Title: Introduction of Multi Myograph System 610M
1welcome
2Introduction of Multi Myograph System 610M
3Background
- The technique was first developed for
investigating small vessels with internal
diameters down to around 100 µm by Mulcany
Halpern , in 1976,1977 - Multi Myograph System 610M is suited to vessels
with internal diameters of 100-400 µm - At least 50 of the precapillary pressure drop
occurs in the vessels with internal diameters
than 100 µm ( proximal resistance vessels)
4General Principles
- Opening order
- powerlab myograph software
- Closing order
- software myograph powerlab
- Line connection
- one-to-one correspondingly
5General Principles
- Buttons function
- main options (zero/calibrate/temp
erature, etc.) - number changing
(decrease/increase) - F1 channel selecting
- F2 detailed operation
6General Principles
Calibration bridge
- Zero before experiments
- F 9.81 m N
tip
wire
7solutions
NAK KPSSNE 10 µ M
NAPSS PSSNE 10 µM
8dissection
Example
- The dissection of mesenteric small ateries from
rats at autopsy - mid-line laparotomy to exteriorize mesenteric bed
- Use scissors to remove 10cm of intestine together
with its feeding vasculature, including part of
the superior mesenteric artery. Proximal end of
intestinal section should be 10cm from pylorus
- Place excised section in Petri dish containing
PSS. Petri dish contains a 5 mm thick layer of
Sylgaard to hold fixing pins.
9dissection
- Pin down the proximal end of the intestine on
the left hand side of the dish, and pin out the
remainder of the intestine in an anticlockwise
direction - Dissect out the test segment (normally the third
branch) together with a piece of the proximal
branch
- Using ocular dissection scissors, cut through
the mesenteric membrane along either side of
vessel, about 1-2mm from the vessel
10dissection
- Dissect away the vein (distinguish from the
pattern of the branch points) - Clean artery of connective tissue by holding the
connective tissue with forceps and pulling gently
away from the artery - Cut distal end of vessel to be investigated and
cut unwanted arcade 2mm from feeder artery, which
are to be used as a handle to move the vessel - Cut proximal end of feeder artery
11mounting
- Caution! The procedure involves attaching the
mounting wires to jaws which are in turn mounted
on the force transducer (sensitivity- 0.01 mN
(1mg) upper limit- 1N (100g) ) - So ! Do not press the jaws too hard together a
movement of 50mm after they have touched is
sufficient to hold the wires clamped
12mounting
jaw
screw
- Attaching the first mounting wire
- Clamp the wire between two jaws
- Wrap the far end of the wire around under fixing
screw clockwise and tighten the screw - Fill myograph with PSS at room temperature
13mounting
- Mounting of the vessel
- Hold the handle segment of the excised arcade
and try to mount the vessel onto the proximal end
of wire. - Pull the proximal end of excised arcade along
wire until the vessel to be investigated is
situated between the mounting jaws - Screw jaws together to clamp the wire. Wrap the
near end of wire around under near fixing screw
again in a clockwise direction
14mounting
- Using forceps, gently rub vessel on the far side
of the jaw and sever the vessel so that the
portion of the excised arcade on the far side of
the jaw can be pulled away - Screw jaws apart, take a second wire and align it
parallel with the vessel, passing into the far
end of the lumen - Screw jaws together. Ensure that the second wire
moves under the first. Wrap the ends of wire like
the former - Screw jaws apart so that the wires should be
levelled - Connect oxygen line to chamber and start heating
15Normalization
T the myograph vessel supports had been moved
together (the mounting wires just touched)
B the vessel supports were moved slightly
apart, so that the record at B shows zero tension
1, 2, 3, and 4 the vessel was slowly
stretched, so that the records show the forces 1
min after each stretch. At 4, the vessel supports
were again moved together to release some of the
resting tension
N the vessel internal circumference was set to
IC1 (the internal circumference at which the
active force production of the vessel is maximal)
16Normalization
IC100 the vessels internal circumference when
it is fully relaxed and under a transmural
pressure of 100mmHg
IC1 0.9 IC100
Normalized lumen diameter is taken as I1 IC1/p
T wall tension, T pi(IC/2p)
Pi effective pressure, pi wall
tension/(IC/2p)
17Measurement of responses
- The response of a vessel to an agonist is
normally presented in terms of the change in wall
force above the resting force. -
- F resting alpha(reading resting-reading
baseline) - F agonist alpha(reading agonist-reading
baseline) - Alpha is the force transducer calibration
(mN/recorder division)
i.e. the force response ?FF agonist F resting
?F alpha(reading agonist-reading resting)
18removing Endothelium
- Purpose
- To remove the endothelium from an artery,
trying to eliminate endothelial effects on the
rest of the vessel - To rule out the influence of the endothelium on a
certain mechanism - To confirm the role of the endothelium in the
mechanism
19removing Endothelium
- Procedure
- First , fix the vessel and check the endothelium
function - Set the vessel to a small tension (0.5mN)
- Use a coarse pair of forceps to hold a human hair
(clean, round, straight, thick), with a large
microscope-magnification to look closely at the
artery - Enter the tip of hair into the vessel lumen, and
work down through the hole inner surface of the
vessel rubbing forwards and backwards
20removing Endothelium
- Procedure (cont.)
- Change to fresh solution in the chamber, and let
the vessel equilibrate for 5-10 min - Stretch the vessel to its normalized micrometer
setting and let it relax for 5 min - Check the endothelium-function again
21Checking endothelium-function
- Purpose
- To check whether the relaxing function of the
endothelium is intact after mounting a vessel - To check the success of an endothelium-removal
procedure which you have applied to a vessel
22Checking endothelium-function
- Principle
- By stimulating a vessel with Acetylcholine
(Ach.), the vessel will relax, following the
release of EDRF (Endothelium Derived Relaxing
Factor) from the endothelium. - The procedure can be performed at any time after
the heating and equilibration of the vessel.
23Checking endothelium-function
- Protocol 1 Quick check-single dose
- Stimulate 4 min with NA3PSS (10ml)
- Add 10 ml stock B to the chamber to give
510(-6)M Ach - Wait 2 min
- Wash out 4 times with PSS2.5
24Checking endothelium-function
- Protocol 2 Ach-concentration response curve
(Ach-CRC) - Stimulate 4 min with NA3PSS (10ml)
- Add 20 ml stock E to the chamber to give 10(-8)M
Ach - After 2 min, add 18 ml stock D to the chamber to
give 10(-7)M Ach - Repeat 3) with stock C and B to give 10(-6) and
10(-5) M Ach - After 2 min, add 90 ml stock A to the chamber,
to give 10(-4)M Ach - Wait 2 min
- Wash out 4 times with PSS2.5
25Checking endothelium-function
- Solutions
- NA3PSSPSS2.5noradrenaline 3 mM
- Stock AAcetylcholine (Ach) 10(-2)M (stock
solution in freezer) - Stock B Ach 5 10(-3)M . Made by mixing stock A
and water 11 - Stock C Ach 5 10(-4)M . Made by diluting stock
B 10-fold with water - Stock D Ach 5 10(-5)M . Made by diluting stock
C 10-fold with water - Stock E Ach 5 10(-6)M . Made by diluting stock
D 10-fold with water
26Thats all, thank you