Title: ALICE PMD
1ALICE PMD Pre-Production Prototype
test September 1-17, 2003
- OUTLINE
- Goals for the test
- Prototype Detectors
- T10 test setup
- Results for the pion beam
- Results for the electron beam
- Simulation results
- Calibration Curve
PMD Group (Analysis by Raghu, Rashmi, Prem and
Basanta) November 22, 2003
2Goals for the PPP test
- Optimization of Fabrication Procedure
- Final cell design optimized from previous
tests - Fabrication procedure close the real detector
- One Super Module (SM) made containing 2 UMs
- HV isolation between the two UMs
- Quality Assurance
- Optimization during Fabrication testing
- Lab tests before shipping to CERN
- Test at CERN
- First time test of both MANAS and GASSIPLEX
- Response to hadrons
- Response to electrons
- To obtain the Calibration Curve (MeV-ADC)
3Prototype Fabrication
Gas OUT
Insulation strips
HV block
32-pin FPC Connector
Gas IN
Unit Module 1
Unit Module 2
- Unit Module size is half of the actual UM
- Boundary channels are the actual ones.
4A bit more detail about the Prototypes
Number of unit modules
2 Number of sub-modules in a unit module
4 Number of cells in a unit module
2304 Number of high voltage channels
2 Number of signal (FPC) connectors
96
5PMD - Pre-production prototype test September
1-17, 2003
- T10 Beamline at PS
- Response to hadrons using pion beam at 6GeV/c
- Response to electrons is studied using the setup
shown above. A Cherenkov counter is placed
upstream for the electron Trigger. Data taken at
various energies and for different lead thickness.
6Response to pion Beam
- Pulse height spectra
- pion beam at 6GeV/c
- 1400Volts
- Data taken with both GASSIPLEX and MANAS chips
7Number of cells affected for 6GeV/c pion beam
8Mean Pulse height vs. Operating voltage
Mean Pulse height vs. Operating voltage
Similar gains for both MANAS and GASSIPLEX
9Efficiency vs. Operating voltage GASSIPLEX
MANAS
Efficiency 98.5
10Relative gain of cells
GASSIPLEX
MANAS
26 cells
41 cells
11Transverse shower spread
3GeV/c electron beam and 3X0 lead
12Pulse Height Spectra for different energy
electron beam and different lead thickness
HV1400V
13Simulation Results
Energy Deposition (EDEP) Spectra for different
energy electron beam and different lead thickness
14Calibration Curve
Energy Deposition from Data (ADC)
Energy Deposition from Simulation (EDEP)