Title: EPS20 Cardiac Stimulator Safety


1
EPS20 Cardiac Stimulator Safety
  • Installation issues
  • Safety Warnings and Notes
  • Configuration
  • Stimulator Safety Features
  • Safety in demonstrating system
  • Safety in routine application

2
1. Installation Issues
  • Isolation transformer
  • RF filtered Patient Connection Box
  • In-line RF Filter
  • Stim extension cable
  • Good External ECG channel setup
  • Demonstration of safety features
  • IEC60601-1 test before patient

3
2. Safety Warnings
  • Go through User Manual Warnings Section with
    customer
  • Explain RF Filtering
  • Explain, determine and set safety limits
  • Stimulator is not Life Support Device

4
Causes of Failure to Pace
5
Mitigations for Failure to Pace
6
3. Stimulator Configuration
  • Lower limits for S1
  • Pace
  • Burst Pace
  • Auto-decremented
  • QRS sync timeout
  • Idle safety timeout

7
4. Stimulator Safety Features
  • Lower limit S1 interlocks
  • Data entry validation
  • Lead High Impedance warning
  • Emergency Pace - F12
  • Backup Pace BACKUP ENABLE
  • Emergency Fixed Rate Pacing
  • Warning Messages e.g Defib

8
5. Safety in demonstrating EPS320
  • What harm can you cause patient ?
  • How can you bring about this harm ?

9
5. Safety in demonstrating EPS320
  • Potential Patient harms
  • Asystole
  • VF induction
  • VT acceleration
  • VT failure to Burst Overdrive
  • inadvertent VT termination

10
AsystoleCauses of Failure to Pace in Asystole
  • Asystole may not be noticed
  • Not select PACE protocol - in panic
  • Pace into A instead of V
  • Current too low
  • Leads / Stim Channel not connected
  • EP Recorder Stim setting wrong
  • Failed EP Recorder Stim Relays

11
Asystole Failure to pace -Remedies
Remedies All solved with F12 function
  • Not select PACE protocol -in panic
  • Pace into A instead of V
  • Pace at to low a current
  • EP Recorder stim setting wrong
  • Leads disconnected
  • Failed EP Recorder Stim relays

1. High impedance warning 2. Use EP Recorder
Bypass
12
Ventricular Fibrillation Induction - User Fault
  • Accidental Burst Pace S1lt200ms
  • Unsynchronised Stimulation - R on T
  • Burst Pacing into VT
  • Pace with S2-S7 too short
  • Excessive stimulation current
  • Remedy Use PACE for all non-burst
  • Remedy Set safety limits /vigilance

13
Ventricular Fibrillation Induction - Equipment
fault
  • DC current Stimulation - Stimulator fault or
    external source - e.g. RF Ablation energy
  • Remedy Note alarm and disconnect patient
  • Micro-electrocution - Exposed catheter pins /
    attached equipment non-isolated
  • Remedy check and remove

14
VT Acceleration
  • Accidental Pacing into VT
  • Remedy ready to stop pacing in VT induction
  • Remedy Use StopOnTachy feature
  • Load ATP protocol with S1 too low
  • Remedy ensure good QRS sensing

15
Inadvertent VT termination
  • Continuation of pacing once VT induced
  • Remedy be ready to stop pacing when VT induced
  • Remedy Use StopOnTachy feature, ensuring good
    ECG sensing during sinus arrhythmia