Title: ECSE6963 Introduction to Subsurface Sensing and Imaging Systems
1ECSE-6963Introduction to Subsurface Sensing and
Imaging Systems
- Lecture 17 Spectral Imaging Fluorescence
- Kai Thomenius1 Badri Roysam2
- 1Chief Technologist, Imaging technologies,
- General Electric Global Technology Center
- 2Professor, Rensselaer Polytechnic Institute
Center for Sub-Surface Imaging Sensing
2Recap
- Use of Phase in Imaging
- Doppler ultrasound
- Phase-contrast optical imaging
- Differential interference contrast
- Spectral imaging
- Fluorescence and contrast agents
3Recap EM Interaction with Matter
4Recap Fluorescence Imaging
Courtesy William Shain Lab
5Light Absorption
Absorption can only occur when
If the energy of the probing photon is mismatched
to the energy difference between the quantum
states, the material is transparent to the probe.
6Emission Fluorescence
- Opposite of absorption
- Can be coherent (Raman)/incoherent (Fluorescence)
- Typically, there is some loss in the molecule, so
the emitted energy is lower than the absorbed
energy - The difference is the Stokes Shift
Molar Concentration of fluorophore
Excitation intensity
Path length
Fluorescence intensity
Quantum Efficiency ( molecules that emit)
Molar absorptivity
7Multi-photon Excitation
- Use 2 or more infrared photons to excite
fluorophores ordinarily excitable with higher
frequencies - Much more detail in images collected deeper in
the sample. - No sample photobleaching outside focal plane.
- Dramatic improvement in longevity of living
cells, tissues and organisms. - Ready determination of co-localising fluorescent
probes. - No need for confocal apertures.
- Ability to image autofluorescence
- UV flourophores may be excited using a lens that
is not corrected for UV as these wavelengths
never have to pass through the lens. - MPE also offers enhanced photoselection in
spectroscopy.
3-photon works the same way
8Multi-photon Microscopy
- Practical issues
- The two photons need to arrive simultaneously a
low probability event - Use ultrafast, mode-locked near-infrared lasers
(i.e. Tisapphire 100-200 femtosecond pulse
duration, 76MHz repetition rate). - Under the appropriate conditions, these lasers
produce short duration pulses with the high peak
power required for a multi-photon effect and an
average power low enough to make specimen damage
negligible. - Such lasers are tunable over a range of
700nm-1000nm which permits optimal wavelength
selection to elicit an efficient multi-photon
effect. - Probability of simultaneous absorption falls off
steeply away from the focal volume
9Skin Tumor Example
10Observing Changes Over Time
Successive images are 24 hours apart
11Limitations of Multi-photon
- Slightly lower resolution with a given
fluorophore when compared to confocal imaging. - This loss in resolution can be eliminated by the
use of a confocal aperture at the expense of a
loss in signal. - Thermal damage can occur in a specimen if it
contains chromophores that absorb the excitation
wavelengths, such as the pigment melanin. - Only works with fluorescence imaging.
- Currently rather expensive.
12Second-harmonic Generation
- Similar to 2-photon, but essentially lossless
- Great for live imaging
- E.g., collagen
- Great for
- Deep-tissue imaging
- quantification along the frequency axis!
Virtual state
13SHG Example
Extra-cellular matrix
Muscle filament lattice structure
Extra-cellular and intracellular structures
within native muscle tissue
An optical section at a depth of 250 µm into the
sample
http//www.biophysj.org/cgi/content/full/82/1/493
14Improving Light Microscopy
- Its Not Just About Resolution
- Resolution Limited by Diffraction
- Its About What Is Measured
- Transmission, Reflection, Phase, Fluorescence,
Polarization, Non-Linear Properties - Minimizing specimen damage
- And About How Data Are Processed
- Registration, Deconvolution, Tomography,
Parameter Estimation - And About Measuring Everything at Once
15What is Sensed by the Different Modes?
LSCM
SHG
Epi F
DIC
TPLSM
QTM
Staring
Scanning
Phase (optical path length)
16What can be measured
Morphological Dynamics
- Sensor Fusion
- Different sensors have different strengths and
weaknesses - Combine them to take advantage of each
Space (x, y, z)
Time (t)
Structure
Dynamics
Function
Signaling Paths, Molecular Transport
Chemical Dynamics
Fluorescent labels
Other Physical Properties
e.g., Refractive index, strain
17(Phase Fluorescence) Example
Data courtesy Gary Banker (U. Oregon)
18Staring Modes Similar Embryos, Diverse Images
Thanks to Bill Warger
19Multi-Modal Imaging of Mouse Embryos
2-photon
Confocal Fluorescence
Differential Interference Contrast
Quadrature Microscopy
20Multi-Modal Imaging of Mouse Kidney Cells Using
a 3D Fusion Microscope
2-photon DAPI
2-photon Alexa Fluor 488
QTM
DIC
Thanks to Dan Townsend
21Accessing Wider Regions
22Current State of the Art
- Fusion Microscopy Put Multiple modalities on
same platform
23Current Trends
- Ever increasing numbers of structural and
functional endpoints can be observed
simultaneously in 3-D - Growing libraries of organic fluorophores
quantum dots - Multi- and hyper-spectral microscopes
- Spectral unmixing tools
- Support for complex fluorescence phenomena
- Easier to work with live cells
- Sensitive, high-resolution, 3-D imaging
- Minimally-damaging (MP, SHG), time-resolved
imaging - Better instrumentation better understanding of
biology - Fusion of multiple microscopy modalities
- High-extent high-resolution high-throughput
imaging - High-throughput tissue prep imaging hardware
24Summary
- Spectral response provides substance-specificity
to imaging - Fluorescence and Multi-photon imaging are
powerful tools - Thanks to Prof. Charles DiMarzio for slides
regarding QTM and Fusion Microscopy - Lets talk about Projects!!
25Instructor Contact Information
- Badri Roysam
- Professor of Electrical, Computer, Systems
Engineering - Office JEC 7010
- Rensselaer Polytechnic Institute
- 110, 8th Street, Troy, New York 12180
- Phone (518) 276-8067
- Fax (518) 276-8715
- Email roysam_at_ecse.rpi.edu
- Website http//www.ecse.rpi.edu/roysabm
- NetMeeting ID (for off-campus students)
128.113.61.80 - Secretary Laraine, JEC 7012, (518) 276 8525,
michal_at_.rpi.edu
26Instructor Contact Information
- Kai E Thomenius
- Chief Technologist, Ultrasound Biomedical
- Office KW-C300A
- GE Global Research
- Imaging Technologies
- Niskayuna, New York 12309
- Phone (518) 387-7233
- Fax (518) 387-6170
- Email thomeniu_at_crd.ge.com, thomenius_at_ecse.rpi.edu
- Secretary TBD
-