Title: On the use of VIP Gel dosimetry
1On the use of VIP Gel dosimetry
in HDR brachytherapy
- L Petrokokkinos1,2, A Moutsatsos2, P Karaiskos2,
- V Kouridou3, E Pantelis4, P Papagiannis2 and I
Seimenis5
1. Nuclear Particle Physics Section, Physics
Department, University of Athens, Greece 2.
Medical Physics Laboratory, Medical School,
University of Athens, Greece 3. Anticancer
Hospital of Athens St. Savvas, Athens,
Greece 4. CyberKnife centre, Iatropolis -
Magnitiki Tomografia, Athens, Greece 5. Medical
Diagnostic Centre Ayios Therissos, Nicosia, Cyprus
2Gel dosimeters an absolute dosimeter?
3Gel preparation irradiation
Irradiation?192Ir Varisource (5 mm)?Sk
12861.42 cGycm2/h (3 unc.)?Varian HDR remote
afterloader?Single dwell position?tir 283.5
s?11.15 Gy _at_ 1 cm
VIP formulation?8 w/v N-Vinylpyrrolidone?7.5
w/v gelatin?4 w/v N,N-methylenebisacrylamide?
0.0008 w/v Copper Sulfate ?0.007 w/v Ascorbic
acid?Hyperpure water (gt18 MO?cm)
4MR readout
Scanning parameters?1.5 T Philips Gyroscan
NT?Quadrature RF receiver coil?3D CPMG
sequence?20 equidistant echoes, 40 ms
step?TE120 ms, TR2000 ms?Axially
reconstructed slices?0.75 mm slice
thickness?(0.73864 mm)2 in-plane pixel
5Calibration procedure
Reference data?MC source dosimetric data ?(lt1
uncertainty)?Source Sk (3 uncertainty)?Irradiat
ion time
Response Ref.Dose matching?3 mm lt
ri lt 22 mm away distances?0.74 mm ri
steps?R2imeanR2mean(ri)
sR2isR2mean(ri)
Weighted least squares regression?wi1/sR2i2?R
2a1D2a2Da3?a1 -(0.000280.00004) Gy-2?s-1
?a2(0.07920.0018) (Gy?s)-1 ?a3(1.3410.011)
s-1
6Calibration procedure induced uncertainty
7Measurement procedure induced uncertainty
8Total experimental uncertainty to dose
13 mm
5.4 mm
9Along dose profile
10On the use of VIP Gel dosimetry
in HDR brachytherapy
?
Questions
THE END