Title: Edit Varga
1Heart Activity Measurements
2Heart Activity Measurements
- Pulse rate
- ECG/EKG electrocardiogram
- Impedance cardiogram, IKG -- impedance
cardiography - EchoKG echocardiogram
- Magnetocardiography (MCG)
3Arterys pulsation
4Heart Activity Measurements
- Pulse rate
- ECG/EKG electrocardiogram
- Impedance cardiogram, IKG -- impedance
cardiography - EchoKG echocardiogram
- Magnetocardiography (MCG)
5(No Transcript)
6EKG/ECG -- electrocardiogram
Williem Einthoven in 1924 Nobel prize
7(No Transcript)
8ATRIUM
VENTRICLE
9 The conduction system of the heart.
10Animation of a normal EKG wave
11Different waveforms for each of the specialized
cells found in the heart
12Information from EKG
- If the electric or muscular function of the heart
is disturbed for some reason, it will affect how
the electric signals spread through the heart
muscle.
13- The heart's electrical axis refers to the general
direction of the heart's depolarization wavefront
(or mean electrical vector) in the frontal plane.
Lead I.
Lead II.
Lead III.
14EKG
- These three bipolar limb leads roughly form an
equilateral triangle (with the heart at the
center) that is called Einthoven's triangle in
honor of Willem Einthoven who developed the
electrocardiogram in 1901.
15Negative electrode on the right arm and the
positive electrode on the left arm
Different directions during the EKG measurement
Negative electrode on the right arm and the
positive electrode on the left leg.
Negative electrode on the left arm and the
positive electrode on the left leg.
Maximal positive deflection is obtained in lead
III when the depolarization wave travels
parallel to the axis between the left arm and
left leg.
16(No Transcript)
17The realtively small P wave is produced by
electrical currents generated just before
contraction of the atria
EKG
Accoustic
QRS complex is caused by currents generated in
the ventricles during depolarization just prior
to venticular contraction. R is the most
prominent component of this wave complex.
18160 ms
300 ms
370 ms
830 ms
19positive deflection on the ECG
negative deflection on the ECG
equiphasic complex deflection on the ECG
20(No Transcript)
21The ECG system
Instrumentation amplifier, which has a very high
CMRR (90dB) (common-mode rejection ratio) and
high gain (1000), with power supply 9V and -9V.
An opto-coupler to isolate the In-Amp and Output.
A transducer AgCl electrode, which convert ECG
into electrical voltage. The voltage is in the
range of 1 mV 5 mV
Bandpass filter of 0.04 Hz to 150 Hz filter.
Its implemented by cascading a low-pass filter
and a high pass filter
22Heart Activity Measurements
- Pulse rate
- ECG/EKG electrocardiogram
- Impedance cardiogram, IKG -- impedance
cardiography - EchoKG echocardiogram
- Magnetocardiography (MCG)
23Impedance cardiogram
Impedance changes
Differentiatior
Amplifier
dZ/dt
Z0 value is the total impedance between the two
inner leads
High-frequency 40-100 KHz Constant current
4 mA
24Impedance cardiogram
R-Z is the time interval from the R wave of the
ECG to maximum ejection as indicated by the
peak of dZ/dt
ventricular ejection time
25- Part of automated external defibrillator
- Investigate circulatory system problems valve
defects, right-left shunting, congestive failure - Impedance of the thorax can be considered to be
divided into two parts - the impedance of both tissue and fluids
- the amount and distribution of blood The amount
of blood in the thorax changes as a function of
the heart cycle. The changes in the distribution
of blood in the thorax as a function of the heart
cycle can be determined by measuring the
impedance changes of the thorax.
Band electrodes
26Heart Activity Measurements
- Pulse rate
- ECG/EKG electrocardiogram
- Impedance cardiogram, IKG -- impedance
cardiography - EchoKG echocardiogram
- Magnetocardiography (MCG)
27EchoKG Echocardiogram
Ultrasound waves 2.518 MHz
28(No Transcript)
29EchoKG -- Echocardiogram
30Transducer
3140 kHz Ultrasound receiver
A X100 transistor amplifier is followed by a zero
cross detector circuit, using a voltage
comparator. The output is a TTL logic signal,
corresponding to the received 40KHz signal.
32Medium Power 40KHz Ultrasound Transducer Driver
- This crystal controlled circuit drives a 40KHz
piezoelectric transducer with a 30v peak to peak
signal.
33Heart Activity Measurements
- Pulse rate
- ECG/EKG electrocardiogram
- Impedance cardiogram, IKG -- impedance
cardiography - EchoKG echocardiogram
- Magnetocardiography (MCG)
34(No Transcript)
35dc SQUIDs
- Scanning electron micrograph of the SQUID ring
with step-edge Josephson junctions.
- Magnetometers based on dc SQUIDs are currently
the most sensitive sensors for magnetic fields,
achieving a magnetic field resolution which is
about a billion times below the earth's magnetic
field. - A dc SQUID basically consists of a
superconducting ring interrupted by two weak
links called Josephson junctions. - SQUID can be viewed as a flux-to-voltage
converter.
36(No Transcript)
37Noise spectrum of the magnetic field
noise in an industrial environment, measured
with an unshielded multiloop magnetometer
The magnetometer's intrinsic noise level is
several orders of magnitude lower.
38dc SQUIDs
- A multiloop SQUID magnetometer. The diameter of
this device is 8.5 mm.
- The SQUID ring itself is enlarged and it consists
of several identical pickup loops, which are
connected in parallel to reduce the inductance.
The 16 loops are arranged to the cart-wheel like
shape of the device. A multilayer technology is
needed for the preparation.
39Block diagram of a SQUID magnetometer
transform the applied flux into a room
temperature voltage output
senses changes in the external magnetic field
and transforms them into an electrical current
acquiring, storing analyzing data
transforms the resulting current into a magnetic
flux in the SQUID sensor
4068-channel dc-SQUID system
SQUID maps the axial (BZ) component
41Real-time magnetocardiogram recorded with a
multiloop magnetometer.
42Magnetocardiography (MCG)
- Magnetic field x, y, z component
- Grid measurement
- Similar sensitivity to EKG
- Higher SNR than EKG
43magnode
Solenoid coil
Ir reciprocal current FLM reciprocal magnetic
scalar potential field HLM reciprocal magnetic
field BLM reciprocal magnetic induction field
ELM reciprocal electric field JLM lead
field VLM voltage in the lead due to the
volume source i in the volume conductor
44Measurement of the x-component of the magnetic
heart vector
The general construction of the measurement
system
Baule-McFee lead system
Measurement of the z-component of the magnetic
heart vector
Measurement of the y-component of the magnetic
heart vector
45Unipolar Bipolar measurement locations on the
anterior and posterior sides
46Schematic illustration of the generation of the x
component of the MCG signal.
47Advantages
- Insulating barriers such as the skull, varying
layers of tissue, anatomical open spaces, do not
attenuate or distort magnetic fields. The
magnetic permeability of the tissue free space.
Therefore the sensitivity of the MCG is not
affected by the high electric resistivity of lung
tissue. - different sensitivity distribution with EKG
- the magnetic detector is not in contact with the
subject - SQUID magnetometer is readily capable of
measuring DC signals. Such signals can be
obtained electrically only with great difficulty.
48Disadvantages
- ECG is easier to use
- Technologically more complicated, requires
complex and expensive equipment SQUID
magnetometer, liquid helium, and a low-noise
environment - Because of the development of the SQUID
technology, a shielded room is no longer needed
in magnetocardiography. - Future at the liquid nitrogen temperature which
decreases the operational costs
49References
- MCG
- http//butler.cc.tut.fi/malmivuo/bem/bembook/20/2
0.htm - http//www.kreynet.de/asc/squids.html
- EchoKG
- http//www.heartsite.com/html/echocardiogram.html
what_US - http//www.discovercircuits.com/DJ-Circuits/ultra4
0khzxtr1.htm - EKG
- http//en.wikipedia.org/wiki/Electrocardiogram
- http//www.ecglibrary.com/
- http//www.americanheart.org/presenter.jhtml?ident
ifier3005172 - http//www.medmovie.com/mmdatabase/MediaPlayer.asp
x?ClientID68TopicID600 - John L. Andreassi Psychophysiology
- http//www.cisl.columbia.edu/kinget_group/student_
projects/ECG20Report/E600120ECG20final20report
.htm