Title: Shpolskii Spectroscopy
1Shpolskii Spectroscopy
- Analytical potential of fluorescence spectroscopy
often limited by unresolved band structure (5-50
nm) - homogeneous band broadening depends directly on
radiative deactivation properties of the excited
state (usually 10-3 nm) - inhomogeneous band broadening various analyte
microenvironments yields continuum of bands
(usually few nm) - Solution Incorporate molecules in rigid matrix
at low temperature to minimize broadening - Result Very narrow luminescence spectra with
each band representing different substitution
sites in the host crystalline matrix
2Shpolskii Spectroscopy
- Requirements
- T lt 77K with rapid freezing rate
- Matrix with dimension match
- Low analyte concentration
- Instrumentation
- Xe lamp excitation
- Cryogenerator with sample cell
- High resolution monochromator with PMT
Analytes polycyclic aromatic compounds in
environmental, toxicological, or geochemical
systems
Garrigues and Budzinski, Trends in Analytical
Chemistry, 14 (5), 1995, pages 231-239.
3Shpolskii Spectroscopy
Garrigues and Budzinski, Trends in Analytical
Chemistry, 14 (5), 1995, pages 231-239.
4Fluorescence Microscopy
Need 3 filters Exciter Filters Barrier
Filters Dichromatic Beamsplitters
http//microscope.fsu.edu/primer/techniques/fluore
scence/filters.html
5Are you getting the concept?
You plan to excite catecholamine with the 406 nm
line from a Hg lamp and measure fluorescence
emitted at 470 15 nm. Choose the filter cube
you would buy to do this. Sketch the transmission
profiles for the three optics.
http//microscope.fsu.edu/primer/techniques/fluore
scence/fluorotable3.html
6Fluorescence Microscopy Objectives
Image intensity is a function of the objective
numerical aperture and magnification
Fabricated with low fluorescence glass/quartz
with anti- reflection coatings
http//micro.magnet.fsu.edu/primer/techniques/fluo
rescence/anatomy/fluoromicroanatomy.html
7Fluorescence Microscopy Detectors
No spatial resolution required PMT or
photodiode Spatial resolution required CCD
http//micro.magnet.fsu.edu/primer/digitalimaging/
digitalimagingdetectors.html
8Epi-Fluorescence Microscopy
- Light Source - Mercury or xenon lamp (external
to reduce thermal effects) - Dichroic mirror reflects one range of
wavelengths and allows another range to pass. - Barrier filter eliminates all but fluorescent
light.
http//web.uvic.ca/ail/techniques/epi-fluor.jpg
http//micro.magnet.fsu.edu/primer/techniques/fluo
rescence/fluorosources.html
9Special Fluorescence Techniques
TIRF
Langmuir 2009, 25, 2563-2566
http//microscopy.fsu.edu/primer/techniques/fluore
scence/tirf/tirfintro.html
10Fluorescence Resonance Energy Transfer (FRET)
11Photoactivated Localization Microscopy
Left Viewing a mitochondrion using conventional
diffraction-limited microscopy offers a
resolution (200 nanometers) barely sufficient to
visualize the mitochondrial internal membranes.
Right Viewing the same mitochondrion by imaging
sparsely activated fluorescent molecules one at a
timeusing PALMprovides much better resolution
(20 nanometers), producing a detailed picture of
the mitochondrions internal membranes.
http//www.hhmi.org/news/palm20060810.html
- http//www.hhmi.org/bulletin/nov2006/upfront/image
.html