Title: Recent Technological Developments
1Recent Technological Developments
- Tissue Harmonics
- Spatial Compounding
- Extended Field of View Imaging
- 3D 4D Imaging
2These new technologies are reviewed in excellent
detail in the 4th Ed of HedrickPages 140 - 152
Make sure you read it!
3Harmonic Imaging
-Be Careful!!! The term Harmonic Imaging is used
to describe two separate modes of imaging. 1.
Associated with contrast agents use. 2. Tissue
harmonics.
4Harmonic Imaging Associated with Contrast Agents
- U/S Contrast Agents generally consist of
encapsulated microbubbles that boost the
echogenicity of the blood and perfused tissues.
More importantly resonance phenomena within the
microbubbles give rise to harmonic frequencies
which can be selectively received.
3 MHz
3 MHz
6 MHz
6 MHz
Microbubble
3 MHz
RBC
3 MHz
If you tune the receiver to receive at twice the
transmitted frequency then the harmonic echoes
from the microbubbles will dominate the returning
signal.
5The history of Tissue Harmonics
- The fact that harmonics frequencies arise from
soft tissue even without the injection of
microbubble contrast agents was discovered by
mistake! - Researchers working on the development of
ultrasound contrast agents found that an image of
the tissues would still be produced when the
receiver was tuned to receive at twice the
transmitted frequency. In fact some of the
features of the image were superior to the
fundamental image. - From there they had to set about explaining why
these natural tissue harmonics occurred!
6Tissue Harmonic Processing
Acoustic field
Transmitted signal
After Processing
7Tissue Harmonics- The Basics
- Wave travels faster in compressive phase (NB
Increased particle density) and slower during
rarefaction. - This Nonlinear Propagation shifts a portion of
the beams energy to harmonic frequencies. - Harmonics are not produced in the first few cm of
tissue. - Harmonics are most prominent in the central
portion of the transmitted beam. Therefore
lateral resolution is typically superior in
harmonic mode. - Most Harmonic Energy is in the Second Harmonic.
This has a frequency which is twice the
fundamental frequency.
8From 4th Ed of Hedrick
9Pros And Cons of Harmonics
- Pros
- Many Fundamental Image artefacts are reduced
- Side Grating Lobes
- Reverberation
- gtgt Better Contrast Resolution due to less
noise. - Better Lateral Resolution.
- Cons
- Less Penetration
10(No Transcript)
11(No Transcript)
12Fundamental Image
Harmonic Image
13Spatial Compounding
- Basics- Angle of incidence to an interface
influences the amplitude of the echo returning.
If you produce an image with transmitted beams at
multiple angles then border definition will
improve. - Most of the major ultrasound equipment companies
have adopted this technique. It is given
different proprietary names such as SonoCT. - Beam steering principles are employed to steer
beams at multiple angles through the FOV. The
data stored at each pixel is the average of the
multiple angles. - As well of better border definition there is
noise reduction achieved because of the data
averaging at each point. - As the technique is performed in real time there
is no effect on frame rate. Just like frame
averaging has no effect on FR, except in this
case the previous frames are taken from different
beam steering angles.
14From 4th Ed of Hedrick
15Spatial Compounding Clinical Cases
Spatial Compound Image
Note reduced artefact in the anechoic renal cysts
Standard Image
16Spatial Compound Image
Note greater continuity in appearance of muscle
fibres and more defined margins to the partial
tear of this supraspinatus tendon.
Standard Image
17Extended Field of View Imaging
Normal Calf Muscles
This technique can produce FOVs over 500mm wide
(a big improvement from the 40mm offered by the
average linear array transducer). It is
particularly useful for demonstrating MSK anatomy
and pathology.
Large Calf Muscle Tear
18Extended Field of View Imaging
- Different Proprietary names exist for this
technique Siescape, Panoramic Imaging, Logiqview.
- Successive frames are combined to produce the
extended FOV. - Computer analysis of successive frames for common
image features is employed to determine the
degree of transducer movement between frames. - Careful and smooth transducer movement in a
single plane (and lots of practice) is required
to avoid image artefacts caused by rapid movement.
193D 4D Imaging
- In 3D imaging a volume of data is scanned and
displayed as a static image. - In 4D imaging the volume is scanned in real
time such that a moving 3D image of the volume
is shown.
203D 4D Basics
By moving the Transducer perpendicular to the
scan plane it is possible to produce a volume of
ultrasound data. Loaf of Bread Analogy. This
data can be used to create a 3D image or
reconstruct a 2D image not in the original scan
plane.
If the volume of data is reproduced many times a
second. i.e. By having the transducer
mechanically steering through the volume, then a
real time 3D image or 4D image can be created.
21Stilled 4D Images I have produced of two Facial
Clefts of different severity. There is a
counseling benefit in the parents better
understanding the severity of the cleft with the
4D image.
22The non-medical use of 4D ultrasound in pregnancy
is causing much ethical debate in the profession.