Title: Evaluation of Candidate Standard XX (97/650)


1
Evaluation of Candidate Standard XX (97/650)
  • Clare Davis
  • Division of Retrovirology
  • National Institute for Biological Standards and
    Control
  • (NIBSC)
  • UK

2
1st International Standard for HIV-1 RNA (NIBSC
code 97/656)
  • Established as International Standard in 1999
    (Holmes et al, 2001).
  • Assigned a concentration of 100,000 IU per
    vial/ml.
  • Based on HIV-1 PCR positive, antibody negative
    plasmapheresis donation.
  • Virus diluted 1312.5 in base matrix
    (defibrinated plasma).
  • 2400 vials freeze-dried. Held at -200C.
  • Preparation showed good stability and correlation
    amongst assays.
  • Discussions at SoGAT Jul 2003 (PEI) raised
  • Low Stocks ( 650).
  • Presence of HBV DNA problematical for multiplex
    assays.

3
--- bDNA --- Monitor --Nuclisens --- Abbott ---
in-house
4
Re-evaluation of Candidate XX
  • Non-syncitium inducing HIV-1 genotype B field
    isolate field isolate (donated by Dr. P.
    Simmonds).
  • Supplied as a low passage virus.
  • Further propagated on human PBMCs, supernatant
    stored in LN2.
  • Virus diluted 1250 in pooled human plasma.
  • 2200 vials of material freeze-dried.

5
Study Details
  • Study purpose-
  • Calibrate XX against HIV-1 RNA International
    Standard.
  • Confirm stability of XX
  • Determine values in current assays
  • 9 Laboratories participating.
  • International Standard and Candidate XX randomly
    coded.
  • Participants asked to test samples in 3
    independent assays.
  • Initially 10 fold dilution
  • Then 2 subsequent assays half log dilutions
    around the end point.
  • NIBSC to analyse data.

6
Preliminary analysis of data so
far..Quantitative
Lab Number Assay Used Sample A (copies/ml Log 10) Sample B (copies/ml Log 10)
7 COBAS Taq Man 5.50 5.07
22 Amplicor 5.55 4.96
58 bDNA V3 5.21 4.69
73 COBAS Amplicor 5.84 5.26
Overall Mean 5.52 4.99
Values obtained in previous study 5.31
4.73
7
Qualitative
Lab Number Assay Used Sample A (log 10 detectable units/ml) Sample B (log 10 detectable units/ml)
70 In house 4.69 4.69
71 In house 5.36 4.11
72 In house 5.15 5.29
Overall Mean 5.06 4.70
Values obtained in previous study 4.70
4.22
8
Conclusion
  • Preliminary analysis of data shows good
    correlation with previous study.
  • Hope to submit report to ECBS Aug 04.
  • Please return any outstanding data to us by end
    Jun 04.