Title: x ray fluorescence
1X-ray fluorescence
-
By
Prateek -
Instrumentation 3rd
year
2Fluorescence
- What is Fluorescence ?
- Many solutions when irradiated with visible or
ultraviolet light, not only absorb this light ,
but re-emit light of a different wavelength. This
effect is known as fluorescence.
3Principle of fluorescence
- The ground state of molecules, molecular energies
are constant and at minimum values. - When a quanta of light impinges on a molecule,
it is absorbed and an electronic transition to a
higher electronic state take place. - In a ground state of most molecules, each
electron in the lower energy levels is paired
with another electron, whose spin is opposite to
its own spin. This state is called a singlet
state.
4Principle of Fluorescence
- When the molecules absorb energy, they are raised
to an energy level of an upper excited state s1
or s2 or s3. - These singlet transition are unstable and the
molecular energy tends to revert immediately to
a lower level. - In going to a lower energy level, some energy
will be lost responsible for visible
ultraviolet absorption spectra.
5Principle of Fluorescence
- The excited singlet state persist for a time ,
which is of order of 10 -8 to 10 -4 s. - During this time interval, some energy in excess
of the lowest vibrational energy level is rapidly
dissipated. - In case all the excess is not further dissipated
by collisions with other molecules , the electron
would return to the ground state with the
emission of energy. This phenomenon is called
fluorescence.
6(No Transcript)
7 Characteristic of fluorescent molecule
- Fluorescent molecule has two characteristic
spectra - The excitation spectrum
- The emission spectrum
- The excitation spectrum shows the relative
efficiency of different wavelengths of exciting
radiation emitted to cause fluorescence. - The shape of the excitation spectrum should be
identical to that of the absorption spectrum of
the molecule.
8 Characteristic of fluorescent molecule
- The emission spectrum would indicate the relative
intensity of radiation emitted at various
wavelengths. - The shape of emission spectrum is independent of
the wavelength of exciting radiation.
9Relationship between conc. fluorescence
intensity
- The total fluorescence intensity is equal to the
rate of light absorption multiplied by the
quantum efficiency of fluorescence(n),that is - Fn(Po -P)_________________(
1) - Po power of the beam incident upon the
solution - Ppower after traversing a length b of the
medium
10Relationship between conc. fluorescence
intensity
- Acc. To beers law
- P/Po ??
-?????? _________________(2) - ???????????? ???????????????????????? of the
fluorescent molecule - ?????????????????????????? ??
- (1-P/P0)1- ??
-?????? - (Po -P)Po (1- ??
-?????? )_______________(3) - Substituting eq(3).in eq(1)
- Fn Po (1- ??
-?????? ) - Fk . n .
P(2.3??????) - FK Cm
11Measurement of fluorescence
- Fluorescence measurement are carried out in
instruments called fluorometers or fluorimeters. - The various components of these instruments are
similar to photometers. - Fluorescence measuring instruments making use of
monochromators like spectrophotometers are called
spectrofluorimeters.
12Single-beam filter fluorimeter
- The source of light is a high pressure mercury
discharge lamp, with glass or fused silica
envelope. - Selection of the exciting wavelength is made by
inserting a primary filter in the incident beam. - Low pressure lamps, equipped with a silica
envelope can produce intense radiation at 254um.
13Single-beam filter fluorimeter
- The fluorescence emitted by the sample is
measured by selecting the wavelength of
fluorescence radiant energy by a second filter,
called secondary filter. - This is placed between the sample and a photo
detector located at s 90 ?????????? from the
incident optical path. - The fluorescent signal is generally of low
intensity require large amplification for
carrying out measurement. - The detector used in fluorescence measurements
are the sensitive photomultiplier tubes.
14Diagram of single beam fluorimeter
15Double beam filter fluorimeter
- Double-beam instruments are direct reading type
and are generally preferred because of this
several advantages. - In this instruments, the circuit is so arranged
that the current from the reference detector
opposes that generated by the measuring
photocell. - The reading on the calibrated scale, at which the
two currents neutralise each other directly gives
the fluorescence reading.
16Double beam filter fluorimeter
- A mercury lamp is used as a source of radiation.
- A collimated beam is passed through the primary
filter, which then falls on a cuvette containing
the sample. - The measuring photocell is mounted at right
angles to the excitation beam. - Fluorescent radiation passes through secondary
filters, which remove scattered ultraviolet
radiation, but pass visible light.
17Double beam filter fluorimeter
- The output of the reference the measuring
photocells is compared and a sensitive
galvanometer is used to check the null point. - The circuit is arranged, so that the current from
the reference cell opposes that generated by the
measuring photocell, which respond to fluorescent
radiation.
18Diagram of double beam fluorimeter
19