Title: Sensors in Wireless Sensor Networks
1Sensors in Wireless Sensor Networks
MentorProf. Dr. Veljko Milutinovic
vm_at_etf.bg.ac.yu
Goran Rakocevic goxy_83_at_yahoo.com
1/67
2Agenda
- Introduction
- Microsensing principles
- Off the shelf (available sensors)?
- In research (not yet available, but in
preparation)? - On the wish list (for someone to make)?
2/67
3(No Transcript)
4Classification of sensors
- Based on the function
- that the sensor performs
- Mechanical
- Thermal
- Electrical
- Magnetic
- Radiant
- Chemical (including bio-chemical)?
4/67
5Trends in sensor technology
- Miniaturization
- Integration (sensor, signal processing and
actuator)? - sensor with signal processing circuits for
linearising sensor output, etc. - sensor with built-in actuator for automatic
calibration, change of sensitivity - etc.
- Sensor arrays
- one-function units (to improve reliability)?
- multiple-function units
5/67
6Microsensors
- Microsensor A miniature electronic device that
detects information about a specific variable - Why microsensors?
- lower manufacturing cost (mass-production, less
materials)? - wider exploitation of IC technology
(integration)? - wider applicability to sensor arrays
- lower weight (greater portability)?
6/67
7Two Families of Microsensors
- Those which were Born with semiconductors
- Photodetectors UV, Vis, IR,
- Hall effect
- Optical
- Magnetic
- Those which are feasible without
microtechnologies - Pressure, accelero, gyro
- Gas, Ph, Ions
- DNA analyses
- Mechanical
- Chemical
- Biological
- (MEMS)?
7/67
8Pressure sensors
- First microsensors developed by the industry
- Low production costs, high sensitivity
- Piezoresistive
- Membrane sensors
- deflection of the membrane
- change in the resonance frequency
- Planar comb structures
- Optical methods 33(Mach-Zehnder interferometer)?
8/67
9Piezoresistive pressure sensor
- Piezoresistors integrated in the membrane
- Pressure deflects the membrane
- Resistance changes proportional to deflection and
thus to pressure - Resistance change measured with Wheatstone bridge
9/67
10Capacitive membrane pressure sensor
- Membrane deflects when pressure is applied
- Distance between the electrodes changes
- Capacitance changes
- Capacitive sensors have
- no hysteresis
- better long-term stability and
- higher sensitivity
- but higher production costs
10
11Capacitive pressure sensor, based on comb
structure
- Utilizes parallel comb structure
- Force is applied parallel to the sensor surface
- Force is transformed into displacementgt change
in capacitance - On one side capacitance increases and on the
other side decreases gt higher linearity and
sensitivity
11/67
12Mach-Zehnder interferometer
- Laser light brought into the sensor by optical
fiber - Light is split to two beams
- One light beam crosses a micromembrane which is
deformed by pressure - The deformation changes light properties
- The beams are combined and brought a photodiode
- Different propagation speeds result in phase
shift
12/67
13Position and speed microsensors
- The most significant are
- Contact-free optical
- magnetic methods are
13/67
14Magnetic sensor to measure angular displacement
- Hall sensor based measurement of angular
displacement - Rotor with a row of teeth
- Stator contains Hall sensors
- Permanent magnet located under the sensors
- Teeth passing by a Hall sensor change magnetic
field
14/67
15Capacitive angular speed sensor
- The fork arrangement is used as a resonator
- The resonator starts to oscillatewhen magnetic
field and alternating current are
applied(Lorentz force)? - The amplitude of the swing angleis detected by
the capacitance change between movable and fixed
electrodes
15/67
16Acceleration microsensors
- Usually detected with capacitive and
piezoresistive methods - An elastic cantilever where a mass is attached is
mostly used - Under acceleration mass displaces the cantilever
- Deflection of the cantilever is detected
- By increasing the mass sensitivity can be
increased
16/67
17Capacitive cantilever microsensor
- Sensor consists of cantilevers acting as one
electrode, an electrode strip and a contact
strip - Sawtooth voltage applied to gradually increase
the electrostatic force - Finally cantilever touches the contact strip
- Acceleration affects the magnitude of the voltage
that is required for contact
Cantilever length 120 - 500 µm Sensitivity 0.6
- 100 mV/g Fabrication dry etching
17/67
18Piezoresistive microsensorwith oil damping
- Sensor consists of cantilever beams, a seismic
mass and oil. - Oil dampens the resonance of the suspended mass
18/67
19Temperature sensors
- Thermoresistive sensors
- Resistive Temperature Devices (RTD)?
- Thermistors
- Thermoelectric sensors
- The Seebeck effect
- The Peltier effect
- Thermocouples
- The p-n junction sensors
19/67
20Applications of temperature sensors
- Important role in monitoring systems
- process industry
- environmental protection
- medicine
- body temperature (oral, nasal, rectal,
surface,catheter and tympanic, ventilator
airway, and needle temperature probes)? - Heating and air conditioning systems
- Indirect measurement of other parameters, e.g.
in flow sensors - Error compensation for temperature dependent
sensors and actuators
20/67
21Thermoresistive sensors
- Based on materials whose resistance changes in
accordance with temperature - Resistance Temperature Detectors (RTDs)?
- The material is a metal
- Platinum, Nickel, Copper are typically used
- Thermistors (thermally sensitive resistor)?
- The material is a semiconductor
- A composite of a ceramic and a metallic oxide
(Mn, Co, Cu or Fe)? - Typically have negative temperature coefficients
(NTC thermistors)?
21/67
22Thermoelectric sensors
- The Seebeck effect
- When a pair of dissimilar metals are joined at
one end, and there is a temperature difference
between the joined ends and the open ends,
thermal emf is generated, which can be measured
in the open ends - The Peltier effect
- When a current passes through the junction of two
different conductors, heat can be either
absorbed or released depending on the direction
of current flow - Thermocouples
- Based on the Seebeck effect
- Open ends must be kept at a constant reference
temperature TREF - A number of standard TCs are used
- These are denominated with different letter
codes T, J, K, S, R - i.e, type J (the most popular) is made of Iron
and Constantan (Cu/Ni alloy 57/43)?
22/67
23Chemical sensors
- Detect presence or concentrationof a chemical
substance - Applications
- medical diagnostics
- nutritional science
- environmental protection
- automobile industry
- About 60 of chemical sensorsare gas sensors
23/67
24Chemical sensors
- Conventional measurement methods are often very
complicated and expensive, require laboratory
conditions, etc. - Objectives of microsensors
- small and inexpensive
- mass-produced
- accurate and robust
- use only small amount of reagents
- short response times
24/67
25Chemical sensors
- Research trends (in addition to the development
of sensor units) - integration of sensors into measurement systems
- integration of several types of sensors
- microsystems with several identical sensors
(local analysis of a substance,distribution of
a parameter over a certain domain)? - Sensor principles
- potentiometer principle in connection with FET
- acoustic sensors
- optical sensors
25/67
26Structure of a chemical sensor system
- A sensitive layer is in contact with the
substance - Chemical reaction occurs on the sensitive layer
- Due to the reaction physical, optical, acoustic
or dielectric properties are changed - Transducer transforms the signal into electrical
form
26/67
27Interdigital transducer sensors
- Interdigital transducers using capacitive
measurement are often used in chemical sensors - The capacitance can be adjusted by changing the
dielectric properties of the sensitive layer - E.g. resistance of SnO2 sensitive layer changes
when it interacts with certain substances
27/67
28Optical sensor principle
- Optical sensors are inexpensive,easy to
sterilize, can handle small samples and are
highly sensitive - Coupling grid detector
- substance to be analyzed is in directcontact
with the waveguide - depending on the concentration of the substance
its index of refraction variesgt amount of light
striking the sensor depends on the
concentration
28/67
29Ion sensitive FET sensor
- For continuous measurement ofpH value and gases
in blood (O2, CO2)? - A device for external use to makeon-line
diagnosis of a patient - Consists of a sensor, a bloodsampling and
processing part - Uses ion sensitive FET gatepotential is
proportional to gas concentration
29/67
30Conductivity sensors
- Absorption of gases modifies the conductivity of
sensing layer - Sensing layer types
- Metal Oxide
- Typically SnO2 doped with Pt or Pd
- Operate at high temperatures (300-5000C)?
- Particularly suitable for combustible gases
- Conducting Polymers
- Based on pyrrole, aniline or thiophene
- Operate at room temperature
- CPs vs MOXs
- CP advantages
- Large number of polymers available with various
selectivities - Sensitivity to wide number of VOCs
- Low power consumption
- Faster response and recovery times
- CP Limitations
- Cross-sensitivity to humidity
- Lower sensitivity than MOXs
30/67
31Piezo-electric chemical sensors
- Piezo-electric effect
- The generation of an electric charge by a
crystalline material upon subjecting it to
stress (or the opposite)? - A typical piezo-electric material is Quartz
(SiO2)? - Piezo-electric sensors
- Thin, rubbery polymer layer on a piezo-electric
substrate - Sensing principle mass and viscosity changes in
the sensing membrane with sorption of VOCs - Surface Acoustic Wave (SAW)?
- AC signal (30-300MHz) applied to interdigitated
input electrode generates a surface (Rayleigh)
wave - Propagation delays to output electrode are
affected by changes in the surface properties - Phase shifts of the output electrode signal are
used as a response - Quartz Crystal Microbalance (QMB)?
- Also known as Bulk Acoustic Wave (BAW) devices
- Device is operated in an oscillator circuit
- Changes in the sensing membrane affect the
resonant frequency (5-20MHz) of the device
31/67
32Biosensors
- Measurement principle is similar as with chemical
sensors - Sensitive layer is biologically sensitive,
containing e.g. enzymes or antibodies - Interaction between the molecules of the
bioelementand the molecules of the substance
changes a physical or chemical parameter - Parameter change is converted into electrical
signal - Signal represents concentration to be measured
32/67
33Applications of biosensors
- Biological and nutritional research
- to detect e.g. heavy metals or allergens
- Medical applications
- patient data recording for correct and quick
diagnosis during surgery - Integration of biosensors with microfluidic
componentsgt very small analyzers - Difficulties
- immobilization of proteins
- proteins are not stable for a very long time
33/67
34Metabolism sensors
- Use bio-sensitive enzymes to catalyze a chemical
reaction - Phosphate measurement
- enzyme NP detects phosphate and triggers
chemical reaction - one product of the reaction HX is transformed
into XO in another chemical reaction after
consuming oxygen - amount of oxygen can be measured using a
chemical sensor - phosphate concentration is proportionalto the
amount of consumed oxygen
34/67
35Immuno-sensors
- Antibody is a bio-sensitive element
- Immobilized antibody moleculesbond with antigen
molecules in the substance (lock and key)? - The concentration of antigens can be measured
using for example interferometric method (light
intensity changes)?
35/67
36Pulse oximetry
- Based on the red and infrared light absorption
characteristics of oxygenated and deoxygenated
hemoglobin - Oxygenated hemoglobin absorbs more infrared light
and allows more red light to pass through - Deoxygenated (or reduced) hemoglobin absorbs more
red light and allows more infrared light to pass
through
- The emitter and photodetector are opposite of
each other with the measuring site in-between - After the transmitted red (R) and infrared (IR)
signals pass through the measuring site and are
received at the photodetector, the R/IR ratio is
calculated - The R/IR is compared to a "look-up" tables that
convert the ratio to an SpO2 value
36/67
37Pulse oximetry
- A major advancement from the oximeters of the
'70s was the inclusion of arterial pulsation to
differentiate the light absorption in the
measuring site due to skin, tissue, and venous
blood from that of arterial blood - Constant light absorbers at the measuring site
(skin, tissue, venous blood, arterial blood)? - With each heart beat there is a surge of arterial
blood, which momentarily increase arterial blood
volume - Result (when received signals are looked at 'as a
waveform') peaks with each heartbeat and
troughs between heartbeats - The light absorption at the trough is subtracted
from the light absorption at the peakthe
resultants are the absorption characteristics due
to added volume of blood only which is arterial - The advent of "Next Generation" pulse oximetry
technology has demonstrated significant
improvements in the ability to read through
motion and low perfusion
37/67
38Pulse oximetry- limitatons
- Not a complete measure of respiratory
sufficiency. A patient suffering from
hypoventilation given 100 oxygen can have
excellent blood oxygen levels while still
suffering from respiratory acidosis due to
excessive carbon dioxide - Not a complete measure of circulatory
sufficiency. If there is insufficient bloodflow
or insufficient hemoglobin in the blood, tissues
can suffer hypoxia despite high oxygen saturation
in the blood. - A higher level of methemoglobin will tend to
cause a pulse oximeter to read closer to 85
regardless of the true level of oxygen
saturation. - Carbon monoxide concentrations in human blood,
are normally insignificant, but significant
levels will follow smoke inhalation.Carboxyhemogl
obin prevents oxygen binding to hemoglobin, yet
being bright red, causes over reading of oxygen
saturationPulse oximetry should be avoided where
significant amounts of carbon monoxide have been
inhaled - A CO-oximeter measures absorption at additional
wavelengths to distinguish CO from O2 and
determines the blood oxygen saturation more
reliably.
38/67
39Continuous blood glucose monitors
- A continuous blood glucose monitor determines
blood glucose levels on a continuous basis - A disposable glucose sensor placed just under the
skin is worn for a few days until replacement. - A link from the sensor to a non-implanted
transmitter communicates to a radio receiver. - An electronic receiver is worn like a pager that
displays blood glucose levels, as well as
monitors rising and falling trends in glycemic
excursions - Continuous blood glucose monitors measure the
glucose level of interstitial fluid. - Disadvantages
- continuous systems must be calibrated with a
traditional blood glucose measurement and do not
yet fully replace "fingerstick" measurements. - glucose levels in interstitial fluid lag
temporally (aprox. 5 minutes )behind blood
glucose values. - Patients therefore require traditional
fingerstick measurements for calibration
(typically twice per day) and are often advised
to use fingerstick measurements to confirm hypo-
or hyperglycemia before taking corrective action.
39/67
40Wireless sensor networks
- A wireless network consisting of spatially
distributed autonomous devices using sensors to
cooperatively monitor physical or environmental
conditions, such as temperature, sound,
vibration, pressure, pollutants, etc. - Sensors in wireless sensor networks
- usually microsenzors
- must not require too much power
40/67
41Off the shelf
- Wireless sensor systems are commercially
available from - Crossbow Berkeley
- Microstrain
- A number of sensors has been developed
- for these systems
- some are integrated in the basic versions
- others come as add-ons
-
41/67
42(No Transcript)
43Crossbow Berkeley
- 51-pin sensor boards for MICA2, MICAz and IRIS
modules - MTS series Sensor Boards
- MDA Data Acquisition Boards
- ITS Sensor Board
43/67
44(No Transcript)
45(No Transcript)
46Crossbow Berkeley - MDA
MDA300CA is an extremely versatile data
acquisition board that also includes an onboard
temperature/ humidity sensor 7 single-ended or
3 differential ADC channels 4 precise
differential ADC channels 6 digital I/O
channels with event detection interrupt
- MDA320CA is a high-performance data acquisition
board with up to 8 channels of 16-bit analog
input - 8 single-ended 0-2.5V inputs, or 4
differential 0-2.5V ADC channels - 8 digital 0-2.5V I/O channels with event
detection interrupt -
46/67
47Microstrain
- V-LINK
- SG-LIN
-
- G-LINK
- TC-LINK
- EmbedSense Wireless Sensor
47/67
48Microstrain
- V-LINK, SG-LIN, G-LINK, TC-LINK
- Come with integrated temperature sensors
- compatible with a wide range of analog sensors,
including -strain gauges, -displacement
sensors, -load cells, -torque transducers,
-pressure sensors, -accelerometers,
-geophones, -temperature sensors,
-inclinometers, -etc.
48/67
49Microstrain FAS-G gyro enhanced Inclinometer
- Inclination Angle Range 360 degrees full scale
(FS)? - Angular Velocity Range /- 300 degrees/second
(max.)? - Dynamic Compensation Closed loop digital control
(0 to 50 Hz)(angle resolution specs. taken at
most aggressive filter setting)? - Temperature Drift Single axis 0.025 /deg.C
- Output Data Rate 150 Hz (digital analog)?
- Inclination Resolution lt 0.1 degrees
- Nonlinearity 0.23 full scale (static)?
- A/D Resolution 12 bits
- D/A Resolution 12 bits
49/67
50Microstrain 3DMsolid state pitch, roll, and yaw
module
- Range yaw 180 degrees
- Angle resolution Pitch lt 0.1 degrees
- Update rate (angle mode) 30 Hz/ 3 channels
- (raw mode) 70 Hz/ 6 channels
- A/D resolution 12 bits
- Roll lt 0.1 degrees
- Yaw lt 0.1 degrees
- pitch 180 degrees
- roll 70 degrees
50/67
51EmbedSense Wireless Sensor
- Batteries are completely eliminated
- uses an inductive link to receive power from an
external coil - Sensor types
- Piezoresistive bonded foil
- semiconductor strain gauges,
- pressure/load/torque transducers,
- thermocouples
51/67
52- Microsensors that are available,
- but not yet implemented in wireless sensor nodes
- Chemical sensors
- Biosensors
- Radiation sensors
- Both the microsensor and wireless node
manufacturers - offer custom-building,
- so many of these sensors might be specially
ordered
52/67
53Chemical sensors
- Oxygen Sensors
- Sensitivity 0 - 150 mmHg
- Time Constant or Response Rate (0 - 63 of
response) 4 sec - Bias Voltage -0.7 V
- Current Response for ambient pO2 18 nA
- Thermal Drift 4 /C
- pH Sensors
- Response Time 10 sec
- Temperature Range -4?C - 75?C
- pH Range 0 14
- Resolution better 0,01 pH
- Accuracy 0,02 pH
53/67
54Chemical sensors
- CO, CO2 Sensors
- Response Time (T1/e) 20 sec diffusion time
- Measurement Range 0 - 5 000 ppmvol.
- Sensitivity 20 ppm 1 of measured value
- Accuracy 30 ppm 5 of measured value
- Pressure Dependence 1.6 reading per kPa
deviation from normal pressure, 100 kPa - NOx Sensors
- Measurement Range 0 to 100 PPM
- Operating Temperature -20 to 50o C
- Response Time lt 40s typical at 20o C
- Estimated Service Life 24 months Depends on
application - Lower Detection Limit 0.2 PPM (depends on
circuitry) - Resolution 0.2 PPM (depends on circuitry)?
54/67
55Chemical sensors
- H2S Sensors
- Measurement Range 0 to 200 PPM
- Response Time lt 60 s typical at 20o C
- Estimated Service Life 24 month (Depends on
application)? - Lower Detection Limit 2 PPM depends on circuitry
- Resolution 2 PPM depends on circuitry
- And others sucha as
- Methane Sensors
- Conductivity Sensors
- Water Quality Sensors
- Air Quality Sensors
55/67
56Example POCKET CO
- CO Digital Dosimeter Detector
- Digital carbon monoxide dosimeter detector
- Alerts users of CO exposure from 5 to 1000 PPM
- Dosimetry variables include TWA time weighted
average, - TE total exposure, MAX maximum ppm, TIME when max
occurred - 5 minute inspection mode, up to 8 hr collection
mode - 2 yr limited instrument, sensor warranty
- 85 decibel alarm at 2 feet
56/67
57Biosensors
- Detection of (bio-)molecules
- Proteins
- Allergens
- nucleic acids
- Carbohydrates
- viral particles
- Glucose monitors
- Detection of pathogens
- Example
- HiPerChip HIGH PERFORMANCE CHIP
57/67
58Radiation microsensors
- Nuclear Radiation Microsensorsfor detecting a,
and ß particles, ?-rays, x-rays - residual radiation (0.001 cGy/hr to 999 cGy/hr)?
- prompt radiation
- personal dosimeters
- Infrared Radiation Microsensors
- responsivity1,2 gt800 V/W
- noise1 lt370 nV/vHz
- specific detectivity1,2 gt2.5108 cm vHz / W
- operating temperature 20 to 70 C
58/67
59Radiation
- UV Radiation Microsensors
- Broad band UVA(315-400 nm), UVB(280-315 nm), and
UVC(100-280 nm) photodiodes - UVI type sensor (for accurate sun-UV dosimetry)?
- precision up to /- 0.5UVI
- operating temperature 20 to 70 C
- Visible and NIR Radiation Microsensors
- Illumination detectors (400-700nm)?
- Blue light detectors measure the blue light
radiation hazard to eyes - Flash detectors (detect flashes with a width from
0.1ms to 100ms,measure peak irradiance and total
effective energy)? - Micro Cameras (resolutions up to 628 X 582)?
- Thermal radiation Microsensors
- Responsivity 90.0 V/W
- Time Constant 6ms
- Operating Temperature -50-85 C
- Up to 32X32 resolution
-
59/67
60Pulse oximetry
- Oxygen Saturation Range (Sp02) 0 to 100
- Pulse Rate Range 18 to 300 pulses per minute
- Accuracy
- Blood Oxygen Saturation 70 - 100 2 digits
- Pulse Indicator 3
- Operating temperature 0 C to 50 C
- Also available (but devices somewhat larger in
size and more battery-consuming) - Carboxyhemoglobin (SpCO)?
- Methemoglobin (SpMet)?
- Perfusion Index
60/67
61Continuous blood glucose monitors
- REAL-Time display , 24 hours a day
- Accuracy (up to, depending on the system)
- Consensus Error Grid 98.9 AB
- MARD (Mean) - 19.7
- MARD (Median) - 15.6
- Sensor life FDA approved for up to 7 days
- Start-up Initialization Time 2 hours
- Waterproof transmitters, up to 8 feet for 30
minutes - Calibration usually every 12 hours
- Also available
- A system with an integrated insulin pump
- The glucose sensor transmits data directly to
the insulin pump - Helps keep blood glucose levels within
personalized target range byrecommending
necessary correction doses
61/67
62Pollen sensors
- Much smaller than conventional methods
- Pollen discrimination and count, by "Degree of
Polarization - Built-in Suction Fan
- Pollen count information can be made available
without delay by taking advantage of the real
time analysis - Compact, Light and Low Cost
62/67
63Contact Microphones
- Broad bandwidth (8Hz 2.2 KHz)?
- High sensitivity (40 V/mm)?
- Excelelent impact resistance
- Operating Temperature 5 60 C
- Light Weight
- Low Cost
- Ideal for detecting body sounds.
63/67
64In research
- Advanced gas sensors
- Micro-scale cell analysis device
- Sensors that detect chemical warfare agents,such
as sarin, tabun, sulfur, and mustard - Sensors that detect Explosives such as TNB, TNT,
and 2,4-DNT - Sensors that detect pesticides and insecticides
- Improvements on the existing microsensors(better
precision, less power consumption, smaller in
size )?
64/67
65In research
- Glucose sensing bio-implants
- longer term solutions to continuous monitoring
under development use a long-lasting
bio-implants - a minor surgical implantation of the sensor
- should last from one year to more than five years
- Non-Invasive glucose sensing
- new technologies to monitor blood glucose levels
will not require access to blood - near IR detection
- ultrasound
- dielectric spectroscopy
65/67
66On the wish list
- Detection of bio-currents
- Micro-sized sensor nodes with integrated
short-distance communication equipment
66/67
67Sensors in Wireless Sensor Networks
- Goran Rakocevic goxy_83_at_yahoo.com
- Zoran Babovic zbabovic_at_verat.net
- Mentor
- Prof. Dr. Veljko Milutinovic vm_at_etf.bg.ac.yu
67/67