Title: Laboratory QAQC Quality AssuranceQuality Control
1Laboratory QA/QC (Quality Assurance/Quality
Control)
2Course Outline
- Week 1 Fundamental Statistics (Normal
distribution, outlier and examples) ?Glossary and
explanation - Week 2 Detection limit (method detection limit,
instrument detection limit) and calibration curve - Week 3 Quality assurance/Quality control
(definition, standard curve, control chart) - Week 4 Quality assurance/Quality control
(definition, standard curve, control chart) - Week 5 Result expression, sample correction and
others - Week 6 Examination
3?????????????
4Fundamental Statistics
- Normal or Gaussian distribution If a measurement
is repeated many times under essentially
identical conditions, the results of each
measurement, x, will be distributed randomly
about a mean value (arithmetic average) because
of uncontrollable or experimental error. - Standard deviation? is denoted by s (number of
estimate is finite)
5Fundamental Statistics
- Therefore, 95 of the measurement lie between ?
2? - Variance of the population (s2)
- Standard error of the mean Standard deviation
divided by the square root of the number of the
value (s/ ) - Confidence limit 104, 6 and 14 are the limit
- Confident interval from 6 to 14
6Normal distribution
?average, ?standard deviation
7(No Transcript)
8Outlier (Rejection of data)
- Theoretically, no result should be rejected
- First, calculate the statistic T
- T (XH - X)/s for a high value, or
- T (X - XL)/s for a low value.
- Second, compare the value of T with the value
from Table 1010I for either a 5 or 1 level of
significance
9Examples for rejecting outliers
Find the outliers from the following data for 5
or 1 level of significance
10Glossary
- Accuracy combination of bias and precision of
an analytical procedure, which reflects the
closeness of a measured value to a true value. - Precision measure of the degree of agreement
among replicate analyses of a sample, usually
expressed as the standard deviation. - Bias consistent deviation of measured values
from the true value, caused by systematic errors
in a procedure.
11Glossary Instrumental detection level (IDL)
- the constituent concentration that
- Produce a signal greater than three standard
deviation of the noise level (close to 99
probability that it is different from the blank) - produces a signal greater than five times the
signal/ noise ratio of the instrument. - For seven replicates of the sample, the mean must
be 3.14s above the blank where is the standard
deviation of the seven replicates. - Referred to new_niea_pa107.doc
12Student T test
13(No Transcript)
14Graphical definition of IDL
Note Determined from the blank which would be
near zero concentration
15Glossary Method detection level (MDL)
- the constituent concentration that, when
processed through the complete method, produces a
signal with a 99 probability that it is different
from the blank (zero). - For seven replicates of the sample, the mean must
be 3.14s above the blank where is the standard
deviation of the seven replicates. Compute MDL
from replicate measurements one to five times the
actual MDL.
16Glossary Method detection level (MDL)
- Prepare the seven portions over a periods of at
least three days - The initial estimated concentration of prepared
solution is about one to five times of the MDL. - Analyze the seven portions twice
- Compare the two population variances by F-test
- Referred to handout and new_niea_pa107.doc
17Graphical definition of MDL
Note Determined from the concentration which
is near the expected limit of detection
18Glossary
- Level of quantitation (LOQ)/minimum quantitation
level (MQL)the constituent concentration that
produces a signal sufficiently greater than the
blank that it can be detected within specified
levels by good laboratories during routine
operating conditions. - Typically, it is the concentration that produces
a signal 10 time the LDL or two to three times of
the MDL.
19Glossary
- Quality assessmentprocedure for determining the
quality of laboratory measurements by use of data
from internal and external quality control
measures, mostly referred for external quality
control - Quality assurancea definitive plan for
laboratory operation that specifies the measures
used to produce data of known precision and bias.
- Quality controlset of measures within a sample
analysis methodology to assure that the process
is in control.
20Glossary
- Surrogate standarda pure compound added to a
sample in the laboratory just before processing
so that the overall efficiency of a method can be
determined. - Internal standarda pure compound added to a
sample extract just before instrumental analysis
to permit correction for inefficiencies.
21Glossary Type I and Type II errors
- Type I errorWe make the mistake of rejecting
the null hypothesis when it is true.?
P(rejecting H0 when it is true). - Type II errorWe make the mistake of failing to
reject the null hypothesis when it is false.?
P(failing to reject H0 when it is false).
22Glossary Relative standard deviation
- Also known as the coefficient of variation (CV),
which commonly is expressed as a percentage. - If analyses at low concentrations yield a result
of 10 1.5 mg/L and at high concentrations 100
8 mg/L. the standard deviations do not appear
comparable. However, the percent relative
standard deviations are 100 (1.5/10) 15 and
100 (8/100) 8. which indicate the smaller
variability obtained by using this parameter.
23Homework 1
- Grab your old statistics book, (if you never
took, go to library to find it), explain the
meaning of - (1) t-distribution
- (2) typically, people say the reasons of 99 of
the total measurement for normal distribution is
when ??3? large (unlimited) or small number of
samples (30 samples) - (3) The meaning of F-distribution
- (4) Bring your statistic book to the class
24Homework 2
- In the previous example, if eight aliquots are
used. The eighth sample concentration is same as
the first one for both sets. Find the new MDL.
25F-distributioncompare two population variance
26(No Transcript)
27Calibration curve preparation
- As a minimum, measure three different dilutions
of the standard when an analysis is initiated.
(standard method). EPA Taiwan requires five
different dilutions of the standard - The lowest concentration should be Level of
quantitation (LOQ) - Subsequently, verify the standard curve daily by
analyzing one or more standards within the linear
range, as specified in the individual method.
(preferably by different source of standard) - Reportable analytical results are those within
the range of the standard dilutions used.
(standard method). Within 20 to 80 of the
highest concentration is preferable (EPA Taiwan).
28Calibration curve preparation
- Do not report values above the highest standard
unless an initial demonstration of greater linear
range has been made, no instrument parameters
have been changed, and the value is less than 1.5
times the highest standard. - The lowest reportable value is the MDL, provided
that the lowest calibration standard is less than
10 times the MDL (It should be). If a blank is
subtracted report the result even if it is
negative.
29Reporting
- Apply the MDL to reporting sample results as
follows Report results below the MDL as "not
detected." - Report results above the LOQ with a value and its
associated error
30Calibration curve determination
- Linear model and non-linear model (rare).
- Least square method. For linear model, R2 should
be greater than 0.995. - Calibration factor (Internal standard) and
Response factor (external standard). Mostly use
for individual organic compounds (VOC, pesticide,
GC, GC/MS, etc). The ratio of response (peak
area) for standard(As)and amount of compound for
standard(Ws)called calibration factor (CF, EPA
Taiwan) or generally called response factor (RF,
Standard method)
31Quality Assurance
- Quality assurance (QA) is the definitive program
for laboratory operation that specifies the
measures required to produce defensible data of
known precision and accuracy. - Quality assurance (QA) includes Quality control
and quality assessment
32Quality Assurance Plan
- Establish a QA program and prepare a QA manual of
plan. Include in the QA manual and associated
documents the following items - cover sheet with plan approval signatures,
- staff or organization responsibilities
- Sample control and documentation
- standard operating procedure for each analytical
method (SOP)
33Quality Assurance Plan
- analyst training requirements
- equipment preventive maintenance procedures
- calibration procedures
- corrective actions
- internal quality control activities
- performance audits
- data assessment procedures for bias and
precision, and data reduction, validation, and
reporting.
34Quality Control
- Quality control (QC) may be either internal or
external. - Internal QC is the subject of this section All
analysts use some QC as an intuitive effort to
produce credible results. However, a good quality
control program consists of at least seven
elements - certification of operator competence,
- recovery of known additions,
- analysis of externally supplied standards,
- analysis of reagent blanks,
- calibration with standards,
- analysis of duplicates.
- maintenance of control charts.
- ???????????????????????????????
35Quality Control
- external QC, also known as "quality assessment,"
is discussed in 1020C.
361. Certification of Operator Competence
- Before an analyst is permitted to do reportable
work, competence in making the analysis is to be
demonstrated. - Make a minimum of four replicate analyses of an
independently prepared check sample having a
concentration between 5 and 50 times the method
detection limit (MDL) for the analysis in that
laboratory. General limits for acceptable work
are shown in the following Table.
371. Certification of Operator Competence
382. Recovery of Known Additions
- Use the recovery or known additions as part of a
regular analytical protocol. Use known additions
to verify the absence of matrix effects. When a
new matrix type is to be analyzed, verify the
amount of interference. - The sum of the duplicates and known additions
must equal at least 10 of the number of samples.
- Make the known addition between 5 and 50 times
the MDL - or between 1 and 10 times the ambient level
392. Recovery of Known Additions
- Do not use a known addition above the
demonstrated linear range of the method - use concentrated solutions so volume change in
sample is negligible. (new_niea_pa104.doc, P.7) - Or calculate recovery by mass balance
- See Table 10201 for acceptable limits
- Referred to new_niea_pa104.doc
40Example for spiked recovery
- A sample with 100 mg/L of calcium. 1000 mg/L
standard of calcium, 5 ml, was used to spiked
into 100 ml water sample. The result showed
150mg/L. Find your recovery.
The recovery will be 100 if added volume is
ignored.
413. Analysis of Externally Supplied Standards
- As a minimum, analyze externally supplied
standards whenever analysis of known additions
does not result in acceptable recovery or once
each day, whichever is more frequent - Use laboratory control standards with a
concentration between 5 and 50 times the MDL or
near sample ambient levels, whichever is greater.
423. Analysis of Externally Supplied Standards
- Where possible, use certified reference materials
as laboratory control standards. National
Institute of Standards and Technology (NIST) or
Standard Reference Materials are preferred, if
available. - Prepare them independently from the standards
used for calibration
434. Analysis of Reagent Blanks
- Analyze reagent blanks whenever new reagents are
used - Analyze a minimum of 5 of the sample load as
reagent blanks - Analyze a reagent blank after any sample with a
concentration greater than that of the highest
standard or that might result in carryover from
one sample - Acceptance level lt 2 MDL (general used) or lt 5
of the regulation allowed (EPA Taiwan)
445. Calibration with Standards
- As a minimum, measure three different dilutions
of the standard when an analysis is initiated.
(same as previously mentioned) - Subsequently, verify the standard curve daily by
analyzing one or more standards within the linear
range, as specified in the individual method.
456. Analysis of Duplicates
- When most samples have measurable levels of the
constituent being determined, analysis of
duplicate samples is effective for assessing
precision. - Analyze duplicates and known additions in
matrices representative of the samples analyzed
in the laboratory. - See Table 1020I for acceptable limits for
duplicate analyses.
467. Control Chart
- Referred to new_niea_pa105
- Precision (range) (???????)
- Mean Chart (?????????) The mean chart for QC
sample is constructed from the average and
standard deviation of a specified number of
measurements of the analytes of interest. - Accuracy Chart (?????????)The accuracy chart for
QC sample is constructed from the known spike and
standard deviation of a specified number of
measurements of the analytes of interest.
47Precision (range) Chart(???????)
48Precision (range) Chart(???????)
- Every 10 samples to analyze a duplicate.
- 15 pairs of duplicates are needed to constructed
the precision chart - For every RPD ()
49Precision (range) Chart(???????)
- Average
- Standard deviations of those 15 pairs
- ?2s and ?3s to represent the warning level and
control level
50Precision (range) Chart(???????)
51Mean Chart (?????????)
- 15 pairs of duplicates are needed to constructed
the mean chart - ?2s and ?3s to represent the warning level and
control level - S is calculated from calculated value for mean,
(mean chart) or using percentage if the
concentration varies (accuracy chart) .
52Mean Chart (?????????)
53Mean Chart (?????????)
54Accuracy Chart (?????????)
- 15 pairs of duplicates are needed to constructed
the mean chart - or
- based on ways of calculation
- ?2s and ?3s to represent the warning level and
control level
55Accuracy Chart (?????????)
56Accuracy Chart (?????????)
57Chart analyses
- Control limitIf one measurement exceeds a CL,
repeat the analysis immediately. If the repeat
measurement is within the CL, continue analyses
if it exceeds the CL, discontinue analyses and
correct the problem.(????) - Warning limitIf two out of three successive
points exceed a WL, analyze another sample. If
the next point is within the WL, continue
analyses if the next point exceeds the WL,
evaluate potential bias and correct the problem.
58Chart analyses
- Standard deviationIf four out of five successive
points exceed l.s, or are in decreasing or
increasing order, analyze another sample. If the
next point is less than 1s. or changes the order,
continue analyses otherwise, discontinue
analyses and correct the problem.(????) - Trendingif seven successive samples are on the
same side of the central line, discontinue
analyses and correct the problem.(????)
59Example of Chart analyses
60Possible ways of correction action(example of a
commercial lab)
61?????????
62Homework
- Glucose BOD Standards. The data below are 30
measurements on a standard glucose/glutamate
mixture that has a theoretical BOD of 200 mg/L.
Use these data to construct a Mean chart. - 203 213 223 205 209 200 200 196 201 206 192 206
185 199 201 206 196 214 189 205 201 226 207 214
210 207 188 199 198 200
63Quality assessment(External quality
control)(P.1-12)
- Example To evaluate the performance for a
commercial laboratory - Laboratory Check Samples (internal Proficiency)
- Use samples with known amounts of the
constituent of interest supplied by an outside
agency or blind additions prepared in-dependently
within the laboratory to determine recovery
achieved by an analyst. - In general, method uncertainty will have been
established beforehand acceptable recovery falls
within the established uncertainty. For example,
if the acceptable range of recovery for a
substance is 85 to 115, then the analyst is
expected to achieve a recovery within that range
on all performance evaluation samples.
64Quality assessment(External quality control)
- 2. Laboratory inter-comparison samples
- 3. Compliance Audits
- Compliance audits are conducted to evaluate
whether the laboratory meets the applicable
requirements of the SOP or consensus method
claimed as followed by the laboratory. - A recommended format with a few initial items in
the check list is shown in Table 1020III.
65Quality assessment(External quality control)
- 4. Quality System audits
- Quality system audits should be conducted by a
qualified auditor who is knowledgeable about the
section or analysis being audited. - 5. Management review
66Easier ways to checkChecking Correctness of
Analyses
- Anion-cation balance
- Measured TDS calculated TDS (mg/L)
- Calculated TDS to EC (electric conductivity as
?s) ratio (Siemen, S1/ohm-cm)
67Expression of Results
- ????????????????ppm(10-6,?????)?ppb(10-9,?????)??
???????,????????,???? - ????mg/L?µg/L
- ????mg/kg?µg/kg
- ????ppm?mg/Nm3?
68Significant Figures
- Reporting requirement
- All digits in a reported results are expected to
be known definitely except for the last digit - Rounding off (4?6?5????,?5????????)
- Calculation
- Several numbers are multiplied or divided (as
few significant figures as the numbers are
present) - Several numbers are added or subtracted, the
number that has the fewest decimal places is the
one to follow
69Sample collectionGeneral Requirement
- Bake at 450oC all bottles to be used for
organic-analysis sampling, especially in ?g/L
level. - Fill the container full or leave space?(P.1-35)
70Sample collectionTypes of Samples
- Grab sample ????(for example, influent)
- Composite sample ????
- Samples are preferred to analyze on site DO,
residual chlorine, soluble sulfide, temperature
and pH.
71Sample collection Chain of Custody procedures???
- A. Sample labels
- B. Sample seals
- c. Field log book Record
- all information pertinent to a field survey or
sampling in a bound log book. As a minimum,
include the following in the log book purpose of
sampling location of sampling point name and
address of field contact producer of material
being sampled and address, if different from
location and type of sample.
72Chain of Custody procedure
- d. Chain-of-custody record
- e. Sample analysis request sheet
- f. Sample delivery to laboratory
- g. Receipt and logging of sample
- h. Assignment of sample for analysis
73Apparatus Containers
- Borosilicate glass bottle is commonly used.
- Silica and sodium maybe leached from glass but
not plastic - Trace metals may absorb onto the walls of glass
containers - Use glass containers for all the organics. (VOC,
pesticide, oil, etc) - Amber bottle for minimization of photodegradation
- Referred Table 1060I
- To contain or to deliver? (NIEA PA-106, P.2)
74To contain or to deliver?
- TC (to contain) and TD (to deliver).
- Graduated cylinders (??) and volumetric
flasks(??), are usually marked with a TC. When
liquid is poured from a piece of glassware a
small amount remains behind, clinging to the
sides of the vessel. - Pipets (???) and burets(???), are marked with a
TD.
75Sample storage (some keys)
- Follow 1060I
- Purposes To retard chemical and biological
changes that inevitably continue after sample
collection. - Certain cations are subject to loss by adsorption
or ion exchange with the walls of glass
containers. - These include aluminum, cadmium, chromium,
copper, iron, lead, manganese, silver, and zinc,
which are best collected in a separate clean
bottle and acidified with nitric acid to a pH
below 2.0 to minimize precipitation and
adsorption on container walls.
76Sample storage (some keys)
- determine temperature, reduction-oxidation
potential (ORP) and dissolved gases in situ and
pH, specific conductance, turbidity, and
alkalinity immediately after sample collection. - Zero head-space is important in preservation of
samples with volatile organic compounds and
radon. Avoid loss of volatile materials by
collecting sample in a completely filled
container.
77Sample preservation (some keys)
- Follow 1060I
- keep samples as cool as possible without
freezing. - Analyze samples as quickly as possible on arrival
at the laboratory. If immediate analysis is not
possible, preferably store at 4C. - Use chemical preservatives only when they do not
interfere with the analysis being made.
78Reagent-Grade Water
What is the processes in our laboratory?
79Reagent water quality guideline
What is the water quality in our laboratory?
80Sample volume needed
- Referred Table 1060I.
- Most of the time, 2-L sample is needed for most
physical and chemical analyses.
81Safe laboratory practice
82Supplements for statistics
83Student-t Distribution
f(t)
?/2
?/2
(???)
0
-tc
t
tc
See any Student-t Distribution Table
84Student-t vs. Normal Distribution
- 1. Both are symmetric bell-shaped
distributions. - 2. Student-t distribution has fatter tails than
the normal. - 3. Student-t converges to the normal for
infinite sample. - 4. Student-t conditional on degrees of freedom
(df). - 5. Normal is a good approximation of Student-t
for the first few decimal places when df gt 30 or
so.
85Probability statements
P( t lt -tc ) P( t gt tc ) ???
P(-tc lt t lt tc ) 1 ? ?
86Deriving a Confidence Interval
87Hypothesis Test Procedures
- 1. A null hypothesis, H0.
- 2. An alternative hypothesis, H1.
- 3. A test statistic.
- 4. A rejection region.
88The Null hypothesis
This hypothesis is a belief we maintain until
proven otherwise
89The Alternative Hypotheses
90The Test Statistic
For Z, T or F Test Statistic
If null hypothesis true
If null hypothesis not true t has some other
probability distribution
91Two-Tailed Test
f(t)
reject
reject
do not reject
?/2
?/2
0
-tc
t
tc
Probability of a Type I error ?
92One-Tailed Test
f(t)
reject
do not reject
?
0
t
tc
Note ignore left tail
Probability of a Type I error ?
93Rejection Rules
- 1. Two-Sided Test If the value of the test
statistic falls in the critical region in either
tail of the t-distribution, then we reject the
null hypothesis in favor of the alternative.
Otherwise, we do not reject the null hypothesis. - 2. Left-Tail TestIf the value of the test
statistic falls in the critical region which lies
in the left tail of the t-distribution, then we
reject the null hypothesis in favor of the
alternative. Otherwise, we do not reject the null
hypothesis - 3. Right-Tail TestIf the value of the test
statistic falls in the critical region which lies
in the right tail of the t-distribution, then we
reject the null hypothesis in favor of the
alternative. Otherwise, we do not reject the null
hypothesis
94Type I and Type II errors
- Type I errorWe make the mistake of rejecting
the null hypothesis when it is true.?
P(rejecting H0 when it is true). - Type II errorWe make the mistake of failing to
reject the null hypothesis when it is false.?
P(failing to reject H0 when it is false).
95Failure to Reject does not mean Hypothesis is
True!
When we fail to reject the null hypothesis, that
does not mean that we can conclude that the null
hypothesis is true.
Failure to reject the null hypothesis is a rather
weak conclusion since it only means that the data
are compatible with the null hypothesis.
If the null hypothesis ?20 is not rejected, then
the null hypotheses ?20.1, ?2-.2, etc. may well
be compatible with the data too, so tests of
these hypotheses would not be rejected either.
96Format for Hypothesis Testing
- 1. Determine null and alternative hypotheses.
- 2. Specify the test statistic and its
distribution as if the null hypothesis were
true. - 3. Select ? and determine the rejection region.
- 4. Calculate the sample value of test
statistic. - 5. State your conclusion.