Calibration MINOS and Minerva - PowerPoint PPT Presentation

1 / 17
About This Presentation
Title:

Calibration MINOS and Minerva

Description:

Studies were done of the calibration necessary to extract Dm2 ... LEDs are flashed at a constant height every hour. From CalDet (R. Nichols) Minerva Calibration ... – PowerPoint PPT presentation

Number of Views:19
Avg rating:3.0/5.0
Slides: 18
Provided by: KevinMc61
Category:

less

Transcript and Presenter's Notes

Title: Calibration MINOS and Minerva


1
Calibration (MINOS and Minerva)
2
MINOS Calibration Requirements
  • Studies were done of the calibration necessary to
    extract Dm2 and sin22q from nm CC events with
    10 precision
  • Energy calibration ND vs. FD consistent to 2
  • Absolute energy 5

En,cc Ehad_shower Em
3
Energy reconstruction in MINOS
  • ADC ? light
  • PMT gain singles data or low-light runs with
    light-injection (LI) system
  • Gain changes LI drift-point calibration
  • Linearity of PMTs readout LI system charge
    injection (for PIN)
  • Electronics calibration (pedestals linearity)
  • Light ? MIP Units
  • Attenuation correction, from production mapping
    of scintillator
  • Strip-to-strip calibration with cosmic muons
  • Path-length corrections to establish MIP energy
    scale
  • Near-far differences must be corrected with
    cosmic muons
  • MIP Units ? Hadron energy
  • From the calibration detector (1m x 1m
    replica of the MINOS detector)

4
MINOS Light Injection System
  • The MINOS light injection system has 4 main
    purposes
  • To linearize the response of the PMT /
    electronics
  • To monitor short timescale drifting of the
    response
  • To accurately determine the position of the
    single photoelectron peak
  • Is everything plugged in?

5
LI system Basics
Light from ultra-violet LEDs illuminates up to 70
fibers in a pulser box. Light is transported to
the light injection manifold where it is
dispersed uniformly over 8-10 WLS fibers.
Scintillator module
PMT
Pulser box (UV LEDs)
Light injection manifold
Calibrates the PMT /electronics chain, but not
the scintillator/ fiber.
Linear PIN diode (stable to 0.5 over several
months. Charge injection system linearizes the
readout electronics for the PIN)
6
Linearizing PMT / Electronics
  • Linearizing PMT / electronics response
  • Light is injected over a range of intensity. PIN
    diode monitors the light output and is used to
    convert the ADC scale to light.
  • Full LI runs are taken on the order of once a
    month.

7
2 Drifting in response
  • CR rate at the far detector is only 500 / strip /
    month.
  • LEDs are flashed at a constant height every hour.

From CalDet (R. Nichols)
8
Minerva Calibration
  • In my opinion the important first task is
    defining the calibration requirements imposed on
    us by our physics mission
  • Calibrate Electronics, scintillator/fiber
    system, timing, B-field, single particle
    response.
  • I am assuming we will require a test beam run in
    e/p/p/m beams from several hundred MeV to a few
    GeV.

9
Minerva Calibration
  • I am assuming we will require a test beam run in
    e/p/p/m beams from several hundred MeV to a few
    GeV.
  • Where?
  • When?
  • What subsystems?

For MINOS, hadron comparisons to GHEISHA,
GFLUKA, GCALOR and SLAC-GHEISHA show large
differences
10
Calibration Detector (2 GeV)
MIPs
11
Calibration Detector
12
Minerva Calibration
  • One clear physics strength
  • is particle ID.
  • Requires good absolute
  • calibration of the strips
  • as well as good strip to
  • strip calibration.
  • Repeat Daves previous studies
  • to determine our requirements on
  • strip to strip calibration?

13
Calibration Discussion
  • From a quick calculation the muon rates seen by
    Minerva in the near hall are approximately
  • 400k beam muons / month
  • 500k stopping CR muons / month
  • 6k CR hits / strip / month
  • Muon rates are probably sufficient to monitor
    gain drifts.
  • What will we use muons to calibrate?
  • Strip positions (need to know at mm level)
  • Strip-to-strip calibration
  • Attenuation correction for full strip after fiber
    insertion?
  • What timescale will be necessary.

14
MINOS m light output
Large corrections are also related to the
attenuation in the fiber.
15
Calibration in Assembly - MINOS
5mCi Cs-137 (662keV g) Determine Light
output Strip position to 1mm
16
Module Mapper Results
  • Tests EVERY strip
  • Both ends
  • 8cm steps
  • Maps light output, reproducible to 1
  • 40 minutes/module

17
Module Uniformity
11 Variation (s) with over 90 of Production
179,000 Strip Ends Completed Only 290 show any
damage lt0.2
Write a Comment
User Comments (0)
About PowerShow.com