Title: Xray Timing Parameter Relationship: one Acquisition
1Xray Timing Parameter Relationship one
Acquisition
Frame 2 Acquisition Datetime
Frame i Acquisition Datetime
Frame N Acquisition Datetime
Frame 1 Acquisition Datetime (0018,9074)
Content Date (0008,0023) Content Time (0008,0033)
Acquisition Datetime (0008,002A)
Frame 1 Reference Datetime (0018,9151)
Frame 2 Reference Datetime
Frame i Reference Datetime
Frame N Reference Datetime
time
FRAME 2
FRAME N
FRAME 1
FRAME i
Frame 1 Acquisition Duration (0018,9220)
Frame N Acquisition Duration
Acquisition Duration (xxxx,xxxx)
If Acquisition is synchronized with external time
reference then Acquisition Time Synchronized
(0018,1800) YES
Exposure Time (0018,1150) SUMi(Frame i
Acquisition Duration) Average Pulse Width
(0018,1154) SUMi(Frame i Acquisition
Duration) / N
2Xray Timing Parameter Relationship one Frame
Frame Reference Datetime (0018,9151)
Frame Acquisition Datetime (0018,9074)
kV, mA
kV peak (0018,0060)
mA peak (0018,1152)
X-ray FRAME i Frame Acquisition Number
(0020,9156) i
PRE-FRAME X-ray
time
Frame Acquisition Duration (0018,9220)
R
Detector read-out
Cardiac Trigger Delay Time (0020,9153)
Last R-peak prior to the X-rayFRAME i
T
Q
S