of
    Orthopaedic Surgery, En Chu Kong Hospital
  • Liu Dah-Hsiang , MD, Visiting Staff, Department
    of Orthopaedic Surgery, En Chu Kong Hospital

15
Institutions
  • En Chu Kong Hospital
  • National Taiwan University Hospital
  • Department of Computer Science and Information
    Engineering, National Taiwan University

16
Reference
  • Ebramzadeh E, Sangiorgio SN, Lattuada F, Kang JS,
    Chiesa R, McKellop HA, Dorr LD. Accuracy of
    measurement of polyethylene wear with use of
    radiographs of total hip replacements. Journal
    Article Journal of Bone Joint Surgery -
    American Volume. 200385-A(12)2378-84.
  • Livermore J, Ilstrup D, Morrey B. Effect of
    femoral head size on wear of the polyethylene
    acetabular component. J Bone Joint Surg Am.
    199072518-28.
  • Hardinge K, Porter ML, Jones PR, Hukins DW,
    Taylor CJ. Measurement of hip prostheses using
    image analysis. The maxima hip technique. J Bone
    Joint Surg Br. 199173724-8.
  • Shaver SM, Brown TD, Hillis SL, Callaghan JJ.
    Digital edge-detection measurement of
    polyethylene wear after total hip arthroplasty. J
    Bone Joint Surg Am. 199779690-700.
  • Martell JM, Berdia S. Determination of
    polyethylene wear in total hip replacements with
    use of digital radiographs. J Bone Joint Surg Am.
    1997791635-41.
  • Collier MB. Kraay MJ. Rimnac CM. Goldberg VM.
    Evaluation of contemporary software methods used
    to quantify polyethylene wear after total hip
    arthroplasty. Evaluation Studies. Journal
    Article Journal of Bone Joint Surgery -
    American Volume. 85-A(12)2410-8, 2003 Dec.
  • Stuart R. Peter N. Artificial Intelligence A
    Modern Approach. book 111-114.
Write a Comment
User Comments (0)
About PowerShow.com