Title: Introducing the DSQII
1- Introducing the DSQII
- Terry Jeffers
-
2Agenda
- Update on Thermos Environmental Initiatives
- Discussion on new approaches to the challenges of
8270 - Introducing Thermo Electrons Productivity Kits
- EnviroLab Forms 2.0 High Throughput
Quatification - New application note Rapid Analysis of 8270
3Thermo Electron Environmental Initiative
- Focus on Environmental Laboratory needs
- Requirement for Productivity
- Need for good work flow in Generating, Reviewing
and Reporting data - Benefit from guides for operation, maintenance
and troubleshooting specific Methodologies
4Thermo Electron Environmental Initiative
Product Sales Specialist
Environmental Demo Chemist
Environmental Seed Chemist
New hardware DSQ II
Productivity Kits
8270 and 525
8260 and 524
5Pitt Falls for Method 8270
- Those little things that stop everything
6Activity in the GC/MS system
- Activity
- These sites can be everywhere
- Injector, column, source
- SPCC compounds are used to show good system
- Nitrophenols are some of the best compounds to
find active sites - Minimizing Effects of Activity
- Deactivated liners
- Good column that shows no active sites
- High flow rates to reduce the residence time of
any active compound. - Balance temperature zones to reduce activity and
thermal break down.
7Chromatography
- How important is it really?
8GC and Autosampler
- Separations accomplished with slow chromatography
- Common GC methods for 8270
- Fast injection
- 250 - 300 0 C injector temperature
- 1 mL/min flow rate
- 30 m x 0.25 mm ID column
- Oven ramps between 5 0 and 10 0 C/min from 35 0
to 300 0 C - Separations accomplished with fast chromatography
- Alternate GC method for 8270
- Hot needle injection
- 250 0 C injector temperature
- 3 mL/min flow rate
- 30 m x 0.25 mm ID x 0.5 µm column
- Oven stepped ramps to provide the critical pair
separations.
9Hot Needle Injection Run at Lower Inlet
Temperature Fewer Replacements of Liner
10Increase the Mobile Phase 3 mL/min vs 1mL/min
- Increase capacity of stationary
- No fronting of peak shape for 160 ng/uL PNA
- Decrease the irreversible absorption of active
species - Pentachlorophenol lt 10RSD
- (5-160 ng/uL)
- 0.1 ng/uL MDL
- replicates at 1 ng/uL 4.7RSD
- Minimal breakdown for endrin or 4,4-DDT
- Provide Gaussian shape peaks for the most
volatile - No solvent skewing of NDMA and pyridine
- Shorter run times 17 minutes
11Effects of increased flow rates
- Why increase the flow rate to 3.0 mL/min
- Learned from applications developed for the
analysis of explosives. - Decrease the residence time of the compounds in
potential active areas. - Decreases the run time without higher
temperatures that could potentially damage
thermally labile compounds. - Do not need to expose the column to higher
temperatures that will damage it. - Provides better separations and narrower peak
widths. - This is balanced with a 201 split flow ratio.
12Injection Choice of Liner
13PNAs ( 5 ng and 160 ng)
14NDMA and Pyridine
Split
Splitless
15Two Methods Splitless and Split
16TRACE Ultra DSQ II GC/MS
17Source materials and geometry
- New DuraBriteTM Source Open Geometry
- Passive Collector
- Extends Dynamic Range
- Improves Robustness
- Optional Closed EI Ion Volume
- Constructed of new proprietary material for
enhanced inertness - Protects Lens 1 from ion burn
18New DynaMax XR Ion Detection System
- New Linear/Log Hybrid Electrometer
- Extends output current to 70µA
- Provides dynamic range to gt106
- No loss of response or precision at the lower
- sample/analyte levels
- SGE ETP AF635 Electron Multiplier
- Detector linear range
- output current gt30µA
- Simple snap out replacement of
- multiplier assembly
19Expected Results
- Results I can get, you can get them too.
20Calibration Curve
- First page of the calibration curve report
- All SPCC and CCC requirements met
- Average of all RSD calibration curves of 7.46
21SPCCs (5ng to 160 ng min gt0.05)
22CCCs lt 30 RSD for Initial Calibration Curve
23Quantitated Data File
- First page of the Quantitation Report
- This shows the quantitation of the 20 ng/uL
sample for the calibration curve - Meets the requirements of the internal standards
and surrogates - Meets all ion ratio requirements
- No manual integrations were required
24Robustness Split (2gr sediment/2mL methylene
chloride 25 uL diesel fuel)
25How can we prove it
26Productivity Solutions
- Integrating the DSQ II with Software, Methods,
and Consumables to Get Productive, Quickly
27Productivity Solutions
- Provide all of the necessary material to run the
method - Proven Thermo Column
- Deactivated Liners
- Standards
- Snap and shoot prediluted for entire method
validation - All other consumables for about 2 weeks
- Provide all of the software
- EnviroLab Forms 2.0
- Interactive Reference CD
- All of the methods, sequences,
- Validation data used to generate all of what you
see here and in the manuals - Two Boxes full of materials, manuals, and CDs.
28Productivity Kit
- SOP manual for method
- How to Guide
- Quick Start Guide
- Provide consumables for running method
- Standards, liners, septa, syringes, column,
- Instrument method, processing method, reporting
package - CD deliverable
- SOP
- Data from method validation
- Instrument method
- Processing method
- Reporting software with templates
29Pieces of the Productivity Kit
30Quick Start Guide
318270C/D Productivity Solution
- Customers system can be taken from installation
sign off to method IDLs in as little as 2 days! - 8270 is a value added product which will sell for
3000 US - Add on service provided by Thermo FSEs
- US Field Service Engineers are being trained on
the 8270 PS - Kits serve as a great tool for training aides for
the lab
32Productivity Solutions are a Win-Win
- Good for customers
- Improves overall experience
- Increases satisfaction
- Eases transition to Thermo GC/MS
- Gives optimal performance for the DSQ II
- Good for Thermo
- Easy to support these solutions
Helps the customer and support by establishing a
known set of criteria to achieve success. Address
the new lab environment less highly skilled users
with higher throughput requirements
33Redefining the Way We Look at GC/MS
Productivity Solutions target routine GC/MS
applications to offer value to the customer
34EnviroLab Forms v 2.0High Throughput
Quantification
35EnviroLab Forms v 2.0
- An entirely new data processing and reporting
system running as a layer over Xcalibur 1.4 - While targeted towards environmental laboratories
and applications, it has been shown to be
valuable for any laboratory performing
high-throughput quantification - Adds a significant complement of capabilities
that Xcalibur - Ion ratio calculations
- Peak integration
- Reporting
- Integrated data review and reporting
- No more batch reprocessing to re-generate reports
of edited results - Fully automated while still allowing
sophisticated manual review
36Data Review
37Report View
38EnviroLab Forms v 2.0
- Brand new workflow oriented user interface
- Substantially improved ion ratio review and
editing support - Comprehensive set of textual and graphical
reports - Fully automated, fully manual or any combination
in between - Can be used as a high-throughput quantification
engine for any application - Simply turn off or dont use those options which
arent relevant for a given user
39Conclusion
- The DSQ II system provides updates to hardware
and software that make it easy to use - The Productivity Solution for method 8270
provides a method that has been validated across
multiple DSQ II systems. - All of this allows a new operator to be
productive immediately - The proven changes to the method allow an
increased amount of productivity and throughput
for the analyst and the laboratory.
40Thank You