x ray fluorescence - PowerPoint PPT Presentation

About This Presentation
Title:

x ray fluorescence

Description:

principle,characteristic,measurement of fluorescence – PowerPoint PPT presentation

Number of Views:742
Slides: 20
Provided by: sushilkumar15
Tags:

less

Transcript and Presenter's Notes

Title: x ray fluorescence


1
X-ray fluorescence

  • By
    Prateek

  • Instrumentation 3rd
    year

2
Fluorescence
  • What is Fluorescence ?
  • Many solutions when irradiated with visible or
    ultraviolet light, not only absorb this light ,
    but re-emit light of a different wavelength. This
    effect is known as fluorescence.

3
Principle of fluorescence
  • The ground state of molecules, molecular energies
    are constant and at minimum values.
  • When a quanta of light impinges on a molecule,
    it is absorbed and an electronic transition to a
    higher electronic state take place.
  • In a ground state of most molecules, each
    electron in the lower energy levels is paired
    with another electron, whose spin is opposite to
    its own spin. This state is called a singlet
    state.

4
Principle of Fluorescence
  • When the molecules absorb energy, they are raised
    to an energy level of an upper excited state s1
    or s2 or s3.
  • These singlet transition are unstable and the
    molecular energy tends to revert immediately to
    a lower level.
  • In going to a lower energy level, some energy
    will be lost responsible for visible
    ultraviolet absorption spectra.

5
Principle of Fluorescence
  • The excited singlet state persist for a time ,
    which is of order of 10 -8 to 10 -4 s.
  • During this time interval, some energy in excess
    of the lowest vibrational energy level is rapidly
    dissipated.
  • In case all the excess is not further dissipated
    by collisions with other molecules , the electron
    would return to the ground state with the
    emission of energy. This phenomenon is called
    fluorescence.

6
(No Transcript)
7
Characteristic of fluorescent molecule
  • Fluorescent molecule has two characteristic
    spectra
  • The excitation spectrum
  • The emission spectrum
  • The excitation spectrum shows the relative
    efficiency of different wavelengths of exciting
    radiation emitted to cause fluorescence.
  • The shape of the excitation spectrum should be
    identical to that of the absorption spectrum of
    the molecule.

8
Characteristic of fluorescent molecule
  • The emission spectrum would indicate the relative
    intensity of radiation emitted at various
    wavelengths.
  • The shape of emission spectrum is independent of
    the wavelength of exciting radiation.

9
Relationship between conc. fluorescence
intensity
  • The total fluorescence intensity is equal to the
    rate of light absorption multiplied by the
    quantum efficiency of fluorescence(n),that is
  • Fn(Po -P)_________________(
    1)
  • Po power of the beam incident upon the
    solution
  • Ppower after traversing a length b of the
    medium

10
Relationship between conc. fluorescence
intensity
  • Acc. To beers law
  • P/Po ??
    -?????? _________________(2)
  • ???????????? ???????????????????????? of the
    fluorescent molecule
  • ?????????????????????????? ??
  • (1-P/P0)1- ??
    -??????
  • (Po -P)Po (1- ??
    -?????? )_______________(3)
  • Substituting eq(3).in eq(1)
  • Fn Po (1- ??
    -?????? )
  • Fk . n .
    P(2.3??????)
  • FK Cm

11
Measurement of fluorescence
  • Fluorescence measurement are carried out in
    instruments called fluorometers or fluorimeters.
  • The various components of these instruments are
    similar to photometers.
  • Fluorescence measuring instruments making use of
    monochromators like spectrophotometers are called
    spectrofluorimeters.

12
Single-beam filter fluorimeter
  • The source of light is a high pressure mercury
    discharge lamp, with glass or fused silica
    envelope.
  • Selection of the exciting wavelength is made by
    inserting a primary filter in the incident beam.
  • Low pressure lamps, equipped with a silica
    envelope can produce intense radiation at 254um.

13
Single-beam filter fluorimeter
  • The fluorescence emitted by the sample is
    measured by selecting the wavelength of
    fluorescence radiant energy by a second filter,
    called secondary filter.
  • This is placed between the sample and a photo
    detector located at s 90 ?????????? from the
    incident optical path.
  • The fluorescent signal is generally of low
    intensity require large amplification for
    carrying out measurement.
  • The detector used in fluorescence measurements
    are the sensitive photomultiplier tubes.

14
Diagram of single beam fluorimeter
15
Double beam filter fluorimeter
  • Double-beam instruments are direct reading type
    and are generally preferred because of this
    several advantages.
  • In this instruments, the circuit is so arranged
    that the current from the reference detector
    opposes that generated by the measuring
    photocell.
  • The reading on the calibrated scale, at which the
    two currents neutralise each other directly gives
    the fluorescence reading.

16
Double beam filter fluorimeter
  • A mercury lamp is used as a source of radiation.
  • A collimated beam is passed through the primary
    filter, which then falls on a cuvette containing
    the sample.
  • The measuring photocell is mounted at right
    angles to the excitation beam.
  • Fluorescent radiation passes through secondary
    filters, which remove scattered ultraviolet
    radiation, but pass visible light.

17
Double beam filter fluorimeter
  • The output of the reference the measuring
    photocells is compared and a sensitive
    galvanometer is used to check the null point.
  • The circuit is arranged, so that the current from
    the reference cell opposes that generated by the
    measuring photocell, which respond to fluorescent
    radiation.

18
Diagram of double beam fluorimeter
19
  • Thanks
Write a Comment
User Comments (0)
About PowerShow.com