Overall PACS EQM IMT Analysis Contributions by PACS ICC - PowerPoint PPT Presentation

1 / 18
About This Presentation
Title:

Overall PACS EQM IMT Analysis Contributions by PACS ICC

Description:

generated corrupted output only re-installed AVM1 ... show-stopper (Corrupted packet checksums, duplicate HK packets, SSC. discontinuities) ... – PowerPoint PPT presentation

Number of Views:128
Avg rating:3.0/5.0
Slides: 19
Provided by: HelmutFeu2
Category:

less

Transcript and Presenter's Notes

Title: Overall PACS EQM IMT Analysis Contributions by PACS ICC


1
Overall PACS EQM IMT AnalysisContributions by
PACS ICC
2
PACS EQM IMT Contents
  • Short Functional Test (SFT) warm (3 x) 1h
  • Short Functional Test (SFT) cold HeII (not
    done) ( 1h)
  • PACS Integrated Module Test
  • Full Functional Test (FFT) 7.7h
  • Short Performance Test (SPT)
    12.8h
  • AOT Test 9.7h
  • EMC Test (SpecPhot) 5d
  • Separate, joint SPIRE/PACS procedure
  • PACS/SPIRE Parallel Mode Test (PACS
    part) 4.7h
  • Extra straylight tests (3d)
  • Short Functional Test (SFT) cold HeI 1h

21-22 Nov 2005 H-Field Tests incl. threshold
measurements
23-25 Nov 2005 E-Field Tests
3
Available Reports/Notes
  • 3 reports on short functional tests warm
  • 1 report on short functional test HeI
  • Performance test of PACS chopper and duty cycle
  • Internal calibration sources performance tests
  • Cryostat background adjustment
  • Emissivity/RSRF on internal calibration sources
  • Function and performance of PACS grating
  • Performance test PACS Photometer filter wheel
  • GeGa detector dark current test on cold
    internal calibration sources
  • GeGa bias loop tests and detector performance
  • Dead/suspicious pixel maps from RSRF, bias loop
    and dark current tests
  • GeGa time constants for flux changes
  • Low frequency noise measurements of the
    bolometer
  • Photometer FOV maps
  • Summary note on AOT related tests

4
Warm Electronics and OBSW
  • SPU-L failure loose FPGA
  • ? removed, repaired and re-installed
  • DPU AVM1 failure broken inductor
  • ? removed and repaired, exchange with CFM2 but
    that one
  • generated corrupted output only ?
    re-installed AVM1
  • SPU OBSW upgraded from V11.1 to 11.7 to prevent
    IMT data
  • losses due to a known bug, detected during ILT
  • SPU OBSW Data rate / CPU workload too high (as
    expected from ILT)
  • DPU HW/OBSW Few known and new bugs detected,
    but no major
  • show-stopper (Corrupted packet checksums,
    duplicate HK packets, SSC
  • discontinuities)
  • BOLC 20Hz command had unexpected impact on
    supply groups


5
Test Results (1)
  • Chopper
  • duty cycle comparable to ILT (but too small)
  • noise slightly worse than in ILT, but still in
    spec.
  • Slight angular offset detected contact with
    housing has been
  • removed after the ILT


6
Test Results (2)
  • Calibration sources
  • Nominal performance total heating time and
    stability in spec


Fluctuations ? 50 units Spec. 7mK ? 70 units
7
Test Results (3)
  • Grating Suffered from a DMC FPGA problem
    (introduced after ILT)


Excessive noise on position readback
Spontaneous grating controller self-disable
8
Test Results (4)
  • Photometry Filter Wheel
  • nominal performance, transition time even
    slightly shorter
  • than in ILT


Position A
Position B
9
Test Results (5)
  • Temperature sensors
  • nominal performance for all but the two blue
    detector sensors
  • 1 sensor was recovered, apparently a contact
    problem on
  • a DMC connector
  • Blue detector heater
  • - o.k. (but thermal shimming to be
    improved)
  • Optical performances
  • - no particular tests done, since no
    telescope/OGSE available
  • - straylight
  • - FoV maps do not indicate any other problem


10
Test Results (6)
  • GeGa Detector performance
  • IMT performance similar to ILT
  • Differences due to other background conditions
    absolute
  • level and stability
  • - Lost a number of pixels


Rint64, Cint0.3pF
Rint64, Cint0.3pF
11
Test Results (6)
  • GeGa Detector performance
  • - Lost a number of pixels


Blue array   - 2 new high noise pixels
Red array   - 7 new dead pixels   - 5 new high
noise pixels   - 1 dead pixel in ILT seems to
react better to flux again, but very
noisy.
12
Test Results (7)
  • Bolometer performance
  • IMT performance similar to ILT
  • Differences due to shorter test time and other
    background
  • conditions absolute level and stability


ILT
IMT
13
Test Results (8)
  • Bolometer performance
  • IMT performance similar to ILT
  • Differences due to shorter test time and other
    background
  • conditions absolute level and stability


ILT
IMT
14
Further Tests
  • Not reported in this presentation
  • - Optimum GeGa detector Bias
  • - S/N as a function of reset interval
  • - Emissivity of calibration sources
  • - Quick wavelength check (several executions
    failed, one was
  • successful)
  • - Memory Management
  • - SAFE mode transition
  • - Non-Prime PACS Switch-on
  • - Bolometers Saturation check
  • - Detector selection table test
  • - Compression/Reduction Modes Test (Phot Spec)
  • - Time constants for flux changes in
    spectroscopy
  • Failed Tests (due to grating problem)
  • - Grating relative spectral response function
  • - Emissivity of calibration sources (failed
    partly)
  • - Medium sampling grating test


15
AOT Tests (1)
  • Overview
  • Spectroscopy TuttiFrutti AOT (? o.k.)
  • Photometry TuttiFrutti AOT (? o.k.)
  • Test of all AOT relevant OBCPs, verification of
    PACS sub-unit synchronisation and building blocks
    for real AOTs
  • Spectrometer Line Scan AOT
  • - Calibration blocks concepts
  • - Line scan concepts
  • - Frequency switching concepts
  • Spectrometer Range Scan AOT
  • Photometer AOT
  • - Calibration block concepts
  • - Nominal photometry concepts
  • - Photometry in PACS/SPIRE parallel mode


16
AOT Tests (2)
  • Consolidated set of AOT performance tests
    established for future
  • test campaigns
  • Although the AOT tests are highly valuable, not
    everything could
  • be tested pointing modes, space environment
    effects, no
  • external sources
  • Report on AOT related test during the EQM IMT
    is in
  • preparation, results to be presented at next
    PACS ICC Feb 23/24


17
PACS/SPIRE Parallel Mode Test (1)
Concept - Shifted (by 25-30min.) cooler
recycling - SPIRE activities - Thermal
behaviour test (with SPIRE on in parallel) -
Two test cases Single and Dual band photometry
- Only Line Scan AOT option tested
- Dual band blue LW, double compression
red, nominal compression - Single band
only red channel, nominal compression - 15
TM packets per sec available for PACS ? data
rates PACS 72 kbit/s
SPIRE 57 kbit/s

18
PACS/SPIRE Parallel Mode Test (2)
  • Results
  • Thermal behaviour o.k.
  • Single band photometry o.k.
  • Dual band photometry Bug in CUS script, double
    reduction in
  • blue channel has been set unfortunately to
    nominal again
  • ? Data loss due to violation of bus
    bandwidth, however correctly
  • handled by DPU.
  • Overall Successful as seen by PACS
  • ? next steps to be carried out
    elsewhere (CUS, HSPOT)