Title: Neutron Interactions and Dosimetry II
1Neutron Interactions and Dosimetry II
- Paired Dosimeters
- Calibration of the Low-Neutron-Sensitivity
Dosimeter
2Separate Measurement of Neutron and ?-Ray Dose
Components by Paired Dosimeters
- If a mixed n ? field is measured by means of
two dosimeters having different values of B/A,
Eq. (2) can then be applied to each one and
solved simultaneously to obtain D? and Dn, so
long as B and A have known values - The best dosimeter pair is a TE-plastic ion
chamber containing TE gas (for which B/A ? 1) to
measure the total n ? dose, and a
nonhydrogenous dosimeter having as little neutron
sensitivity as possible to measure the ? dose - Ideally this dosimeter should measure only ?-rays
3Dosimeters with Comparable Neutron and ?-Ray
Sensitivities (B/A ? 1)
- A-150 TE plastic ion chambers (B/A ? 1)
- Rossi TE proportional counter (B/A ? 1)
- Tissue-equivalent plastic calorimeters (B/A ? 1)
- Aqueous chemical dosimeters (B lt A)
- Organic or plastic scintillators (B lt A)
4(No Transcript)
5Neutron Dosimeters Insensitive to ?-Rays (A B)
- Activation of metal foils (A ? 0)
- Fission foils (A 0)
- Etchable plastic foils (A ? 0)
- Damage to silicon diodes (A ? 0)
- Hurst proportional counter (A ? 0)
- Rem meters
- Long counters
- Bubble detectors
6Gamma-Ray Dosimeters with Relatively Low Neutron
Sensitivity (B lt A)
- There are no known dosimeters for which B 0
while A ? 0 - The primary means available for minimizing the
value of B is the avoidance of hydrogen in a
dosimeter, including its CPE buildup layer, since
elastic scattering of H nuclei accounts for most
of the absorbed dose in the interaction of fast
neutrons in tissue and other hydrogenous media
7Non-hydrogenous Ion Chambers
- Graphite-walled ion chambers through which CO2
gas is flowed at 1 atm have the advantage of
being low in atomic number, thus avoiding
overresponse for low-energy ? rays due to the
photoelectric effect - However, the discrimination against neutrons is
only moderate, with B/A ? 0.30 at 15 MeV for a
0.3 cm3 cylindrical chamber, decreasing gradually
as the neutron energy is increased
8Non-hydrogenous Ion Chambers
- Somewhat better neutron discrimination can be
achieved with a magnesium chamber containing
argon, because of the decrease in the energy
transferred to the heavier nuclei by neutron
elastic scattering - For a 2.4-cm3 spherical Mg-Ar chamber the B/A
value for 14.8-MeV neutrons is about 0.17
9Thermoluminescent Dosimeters
- 7LiF (TLD-700) and CaF2Mn TLDs both have B/A
values comparable to that of the Mg-Ar ion
chamber - Thus either of these TLDs can be employed as the
neutron-insensitive member of the
paired-dosimeter method - 7LiF, at least, has been shown to have a B/A
value that is nearly proportional to the energy
of the fast neutrons below 15 MeV
10Thermoluminescent Dosimeters
- A LiF (TLD-100) or 6LiF (TLD-600) TLD can be
employed as an indirect fast-neutron dosimeter by
coupling it with a large moderating mass, for
example, by wearing it in a personnel badge on
the body - The incident fast neutrons become thermalized by
multiple elastic collisions in the body and some
of them diffuse back out to the dosimeter - This is called an albedo dosimeter because its
reading depends on the ability of the body to
reflect the thermalized neutrons
11(No Transcript)
12Thermoluminescent Dosimeters
- Since 6LiF and LiF (containing natural lithium
with 7 6Li content) both are sensitive to ? rays
also, it is usually necessary to provide a second
TLD in the dosimeter package that is insensitive
to thermal neutrons - Both dosimeters in the pair require ?-ray
calibration, as their ?-ray sensitivities are
seldom identical
13Thermoluminescent Dosimeters
- An alternative to using 7LiF as a separate ?-ray
dosimeter in the albedo package is offered by the
fact that LiF and 6LiF show an extra TLD glow
peak at about 250 300 C, produced by the
thermal-neutron dose deposited by the secondary
?-particle and the triton
14(No Transcript)
15X-Ray Film
- Nuclear-track emulsions are thick enough to allow
fast neutrons to scatter protons elastically, and
to allow them to spend their energy internally in
producing chemically developable tracks - An x-ray film has an emulsion thickness of 2 5
mg/cm2, which is comparable to the range of a
1-MeV proton - If the film is sandwiched between Pb foils to
keep out protons from the films surroundings,
B/A can be reduced to even lower levels than
those exhibited by 7LiF
16Miniature G-M Counters
- A miniature stainless-steel G-M counter with a
high-Z filter to flatten the energy dependence of
the ?-ray response has been found to have the
lowest B/A ratio of any known ?-ray dosimeter
approximately 0.02 for 15-MeV neutrons,
decreasing gradually with decreasing neutron
energy
17Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- The ?-ray calibration factor A is first obtained
from a 60Co ?-ray beam for which the free-space
exposure rate is known - The absorbed dose at the center of an equilibrium
sphere of tissue, 0.52 g/cm2 in radius, for a
free-space exposure X (C/kg) at the same
location, is given (in grays) by -
18Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- where ? ? 1.003,
- Aeq attenuation of photons in
penetrating to the center of the tissue sphere ?
0.988, - 33.97 J/C, and
- the ratio of mass energy
absorption coefficients for tissue/air,
0.0293/0.0266 1.102
19Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- Eq. (3) thus reduces to
- If (Q?)TE is the charge (C) produced in the TE
ion chamber when it is given the same
?-irradiation that deposits D? (Gy) in the tissue
sphere, then
20Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- The absorbed dose D? in muscle tissue can be
related to the dose (D? )TE in the TE plastic
chamber wall under TCPE conditions by
21Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- The B-G relation, assumed to be valid here,
allows one to write - Substituting gives
22Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- The neutron calibration factor BTE for the TE ion
chamber can next be expressed in a form similar
to that of ATE
23Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- Applying the B-G relation to the neutron case
- Now substituting gives
24Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- (B/A)TE for the TE chamber is the ratio
- The compositions of the TE gas and TE-plastic
wall are sufficiently similar that the stopping
power ratios are both close to unity, as is their
ratio
25Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- The ratio is also nearly unity,
since differences in the carbon and oxygen
content in the gas and wall have no effect - These elements have practically identical ?en/?
values over the wide range of ?-ray energies
where the Compton effect dominates - Therefore, for the TE-gas-filled TE-plastic
chamber -
26Calibration of a Tissue-Equivalent Ion Chamber
for n ? Dosimetry
- The value of the ratio is obtained from
tables such as those in Appendix F, entered at
the appropriate neutron energy for A-150 plastic
and ICRU muscle - The reciprocal of the W -ratio has been computed
as a function of neutron energy by Goodman and
Coyne for methane-based TE gas
27(No Transcript)
28Calibration of the Low-Neutron-Sensitivity
Dosimeter
- In principle one could use
- to calculate B/A for a graphite-CO2 or Mg-Ar
ion chamber to be employed in the
paired-dosimeter method - The resulting B/A value so obtained is seldom
accurate enough to be useful, especially where
the ?-ray content is fairly low
29Calibration of the Low-Neutron-Sensitivity
Dosimeter
- The most practical approach to determining B/A is
an experimental one employing a narrow neutron
beam of the desired spectrum - The method makes use of a Pb filter to remove the
?-ray contamination from the beam, while passing
most of the neutrons, which have a smaller
attenuation coefficient - Secondary radiation produced in the filter
escapes from the narrow beam
30Calibration of the Low-Neutron-Sensitivity
Dosimeter
- A previously calibrated TE chamber is used to
calibrate the beam in terms of neutron tissue
dose Dn - The low-neutron-sensitivity dosimeter (x) for
which the value of (B/A)x is to be determined is
given an identical irradiation, yielding the
reading Qx - Bx is simply equal to Qx/Dn, assuming D? to be
zero - Ax for that dosimeter is obtained from a 60Co
?-ray exposure
31Calibration of the Low-Neutron-Sensitivity
Dosimeter
- In practice one does not know the degree to which
the beam is initially contaminated with ?
radiation, how much Pb filtration is needed to
purify the beam adequately, or how much of the
?-ray contamination may have come from elsewhere
than the beam port - Gamma rays from the face of the shield would, for
example, not be removed by a beam filter
32Calibration of the Low-Neutron-Sensitivity
Dosimeter
- A solution to this problem was devised which uses
the narrow-beam Pb-filtration method for
determining (B/A)x - The neutron beam was generated by 35-MeV
deuterons on Be its average energy was 15 MeV - It was collimated by a 2-cm hole through a large
Benelex (pressed wood) shield, as shown in the
following diagram
33(No Transcript)
34Calibration of the Low-Neutron-Sensitivity
Dosimeter
- The dosimeters were a TE-plastic-TE-gas chamber
and an air-filled graphite chamber - The three beam filtrations chosen were open beam,
7.6-cm Pb, and a steel plug 66 cm long filling
the entire bore hole - The six measurements and response equations are
listed in the following table
35(No Transcript)
36Calibration of the Low-Neutron-Sensitivity
Dosimeter
- was the absorbed dose at the measurement
point in the open beam due to ? rays coming out
of the beam port - was the dose contributed by ? rays from
elsewhere mostly H-capture ? rays emitted from
the face of the Benelex shield - Dn is the open-beam neutron dose, is that
with the Pb filter, and that with the plug
in place
37(No Transcript)
38Calibration of the Low-Neutron-Sensitivity
Dosimeter
- This experimental approach to determining (B/A)x
for a low-neutron-sensitivity dosimeter provides
a value that is consistent with the (B/A)TE of
the tissue-equivalent chamber with which it is
compared, and is relevant to the neutron spectrum
of the beam used - The method works as well with TLDs, G-M counters
or other nonhydrogenous dosimeters as it does
with ion chambers
39Calibration of the Low-Neutron-Sensitivity
Dosimeter
- Narrow-beam geometry is required for this
calibration procedure - The beam must be narrow enough, and the
measurement location distant enough from the
filters, so that significant amounts of secondary
radiation from the filters cannot reach the
dosimeters - The method therefore requires a collimatable beam
of neutrons