1 / 40
About This Presentation
Title:

Introducing the DSQII

Description:

Discussion on new approaches to the challenges of 8270. Introducing Thermo Electrons ... New Linear/Log Hybrid Electrometer. Extends output current to 70 A ... – PowerPoint PPT presentation

Number of Views:239
Avg rating:3.0/5.0
Slides: 41
Provided by: terryj9
Category:

less

Transcript and Presenter's Notes

Title: Introducing the DSQII


1
  • Introducing the DSQII
  • Terry Jeffers

2
Agenda
  • Update on Thermos Environmental Initiatives
  • Discussion on new approaches to the challenges of
    8270
  • Introducing Thermo Electrons Productivity Kits
  • EnviroLab Forms 2.0 High Throughput
    Quatification
  • New application note Rapid Analysis of 8270

3
Thermo Electron Environmental Initiative
  • Focus on Environmental Laboratory needs
  • Requirement for Productivity
  • Need for good work flow in Generating, Reviewing
    and Reporting data
  • Benefit from guides for operation, maintenance
    and troubleshooting specific Methodologies

4
Thermo Electron Environmental Initiative
Product Sales Specialist
Environmental Demo Chemist

Environmental Seed Chemist
New hardware DSQ II
Productivity Kits
8270 and 525
8260 and 524
5
Pitt Falls for Method 8270
  • Those little things that stop everything

6
Activity in the GC/MS system
  • Activity
  • These sites can be everywhere
  • Injector, column, source
  • SPCC compounds are used to show good system
  • Nitrophenols are some of the best compounds to
    find active sites
  • Minimizing Effects of Activity
  • Deactivated liners
  • Good column that shows no active sites
  • High flow rates to reduce the residence time of
    any active compound.
  • Balance temperature zones to reduce activity and
    thermal break down.

7
Chromatography
  • How important is it really?

8
GC and Autosampler
  • Separations accomplished with slow chromatography
  • Common GC methods for 8270
  • Fast injection
  • 250 - 300 0 C injector temperature
  • 1 mL/min flow rate
  • 30 m x 0.25 mm ID column
  • Oven ramps between 5 0 and 10 0 C/min from 35 0
    to 300 0 C
  • Separations accomplished with fast chromatography
  • Alternate GC method for 8270
  • Hot needle injection
  • 250 0 C injector temperature
  • 3 mL/min flow rate
  • 30 m x 0.25 mm ID x 0.5 µm column
  • Oven stepped ramps to provide the critical pair
    separations.

9
Hot Needle Injection Run at Lower Inlet
Temperature Fewer Replacements of Liner
10
Increase the Mobile Phase 3 mL/min vs 1mL/min
  • Increase capacity of stationary
  • No fronting of peak shape for 160 ng/uL PNA
  • Decrease the irreversible absorption of active
    species
  • Pentachlorophenol lt 10RSD
  • (5-160 ng/uL)
  • 0.1 ng/uL MDL
  • replicates at 1 ng/uL 4.7RSD
  • Minimal breakdown for endrin or 4,4-DDT
  • Provide Gaussian shape peaks for the most
    volatile
  • No solvent skewing of NDMA and pyridine
  • Shorter run times 17 minutes

11
Effects of increased flow rates
  • Why increase the flow rate to 3.0 mL/min
  • Learned from applications developed for the
    analysis of explosives.
  • Decrease the residence time of the compounds in
    potential active areas.
  • Decreases the run time without higher
    temperatures that could potentially damage
    thermally labile compounds.
  • Do not need to expose the column to higher
    temperatures that will damage it.
  • Provides better separations and narrower peak
    widths.
  • This is balanced with a 201 split flow ratio.

12
Injection Choice of Liner
13
PNAs ( 5 ng and 160 ng)
14
NDMA and Pyridine
Split
Splitless
15
Two Methods Splitless and Split
16
TRACE Ultra DSQ II GC/MS
17
Source materials and geometry
  • New DuraBriteTM Source Open Geometry
  • Passive Collector
  • Extends Dynamic Range
  • Improves Robustness
  • Optional Closed EI Ion Volume
  • Constructed of new proprietary material for
    enhanced inertness
  • Protects Lens 1 from ion burn

18
New DynaMax XR Ion Detection System
  • New Linear/Log Hybrid Electrometer
  • Extends output current to 70µA
  • Provides dynamic range to gt106
  • No loss of response or precision at the lower
  • sample/analyte levels
  • SGE ETP AF635 Electron Multiplier
  • Detector linear range
  • output current gt30µA
  • Simple snap out replacement of
  • multiplier assembly

19
Expected Results
  • Results I can get, you can get them too.

20
Calibration Curve
  • First page of the calibration curve report
  • All SPCC and CCC requirements met
  • Average of all RSD calibration curves of 7.46

21
SPCCs (5ng to 160 ng min gt0.05)
22
CCCs lt 30 RSD for Initial Calibration Curve
23
Quantitated Data File
  • First page of the Quantitation Report
  • This shows the quantitation of the 20 ng/uL
    sample for the calibration curve
  • Meets the requirements of the internal standards
    and surrogates
  • Meets all ion ratio requirements
  • No manual integrations were required

24
Robustness Split (2gr sediment/2mL methylene
chloride 25 uL diesel fuel)
25
How can we prove it
  • Productivity Solutions

26
Productivity Solutions
  • Integrating the DSQ II with Software, Methods,
    and Consumables to Get Productive, Quickly

27
Productivity Solutions
  • Provide all of the necessary material to run the
    method
  • Proven Thermo Column
  • Deactivated Liners
  • Standards
  • Snap and shoot prediluted for entire method
    validation
  • All other consumables for about 2 weeks
  • Provide all of the software
  • EnviroLab Forms 2.0
  • Interactive Reference CD
  • All of the methods, sequences,
  • Validation data used to generate all of what you
    see here and in the manuals
  • Two Boxes full of materials, manuals, and CDs.

28
Productivity Kit
  • SOP manual for method
  • How to Guide
  • Quick Start Guide
  • Provide consumables for running method
  • Standards, liners, septa, syringes, column,
  • Instrument method, processing method, reporting
    package
  • CD deliverable
  • SOP
  • Data from method validation
  • Instrument method
  • Processing method
  • Reporting software with templates

29
Pieces of the Productivity Kit
30
Quick Start Guide
31
8270C/D Productivity Solution
  • Customers system can be taken from installation
    sign off to method IDLs in as little as 2 days!
  • 8270 is a value added product which will sell for
    3000 US
  • Add on service provided by Thermo FSEs
  • US Field Service Engineers are being trained on
    the 8270 PS
  • Kits serve as a great tool for training aides for
    the lab

32
Productivity Solutions are a Win-Win
  • Good for customers
  • Improves overall experience
  • Increases satisfaction
  • Eases transition to Thermo GC/MS
  • Gives optimal performance for the DSQ II
  • Good for Thermo
  • Easy to support these solutions

Helps the customer and support by establishing a
known set of criteria to achieve success. Address
the new lab environment less highly skilled users
with higher throughput requirements
33
Redefining the Way We Look at GC/MS
Productivity Solutions target routine GC/MS
applications to offer value to the customer
34
EnviroLab Forms v 2.0High Throughput
Quantification
35
EnviroLab Forms v 2.0
  • An entirely new data processing and reporting
    system running as a layer over Xcalibur 1.4
  • While targeted towards environmental laboratories
    and applications, it has been shown to be
    valuable for any laboratory performing
    high-throughput quantification
  • Adds a significant complement of capabilities
    that Xcalibur
  • Ion ratio calculations
  • Peak integration
  • Reporting
  • Integrated data review and reporting
  • No more batch reprocessing to re-generate reports
    of edited results
  • Fully automated while still allowing
    sophisticated manual review

36
Data Review
37
Report View
38
EnviroLab Forms v 2.0
  • Brand new workflow oriented user interface
  • Substantially improved ion ratio review and
    editing support
  • Comprehensive set of textual and graphical
    reports
  • Fully automated, fully manual or any combination
    in between
  • Can be used as a high-throughput quantification
    engine for any application
  • Simply turn off or dont use those options which
    arent relevant for a given user

39
Conclusion
  • The DSQ II system provides updates to hardware
    and software that make it easy to use
  • The Productivity Solution for method 8270
    provides a method that has been validated across
    multiple DSQ II systems.
  • All of this allows a new operator to be
    productive immediately
  • The proven changes to the method allow an
    increased amount of productivity and throughput
    for the analyst and the laboratory.

40
  • Questions?

Thank You