TEST SETUP FOR HF PMT AT IOWA LAB - PowerPoint PPT Presentation

1 / 26
About This Presentation
Title:

TEST SETUP FOR HF PMT AT IOWA LAB

Description:

Storage Cupboards. Electronics. Racks. Scope. UI CMS HF Test Lab. 6 m. 5 m. The hardware system for the quality control of HF PMT at delivery will be ... – PowerPoint PPT presentation

Number of Views:59
Avg rating:3.0/5.0
Slides: 27
Provided by: omegaPhys
Category:
Tags: for | iowa | lab | pmt | setup | test | cupboards

less

Transcript and Presenter's Notes

Title: TEST SETUP FOR HF PMT AT IOWA LAB


1
TEST SETUP FOR HF PMT AT IOWA LAB
  • CMS HCAL Meeting
  • FNAL, November 16-18
  • Y. Onel
  • U. of Iowa
  • W. Anderson
  • Iowa State U.
  • D. Winn
  • Fairfield U.

2
 Hardware and Software Tools for Test and
Calibration of HF PMT 
  • The hardware system for the quality control of HF
    PMT at delivery will be installed at UI CMS lab.
  • Test Station Components
  •  
  • 1 - Light sources LASER, LED, Radioactive
    sources Scintillators
  •  
  • 2 - Light-tight Test Box
  •  
  • 3 - Readout Electronics
  •  
  • 4 - Data Acquisition, Analysis and Storage PC -
    controlled System
  •  

UI CMS HF Test Lab
Electronics
Light-tight
Laser Bench
Racks
Test Box
Scope
5 m
PC
Storage Cupboards
6 m
3
Tasks of the Test System
  • HF PMT Quality Control, Calibration and
    Monitoring System should address the following
    items
  • label and catalogue each PMT at delivery and
    storage (StorageIN)
  • mechanical assembly with HV power supply and
    base
  •  installation in Test Box individually or in
    groups (6 - 9)
  • measure (a) - gain vs HV G-HV
  • (b) - pulse shape PS
  • (c) - single photoelectron response SPR
  • (d) - noise and dark current N/DC
  • (e) - photocathode uniformity PCU
  • (f) - linearity of response LR
  • (g) - pulse rate dependence PRD
  • (h) - quantum efficiency vs wave-length
    QE-WL
  • (i) - timing response and intrinsic delay
    TR-D.

4
Procedures for measurements
  • The following sequence of measurements will be
    performed for each PMT
  • 1 - PMT's installed in Test-Box are let to
    stabilize at standard HV
  • 2 - Check of normal operating conditions
    source scint.
  • 3 - Noise and dark current measurements vs.
    HV
  • 4 - Gain vs. HV laser
  • 5 - Single photoelectron level
  • 6 - Linearity for 1- 2000 p.e.
  • 7 - Rate dependence for 0.1 - 40 Mz LED
  • 8 - Photocathode uniformity
  • 9 - Quantum efficiency (300-600 nm) dye
    laser
  • 10 - Pulse shape measurements at nominal HV.
  • According to specifications of the PMT
    (manufacturer's data sheet and preliminary
    measurements on a test sample) and requirements
    of HF application (Nphe/GeV, dynamic range, etc.)
    the test setup working conditions will be
    adjusted in a range of light yield and
    sensitivity appropriate for the standard test
    procedure. Three light sources will be used for
    the specific measurements
  • - laser
  • - LED
  • - Rad. sources radiator

5
(a) Gain vs HV G-HV
  • The signal charge QA at the PMT anode for a pulse
    of Nph photons of w. l. l hitting the
    photocathode of quantum efficiency h(l) is
  • QA G(V) h(l) Nph qe
  • where G(V) is the PMT gain, h(l) Nph Nphe is
    the number of photoelectrons (charge qe) at the
    cathode. For fixed illumination (ltNphegt
    const.), the relation ltQAgt A Vk determines k.
  • At any voltage V, and fixed illumination (ltNphegt
    const.), A is related to ltQAgt/ltNphegt qe
    DQA2/ltQAgt, where ltNphe gt is estimated as
  • ltNphe gt (1d) (ltQAgt/DQA ) 2
  • d 0 for pure Poisson photoelectron distribution,
    d 0.25 taking into account first dynode
    collection efficiency and SER DQA is the RMS
    spread (0.85 HWHM) of the anode charge QA
    distribution (see SER measurement).
  • Measurements of gain vs HV can be performed with
    laser, LED, or source radiator. The HV supply
    settings are stepped through the useful HV range
    for the PMT via computer control and pulse
    amplitude distribution measured with ADC.

6
(a) Gain vs. HV G-HV
7
(b) Pulse Shape PS
  • The pulse shape (characterized by rise-time TR
    and FWHM of anode signals) is measured on a
    digital scope with PMT illuminated with a
    radiator and b-source. The digitized pulse shape
    is recorded.

8
(c) Single Photoelectron Response SPR
  • The light yield is set to a low level, in such a
    way that there is no count in the majority of
    cases for instance, with 80 missing counts,
    according to Poisson statistics, approximately
    90 of the residual counts in these conditions,
    should correspond to single phe.

9
(d) Noise and Dark Current N/DC
  • Dark current is measured at various HV settings
    with a picoamperometer, recorded with the PMT in
    stable temperature conditions and after a period
    ( 1/2 hour) in dark. Noise spectrum is obtained
    in same situations by measuring ADC pedestal
    spectrum, and subtracting the electronic noise
    contribution.

10
(e) Photocathode Uniformity PCU
  • The uniformity of response for the photocathode
    is measured by having an illuminating spot (typ.
    1mm diam.) spanning the photocathode surface (via
    a fiber moved with a scanning table) and
    recording pulse amplitude at each position.

11
(f) Linearity of Response LR
  • The light yield on the photocathode is varied
    over a wide range and PMT response measured as a
    function of light intensity various methods can
    be used to provide different light levels
    calibrated filters, diaphragms, etc. the
    variation of light yield is monitored with linear
    (photodiode) devices the effective steps in
    illumination can be checked and adjusted at low
    light levels, using wide dynamic range ADC (or
    proper attenuation/amplification).

12
(g) Pulse Rate Dependence PRD
  • The pulse rate dependence is measured with (a
    group of) LED (of equalized yield) mixed in a
    light manifold over a certain range of pulse
    rates, the frequency of LED can be varied
    individually without change of the light pulse
    amplitude or duration for higher frequencies few
    LED, feeding a light mixer, can be pulsed
    together, with short relative delays, thus
    producing a high frequency (defined by the delay)
    pulse train. Alternatively one LED can be pulsed
    for long times with an average light yield
    reproducing the amount of photons of a high
    frequency burst, and short pulses from another
    LED are analyzed as a function of the average
    current load produced by the first LED.

13
(h) Quantum Efficiency vs. Wavelength QE-WL
  • The response of PMT is measured with definite WL
    of light (of known intensity).
  •  
  •  

14
(i) Timing Response and Intrinsic Delay TR-D
  • The PMT is illuminated with radiator/scintillator
    excited with rad. source the time response of
    the PMT and its transit time are measured with
    respect to a calibrated counter viewing the same
    light source.

15
  • Measurements will be performed at stable
    (controlled) temperature. For each delivery
    batch, sample measurements will be performed for
    temperature and magnetic field dependence.
  • The data for each PMT will be stored in
    appropriate archive files on disk and copied to
    permanent storage media. For each PMT an entry
    will be printed and logged to a general PMT
    directory and test logbooks.

16
  • The PMT's, conforming to acceptance criteria,
    will be sorted in classes and stored (Storage
    OUT), ready for installation. Those not
    conforming will be returned to the manufacturer.
  • All measurement procedures will be automated and
    computer-controlled, to minimize individual
    biases and interventions daily test shifts will
    be supervised by an expert, who will also review
    the archived data of the day and certify their
    validity.

17
Electronics
  • The electronic system for the Test Setup
    includes
  • 1 - Pulsers for the Laser, LED and calibration
    pulses
  • 2 - Digital control circuits for
  • - HV and LV power supplies,
  • - Laser parameters and settings,
  • - Stepping motors for moving parts in
    Test Box,
  • - Probes for temperature, etc., in Test Box
  • - Safety circuits on door, sliding
    mechanics, etc.
  • 3 - Trigger logic for DAQ, monitors and gates
  • 4- ADC and TDC for amplitude and timing
    measurements
  • 5- Digital Oscilloscope for measuring and
    recording pulse shapes

Electronics block diagram
PMT's, Monitors, etc.
PMT's, Monitors, etc.
18
Test Box
  • The Test Box assembly consists of  
  • 1 - PMT holder support with position remote
    control
  • 2 - Optical Manifold (O. M.) providing fiber
    coupling to individual PMT's and receiving light
    from
  • a - laser system with variable wavelength
    (dye),
  • b - LED system pulsed at high
    frequencies,
  • c - Radioactive sources scintillators
  • 3 - Monitoring and control devices.

19
  • The Test Box is a light-tight and
    temperature-stabilized chamber 60 (H) x 60 (W) x
    120 (L) cm3 containing precision
    remotely-controlled slides, supporting the PMT's
    holder and the assembly for O. M. and fibers
    coupling to the PMT's it provides a number of
    sealed cable inlets and outlets for PMT signals,
    HV and LV and control signals as well as for
    fibers from the Laser system.

20
DAQ System and Measurement Strategy
  • The following sequence of measurements will be
    performed for each PMT
  • 1 - PMT's installed in Test-Box are let to
    stabilize at standard HV
  • 2 - Check of normal operating conditions source
    scint.
  • 3 - Noise and dark current measurements vs. HV
  • 4 - Gain vs. HV laser
  • 5 - Single photoelectron level
  • 6 - Linearity for 1- 2000 p.e.
  • 7 - Rate dependence for 0.1 - 40 MHz LED
  • 8 - Photocathode uniformity
  • 9 - Quantum efficiency (300-600 nm) dye laser,
    source scint.
  • 10 - Pulse shape measurements at nominal HV.

21
Quality Assurance
  • At the manufacturer
  • testing/preselection as they arrive
  • beam/calibration tests during the installation
    period
  • we will eventually have two test stations for
    testing
  • PMT can be replaced
  • any PMT will function within 20 of any other PMT

22
Data Acquiaition System (DAQ)
  • We have VAx3200 (with backplane) which is a
    functional DAQ system with CAMAC
  • We have also purchased the interface cards and
    software for the mobile DAQ which is LabView
    based
  • We have U. of Iowa matching funds to purchase the
    components of the fixed PMT test station

23
Manpower and expert team to install the test
system
  • UI
  • U. Akgun
  • A. Ayan
  • P. Bruecken
  • J.P. Merlo
  • M. Miller
  • Y. Onel
  • I. Schmidt
  • A. Tauke
  • R. Vogel
  • post-doc (to be named)
  • ISU
  • W. Anderson
  • Fairfield U
  • P. Winn
  • International Team
  • E. Gulmez Turkey
  • A. Penzo Italy
  • M. Zeyrek Turkey

24
Test Team
  • Post-doc (supervisor)
  • M. Acar (graduate student)
  • P. Bruecken (Quarknet lead teacher)
  • A. Kocbay (graduate student)
  • Bettendorf H.S. seniors (2)
  • UI undergraduate (2)

25
Tests required of the vendor on each tube
  • Determine the cathode luminous sensitivity with a
    tungsten filament lamp with a color temperature
    of 2856K.
  • Determine the cathode blue sensitivity with a
    2856K tungsten light filtered through a Corning
    CS5-58 filter ground to half stock thickness.
  • 2a. Alternatively the vendor may measure and
    report the quantum efficiency of the tube at
    420nm.
  • Determine the voltage at which a current gain of
    5x105 is reached.
  • Measure the dark current at a current gain of
    5x105.

26
Additional tests
  • Our institution/university is interested in
    having additional selection tests performed at
    the vendors site. The vendor is invited to
    submit a proposal for PMTs with the following
    additional measurement made at the vendors site.
  • The vendor shall determine the pulse height
    resolution at a gain of 5x104 using the following
    method or some other method agreed on between the
    University and vendor.
  • The PMT pulse height resolution shall deviate
    less then 30 from the ideal resolution (defined
    as sigma/mean equal to 1/sqrt(t) where Npe is the
    average number of photoelectrons produced by the
    photocathode for a given light pulse intensity)
    at a current gain of 5x104. The vendor and
    University shall agree on an appropriate test to
    determine that this resolution specification is
    met.
Write a Comment
User Comments (0)
About PowerShow.com