Neutron Interactions and Dosimetry II - PowerPoint PPT Presentation

1 / 39
About This Presentation
Title:

Neutron Interactions and Dosimetry II

Description:

( 2) can then be applied to each one and solved simultaneously to obtain ... Rossi TE proportional counter (B/A 1) Tissue-equivalent plastic calorimeters (B/A 1) ... – PowerPoint PPT presentation

Number of Views:145
Avg rating:3.0/5.0
Slides: 40
Provided by: Michae1
Category:

less

Transcript and Presenter's Notes

Title: Neutron Interactions and Dosimetry II


1
Neutron Interactions and Dosimetry II
  • Paired Dosimeters
  • Calibration of the Low-Neutron-Sensitivity
    Dosimeter

2
Separate Measurement of Neutron and ?-Ray Dose
Components by Paired Dosimeters
  • If a mixed n ? field is measured by means of
    two dosimeters having different values of B/A,
    Eq. (2) can then be applied to each one and
    solved simultaneously to obtain D? and Dn, so
    long as B and A have known values
  • The best dosimeter pair is a TE-plastic ion
    chamber containing TE gas (for which B/A ? 1) to
    measure the total n ? dose, and a
    nonhydrogenous dosimeter having as little neutron
    sensitivity as possible to measure the ? dose
  • Ideally this dosimeter should measure only ?-rays

3
Dosimeters with Comparable Neutron and ?-Ray
Sensitivities (B/A ? 1)
  • A-150 TE plastic ion chambers (B/A ? 1)
  • Rossi TE proportional counter (B/A ? 1)
  • Tissue-equivalent plastic calorimeters (B/A ? 1)
  • Aqueous chemical dosimeters (B lt A)
  • Organic or plastic scintillators (B lt A)

4
(No Transcript)
5
Neutron Dosimeters Insensitive to ?-Rays (A B)
  • Activation of metal foils (A ? 0)
  • Fission foils (A 0)
  • Etchable plastic foils (A ? 0)
  • Damage to silicon diodes (A ? 0)
  • Hurst proportional counter (A ? 0)
  • Rem meters
  • Long counters
  • Bubble detectors

6
Gamma-Ray Dosimeters with Relatively Low Neutron
Sensitivity (B lt A)
  • There are no known dosimeters for which B 0
    while A ? 0
  • The primary means available for minimizing the
    value of B is the avoidance of hydrogen in a
    dosimeter, including its CPE buildup layer, since
    elastic scattering of H nuclei accounts for most
    of the absorbed dose in the interaction of fast
    neutrons in tissue and other hydrogenous media

7
Non-hydrogenous Ion Chambers
  • Graphite-walled ion chambers through which CO2
    gas is flowed at 1 atm have the advantage of
    being low in atomic number, thus avoiding
    overresponse for low-energy ? rays due to the
    photoelectric effect
  • However, the discrimination against neutrons is
    only moderate, with B/A ? 0.30 at 15 MeV for a
    0.3 cm3 cylindrical chamber, decreasing gradually
    as the neutron energy is increased

8
Non-hydrogenous Ion Chambers
  • Somewhat better neutron discrimination can be
    achieved with a magnesium chamber containing
    argon, because of the decrease in the energy
    transferred to the heavier nuclei by neutron
    elastic scattering
  • For a 2.4-cm3 spherical Mg-Ar chamber the B/A
    value for 14.8-MeV neutrons is about 0.17

9
Thermoluminescent Dosimeters
  • 7LiF (TLD-700) and CaF2Mn TLDs both have B/A
    values comparable to that of the Mg-Ar ion
    chamber
  • Thus either of these TLDs can be employed as the
    neutron-insensitive member of the
    paired-dosimeter method
  • 7LiF, at least, has been shown to have a B/A
    value that is nearly proportional to the energy
    of the fast neutrons below 15 MeV

10
Thermoluminescent Dosimeters
  • A LiF (TLD-100) or 6LiF (TLD-600) TLD can be
    employed as an indirect fast-neutron dosimeter by
    coupling it with a large moderating mass, for
    example, by wearing it in a personnel badge on
    the body
  • The incident fast neutrons become thermalized by
    multiple elastic collisions in the body and some
    of them diffuse back out to the dosimeter
  • This is called an albedo dosimeter because its
    reading depends on the ability of the body to
    reflect the thermalized neutrons

11
(No Transcript)
12
Thermoluminescent Dosimeters
  • Since 6LiF and LiF (containing natural lithium
    with 7 6Li content) both are sensitive to ? rays
    also, it is usually necessary to provide a second
    TLD in the dosimeter package that is insensitive
    to thermal neutrons
  • Both dosimeters in the pair require ?-ray
    calibration, as their ?-ray sensitivities are
    seldom identical

13
Thermoluminescent Dosimeters
  • An alternative to using 7LiF as a separate ?-ray
    dosimeter in the albedo package is offered by the
    fact that LiF and 6LiF show an extra TLD glow
    peak at about 250 300 C, produced by the
    thermal-neutron dose deposited by the secondary
    ?-particle and the triton

14
(No Transcript)
15
X-Ray Film
  • Nuclear-track emulsions are thick enough to allow
    fast neutrons to scatter protons elastically, and
    to allow them to spend their energy internally in
    producing chemically developable tracks
  • An x-ray film has an emulsion thickness of 2 5
    mg/cm2, which is comparable to the range of a
    1-MeV proton
  • If the film is sandwiched between Pb foils to
    keep out protons from the films surroundings,
    B/A can be reduced to even lower levels than
    those exhibited by 7LiF

16
Miniature G-M Counters
  • A miniature stainless-steel G-M counter with a
    high-Z filter to flatten the energy dependence of
    the ?-ray response has been found to have the
    lowest B/A ratio of any known ?-ray dosimeter
    approximately 0.02 for 15-MeV neutrons,
    decreasing gradually with decreasing neutron
    energy

17
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • The ?-ray calibration factor A is first obtained
    from a 60Co ?-ray beam for which the free-space
    exposure rate is known
  • The absorbed dose at the center of an equilibrium
    sphere of tissue, 0.52 g/cm2 in radius, for a
    free-space exposure X (C/kg) at the same
    location, is given (in grays) by

18
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • where ? ? 1.003,
  • Aeq attenuation of photons in
    penetrating to the center of the tissue sphere ?
    0.988,
  • 33.97 J/C, and
  • the ratio of mass energy
    absorption coefficients for tissue/air,
    0.0293/0.0266 1.102

19
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • Eq. (3) thus reduces to
  • If (Q?)TE is the charge (C) produced in the TE
    ion chamber when it is given the same
    ?-irradiation that deposits D? (Gy) in the tissue
    sphere, then

20
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • The absorbed dose D? in muscle tissue can be
    related to the dose (D? )TE in the TE plastic
    chamber wall under TCPE conditions by

21
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • The B-G relation, assumed to be valid here,
    allows one to write
  • Substituting gives

22
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • The neutron calibration factor BTE for the TE ion
    chamber can next be expressed in a form similar
    to that of ATE

23
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • Applying the B-G relation to the neutron case
  • Now substituting gives

24
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • (B/A)TE for the TE chamber is the ratio
  • The compositions of the TE gas and TE-plastic
    wall are sufficiently similar that the stopping
    power ratios are both close to unity, as is their
    ratio

25
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • The ratio is also nearly unity,
    since differences in the carbon and oxygen
    content in the gas and wall have no effect
  • These elements have practically identical ?en/?
    values over the wide range of ?-ray energies
    where the Compton effect dominates
  • Therefore, for the TE-gas-filled TE-plastic
    chamber

26
Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
  • The value of the ratio is obtained from
    tables such as those in Appendix F, entered at
    the appropriate neutron energy for A-150 plastic
    and ICRU muscle
  • The reciprocal of the W -ratio has been computed
    as a function of neutron energy by Goodman and
    Coyne for methane-based TE gas

27
(No Transcript)
28
Calibration of the Low-Neutron-Sensitivity
Dosimeter
  • In principle one could use
  • to calculate B/A for a graphite-CO2 or Mg-Ar
    ion chamber to be employed in the
    paired-dosimeter method
  • The resulting B/A value so obtained is seldom
    accurate enough to be useful, especially where
    the ?-ray content is fairly low

29
Calibration of the Low-Neutron-Sensitivity
Dosimeter
  • The most practical approach to determining B/A is
    an experimental one employing a narrow neutron
    beam of the desired spectrum
  • The method makes use of a Pb filter to remove the
    ?-ray contamination from the beam, while passing
    most of the neutrons, which have a smaller
    attenuation coefficient
  • Secondary radiation produced in the filter
    escapes from the narrow beam

30
Calibration of the Low-Neutron-Sensitivity
Dosimeter
  • A previously calibrated TE chamber is used to
    calibrate the beam in terms of neutron tissue
    dose Dn
  • The low-neutron-sensitivity dosimeter (x) for
    which the value of (B/A)x is to be determined is
    given an identical irradiation, yielding the
    reading Qx
  • Bx is simply equal to Qx/Dn, assuming D? to be
    zero
  • Ax for that dosimeter is obtained from a 60Co
    ?-ray exposure

31
Calibration of the Low-Neutron-Sensitivity
Dosimeter
  • In practice one does not know the degree to which
    the beam is initially contaminated with ?
    radiation, how much Pb filtration is needed to
    purify the beam adequately, or how much of the
    ?-ray contamination may have come from elsewhere
    than the beam port
  • Gamma rays from the face of the shield would, for
    example, not be removed by a beam filter

32
Calibration of the Low-Neutron-Sensitivity
Dosimeter
  • A solution to this problem was devised which uses
    the narrow-beam Pb-filtration method for
    determining (B/A)x
  • The neutron beam was generated by 35-MeV
    deuterons on Be its average energy was 15 MeV
  • It was collimated by a 2-cm hole through a large
    Benelex (pressed wood) shield, as shown in the
    following diagram

33
(No Transcript)
34
Calibration of the Low-Neutron-Sensitivity
Dosimeter
  • The dosimeters were a TE-plastic-TE-gas chamber
    and an air-filled graphite chamber
  • The three beam filtrations chosen were open beam,
    7.6-cm Pb, and a steel plug 66 cm long filling
    the entire bore hole
  • The six measurements and response equations are
    listed in the following table

35
(No Transcript)
36
Calibration of the Low-Neutron-Sensitivity
Dosimeter
  • was the absorbed dose at the measurement
    point in the open beam due to ? rays coming out
    of the beam port
  • was the dose contributed by ? rays from
    elsewhere mostly H-capture ? rays emitted from
    the face of the Benelex shield
  • Dn is the open-beam neutron dose, is that
    with the Pb filter, and that with the plug
    in place

37
(No Transcript)
38
Calibration of the Low-Neutron-Sensitivity
Dosimeter
  • This experimental approach to determining (B/A)x
    for a low-neutron-sensitivity dosimeter provides
    a value that is consistent with the (B/A)TE of
    the tissue-equivalent chamber with which it is
    compared, and is relevant to the neutron spectrum
    of the beam used
  • The method works as well with TLDs, G-M counters
    or other nonhydrogenous dosimeters as it does
    with ion chambers

39
Calibration of the Low-Neutron-Sensitivity
Dosimeter
  • Narrow-beam geometry is required for this
    calibration procedure
  • The beam must be narrow enough, and the
    measurement location distant enough from the
    filters, so that significant amounts of secondary
    radiation from the filters cannot reach the
    dosimeters
  • The method therefore requires a collimatable beam
    of neutrons
Write a Comment
User Comments (0)
About PowerShow.com