Proteomics: Strategies for protein Identification - PowerPoint PPT Presentation

1 / 60
About This Presentation
Title:

Proteomics: Strategies for protein Identification

Description:

... the N-terminal amino acid of a protein or peptide with phenyl isothiocyanate. ... acid residue is modified, and then fails to react with phenyl isothiocyanate. ... – PowerPoint PPT presentation

Number of Views:219
Avg rating:3.0/5.0
Slides: 61
Provided by: yaote
Category:

less

Transcript and Presenter's Notes

Title: Proteomics: Strategies for protein Identification


1
Proteomics Strategies for protein Identification
  • Yao-Te Huang
  • Oct 12, 2009

2
Two major methods to determine proteins ID
  • Determine proteins ID by traditional chemical
    method (Edman degradation)
  • Determine proteins ID by mass spectrometry

3
  • Part One
  • Determine proteins ID by traditional chemical
    method (Edman degradation)

4
1A complete hydrolysis
  • The first step is to determine the amino acid
    composition of a protein.
  • The protein is hydrolyzed into its constituent
    amino acids by heating it in 6 N HCl at 110C for
    24-72 hours. Amino acids in hydrolysates can then
    be labeled with ninhydrin or fluorescamine, and
    be separated by ion-exchange chromatography on
    columns of sulfonated polystyrene.
  • The identity of the amino acid is revealed by its
    elution volume (which is the volume of buffer
    used to remove the amino acid from the column)
    and the height of the absorption peak is
    proportional to the number of times that
    particular amino acids occurs in the protein.

5
If the unknown protein is A-G-D-F-R-G
Determination of Amino Acid Composition.
Different amino acids in a protein hydrolysate
can be separated by ion-exchange chromatography
on a sulfonated polystyrene resin (such as
Dowex-50).Buffers (in this case, sodium citrate)
of increasing pH are used to elute the amino
acids from the column. The amount of each amino
acid present is determined from the absorbance.
Aspartate, which has an acidic side chain, is
first to emerge, whereas arginine, which has a
basic side chain, is the last.
6
1A complete hydrolysis (contd.)
  • Amino acids treated with ninhydrin give an
    intense blue color, except for proline, which
    gives a yellow color because it contains a
    secondary amino group.
  • The concentration of an amino acid in a solution,
    after heating with ninhydrin, is proportional to
    the optical absorbance of the solution. This
    technique can detect a microgram (10 nmol) of an
    amino acid. As little as a nanogram (10 pmol) of
    an amino acid can be detected by fluorescamine,
    which reacts with the ?-amino group to form a
    highly fluorescent product.

7
(No Transcript)
8
(No Transcript)
9
1A complete hydrolysis (contd.)
  • The complete hydrolysis gives the info of the
    amino acid composition of a protein, not the
    sequence info.
  • However, there are algorithms such as
    AACompIdent, which attempt to predict protein
    sequences on the basis of amino acid compositions
    by searching protein sequence databases for
    entries that would give a similar composition
    profile.

10
Polypeptides have characteristic amino acid
compositions
11
1B protein sequencing by Edman degradation
  • Edman degradation involves labeling the
    N-terminal amino acid of a protein or peptide
    with phenyl isothiocyanate .

12
1B protein sequencing by Edman degradation
(contd.)
  • Mild acid hydrolysis then results in the cleavage
    of the peptide bond immediately adjacent to this
    modified residue, but leaves the rest of the
    protein intact.
  • The terminal amino acid can then be identified by
    chromatography, and the procedure is repeated on
    the next residue and the next, thus building up a
    longer sequence.

13
(No Transcript)
14
1B protein sequencing by Edman degradation
(contd.)
  • It is not suitable for sequencing proteins larger
    than 50 residues in a single run because each
    cycle of degradation is less than 100 efficient.
  • This problem is addressed by cleaving large
    proteins into peptides, using either chemical
    reagents or specific endoproteases.

15
(No Transcript)
16
(No Transcript)
17
1C Edman degradation in proteomics
18
1C Edman degradation in proteomics (contd.)
  • Disadvantages (a) laborious time-consuming
    (sequencing 10 residues per day) (b) in some
    proteins, the ?amino group of the N-terminal
    amino acid residue is modified, and then fails to
    react with phenyl isothiocyanate.
  • Advantages (a) the most convenient method for
    determining the N-terminal sequence of a protein
    (b) also very sensitive method (that can sequence
    0.5-1 pmol of pure protein).

19
  • Part Two
  • Determine proteins ID by mass spectrometry

20
What is a mass spectrometer, and what does it do?
  • A mass spectrometer is an analytical device that
    determines the molecular weight of chemical
    compounds by separating molecular ions according
    to their mass-to-charge ratio (m/z).
  • The ions are generated by inducing either the
    loss or the gain of a charge (e.g., deprotonation
    or protonation).
  • Once the ions are formed they can be separated
    according to the m/z and finally detected.
  • The resulting mass spectrum may provide the info
    about MW of a chemical compound or even about its
    structural information.

21
Major components of a mass spectrometer (1)
(1) The ion source unit (including sample
introduction) in which, molecular ions are
generated, and then electrostatically
propelled into the mass resolution unit (2) the
mass resolution unit (the mass analyzer) in
which molecules ions can be resolved (separated
or filtered) according to their m/z ratios. (3)
the ion detector unit in which the signal is
detected, and transferred to a computer for
further processing.
e.g. MALDI or ESI
e.g., TOF (time-of-flight)
22
Major components of a mass spectrometer (2)
23
MALDI (Matrix-assisted laser desorption-ionization
)
  • We may pump much energy into a solid matrix (in
    which macromolecules are embedded) to ionize and
    desorb (into a vacuum) the macromolecule without
    significant degradation.
  • The best way to pump energy is through a laser
    pulse, and the matrix is chosen as a substance
    that absorbs strongly at the laser wavelength.
  • The power density required to generate a
    significant ion current corresponds to an energy
    flux of 20mJ/cm2.
  • Aromatic molecules such as 2,5-dihydroxybenzoic
    acid, which absorbs in the UV, are favorite
    matrices because of the common use of UV lasers
    in the MALDI method.
  • The pulsed nature of the excitation (from a few
    tens of nanoseconds to a few hundred
    microseconds) simplifies the data analysis
    because all molecules begin their flight in
    nearly a synchronous fashion.

24
MALDI (Matrix-assisted laser desorption-ionization
)
  • In MALDI, the analyte is first co-crystalized
    with a large molar excess of a matrix compound,
    usually a UV-absorbing weak organic acid.
  • Irradiation of this analyte-matrix mixture by a
    laser results in the vaporization of the matrix,
    which carries the analyte with it into the vapor
    phase. That is, both the matrix and any sample
    embedded in the matrix are vaporized.
  • Ionization of the analyte results from exchanges
    of electrons (or protons) with the matrix
    compound.
  • Once in the gas phase, the desorbed charged
    molecules are then directed electrostatically
    into the mass resolution unit (the mass
    analyzer).

25
MALDI (Matrix-assisted laser desorption-ionization
)
26
Commonly used MALDI matrices
27
(No Transcript)
28
ESI (electrospray ionization)
  • Charged microdroplets containing the
    macromolecules to be studied are sprayed into the
    mass spectrometer through a charged nozzle (which
    ionizes the drops that are exciting the tip).
  • As the droplets accelerate away from the tip, the
    solvent evaporates until, at some point, the
    concentration of charges is so high that the
    coulombic forces overcome the surface tension of
    the drop, resulting in dispersion of the drop
    into a spray of smaller droplets.
  • These droplets continue to evaporate and will
    themselves disperse into even finer sprays until
    all the solvent is gone, leaving the macroions
    they contained for analysis.

29
ESI (electrospray ionization)
30
ESI
31
nanoESI
  • The spray needle has been made very small, and is
    positioned close to the entrance to the mass
    analyzer.
  • The flow rates for nanoESI sources are on the
    order of tens to hundreds of nanoliters per
    minute.
  • The end result of this rather simple adjustment
    is increased efficiency, which includes a
    reduction in the amount of sample needed.
  • NanoESI is more tolerant of salts and other
    impurities (because less evaporation means the
    impurities are not concentrated down as much as
    they are in ESI)

32
nanoESI
33
Mass Analyzers having many kinds, including TOF,
quadrupoles, etc
  • Performance characteristics accuracy,
    resolution, mass range, tandem analysis
    capabilities, and scan speed

34
Accuracy
  • It is the ability with which the analyzer can
    accurately provide m/z information and is largely
    a function of an instruments stability and
    resolution.
  • For example, an instrument with 0.01accuracy can
    provide info on a 1000.00 Da peptide to 0.1 Da
    or a 10000 Da protein to 1.0 Da.
  • An alternative means of describing accuracy is
    using part per million (ppm) terminology, where
    1000.00 Da peptide to 0.1 Da could also be
    described as 1000.00 Da peptide to 100 ppm.

35
Resolution
  • It is the ability of a mass spectrometer to
    distinguish between ions of different
    mass-to-charge ratios. ResolutionM/(?M)
  • Where ?M represents the peak width at half
    maximum,
  • And M corresponds to the m/z

36
Resolution (contd.)
37
Mass Range
  • It is the m/z range of the mass analyzer. For
    instance, time-of-flight (TOF) analyzers have
    virtually unlimited m/z range, and quadrupole
    analyzers typically scan up to m/z 3000.

38
Tandem MS analysis
  • It is the ability of the analyzer to separate an
    ion, generate fragment ions from the original
    ion, and then analyze the fragmentation ions.
  • Typically tandem MS experiments are performed by
    generating the ion of interest and selecting it
    with an analyzer. The ion is then collided with
    inert gas molecules such as argon or helium, and
    the fragments generated by the collision are
    analyzed.

39
Tandem MS analysis
  • Information obtained via tandem analysis can be
    used to sequence peptides, or structurally
    characterize carbohydrates, small
    oliogonucleotides, and lipids.

40
Scan speed
  • It refers to the rate at which the analyzer scans
    over a particular mass range. Most instruments
    require seconds to perform a full scan, however
    this can vary widely depending on the analyzer.
    Time-of-flight analyzers, for example, complete
    analyses in milliseconds or less.

41
The principles of a time-of-flight (TOF) mass
spectrometer
42
The principles of a time-of-flight (TOF) mass
spectrometer
43
reflector time-of-flight (TOF)
In reflector time-of-flight(TOF) instruments,
the ions are accelerated to high kinetic energy
and are separated along a flight tube as a
result of their different velocities. The ions
are turned around in a reflector, which
compensates for slight differences in kinetic
energy, and then impinge on a detector that
amplifies and counts arriving ions.
44
Quadrupoles
Quadrupole mass spectrometers select by
time-varying RF fields between four rods, which
permit a stable trajectory only for ions of a
particular desired m/z.
45
Triple Quadrupoles
Again, ions of a particular m/z are selected in
a first section (Q1), fragmented in a collision
cell (Q2), and the fragments separated in Q3.
46
Triple quadrupoles (contd.)
  • The first quadrupole (Q1) is used to scan across
    a preset m/z range or to select an ion of
    interest.
  • The second quadrupole (Q2), also known as the
    collision cell, transmits the ions while
    introducing a collision gas (argon) into the
    flight path of the selected ion. After colliding
    with Ar, the selected ion is fragmented ( a
    process called CID (collision-induced
    dissociation).
  • The third quadrupole (Q3) serves to analyze the
    fragment ions generated in the collision cell
    (Q2).

47
Peptide Mass Fingerprinting (PMF)
  • More recently, MS has been combined with protease
    digestion to enable peptide mass fingerprinting.
  • (1) Sequence specific proteases or certain
    chemical cleaving agents are used to obtain a set
    of peptides from the target protein that are then
    mass analyzed
  • (2) The observed masses of the proteolytic
    fragments are compared with theoretical in
    silico digests of all the proteins listed in a
    sequence database.
  • (3) The matches or hits are then statistically
    evaluated and ranked according to the highest
    probability

48
Protein identification by PMF
49
  • Various databases are available on the web, and
    can be used in conjunction with such computer
    programs such as Profound, ProteinProspector, and
    Mascot.

50
PMF possible causes of incorrect protein
identification
  • An error in the sequence database
  • An inaccurate experimental mass determination
  • Existence of two or more polymorphic variants
    (e.g., SNPs)
  • Post-translational modifications
  • Occasional nonspecific cleavage of the protein by
    trypsin

51
Protein identification using Tandem Mass
Spectrometry
52
Protein identification using Tandem Mass
Spectrometry (contd.)
  • Tandem mass spectrometry has the ability to
    induce fragmentation and perform successive mass
    spectrometry experiments on these ions. It is
    generally used to obtain this structural info.
    (Abbreviated MSn, where n refers to one the
    number of generations of fragment ions being
    analyzed).

53
(No Transcript)
54
CID (collision-induced dissociation)
  • CID is accomplished by selecting an ion of
    interest with the mass analyzer and then
    subjecting that ion of interest to collisions
    with neutral atoms or molecules. The selected ion
    will collide with the collision gas (e.g., Ar,
    He, or Xe), resulting in fragment ions which are
    then mass analyzed. CID can be accomplished with
    a variety of instruments, including triple
    quadrupoles or TOF/TOF mass analyer.

55
CID (contd.)
  • The fragment ions produced in this process can be
    separated into two classes
  • (1) One class retains the charge on the
    N-terminal and occurs at three different
    positions, designated as types an, bn, and cn.
  • (2) The second class of fragment ion ions retains
    the charge on the C-terminal and fragmentation
    occurs at three different positions, types xn,
    yn, and zn.

56
CID (contd.)
  • Most fragment ions are obtained from cleavage
    between a carbonyl and a nitrogen (the amide
    bond).
  • Thus, if the charge is retained on the N-terminal
    end of the molecule, the cleavage is a b-type. If
    the charge is retained on the C-terminal end of
    the peptide, the cleavage is y-type.

57
(No Transcript)
58
(No Transcript)
59
Ladder sequencing by mass spectrometry
The differences in mass between consecutive ions
in either series should correspond to the masses
of individual amino acids
E Glu T Thr
60
Ladder sequencing by mass spectrometry
Two pairs of residues that are hard to
be distinguished by MS (1) Gln (128.13) Lys
(128.17) (2) Leu (113) Ile (113)
Write a Comment
User Comments (0)
About PowerShow.com