Title: Evaluating Medical DeviceUse Safety
1Evaluating Medical Device-Use Safety
- Todd R. Johnson, PhDAssociate Professor
- The University of Texas School of Health
Information Sciences Houston andCenter of
Excellence in Patient Safety Research and
Practice
2Device-Use Error
- An error resulting from the way a device is used,
rather than a device malfunction.
3Example Infusion Pump
- Order Deliver 130.1 ml/hr
- Nurse programs pump by pressing
- 1 3 0 . 1
- Device ignores decimal point keypress for numbers
greater than 99.9 - 1301 appears on display
4Effect of Device-Use Errors
- 82 of all preventable errors involving
anesthesia devices were due to device-use errors
(Cooper et al., 1978) - Patients may be 3 to 10 times more at risk due to
device-use error than to device failure (Grant,
1998)
5Sources of Device-Use Errors
Devices Design
Task
Use Environment
User
6How Can We Select Devices That Minimize the
Chances of Device-Use Error?
7Results of Purchasing Study at 3 Sites
- Very few alternative devices considered
- Device users were not direct participants in the
purchasing decisions - Evaluation process was often used to justify a
decision - Device evaluation was often limited to technical
safety - Usability testing for assessing device safety was
not conducted at any of the sites.
8Major Recommendations
- Involve all device stakeholders in the selection
process - Directly involve active device users
- Analyze strengths and weaknesses of existing
device - Set explicit safety goals
- Consider a wide range of alternative devices
- Explicitly evaluate device-use safety
- Ask for manufacturers human factors engineering
plan and results - Consider the cost of device-use errors
9Challenge
- Challenge
- Device-use safety is best evaluated by usability
experts, butusability experts are rarely
available to purchasing committees. - Solution
- Develop methods for use by healthcare
professionals
10Methods for Evaluating Device-Use Safety
11What are good dimensions for evaluating
device-use safety?
- Must be
- Informative to healthcare professionals
- Allow clinically meaningful comparisons of devices
12Classic Usability Dimensions are Too Abstract
- Speak the users language
- Visibility of system status
- User control and freedom
- Be consistent and follow standards
- Design to prevent errors
- Recognition rather than recall
- Design for flexibility and efficiency of use
- Aesthetic and minimalist design
- Help users recognize, diagnosis, and recover from
errors - Provide relevant, task oriented, easy to access
and search, help and documentation
13Classic Usability Results May Not be Very Helpful
for Clinicians
14Solution The UPCARE model
- Developed by Kaye, North, and Peterson
- Designed to help describe and analyze device use
problems - Uses device-based dimensions familiar to clinical
users
15UPCARE
- Unmet user needs
- Perception
- Cognition
- Actions
- Results
- Evaluation
16Modification of UPCARE for Device-Use Evaluation
- UPC Simplified from 38 to 23 usability items
- ARE not relevant to device-use evaluation
17Usability Dimensions
- System set-up and configuration
- 1. System set-up is easy and efficient
- 2. Start up procedure is intuitive and efficient
- User-device interaction
- Perception
- 3. Device displays, labels, or markings are
clearly visible - 4. Text, number, or status indication is easy to
locate in the visual display
18- Information interpretation
- 1. Labeling on device or overall device
configuration clearly shows identity, operation,
or use - 2. Input, output, level, or calibration values
conform to established standards - 3. Device data are sufficient for user to
diagnose patient situation or adjust treatment
19- Feedback
- 1. Feedback to user actions is timely, adequate,
and easy to understand - 2. Device status, including normal operation,
critical change and device failure, is adequately
indicated - 3. Battery (or charge) end-of-life condition is
clearly indicated - Error-correction
- 4. Device provides safe default/fail-safe mode
- 5. In case of error, device allows user to make
corrections easily and quickly - 6. In case of error, device allows user to
immediately stop device action or process
20Device-Use Survey
- Uses the same dimensions
- Used after user tests or clinical trials
- Neutral items with semantic differentials
21Combining Results
- Nearly every recommended evaluation method uses
the same dimensions - Exception Error reports
- Spreadsheet combines cost, technical specs, and
device-use safety for a variety of devices
22(No Transcript)
23(No Transcript)
24Current Status
- Guidelines are nearly complete
- Testing with mock purchasing groups will begin
soon - Future work
- Test use of each method with healthcare
professionals
25Acknowledgments
- Collaborators
- University of Texas Health Science Center at
Houston - Jiajie Zhang, PhD, James P. Turley, RN, PhD
- Xiaozhou Tang, MD, Juliana Brixey, MSN, MPH, RN
- Danielle Paige, Zhihua Tang, PhD, Kelly
deChermont - Columbia University
- Vimla L. Patel, PhD, DSc
- Alla Kesselman, PhD
- Mark Graham, PhD
- This project was supported by grant number PO1
HS11544 from the Agency for Health Care Research
and Quality