Barrel TRT Straw Leak Test - PowerPoint PPT Presentation

1 / 9
About This Presentation
Title:

Barrel TRT Straw Leak Test

Description:

Straw Leak Test Status. Running now with 2-straw partially automated system ... Basic leak test technology is unchanged. 8/5/09 ... Straw Leak Test Criterion ... – PowerPoint PPT presentation

Number of Views:52
Avg rating:3.0/5.0
Slides: 10
Provided by: kenne120
Category:
Tags: trt | barrel | leak | leaktest | straw | test

less

Transcript and Presenter's Notes

Title: Barrel TRT Straw Leak Test


1
Barrel TRT Straw Leak Test
  • K. McFarlane
  • Hampton University
  • DOE/NSF Review, BNL March 2001

2
Straw Leak Test Status
  • Running now with 2-straw partially automated
    system
  • Can reliably get 125 straws/day (target)
  • New 4-straw system is under construction (in use
    April 2001)
  • Automation of additional steps is being developed
  • Basic leak test technology is unchanged.

3
Leak Test Station
4
Leak Test Head
5
Straw Leak Test Criterion
  • Straw is re-tested if leak rate exceeds 0.020
    water in 20 s (0.7 mbar/bar/min cf. TDR
    specification of 0.1 mbar/bar/min)
  • Straw that fails three times is rejected.
  • Typical re-test results are 0.004, 0.001, 0.010,
    0.002.
  • Leak rate is usually a function of our machine.
    We do not plan to change.

6
Leak Test Partial Automation
Partial automation BridgeView front panel
7
Database Entry
8
New Leak Test Head
9
Conclusions
  • Leak check criterion is satisfactory upgrade of
    equipment will proceed on the basis that existing
    basic method works
  • Production rate now adequate expected to
    increase, releasing technician time for other
    work.
Write a Comment
User Comments (0)
About PowerShow.com