Title: TEST SETUP FOR HF PMT AT IOWA LAB
1TEST SETUP FOR HF PMT AT IOWA LAB
- CMS HCAL Meeting
- FNAL, November 16-18
- Y. Onel
- U. of Iowa
- W. Anderson
- Iowa State U.
- D. Winn
- Fairfield U.
2 Hardware and Software Tools for Test and
Calibration of HF PMT
- The hardware system for the quality control of HF
PMT at delivery will be installed at UI CMS lab. - Test Station Components
-
- 1 - Light sources LASER, LED, Radioactive
sources Scintillators -
- 2 - Light-tight Test Box
-
- 3 - Readout Electronics
-
- 4 - Data Acquisition, Analysis and Storage PC -
controlled System -
UI CMS HF Test Lab
Electronics
Light-tight
Laser Bench
Racks
Test Box
Scope
5 m
PC
Storage Cupboards
6 m
3Tasks of the Test System
- HF PMT Quality Control, Calibration and
Monitoring System should address the following
items - label and catalogue each PMT at delivery and
storage (StorageIN) - mechanical assembly with HV power supply and
base - installation in Test Box individually or in
groups (6 - 9) - measure (a) - gain vs HV G-HV
- (b) - pulse shape PS
- (c) - single photoelectron response SPR
- (d) - noise and dark current N/DC
- (e) - photocathode uniformity PCU
- (f) - linearity of response LR
- (g) - pulse rate dependence PRD
- (h) - quantum efficiency vs wave-length
QE-WL - (i) - timing response and intrinsic delay
TR-D.
4Procedures for measurements
- The following sequence of measurements will be
performed for each PMT - 1 - PMT's installed in Test-Box are let to
stabilize at standard HV - 2 - Check of normal operating conditions
source scint. - 3 - Noise and dark current measurements vs.
HV - 4 - Gain vs. HV laser
- 5 - Single photoelectron level
- 6 - Linearity for 1- 2000 p.e.
- 7 - Rate dependence for 0.1 - 40 Mz LED
- 8 - Photocathode uniformity
- 9 - Quantum efficiency (300-600 nm) dye
laser - 10 - Pulse shape measurements at nominal HV.
- According to specifications of the PMT
(manufacturer's data sheet and preliminary
measurements on a test sample) and requirements
of HF application (Nphe/GeV, dynamic range, etc.)
the test setup working conditions will be
adjusted in a range of light yield and
sensitivity appropriate for the standard test
procedure. Three light sources will be used for
the specific measurements - - laser
- - LED
- - Rad. sources radiator
5(a) Gain vs HV G-HV
- The signal charge QA at the PMT anode for a pulse
of Nph photons of w. l. l hitting the
photocathode of quantum efficiency h(l) is - QA G(V) h(l) Nph qe
- where G(V) is the PMT gain, h(l) Nph Nphe is
the number of photoelectrons (charge qe) at the
cathode. For fixed illumination (ltNphegt
const.), the relation ltQAgt A Vk determines k. - At any voltage V, and fixed illumination (ltNphegt
const.), A is related to ltQAgt/ltNphegt qe
DQA2/ltQAgt, where ltNphe gt is estimated as - ltNphe gt (1d) (ltQAgt/DQA ) 2
- d 0 for pure Poisson photoelectron distribution,
d 0.25 taking into account first dynode
collection efficiency and SER DQA is the RMS
spread (0.85 HWHM) of the anode charge QA
distribution (see SER measurement). - Measurements of gain vs HV can be performed with
laser, LED, or source radiator. The HV supply
settings are stepped through the useful HV range
for the PMT via computer control and pulse
amplitude distribution measured with ADC. -
6(a) Gain vs. HV G-HV
7(b) Pulse Shape PS
- The pulse shape (characterized by rise-time TR
and FWHM of anode signals) is measured on a
digital scope with PMT illuminated with a
radiator and b-source. The digitized pulse shape
is recorded.
8(c) Single Photoelectron Response SPR
- The light yield is set to a low level, in such a
way that there is no count in the majority of
cases for instance, with 80 missing counts,
according to Poisson statistics, approximately
90 of the residual counts in these conditions,
should correspond to single phe.
9(d) Noise and Dark Current N/DC
- Dark current is measured at various HV settings
with a picoamperometer, recorded with the PMT in
stable temperature conditions and after a period
( 1/2 hour) in dark. Noise spectrum is obtained
in same situations by measuring ADC pedestal
spectrum, and subtracting the electronic noise
contribution.
10(e) Photocathode Uniformity PCU
- The uniformity of response for the photocathode
is measured by having an illuminating spot (typ.
1mm diam.) spanning the photocathode surface (via
a fiber moved with a scanning table) and
recording pulse amplitude at each position.
11(f) Linearity of Response LR
- The light yield on the photocathode is varied
over a wide range and PMT response measured as a
function of light intensity various methods can
be used to provide different light levels
calibrated filters, diaphragms, etc. the
variation of light yield is monitored with linear
(photodiode) devices the effective steps in
illumination can be checked and adjusted at low
light levels, using wide dynamic range ADC (or
proper attenuation/amplification).
12(g) Pulse Rate Dependence PRD
- The pulse rate dependence is measured with (a
group of) LED (of equalized yield) mixed in a
light manifold over a certain range of pulse
rates, the frequency of LED can be varied
individually without change of the light pulse
amplitude or duration for higher frequencies few
LED, feeding a light mixer, can be pulsed
together, with short relative delays, thus
producing a high frequency (defined by the delay)
pulse train. Alternatively one LED can be pulsed
for long times with an average light yield
reproducing the amount of photons of a high
frequency burst, and short pulses from another
LED are analyzed as a function of the average
current load produced by the first LED.
13(h) Quantum Efficiency vs. Wavelength QE-WL
- The response of PMT is measured with definite WL
of light (of known intensity). -
-
14(i) Timing Response and Intrinsic Delay TR-D
- The PMT is illuminated with radiator/scintillator
excited with rad. source the time response of
the PMT and its transit time are measured with
respect to a calibrated counter viewing the same
light source.
15- Measurements will be performed at stable
(controlled) temperature. For each delivery
batch, sample measurements will be performed for
temperature and magnetic field dependence. - The data for each PMT will be stored in
appropriate archive files on disk and copied to
permanent storage media. For each PMT an entry
will be printed and logged to a general PMT
directory and test logbooks.
16- The PMT's, conforming to acceptance criteria,
will be sorted in classes and stored (Storage
OUT), ready for installation. Those not
conforming will be returned to the manufacturer. - All measurement procedures will be automated and
computer-controlled, to minimize individual
biases and interventions daily test shifts will
be supervised by an expert, who will also review
the archived data of the day and certify their
validity.
17Electronics
- The electronic system for the Test Setup
includes - 1 - Pulsers for the Laser, LED and calibration
pulses - 2 - Digital control circuits for
- - HV and LV power supplies,
- - Laser parameters and settings,
- - Stepping motors for moving parts in
Test Box, - - Probes for temperature, etc., in Test Box
- - Safety circuits on door, sliding
mechanics, etc. - 3 - Trigger logic for DAQ, monitors and gates
- 4- ADC and TDC for amplitude and timing
measurements - 5- Digital Oscilloscope for measuring and
recording pulse shapes
Electronics block diagram
PMT's, Monitors, etc.
PMT's, Monitors, etc.
18Test Box
- The Test Box assembly consists of
- 1 - PMT holder support with position remote
control - 2 - Optical Manifold (O. M.) providing fiber
coupling to individual PMT's and receiving light
from - a - laser system with variable wavelength
(dye), - b - LED system pulsed at high
frequencies, - c - Radioactive sources scintillators
- 3 - Monitoring and control devices.
19- The Test Box is a light-tight and
temperature-stabilized chamber 60 (H) x 60 (W) x
120 (L) cm3 containing precision
remotely-controlled slides, supporting the PMT's
holder and the assembly for O. M. and fibers
coupling to the PMT's it provides a number of
sealed cable inlets and outlets for PMT signals,
HV and LV and control signals as well as for
fibers from the Laser system.
20DAQ System and Measurement Strategy
- The following sequence of measurements will be
performed for each PMT - 1 - PMT's installed in Test-Box are let to
stabilize at standard HV - 2 - Check of normal operating conditions source
scint. - 3 - Noise and dark current measurements vs. HV
- 4 - Gain vs. HV laser
- 5 - Single photoelectron level
- 6 - Linearity for 1- 2000 p.e.
- 7 - Rate dependence for 0.1 - 40 MHz LED
- 8 - Photocathode uniformity
- 9 - Quantum efficiency (300-600 nm) dye laser,
source scint. - 10 - Pulse shape measurements at nominal HV.
21Quality Assurance
- At the manufacturer
- testing/preselection as they arrive
- beam/calibration tests during the installation
period - we will eventually have two test stations for
testing - PMT can be replaced
- any PMT will function within 20 of any other PMT
22Data Acquiaition System (DAQ)
- We have VAx3200 (with backplane) which is a
functional DAQ system with CAMAC - We have also purchased the interface cards and
software for the mobile DAQ which is LabView
based - We have U. of Iowa matching funds to purchase the
components of the fixed PMT test station
23Manpower and expert team to install the test
system
- UI
- U. Akgun
- A. Ayan
- P. Bruecken
- J.P. Merlo
- M. Miller
- Y. Onel
- I. Schmidt
- A. Tauke
- R. Vogel
- post-doc (to be named)
- ISU
- W. Anderson
- Fairfield U
- P. Winn
- International Team
- E. Gulmez Turkey
- A. Penzo Italy
- M. Zeyrek Turkey
24Test Team
- Post-doc (supervisor)
- M. Acar (graduate student)
- P. Bruecken (Quarknet lead teacher)
- A. Kocbay (graduate student)
- Bettendorf H.S. seniors (2)
- UI undergraduate (2)
25Tests required of the vendor on each tube
- Determine the cathode luminous sensitivity with a
tungsten filament lamp with a color temperature
of 2856K. - Determine the cathode blue sensitivity with a
2856K tungsten light filtered through a Corning
CS5-58 filter ground to half stock thickness. - 2a. Alternatively the vendor may measure and
report the quantum efficiency of the tube at
420nm. - Determine the voltage at which a current gain of
5x105 is reached. - Measure the dark current at a current gain of
5x105.
26Additional tests
- Our institution/university is interested in
having additional selection tests performed at
the vendors site. The vendor is invited to
submit a proposal for PMTs with the following
additional measurement made at the vendors site. - The vendor shall determine the pulse height
resolution at a gain of 5x104 using the following
method or some other method agreed on between the
University and vendor. - The PMT pulse height resolution shall deviate
less then 30 from the ideal resolution (defined
as sigma/mean equal to 1/sqrt(t) where Npe is the
average number of photoelectrons produced by the
photocathode for a given light pulse intensity)
at a current gain of 5x104. The vendor and
University shall agree on an appropriate test to
determine that this resolution specification is
met.