Title: Overall PACS EQM IMT Analysis Contributions by PACS ICC
1Overall PACS EQM IMT AnalysisContributions by
PACS ICC
2PACS EQM IMT Contents
- Short Functional Test (SFT) warm (3 x) 1h
- Short Functional Test (SFT) cold HeII (not
done) ( 1h) - PACS Integrated Module Test
- Full Functional Test (FFT) 7.7h
- Short Performance Test (SPT)
12.8h - AOT Test 9.7h
- EMC Test (SpecPhot) 5d
- Separate, joint SPIRE/PACS procedure
- PACS/SPIRE Parallel Mode Test (PACS
part) 4.7h - Extra straylight tests (3d)
- Short Functional Test (SFT) cold HeI 1h
21-22 Nov 2005 H-Field Tests incl. threshold
measurements
23-25 Nov 2005 E-Field Tests
3Available Reports/Notes
- 3 reports on short functional tests warm
- 1 report on short functional test HeI
- Performance test of PACS chopper and duty cycle
- Internal calibration sources performance tests
- Cryostat background adjustment
- Emissivity/RSRF on internal calibration sources
- Function and performance of PACS grating
- Performance test PACS Photometer filter wheel
- GeGa detector dark current test on cold
internal calibration sources - GeGa bias loop tests and detector performance
- Dead/suspicious pixel maps from RSRF, bias loop
and dark current tests - GeGa time constants for flux changes
- Low frequency noise measurements of the
bolometer - Photometer FOV maps
- Summary note on AOT related tests
4Warm Electronics and OBSW
- SPU-L failure loose FPGA
- ? removed, repaired and re-installed
- DPU AVM1 failure broken inductor
- ? removed and repaired, exchange with CFM2 but
that one - generated corrupted output only ?
re-installed AVM1 - SPU OBSW upgraded from V11.1 to 11.7 to prevent
IMT data - losses due to a known bug, detected during ILT
- SPU OBSW Data rate / CPU workload too high (as
expected from ILT) - DPU HW/OBSW Few known and new bugs detected,
but no major - show-stopper (Corrupted packet checksums,
duplicate HK packets, SSC - discontinuities)
- BOLC 20Hz command had unexpected impact on
supply groups -
5Test Results (1)
- Chopper
- duty cycle comparable to ILT (but too small)
- noise slightly worse than in ILT, but still in
spec. - Slight angular offset detected contact with
housing has been - removed after the ILT
6Test Results (2)
- Calibration sources
- Nominal performance total heating time and
stability in spec
Fluctuations ? 50 units Spec. 7mK ? 70 units
7Test Results (3)
- Grating Suffered from a DMC FPGA problem
(introduced after ILT)
Excessive noise on position readback
Spontaneous grating controller self-disable
8Test Results (4)
- Photometry Filter Wheel
- nominal performance, transition time even
slightly shorter - than in ILT
Position A
Position B
9Test Results (5)
- Temperature sensors
- nominal performance for all but the two blue
detector sensors - 1 sensor was recovered, apparently a contact
problem on - a DMC connector
-
- Blue detector heater
- - o.k. (but thermal shimming to be
improved) - Optical performances
- - no particular tests done, since no
telescope/OGSE available - - straylight
- - FoV maps do not indicate any other problem
10Test Results (6)
- GeGa Detector performance
- IMT performance similar to ILT
- Differences due to other background conditions
absolute - level and stability
- - Lost a number of pixels
Rint64, Cint0.3pF
Rint64, Cint0.3pF
11Test Results (6)
- GeGa Detector performance
- - Lost a number of pixels
Blue array - 2 new high noise pixels
Red array - 7 new dead pixels - 5 new high
noise pixels - 1 dead pixel in ILT seems to
react better to flux again, but very
noisy.
12Test Results (7)
- Bolometer performance
- IMT performance similar to ILT
- Differences due to shorter test time and other
background - conditions absolute level and stability
ILT
IMT
13Test Results (8)
- Bolometer performance
- IMT performance similar to ILT
- Differences due to shorter test time and other
background - conditions absolute level and stability
ILT
IMT
14Further Tests
- Not reported in this presentation
- - Optimum GeGa detector Bias
- - S/N as a function of reset interval
- - Emissivity of calibration sources
- - Quick wavelength check (several executions
failed, one was - successful)
- - Memory Management
- - SAFE mode transition
- - Non-Prime PACS Switch-on
- - Bolometers Saturation check
- - Detector selection table test
- - Compression/Reduction Modes Test (Phot Spec)
- - Time constants for flux changes in
spectroscopy - Failed Tests (due to grating problem)
- - Grating relative spectral response function
- - Emissivity of calibration sources (failed
partly) - - Medium sampling grating test
15AOT Tests (1)
- Overview
- Spectroscopy TuttiFrutti AOT (? o.k.)
- Photometry TuttiFrutti AOT (? o.k.)
- Test of all AOT relevant OBCPs, verification of
PACS sub-unit synchronisation and building blocks
for real AOTs - Spectrometer Line Scan AOT
- - Calibration blocks concepts
- - Line scan concepts
- - Frequency switching concepts
- Spectrometer Range Scan AOT
- Photometer AOT
- - Calibration block concepts
- - Nominal photometry concepts
- - Photometry in PACS/SPIRE parallel mode
16AOT Tests (2)
- Consolidated set of AOT performance tests
established for future - test campaigns
- Although the AOT tests are highly valuable, not
everything could - be tested pointing modes, space environment
effects, no - external sources
- Report on AOT related test during the EQM IMT
is in - preparation, results to be presented at next
PACS ICC Feb 23/24
17PACS/SPIRE Parallel Mode Test (1)
Concept - Shifted (by 25-30min.) cooler
recycling - SPIRE activities - Thermal
behaviour test (with SPIRE on in parallel) -
Two test cases Single and Dual band photometry
- Only Line Scan AOT option tested
- Dual band blue LW, double compression
red, nominal compression - Single band
only red channel, nominal compression - 15
TM packets per sec available for PACS ? data
rates PACS 72 kbit/s
SPIRE 57 kbit/s
18PACS/SPIRE Parallel Mode Test (2)
- Results
- Thermal behaviour o.k.
- Single band photometry o.k.
- Dual band photometry Bug in CUS script, double
reduction in - blue channel has been set unfortunately to
nominal again - ? Data loss due to violation of bus
bandwidth, however correctly - handled by DPU.
- Overall Successful as seen by PACS
- ? next steps to be carried out
elsewhere (CUS, HSPOT)