vaporizer and vaporiser - PowerPoint PPT Presentation

About This Presentation
Title:

vaporizer and vaporiser

Description:

vaporizer- history, physics,classification – PowerPoint PPT presentation

Number of Views:33
Slides: 178
Provided by: suna2223333
Tags:

less

Transcript and Presenter's Notes

Title: vaporizer and vaporiser


1
VAPORIZERS
Sunakshi bhatia SPMC bikaner
2
CONTENTS
  1. What Is Vaporizer?
  2. Ideal Vaporizer
  3. Physics
  4. Classification
  5. Basic Functioning
  6. Factors Affecting Vaporizer Performance
  7. Hazards
  8. Safety Features
  9. Vaporizers Of Historical Importance

3
  • WHAT IS VAPORIZER?
  • Vaporizer is an instrument designed to facilitate
    the change of a liquid anesthetic into its vapor
    and add a controlled amount of this vapor to the
    flow of gases going to the patient.

4
IDEAL VAPORIZER
  • Lightweight, robust and durable
  • Easy to transport
  • Corrosion and solvent resistant
  • Leak proof
  • Economical and safe to use
  • minimal servicing
  • Accurate over a wide range of
  • FGF
  • Liquid agent levels
  • Ambient pressures
  • Ambient temperatures
  • Unaffected by
  • Heat loss due to vaporization
  • Pressure changes downstream of the vaporizer
  • Tilting or tipping
  • Compatible with vaporizer interlock systems.

5
PHYSICS
  1. Critical temperature
  2. Vapor and gas
  3. Saturated vapor pressure
  4. Boiling point
  5. Latent heat of vaporization
  6. Heat of vaporization
  7. Thermal conductivity
  8. Specific heat

6
  • Critical temperature
  • Temperature above which a substance in the
    gaseous phase cannot be condensed into a liquid,
    no matter how much pressure is applied.
  • E.g. CT of oxygen is -119 centigrade

7
  • A vapor is a substance that is in gaseous phase
    below its critical temperature. This means that
    the vapor can be condensed to a liquid or to a
    solid by increasing its pressure without reducing
    the temperature,
  • e.g. water vapor is the vapor form of water.
  • The term gas is used to describe a substance that
    is in the gaseous phase above its critical
    temperature
  • e.g. nitrogen is a gas at room temperature and
    atmospheric pressure. Nitrous Oxide, Cyclopropane
    delivered in cylinders

8
  • When a surface of a liquid is exposed to
    air/gas/vacuum? vapor
  • The remaining liquid now has a lower than average
    kinetic energy and hence temperature of the
    liquid decreases along with the mass.
  • Vapor exerts pressure on the walls ? vapor
    pressure which varies with temperature and nature
    of liquid.
  • When the liquid and its vapor are in
    thermodynamic equilibrium, the vapor pressure is
    now called? saturated vapor pressure

9
Vapor pressure
Saturated Vapor pressure
10
  • saturated vapor pressure
  • SVP at 20 Centigrade
  • Halothane243
  • Isofurane238
  • Sevofurane 159 (lowest)
  • Desfurane669(highest) Most volatile
  • If the temperature of a vaporliquid system is
    increased, the SVP increases.
  • The vapor pressure of a liquid does not depend on
    the amount on the liquid in the container, be it
    one liter or thirty liters at the same
    temperature, both samples will have the same
    vapor pressure.

11
  • Boiling point The temperature at which SVP
    becomes equal to atmospheric pressure and at
    which all the liquid agent changes to the vapor
    phase is known as its boiling point.
  • Halothane243 50.2
  • Isofurane238 48.5
  • Sevofurane 159 (lowest) 58.6
  • Desfurane669 (highest) Most volatile 22.8
  • Changes with atmospheric pressure
  • So lower the atmospheric pressure, lower will be
    the boiling point

12
Latent heat of vaporization the calories
required to convert 1 gram or 1 ml of a substance
from the liquid state to the vapor. For most of
agents 60 Cal/ml Important because as the
vapors form, liquid cools, so heat needs to be
provided continuously (Thermo compensation) Speci
fic heat-quantity of heat required to raise the
temperature of 1 g of the substance by 1
centigrade. Use - to find amount of heat to be
supplied to liquid - to find out best material
for vaporizes e.g. copper has high specific
heat(thermal buffering)
13
Thermal conductivity It is a measure of speed
with which heat flows through a substance.
Substances with a high thermal conductivity can
supply more heat to the liquid anesthetic in
them. Wood- least thermal conductivity. Materi
als with higher thermal conductivity are selected
for vaporizer. Thermal capacity Product of
specific heat and mass (amount of heat stored in
the vaporizer body).
14
Concentration of gas
  • Expressed as Vol (v/v) or Partial pressure
  • Vol it is ml/100 ml . It is a relative value
  • Partial pressure is absolute value. For e.g.
  • Partial pressure of 21 oxygen at
  • 1 tm 760 mmhg --------160 mmhg
  • 500mmhg ---------105 mmhg (cause of
    hypoxia)

15
Partial pressure
  • 760 mmhg 15 PSI 1000cm H2O 100 Kpa (SI unit)
  • Daltons law of partial pressure
  • total pressure by a mixture of gases is equal to
    the sum of the partial pressures of each of the
    constituent gases.
  • 21 oxygen exerts how much partial pressure at
    sea level?
  • PxVol
  • 760 x vol 760x21/100 159.6 mmhg.
  • If the liquid is open to the air, then the vapor
    pressure is seen as a partial pressure along with
    the other constituents of the air.
  • Uptake and anesthetic depth depends on partial
    pressure!!concentration dial is volume percent

16
In a bottle containing halothane, calculate the
of halothane above the liquid level ?
  • SVP243mmhg
  • atm760 mmhg
  • Vol243/760x10032(caution!!)

17
  • Vaporizers concentration Denotes the
    concentration delivered by a vaporizer when fresh
    gas containing no vapor passes through it.
  • Significance
  • In an out-of-system vaporizer, vaporizer output
    equals vaporizer concentration.
  • In an in-system vaporizer, expired gases contain
    some volatile anesthetic, and therefore vaporizer
    output can exceed the vaporizer concentration.

18
  • Vaporizers capability Refers to the maximum
    concentration that can be delivered by a
    vapourizer at the highest setting of the
    concentration dial.
  • Vaporizer capability depends on the MAC of the
    agent.
  • For example, Sevoflurane has a higher MAC than
    Isoflurane and therefore needs a vapourizer with
    a higher capability (maximum 8 in case of
    Sevoflurane as compared to 6 with Isoflurane)
  • Halothane -5
  • Isofurane-6

19
  • Vaporizers efficiency
  • Means the ability of a vaporizer to saturate the
    carrier gas passing into the vaporizing chamber
    at the temperature of the liquid.
  • Vaporizer efficiency is increased by using wicks
    and longer vaporizing chambers to increase the
    surface area available for vaporization.

20
Calculate the amount of vapor produced by 1
gramof its liquid agent. How much liquid agent
does avaporizer use per hour?
  • 1 mole of a gas weighs 1 gram molecular weight.
  • The molecular weight of Isoflurane is 185.
  • Therefore, 1 mole of Isoflurane vapour weighs 185
    g.
  • As per Avogadros hypothesis,
  • 1 mole of a substance 22.4 liters of gas
  • 22.4 liters of isoflurane vapour weighs 185 g.
  • Therefore, density of isoflurane vapour is
    185/22.4 8.25 g/L
  • so 8.25 g in 1 L ie 1000ml,
  • Therefore, 1 gram 1000/8.25 mL of vapour(A)

21
  • Density of Isoflurane liquid is 1.5 g/mL.
  • 1.5 g in 1 ml
  • Therefore, 1 gram 1/1.5 mL of liquid (B)
  • Since 1g1g so AB
  • 1/1.5 mL of liquid 1000/8.25 mL of vapour
  • 1mL of liquid 1000/8.25 1.5 180 mL vapour
  • For most modern agents, 1 mL of liquid volatile
    agent yields
  • about 200 mL vapor.

22
vapor used per minute fresh gas flow rate
time concentration setting For example, to
deliver 1 Isoflurane at FGF of 2 liters
per minute for 60 minutes, one would need 1/100
2000 mL 60 min 1200 mL of Isoflurane
vapor. This would correspond to 1200/180 6.7
mL of liquid isoflurane. So from this we can
calculate time for remaining inhalational agent
to finish.
23
How to calculate liquid vapour use ?(another
method) Ehren werth Eisenkraft formula (
1993) 3 x FGF ( Litres / min) x dial setting
v/v FGF 4 litres dial setting 3 ML of
liquid per hour 3x 4 x3 36 ml/ hr
24
CLASSIFICATION
  • the Dorsch and Dorsch classification
  • Method of regulating output
  • Method of vaporization
  • Temperature compensation
  • Location
  • Caliberation
  • Specificity
  • Resistance

25
  • 1. Acc to Method of regulating output
    concentration
  • Variable bypass(concentration caliberated) Ether
    bottle, TEC vaporizers
  • Draw-over
  • Plenum
  • Both variable and measured flow-Alladin Cassette
  • Measured flow Copper kettle, side-arm vernitrol
  • Dual Circuit-gas/vapor blender, Tec6 Desflurane
    vaporizer
  • Injection- Drager DIVA(component of Drager Zeus
    anaesthesia workstation, The Maquet 950 series
    vapourise

26
  • 2.Method of vaporization
  • Flow over
  • --with wickTEC vaporizers
  • --without wickGoldman bottle
  • Bubble through Copper kettle which has sintered
    bronze disks to increase vaporization
  • Flow-over or bubble-through Ether bottle,
    depending on the position of the plunger
  • Injection Desflurane, Kion though Desflurane
    vapourizer has been described as an injector,
    it is more ideally classified as a gas-vapour
    blender.

27
  • 3.Temperature compensation
  • Thermo-compensated
  • --By altered flowTEC vaporizers
  • --By supplied heatcopper kettle
  • --By both supplied heat and altered flow EMO
  • Non compensated Ether bottle.
  • 4.Location
  • In circuit VIC EMO, Goldman bottle
  • Out of circuit VOC TEC vaporizers.

28
Out-of-system vaporizers All measured flow
vaporizers are located out of the
breathing system and have a dedicated flow meter.
Plenum vaporizers (high-resistance) are also used
out of system. In-system vaporizers They are
those in which the patients inspiratory and
expiratory gases go through the vapourizer.
29
5.Calibration
  • Calibration
  • a. Uncaliberated
  • controls not precise
  • OpenWire mask, Napkin
  • PlenumBoyles bottle
  • b. Caliberated
  • directreading, dialcontrolled, automatic
    plenum, percentagetype, and tectype vaporizers
  • Electronically controlled aladdin casket

30
  • 6.Specificity
  • Agent specific TEC vaporizers
  • Multi agent Goldman bottle.
  • 7.Resistance
  • Plenum ( high resistance) TEC vaporizers
  • Draw over ( low resistance) Goldman bottle, EMO.

31
  • Plenum vaporizers(Plenuma chamber in which there
    is high pressure)
  • Driven by positive pressure from the machine.
  • The performance of the vaporizer does not change
    regardless of whether the patient is breathing
    spontaneously or is mechanically ventilated.
  • The internal resistance of the vaporizer is
    usually high, but because the supply pressure is
    constant the vaporizer can be accurately
    calibrated to deliver a precise concentration of
    volatile anesthetic vapor over a wide range of
    fresh gas flows.
  • ADVANTAGE- which works reliably, without external
    power, for many hundreds of hours, and requires
    very little maintenance.
  • .

32
  • Works by splitting the incoming gas into two
    streams.
  • Plenum vaporizer can only work one way round if
    it is connected in reverse, much larger volumes
    of gas enter the vaporizing chamber that can be
    toxic.
  • Technically, although the dial of the vaporizer
    is calibrated in volume percent (e.g. 2), what
    it actually delivers is a partial pressure of
    anesthetic agent (e.g. 2kPa)
  • The performance of the plenum vaporizer depends
    on the SVP of the agent. SVP is unique to each
    agent, so it follows that each agent must only be
    used in its own specific vaporizer and safety
    features are needed.
  • They have several features designed to compensate
    for temperature changes like metal jacket,
    bimetallic strip

33
  • Drawover vaporizers
  • Driven by negative pressure by the patient, and
    must therefore have a low resistance and Its
    performance depends on the minute volume of the
    patient its output drops with increasing minute
    ventilation.
  • Design is simpler a glass reservoir mounted in
    the breathing attachment.
  • Drawover vaporizers may be used with any liquid
    volatile agent.
  • Because the performance of the vaporizer is so
    variable, accurate calibration is impossible.
    However, many designs have a lever which adjusts
    the amount of fresh gas which enters the
    vaporizing chamber.
  • May be mounted either way round(bi-directional).

34
  • No temperature compensating features.
  • The relative inefficiency of the Drawover
    vaporizer contributes to its safety. A more
    efficient design would produce too much
    anesthetic vapor.
  • ADVANTAGE
  • cheap to manufacture
  • easy to use.
  • Portable design means that it can be used in the
    field
  • Example-G E O
  • Goldman
  • EMO
  • OME

35
Point DRAW OVER PLENUM
Requirement Patients efforts Gas under positive pressure
Pressure produced Negative pressure by the patient, Downstream of vaporizer Positive pressure , Upstream of vaporizer
Allows Bidirectional flow Unidirectional flow
Flow rare Can be up to 60L/min so wide range of fresh gas flows. Depends on flow meter
Resistance to flow Low because pressure has to be produced by patient. high
Can be used as VOC or VIC Only VOC
performance depends on minute volume of the patient its output drops with increasing minute ventilation. does not change regardless of whether the patient is breathing spontaneously or is mechanically ventilated.
36
(No Transcript)
37
BASIC FUNCTIONING
38
Functioning Of Variable-Bypass Vaporizers
  • Variable bypass vaporizers accept the total gas
    flow from the anesthesia machine and deliver the
    gas flow with a predictable concentration of
    vapor to the common gas outlet.
  • The carrier gas flow entering the vapourizer is
    divided into 2 parts
  • one part going through the bypass chamber(major
    part)
  • the other part going through the vaporizing
    chamber where it is saturated with the vapor of
    the liquid anesthetic agent.

Control dial
39
E.g. Sevofurane How much sevofurene collected
vaporizing chamber? SVP/atmospheric
pressure157/760 x 100 21 21 is lethal. So
dilution is required
40
  • SPLITTING RATIO
  • The ratio of the carrier gas flows through bypass
    and vaporizing chambers is called the splitting
    ratio.

The splitting ratio is affected by the
concentration setting on the dial ,SVP and by the
temperature-compensating mechanism in the
vaporizer.
41
  • Can vaporizers be interchanged?
  • Different anesthetic agents have different
    saturated vapor pressures at a given temperature.
  • Therefore, the partial pressure exerted by the
    vapor in a mixture of gases will be different.
  • Hence, splitting ratios are agent-specific and
    vaporizers cannot be interchanged unless 2 vapors
    have the same vapor pressure.

42
Functioning Of A Measured-flow Vapourizer
  • This type of vapourizer utilizes a measured flow
    of carrier gas, usually oxygen, to pick up
    anesthetic vapour.
  • This vapourizer system consists of 3 parts
  • vapourizer fitted with a thermometer
  • flowmeter assembly
  • vapourizer circuit control valve.(diagram)

43
  • A measured amount of carrier gas flows through a
    flow meter and onto the vapourizer circuit
    control valve, which directs it to the
    vapourizer.
  • The gas saturated with vapor is returned to the
    vapourizer circuit control valve, where it is
    diluted by the remaining fresh gas flow and goes
    to the patient.

44
  • It is important to remember that
  • The oxygen flowing through the vapourizer should
    be counted as additional to that going directly
    to the patient.
  • The vapourizer should not be turned on until the
    flows on the other flow meters have been seta
    lethal concentration of agent may be delivered to
    the patient.

45
Problems of Design
  • High flows of fresh gas going through the whole
    vaporiser can affect its output.
  • The temperature of the vaporiser drops with use
    and this can affect its output.
  • Some ventilators transmit a positive pressure
    back into vaporiser which can affect its output.

46
Problem Of Flows
  • If FGF is high, the vaporization process cant
    keep up with excess coming into the vaporization
    chamber.
  • As a result, relative to the high flow of FGF,
    the amount of agent vaporized is inadequate.
  • So this means that at high flows, the basic
    vaporizer delivers less agent concentration than
    is set on the dial.

47
  • 2.Method of vaporization
  • Flow over
  • --with wickTEC vaporizers
  • --without wickGoldman bottle
  • Bubble through Copper kettle which has sintered
    bronze disks to increase vaporization
  • Flow-over or bubble-through Ether bottle,
    depending on the position of the plunger
  • Injection Desflurane, Kion though Desflurane
    vapourizer has been described as an injector,
    it is more ideally classified as a gas-vapour
    blender.

48
Solution?
  • Dip Wicks.
  • Due to capillary action, the agent rises into the
    wicks.
  • This increases the surface area of agent exposed
    to the fresh gas entering the vaporisation
    chamber and thereby improves the efficiency of
    vaporisation.

49
  • Bubbles though
  • Example-Copper Kettle use bubbles to increase the
    surface area for vaporization.
  • In these, some of the FGF is bubbled through a
    disk made out of a special material (sintered
    disk) that is very porous.
  • Sintered --produced by or subjected to sintering
    (the process of coalescing a powdered material
    into a solid or porous mass by means of heating
    without liquefaction).
  • The disk is submerged into the agent and when
    fresh gas is sent through it, tiny bubbles form.
  • The tiny bubbles of fresh gas have a very large
    total surface and thus become fully saturated
    with vapor effciently.

50
(No Transcript)
51
Factors affecting efficiency of vaporization in
a bubble-through vaporizer? 1. Size of the
bubbles The smaller the bubbles, the more are
the surface area available for equilibration. For
example, in the copper kettle, a sintered bronze
plate is used to break the gas flow into small
bubbles to increase efficiency of
vaporization. 2. Depth of the liquid The deeper
the liquid, the more is the time required for
bubbles to reach the surface, allowing more time
for equilibration. For example, in the ether
bottle, depressing the plunger below the level of
the liquid increases vaporization. 3. Velocity
of carrier gas flow The faster the flow of
bubbles, the less is the time available for
equilibration.
52
Problem Of Temperature
  • For vaporisation to occur, the anesthetic
    molecules have to escape from the liquid state
    and become vapor.
  • This process reduces the energy left in the
    remaining liquid.
  • As more and more molecules escape, more and more
    energy is lost from the liquid.

53
Therefore, as the escaping molecules reduce the
energy left in the liquid, the temperature of the
liquid falls. The falling temperature (lowering
energy) of the liquid means that less molecules
are able to escape. i.e. as vaporisation
happens, the temperature of the liquid falls
causing less vaporisation.
54
Solution?
  • Temperature compensation in vaporizers can be by
  • Supplied heat
  • Use of coppercopper kettle
  • Water jacketEMO
  • Electrically heatedTEC 6 vapourizer for
    Desflurane.
  • Altered flow At lower temperatures, a higher
    percentage of carrier gas flows through the
    vapourizing chamber.
  • Eg
  • Bimetallic stripTEC vaporizers.
  • automatic temperature compensating valve

55
Supplied heat
  • In most vaporizers, heat is not actively given.
    That is, we dont electrically heat it
    (complicated and needs a power supply) and nor do
    we light a fire under it.
  • Instead, we make it easy for the vaporizer to use
    heat from the surrounding air.
  • The vaporizing chamber is generally surrounded by
    a lot of metal which helps to minimize the
    temperature drop by two ways.
  • Firstly metal is a very good conductor of heat
    and therefore is able to efficiently transfer
    heat from the surrounding air into the anesthetic
    agent.

56
(No Transcript)
57
  • Secondly, metal acts like a heat store. It
    absorbs heat (green arrows) till its
    temperature equals the temperature of the
    surrounding air.
  • When the anaesthetic agent starts to cool, the
    metal now donates heat ( yellow arrows ) ,
    helping to minimise the temperature drop.
  • However, the metal casing cannot give up heat
    indefinitely and after sometime, its temperature
    also drops.

58
  • Altered flow
  • Increase the flow of gas via the vaporising
    chamber (altering the splitting ratio).
  • One could manually do this by measuring the
    temperature of the liquid with a thermometer and
    increasing the dial setting according to some
    kind of reference chart which is quite tedious as
    we have to do it all the time.
  • Modern vaporisers have removed this hard work.
    When the liquid drops its temperature, the flow
    of gas through the vaporising chamber is
    automatically increased.
  • This is accomplished by an automatic temperature
    compensating valve that influences how much flow
    goes via the vaporising chamber.

59
  • The automatic temperature compensating valve uses
    the physical property that substances (e.g.
    metals and liquids ) become smaller when the
    temperature lowers.
  • A metal rod (shown in black below) shortens as
    the temperature drops.
  • Similarly, a liquid filled in collapsing bellows
    (shown in green below) becomes smaller in volume
    when cooled to a lower temperature.

60
  • This property is used in the design of automatic
    temperature compensating valves in vaporisers.
  • In the design that uses a metal rod, the rod
    offers some resistance to flow into the
    vaporising chamber.
  • As the vaporiser cools, the rod becomes shorter,
    making the valve move away from the opening. This
    reduces the resistance to flow and thus more flow
    occurs into the vaporising chamber.

61
  • Some vaporisers eg. EMO use the expansion or
    contraction property of a special liquid inside
    bellows (shown in green) to control the valve.
  • As the temperature falls, the liquid in the
    bellows contracts into a smaller volume.
  • This makes the bellows shrink, pulling the valve
    away and thereby increase flow.

62
  • Another method uses a bi metallic strip.
  • Different metals expand and contract to differing
    extents when exposed to temperature changes.
  • In the example below, the green metal expands
    and contracts less than the red metal.

63
  • In a bimetallic strip, two metals with very
    different degrees of thermal expansion (
    different coefficients of thermal expansion )
    are fixed together.
  • In the example below, when the temperature drops,
    the green metal contracts much more than the
    red metal.
  • Because they are fixed together, they cannot
    contract independently, like in the diagram
    above.
  • Instead, the green metal tries to drag the
    red metal and causes the bimetallic strip to
    bend.

64
  • This reduces the resistance to flow and thus more
    flow occurs into the vaporising chamber.

65
(No Transcript)
66
FACTORS AFFECTING VAPORIZER PERFORMANCE
  • 1. SURFACE AREA OF THE LIQUID GAS INTERFACE
    Greater the surface area, more will be the
    vaporization. Incorporated with wicks, baffles,
    long vaporizing VC, channels
  • Bubble through gt flow over
  • 2. TEMPERATURE as temperature increase, output
    increase
  • No change between 10 to 40 centigrade
  • 3. TIME Output concentration tend to fall over
    time

67
  • 4. FRESH GAS FLOW RATE
  • At lt250 ml/min-VOC is less due to insufficient
    turbulance.
  • At high flow ratesgt15L/min, the gas leaving
    vaporization chamber is less saturated ( gas
    spends less time in VC- output tend to fall
  • 5. CARRIER GAS COMPOSITION Nitrous oxide is
    more soluble in liquid anaesthesic thus
    decreasing the effective volume passing throug
    Vaporising Chamber. So decrease vapour of agent.
  • 6. AMBIENT PRESSURE
  • - SVP is solely a function of temperature, so if
    ambient pressure is reduced, the constant SVP
    becomes a greater proportion of the total
    pressure -- output increases.

68
  • 7. EFFECT OF REBREATHING
  • causes a difference between the vaporizer
    setting and the inspired concentration with
    significant rebreathing.
  • More the rebreathing , more will be vapor
    concentration.
  • An agent analyzer is needed
  • 8. BOILING POINT Higher boiling point, lesser
    output
  • 9. INTERMITTENT BACK PRESSURE-
  • positive pressure and oxygen flush can cause 2
    situations
  • VOC increasesPumping effect
  • VOC decreases Pressurizing effect

69
PUMPING EFFECT
  • During the positive pressure, there is a pressure
    rise and during expiration, there is a sharp drop
    in pressure.
  • These pressure changes can be transmitted back
    into the vaporiser and can affect the
    concentration of anaesthetic agent delivered.
  • The effect of changing pressure affects the
    output of the vaporiser is called the pumping
    effect.

70
  • When the bag is squeezed pressure is transmitted
    back into the vaporiser
  • This back pressure is transmitted to both, the
    by pass channel and also to the vaporising
    chamber.
  • This back pressure opposes the flow of the
    fresh gas in both the by pass channel and the
    vaporising chamber.
  • The fresh gas entering the vaporiser tries to
    move forward and gets compressed both in the by
    pass channel and the vaporising chamber.
  • However, the vaporising chamber volume is much
    larger than the by pass channel volume, and
    thus, more fresh gas gets compressed into it than
    into the by pass channel.
  • This extra fresh gas that enters the vaporising
    chamber collects anaesthetic vapor.

71
  • when the positive pressure is suddenly released
    (expiration).
  • The previously compressed gases now suddenly
    expands in all directions.
  • Some of the rapidly expanding gas (containing
    vapor) enter the inlet of the vaporiser and cross
    over into the by pass channel.

72
  • Normally, a vaporiser by pass channel does not
    have vapor.
  • So this vapor due the pumping effect is
    additional.
  • When this by pass vapor flows across to the
    exit of the vaporiser, it meets the vapor from
    the vaporising chamber.
  • The addition of the by pass vapor to the vapor
    from the vaporising chamber raises the final
    concentration of anaesthetic delivered.
  • Thus pumping effect increases the delivered
    concentration of anaesthetic agent.

73
Solution?
  • Tricks to reduce the pumping effect
  • Long inlet tube
  • Increased resistance
  • One way valve

74
  • LONG INLET TUBE
  • The vaporiser inlet tube can be made longer.
  • When the back pressure is suddenly released
    during expiration, as discussed before, the extra
    gas in the vaporising chamber will suddenly
    expand.
  • However, due to the long inlet tubing, the extra
    gas containing vapor expands into the long inlet
    tube and doesnt reach the by pass channel.

75
  • INCREASED RESISTANCE
  • The vaporiser can be designed to have a high
    internal resistance to flow.
  • This high resistance resists changes to flow
    caused by the intermittent back pressure of
    positive pressure ventilation.

76
  • ONE WAY VALVE
  • A one way valve or the unidirectional valve can
    be put between the vaporiser outlet and the
    ventilator / breathing system.
  • In the diagram below, the one way valve is
    allowing gases to flow forwards.
  • However, this valve prevents flow from occurring
    in the reverse direction.
  • This reduces the transmission of back pressure
    to the vaporiser.

77
Pressurizing Effect
  • Occurs when carrier gas is compressed in the
    vapourizing chamber due to high flow rate.
  • So no of carrier gas molecule per unit volume is
    excessively more than no of vapor molecules.
  • This results in a dilution of anesthetic in the
    vapourizing chamber and a decrease in vapourizer
    output.
  • There is usually an interplay between the
    pressurizing and pumping effects.
  • Changes in vapourizer output are more with the
    pumping effect and are of greater clinical
    significance.

78
(No Transcript)
79
PUMPING EFFECT PRESSURIZING EFFECT
Increase in vapourizer output Decrease in vapourizer output
More with Low flows, Low dial settings and When the level of liquid agent in the vapourizer is low More at high gas flows Low dial setting Large pressure fluctuations
More effect so is clinically significant Less effect so is clinically insignificant
80
VAPORIZERS OF HISTORICAL ImPORTANCE
81
Materials used in the constructing of
vaporizers Copper Aluminum MRI compatible
vaporisers Brass Steel Glass
82
  • OPEN DROP METHOD
  • Folded Handkerchief
  • Schimmelbusch Mask with Bellamy Gardner Bottle
  • Bellamy Gardener face mask
  • Yankauer face mask
  • Ochsner wire framed face mask
  • Ogden Inhaler
  • DRAW-OVER VAPORIZERS
  • Flaggs Can
  • Epstein Macintosh Oxford Ether Inhaler
  • Oxford Miniature Vaporizer
  • Goldman
  • Rowbotham
  • Triservice Apparatus
  • PLENUM VAPORIZERS
  • Boyles Ether Vaporizer
  • Copper Kettle
  • Siemens KION Vaporizer

83
Mortons ether inhaler
  • WTG Mortan first publically demonstrated
  • ether anesthesia on 16 October 1846
  • Glass sphere with ether soaked sponge
  • Patient inhaled through mouth piece
  • Drawover type
  • Problem-No means of regulating its output
    concentration or compensating for changes in
    temperature caused by vaporization of the liquid
    anesthetic and the ambient environment.

84
Schimmelbusch Mask
  • The Schimmelbusch Mask is a wire frame mask used
    with a suitable drop bottle like the Bellamy
    Gardner bottle.
  • It can be used for chloroform or a mixture.
  • Draw over, non-temperature compensated,
    multi-agent.

GUTTER
HINGE
85
Bellamy Gardner Bottle
It was the most common drop bottle in use. Used
for open-drop method of anesthesia. The dropper
consists of a red rubber stopper with two metal
tubes. The longer tube dips into the liquid and
the shorter acts as an air vent allowing air to
enter the bottle.
86
(No Transcript)
87
Ogston Inhaler
  • It is similar to the above masks with an upright
    wire cage around the periphery. This helps in
    reducing air dilution.
  • Gamgee is wrapped around the cage to increase the
    concentration by decreasing air dilution.

GAMGEE cotton placed between gauge
88
  • Yankauers face mask had mesh and removable
    spiral wire collar holding down a piece of gauze
    onto which agents such as ether or chloroform was
    applied.
  • Bellamy Gardener face mask for inhaling ether had
    hinged wire mesh covered with double layer of
    wool fabric held in place by lever.
  • Difference- no handle.

89
(No Transcript)
90
Modified Yankauer-Gwathemy Mask
Yankauer Mask
Yankauer-Gwathemy Mask
O2 Supply
91
Prevent air dillution
92
Flaggs Can
  • Drawover,
  • flowover without wicks,
  • non-temperature compensated multiple agents.
  • This is an improvised vaporizer designed for use
    in the field (a coffee jar may be used).
  • Hazards of Flaggs can
  • Tipping, leading to direct entry of ether into
    the tube and respiratory tract. If side holes
    were completely obstructed, hypoxic mixture will
    be delivered to the patient.

93
(No Transcript)
94
Boyles Ether Vaporizer(1920)
Cork connected to mental chain.
ETHER 300ml copper
TRILENE lt300 ml steel
95
  • Plenum, flowover and/or bubble through,
    non-temperature compensated, multiple
    agentsether, trichloroethylene, chloroform,
    methoxyflurane and VOC.
  • It is a glass bottle with a copper lid which
    contains a bypass channel, lever and plunger.
  • Copper is an anticatalyst which prevents
    decomposition of ether.
  • The U shaped inlet tube within the vaporizer
    and the cap of the plunger are made from unplated
    copper.

plunger
96
A. Off B.
Onflowover C. Plunger
depressed bubble through
97
(No Transcript)
98
Goldman (1959)
  • Classification
  • Drawover(low resistance)
  • Variable bypss
  • Flowover,
  • Multiple agentshalothane, isoflurane,
  • Non-calibrated,
  • VIC/VOC.
  • Capacity 30 mL
  • Max 3 halothane

99
  • Construction
  • It consists of
  • Small glass bowl.
  • Bowl is attached to head which receives gas and
    divides gas between vaporizing chamber and by
    pass
  • A control lever on the top is used to alter the
    vaporizer output and is turned counterclockwise
    to increase the concentration.
  • DISADVANTGE-
  • Non-temperature compensated(so warm towel is
    used)
  • Shaking can increase concentration. So never
    carry a filled vaporizer.

100
Rowbotham
  • Similar to Goldman but has high resistance
  • modified with steel wire-gauze wick.

101
Oxford Miniature Vaporizer
  • Classification
  • Drawover
  • flowover,
  • non-temperature compensated
  • multiple agentether, trilene, halothane
    (detachable scales for each).
  • Capacity 30 mL

102
  • ADVANTAGE-
  • low resistance
  • Steel wicks increase vaporizaton
  • Has antifreeze
  • Scales are detachable

103
Epstein Macintosh Oxford Ether Inhaler
  • Classification
  • 24cm
  • OMVEM to be used in war field and can cause
    faster induction of ether.

23 cm
104
(No Transcript)
105
Copper Kettle
  • Classification
  • Plenum
  • Measured flow
  • Bubble through
  • Non-temperature compensated but has heat sink
  • VOC
  • Multiple agents

106
  • Principle
  • A small precise volume of carrier gas is
    completely saturated with the anesthetic agent
    and a predetermined amount is added to the FGF.
  • This differs from the percentage v/v of the
    total gas flow which is used in other vaporizers.
  • Every time the FGF is changed the vaporizer
    setting has to be manually changed.
  • This system requires continuous close attention
    and modification.

107
(No Transcript)
108
Tec(Temperature compensted) vaoporizer
Mordern vaporizer Properties- 1. Variable
bypass 2. Flow over with wick 3. Out of system 4.
Temperature compensated by automatic flow
alteration 5. Concentration calibrated 6. Agent
specific.
109
Knob instead of dial
inlet
outlet
Draining port
110
Thymol of halothane damages the bimetalic strip.
So from tec 3 inside bypass chamber
111
Colour coded
Dial instead of knob
112
Tec 2 was notorious for pumping effect and
tilting of 10 degrees
TEC 3
Tec 3 allows 90 degree
MAJOR ADV- prevents pumping effect
113
(No Transcript)
114
Needs both hands to activate same as tec 3
115
Because release button is on posterior side
116
Long inlet reduces pumping effect
Thermocompensatory mechanism is in close contact
with the liquid.
117
Quick fill and easy fill
118
Super six of Desfurane
  1. In chemical structure 6 fluoride atoms
  2. Oil gas partition coefficient -18
  3. Blood gas partition cofficent-0.42
  4. Brain blood partition coefficient - 1.2
  5. SVP at 20 centigrade- 666
  6. Boiling point -22.8
  7. MAC-6
  8. Vaporizer-Tec 6
  9. Maximum dial setting -18
  10. Interim stop -12
  11. No of indicators- 6
  12. Technique of induction-propofol then 6L/min FGF
    and change dial settings every 2 breaths by 1
    till 6 and maintain it at same for 6 minutes

119
Desflurane Vaporizer (Important question)
  • Desflurane has a very low boiling point and even
    at room temperature, has an high vapor pressure.
  • Also, for small changes in temperature, the vapor
    pressure of desflurane changes quite
    dramatically.
  • I.e. desflurane is said to have a very steep
    Vapor Pressure versus Temperature curve.
  • An operating room temperature is not perfectly
    constant.
  • These physical properties creates a big headache
    for vaporiser designers.

120
  • the standard vaporisers try to resist changes in
    temperature (e.g. by having thick metal
    construction).
  • However, these mechanisms are not perfect and in
    practice small changes in vaporiser temperature
    still occur.
  • This is not a big problem with anaesthetic agents
    such as Iso or Sevo which have a relatively less
    steep Vapor Pressure versus Temperature curves.
  • With Des, with its steep curve, even these small
    temperature changes can cause large changes in
    vapor pressure which cannot be compensated for
    with simple devices such a bimetallic strip.
  • So a whole new vaporiser design had to be made
  • DUAL CIRCUIT GAS VAPOR BLENDER

121
Solution?
  • precisely control the temperature.
  • The heated vapor is then injected into the
    fresh gas flow.
  • In this method, the fresh gas flow coming from
    the flow meters does not split into two streams.
  • There is only one stream for the fresh gas flow,
    and into this stream, the anaesthetic agent is
    directly injected.

122
  • There is a tank (sump) which contains desflurane
    which is electrically heated to a highly
    controlled constant temperature (approx. 400C).
  • Because of the heat, the liquid Desflurane
    becomes gaseous Desflurane at a pressure of about
    two atmospheres (about 1500 mmHg or 200 kPa).

123
  • The amount of Desflurane concentration in the
    fresh gas is controlled by the dial setting set.
  • The dial moves a valve which varies the
    resistance to Desflurane flow from the tank to
    the fresh gas.
  • If we want a higher concentration of Desflurane,
    the valve attached to the dial reduces the
    resistance to flow of Desflurane and more of it
    gets injected into the fresh gas.
  • Conversely, if we want a lower concentration of
    Desflurane, the valve attached to the dial
    increases the resistance to flow and less of it
    gets injected into the fresh gas.

124
(No Transcript)
125
  • If we increase the fresh gas flow, but didnt
    increase the injection rate, the emerging mixture
    will now be inaccurate, the concentration being
    lower than before. A similar thing might happen
    if we increase the rate of injectate without
    increasing FGF.
  • The rate of Desflurane gas injection must be
    adjusted to match the fresh gas flow going
    through the vaporiser.

126
Solution?
  • Manually adjust the dial setting to match the
    fresh gas flow. For low flows, we have to reduce
    the dial setting to reduce the rate of Desflurane
    injection, and for high fresh gas flows, we need
    to do the opposite.
  • This would be really tedious in our modern times.
  • Fortunately, the Desflurane vaporiser
    automatically adjusts the rate of injection of
    desflurane to match the flow rate, and thus keeps
    the delivered concentration constant.

127
  • Flow meters deliver the fresh gas flow 1.
  • The fresh gas travels through pipe 2.
  • Note that, unlike other vaporisers, none of the
    fresh gas goes to the vaporising chamber 4.
  • The vaporising chamber is electrically heated
    3.
  • Using sensors for feedback, the temperature is
    kept very constant.
  • The heating causes the Desflurane to become a gas
    under pressure 4 and this travels down pipe 5.

128
  • The flow of Desflurane is resisted by two valves
    6,13.
  • Valve 6 is the valve that we control when we
    set the dial to a particular concentration.
  • When we increase the concentration setting, the
    valve 6 opens a bit and lowers the resistance,
    allowing more Desflurane to flow through.
  • Valve 13 is an electronically controlled valve.

129
  • Computer 12, the vaporisers brain, is able
    to also alter the flow of Desflurane by
    controlling valve 13. i.e. both we and the
    computer can adjust the desflurane injection
    rate.
  • The Desflurane then goes via pipe 7 and meets
    the fresh gas at 8.
  • The Desflurane mixes with the fresh gas 8 and a
    final concentration emerges from the exit of the
    vaporiser 9.

130
  • How the vaporiser, keep the output concentration
    accurate, adjusts the Desflurane flow when the
    fresh gas flow changes?
  • As explained before, the fresh gas flows in pipe
    2.
  • This pipe has a fixed resistance 10 in its
    path.
  • For the fresh gas flow to overcome this
    resistance 10, the pressure in pipe 2 rises.
  • Higher the fresh gas flow in pipe 2, higher
    will be the pressure rise in pipe 2 since more
    flow has to occur through the same fixed
    resistance 10.

131
  • Similarly, when the fresh gas flow is decreased,
    the lesser flow will find it easier to go through
    the fixed resistance and the pressure in pipe 2
    drops.
  • It is important to remember that the pressure in
    pipe 2 is proportional to the fresh gas flow
    going through it.
  • Higher the flow, higher is the pressure in pipe
    2.
  • Lower the flow, lower is the pressure.

132
  • Device 11 is called a differential pressure
    transducer.
  • It has a diaphragm that on one side is exposed to
    the pressure in pipe 2 carrying fresh gas and
    the other side is exposed to the pressure in pipe
    5 carrying Desflurane.
  • When the pressure is equal neutral position
  • If one side of the diaphragm is at a higher
    pressure than the other side, the pressure
    difference makes the diaphragm move.
  • It continuously keeps computer 12 informed
    about pressure difference information.

133
  • For example, when the fresh gas flow is increased
    1.
  • Increased fresh gas flow flows through pipe 2
    and meets fixed resistance 10.
  • The increased flow through the fixed resistance
    10 makes the pressure in pipe 2 to rise and
    this pressure is experienced by differential
    pressure transducer 11.
  • Since the Desflurane pressure in pipe 5 is now
    lower than the fresh gas pressure in pipe 2,
    the diaphragm in the differential pressure
    transducer 11 moves and a signal about the
    pressure difference is sent to the computer 12.

134
  • The computer 12, acts on the information
    provided.
  • It proceeds to increase the flow of Desflurane to
    inject into the increased fresh gas flow.
  • It commands the electronically controlled valve
    13 to reduce the resistance to flow.
  • As the valve 13 opens up and lowers the
    resistance, the Desflurane flow increases.

135
  • The increased flow of Desflurane causes the
    pressure in pipe 5 to rise.
  • This pressure rise pushes the diaphragm of the
    differential transducer back to its neutral
    position 11.
  • The differential transducer 11 informs the
    computer 12 that the diaphragm is in the
    neutral position.
  • The computer 12 is now stops further opening of
    valve 13 .
  • Since the two flows are matched, the output
    concentration 9 does not change despite the
    increased fresh gas flow.

136
Is this necessary?
137
  • Classification of desfurane vaporizer
  • Concentration calibrated
  • Thermocompensated (electrically supplied heat)
  • Agent-specific
  • Out-of circuit
  • Plenum
  • Gas-vapour blender.

138
(No Transcript)
139
  • The various color-coded alarms are
  • Amberwithout audible alarmwarmup,
  • Amber along with audible alarmlow agent level
    below 50 mL
  • Greenoperational status
  • Redno outputdue to
  • Low agent level less than 20 mL
  • Power failure
  • Malfunction
  • Tilted vapourizer beyond 10 degrees
  • A liquid crystal display (LCD) indicates the
    amount of liquid in the vapourizer
  • Once the vapourizer is connected to an electrical
    source, the vapourizer sump gets heated to 39C.
  • During this warm-up phase, the concentration dial
    is in standby mode and the vapourizer cannot be
    used.
  • Once the proper temperature is achieved, the
    green light comes on indicating readiness for
    operation

140
  • Hazards with desfurane vaporser
  • Vapor may leak into the fresh gas when the
    vaporizer is off
  • The bottle may be dropped when released under
    pressure
  • If the machine uses fresh gas decoupling, special
    software needs to be installed
  • The vaporizer should be placed such that its
    power cord does not cause problems with the
    seating and mounting of other vaporizers.

141
(No Transcript)
142
Draer 19.1
143
(No Transcript)
144
(No Transcript)
145
(No Transcript)
146
(No Transcript)
147
(No Transcript)
148
Electronic- automatic record keeping. Magnetic
coding for specificity Light weight
149
HAZARDS and SAFETY FEATURES IN VAPORISERS
  • Incorrect concentration delivered
  • Misfilling
  • Tipping
  • Leaks
  • Overfilling
  • No Output
  • Reversal of flow

150
Incorrect concentration delivered
  • HIGH CONC.
  • LOW CONC.
  • Liquid anesthetic in delivery tube
  • Pumping effect
  • Reversed flow
  • Agitation
  • Cooling of liquid
  • Pressurizing effect
  • Wicks covered with liquid
  • Very low/high flows.

151
(No Transcript)
152
FILLING SYSTEMS in vaporisers
  • These systems reduce the air pollution associated
    with filling and draining vaporizers.
  • Types
  • Screw Fill
  • Key Fill Systems prone to mis-filling
  • Easy-Fil
  • Solution
  • Quik-Fil
  • Saf-T-Fil

153
(No Transcript)
154
Screw Fill
  • The early plenum vaporizers had a filling port at
    the bottom, and a liquid level indicatorsight
    glass.
  • The screw threaded stopper is unscrewed, agent
    filled and stopper replaced.
  • Filling should stop when the level reaches the
    recommended maximum on the indicator.
  • Hazards are filling the wrong agent, overfilling,
    and environmental pollution.

155
Key Fill Systems
The various parts are made to align and fit into
each other like a key in a lock. The design of
the projections and the slots they fit in are
unique for each individual agent.
B
A
Key fill system. A. Halothane B. Isoflurane C.
Isoflurane key filler with vaporizer
156
  • In addition to the physical shapes being
    different, the key fillers are also color coded
  • The Isoflurane bottle has notches in them
    arranged in a way that is specific for
    Isoflurane.
  • The Isoflurane key filler has specific
    corresponding cuts where the notches of the
    bottle will fit. This makes sure that you cannot
    fix the wrong filler key into the wrong bottle.

157
Isoflurane filler (key) has a notch in a corner.
This fits perfectly with the filling hole in the
Isoflurane vaporiser. The filling hole has pin at
the corner over which the notch of the Isoflurane
filler key can pass over. Not possible to fill,
as halothane key has side notch and isoflurane
vaporizer has corner pin.
158
Easy-Fil
  • The principle is similar to the key-fill system.
  • It is easier to use.
  • The port on the vaporizer has been modified with
    corresponding modification to the distal end of
    adaptor.
  • The filler channel and the bottle adaptor have
    grooves and ridges which align with each other.
  • Each agent has a unique combination.

159
Quik-Fil
  • Sevoflurane bottles are sealed and an agent
    specific filling device is fitted permanently on
    to the neck with a tamper proof crimped metal
    seal.
  • The bottle is opened and inverted onto the
    filling port.
  • A valve opens when a slight pressure is applied
    on the bottle and the agent enters the vaporizer.
  • This prevents spilling and minimizes pollution.

160
Saf-T-Fil
  • It has a filler nozzle which fits into the filler
    port of TEC 6.
  • It is crimped onto the neck of the bottle.
  • The nozzle is pushed onto the spring loaded
    aperture of the port and the bottle is inverted.
  • The contents enter the vaporizer. The LCD display
    indicates the level.
  • When filling is complete the bottle is returned
    to the starting position and the spring then
    ejects the filler nozzle.
  • Advantage---The vaporizer can be filled while in
    use.
  • The bottle is pressurized as Desflurane can boil
    at the ambient temperatures encountered.
  • If the vaporizer has been removed from the back
    bar and a tilt is present, overfilling can occur.
  • The shut-off valve prevents the liquid from
    leaving the sump.

161
(No Transcript)
162
VAPORIZER MOUNTING SYSTEMS
  • Permanent
  • Detachable
  • Select-a-tec system
  • Dräger Medical mounting system

163
Permanent Mounting Vaporiser
  • Require tools to remove or install a vaporizer on
    the anesthesia machine.
  • Advantage- less physical damage to vaporizers and
    fewer leaks.
  • Disadvantage-
  • Inadequate number of mounting locations
    to accommodate all the vaporizers that are likely
    to be needed.
  • If there is a malfunction and the
    vaporizer needs to be exchanged, it cannot be
    easily removed especially with a case underway.

164
Detachable Mounting Vaporiser
  • Here the vaporizers can be mounted or removed
    without using tools.
  • They are standard on most new anesthesia
    machines.
  • The Select-a-tec system and a similar system from
    Dräger Medical are widely used.
  • There is no interchangeability.
  • Advantages
  • More compact anesthesia machine as fewer mounting
    locations needed.
  • Replacement is possible during a case.
  • If malignant hyperthermia is a potential problem,
    the vaporizers can be removed altogether.

165
  • Disadvantages
  • Partial or complete obstruction to gas flow can
    occur due to misalignment
  • Potential for leaks is high, a common leak source
    is an absent or damaged O-ring or if the locking
    lever is unlocked
  • Awareness can occur if there is failure to
    deliver agent vapor to the fresh gas due to
    problems with the mounting system
  • If something is pushed under the vaporizer enough
    so that it slightly lifts off the O-ring, a leak
    may result when the vaporizer is turned ON

166
(No Transcript)
167
Dräger Medical mounting system
  • It has different dimensions.
  • The Vapor 2000 vaporizer must be in the T
    (travel) position before it can be unlocked from
    the machine.
  • This position isolates the vaporizing chamber and
    prevents liquid from passing into the bypass
    during the time that the vaporizer is not on the
    machine.

168
INTERLOCK DEVICES
  • These are vaporizer exclusion systems which
    prevent more than one vaporizer from being turned
    on at the same time.
  • These should be checked with the anesthesia
    apparatus checkout procedure.
  • Examples
  • Select-a-tec interlock system
  • Dräger vapor exclusion system
  • Dräger Interlock II system

169
Select-a-tec interlock system(let us delete this
topic)
  • When the dial release on the Datex-Ohmeda/GE
    vaporizers is unlocked and the dial moved a pin
    moves two extension rods which project out and
    the concentration control dials of the
    neighboring vaporizers become inoperational
  • Select-a-tec interlock system.
  • When both vaporizers are off
  • When the first vaporizer is turned on

170
Dräger vapor exclusion system
  • The Dräger vaporizers have a vapor exclusion
    system which consists of notches on the back of
    the concentration control dials.
  • If a vaporizer is turned on, a pin occupies the
    notch of the adjacent vaporizers rendering the
    dial inoperational.
  • In this system, each vaporiser has two pins
    protruding out.
  • When the vaporiser is in use, the pins protrude
    outwards.
  • When the vaporiser is turned off, the pins
    retract back to where they were.
  • On the other hand, if any of the pins are pushed
    in (i.e. by another vaporiser) this locks the
    vaporiser dial in the OFF position.
  • When the pin is no longer pushed in, the dial
    once again becomes unlocked and can be turned.

OFF
ON
171
WHEN ONE IS ON, IT PUSHES THE OTHER PIN AND TURNS
IT OFF
172
Dräger Interlock II system
  • In the Dräger Interlock II system the operator
    moves a rod to release the concentration control
    dial of the vaporizer to be used. This locks the
    dial of the other vaporizer in place

173
Sequence of Vaporizers
  • Vaporizers should not be used simultaneously.
  • Old anesthesia machines had up to three
    variable-bypass vaporizers arranged in series
    such that fresh gas from the flow meters passed
    through each vapourizer to reach the common gas
    outlet of the anesthesia machine.
  • Without the use of an interlock, it was possible
    to turn on all vaporizers simultaneously.
  • This could lead to
  • Overdose of anaesthetic
  • Vapour from one vapourizer could condense in the
    downstream vapourizer leading to contamination.

174
  • Hence, the order in which they are mounted
    becomes important.
  • If the vaporizers are placed downstream of the
    common gas outlet then the use of oxygen flush
    can deliver large quantities of inhaled
    anesthetic to the patient.
  • The volatile anesthetics are added to the gas
    mixture downstream from the flow meters and
    proportioning system. Hence when using high
    concentration of low potency inhaled anesthetics
    (maximum dial setting for desflurane is 18) it
    is possible to create a hypoxic mixture when you
    give desflurane in air.

175
  • If vaporizers are to be placed in series, the
    vapourizer for the more volatile agent should be
    placed upstream (i.e. further from the patient)
    from the vapourizer for the less volatile. If
    this point is neglected, condensation of the less
    volatile agent (which will have a higher boiling
    point) will occur in a cooler downstream
    vapourizer.
  • Vaporizers with les
Write a Comment
User Comments (0)
About PowerShow.com