Title: IVUSbased CFD
1Project 2 Part 1 Technology Development for MR
Assessment of Small Animal PhysiologyBrian
Rutt, PhD
2MR Gradient Coil Insert Design and
ConstructionAndrew Alejski, PhDCarl
Gazdzinski, BSc Yuteng Chen, MSc
3(No Transcript)
4300 mT/m 3-layer rabbit coil (Rcoil)
5Micro-Imaging
7T, 3D Spin Echo 100 micron isotropic 7 hour scan
1.5T, 3D FIESTA 100 micron in-plane 200 micron
slices 30 minute scan
6Next Generation Insert Coil (Scoil)
- 18 cm inner diameter
- Designed specifically for ultra-high-performance
mouse, rat, rabbit imaging using GE Signa Lx and
EXCITE at 1.5T and 3T - Gmax 500 mT/m, SlewRate 2400 T/m/s
- Completed manufacturing of three identical Scoil
prototypes - 1) GE-ASLW / Stanford Research Institute
- 2) GE-GRC / Albany Medical Center
- 3) Robarts
7Installation and Early Results from GE-ASLW /
SRIYuteng Chen, MScElfar Adalsteinsson,
PhDDolf Pfefferbaum, MD
8- Sunday
- Feb. 22 2004
- Cable cutting crimping
9Monday Feb. 23 2004 Twin switch installation
10Monday Feb. 23 2004 Twin filter installation
11- Tuesday Feb. 24 2004 Magnet room custom cable
installation
12Wednesday Feb. 25 2004 Insert coil calibration
13Completed / Installed Scoil (3T _at_ GE-ASLW)
Clinical Insert Peak gradient 40 mT/m 500
mT/m Slew rate 150 T/m/s 2000 T/m/s Resolution 0.8
mm 50-100 microns
14 Thursday Feb. 26 2004 First imaging by Dr.
Pfefferbaum
15(No Transcript)
16Thiamine-deficiency induced lesion in rat thalamus
17Thermal Engineering of Gradient Insert
CoilsYuteng Chen, MSc
18(No Transcript)
19(No Transcript)
20Zigzap cooling line (copper tube) design
21(No Transcript)
22Novel Design Concepts for Gradient CoilsCarl
Gazdzinski, BSc
23Concept The Harmonic Gradient Coil
- A multilayered gradient coil with each layer
contributing a spatial harmonic of field - Drive layers in variable proportions to obtain
variable resulting fields/DSVs - Amounts to a Fourier Series approximation of the
desired field
24Continuously Variable DSVs
DSV 17cm G 80 mT/m
DSV 33cm G 39 mT/m
DSV 48cm G 29 mT/m
Coil Specs ID 40cm Inductance 300 ?H/layer Max
Current 300 Amps Unshielded
25Conclusions To Date
- High performance insertable gradient coils
feasible on clinical 1.5T and 3T GE scanners - 10 minute, simple switch-over
- Performance increase 10 - 20 fold over best
whole-body gradients - Spatial resolution lt 100µm isotropic
- Advanced clinical pulse sequences (eg. 3DFIESTA,
FSE, DTI) immediately available - Advanced hardware (eg. 8-16 fast receivers)
immediately available