Peter J Baugh - PowerPoint PPT Presentation

1 / 46
About This Presentation
Title:

Peter J Baugh

Description:

... to E, B (or V) or linked E/B or B2/E , etc. The accelerating voltage is 2 8 kV. ... to produce molecular ion signals than FAB or SIMS in the RMM range 10, ... – PowerPoint PPT presentation

Number of Views:504
Avg rating:3.0/5.0
Slides: 47
Provided by: mrba2
Category:
Tags: baugh | peter

less

Transcript and Presenter's Notes

Title: Peter J Baugh


1
Analytical Mass Spectrometry
  • Peter J Baugh
  • Quay Pharma Training Progamme
  • Day 4, September 28, 2004

2
Analytical Mass Spectrometry
  • Mass spectrometry is an extremely powerful
    analytical instrumental tool in its own right but
    also, and importantly, when coupled to a range of
    introductory and/or chromatographic systems, such
    as GC (SFC) and LC. It is used by chemists,
    biochemists, biologists and physicists in many
    fields analytical/environmental, molecular
    biological, toxicological, clinical, medical and
    physical
  • MS has been at the forefront of the recent high
    powered research to elucidate/characterise the
    Human Genome and without its capability in
    analysis no advances would have been made in this
    field.
  • This training session is designed to introduce MS
    as a stand alone analytical tool and when
    interfaced to chromatographic systems. The
    following areas will be covered
  • Theory of MS ion production, processes and
    mechanisms
  • Instrumental configurations ionisers,
    analysers, detectors, data systems, scan modes
    and functions
  • Introductory and interfaced systems
  • Applications in selected areas

3
Introduction to MS
  • Background Aston in 1920s designed a mass
    spectrograph using a magnet to mass analyse and
    photographic plate to detect positive ions
    produced by ionisation of gases and vapours.
    Using this mass spectrograph he identified in
    excess of 200 isotopes of a wide range of
    chemical elements and received the Nobel prize
    for this work. Today we use his foundation work
    as a basis for the field of mass spectrometry (a
    term now used more frequently rather than the
    older term of mass spectroscopy the latter
    inferring that a spectroscopic detector is
    utilised).
  • Nowadays, the analysers are more wide ranging in
    type and in combination as follows
  • Magnetic (B) and electrostatic (E) analysers in
    combination, EB or reverse geometry BE
  • Quadrupole (Q) and ion trap (IT) electric
    fields only
  • Time of flight (TOF)
  • Ion cyclotron resonance (Fourier transform MS)
  • In addition both hybrid and tandem analyser
    systems have been designed to conduct analysis
    using more complex scan modes, EBqQ, QTOF,
    QqQ, EBBE

4
  • Theory of Mass Spectrometry
  • Much work has been carried out on the theory of
    ion production and ion processes. Until the
    1970s only positive ion mass analysis could be
    conducted because the ionisation modes , such as,
    EI, FD (FI) produced only positive ions in
    sufficient abundance from solids bombarded by
    electrons emitted from a hot filament, or by
    desorption - electric field application. Sample
    introduction was via a solids probe. Strictly
    speaking electrons generated could be captured by
    molecules in the ionisation source but with
    almost zero efficiency because of the large mean
    free path high vacuum, very low pressure
    atmosphere. In fact, the efficiency of
    ionisation to generate positive ions is only a
    few .
  • A positive ion, M., are generated from a
    molecule of a chemical solid, M, in Electron
    Impact ionisation with highest efficiency by
    employing electrons with Ee 70 eV as follows
  • M e- ? M. 2e-
  • If the radical cation formed is unstable which is
    often the case because it has too much kinetic
    energy imparted to it during the ionisation
    process it can break up in a number ways to
    produce secondary ions and neutral radicals or
    ions and molecular products.
  • M. ? A B. or A. B or C H2
  • The secondary ions, A , B, or C, are
    accelerated from the ion source using an applied
    voltage Vacc (high or low depending on the mode
    of MS) , mass analysed and detected. Some ions
    can also break down in flight metastables, and
    such ions can be analysed also by linked scanning
    (EB or BE instruments only).

5
  • Theory of Mass Spectrometry
  • Stabilisation and fragmentation of positive ions
    produced by EI
  • Mechanisms of positive ion breakdown are numerous
    and vary depending on the stability of the ion
    and the nature of the molecule. The molecular
    ion produced by EI is commonly unstable and the
    breakdown or fragmentation leads to more stable
    ions which vary in abundance and the distinctive
    pattern characteristic of the molecule is termed
    the mass spectrum (relative abundance versus m/z
    ratio). The most stable ion is termed the base
    ion or the base peak in the mass spectrum (100
    relative abundance). The structure of a molecule
    can often be deduced from the mass spectral
    fragmentation pattern and molecules with similar
    structures show similar mass spectra, however,
    knowledge of the mass (m/z) of the molecular ion
    is required for confirmation. An homologous
    series of compounds exhibit similar mass spectra
    and the same base ion (100).
  • For example
  • N-alkanes. m/z, 57, C4H9 , the butyl ion
    N-alkenes, m/z , 55, C4H7, butylene ion
  • fatty acid methyl esters, m/z, 74, a
    stabilised ion via a McClafferty rearrangement,
  • CH3O C CH2
  • OH
  • Aromatic compounds exhibit molecular ions or
    stabilised fragment ions with the aromatic
    structure intact
  • Other ionisation modes
  • FD generates, preferentially molecular ions or
    stabilised ions by elimination of water.
  • Pseudo molecular positive and negative ions and
    adduct ions are form by other ionisation
    processes CI, FAB, TSP and ESI, APCI, MALDI
    (Laser desorption/ionisation), PD (plasma
    desorption 252Californium)

6
  • Mass Spectrometric instrumentation
  • Essentially, a mass spectrometer comprises an ion
    source, mass analyser(s) and a detector in a
    high vacuum system, almost always nowadays
    electronically-interfaced to a data system (DS).
  • Ion source (see schematic diagram 1)
  • Sample introduction for EI in its simplest form
    is via a solids probe heated to several hundred
    0C. This type of introduction is limited to
    organic solids of up to about 600 u RMM.
    Oligomers, Peptides and proteins of high RMM
    undergo excessive fragmentation to yield low RMM
    fragment ions. Excessive decomposition occurs in
    organic solids that decompose on heating, i.e.,
    have no m.pt. The ions are generated in the ion
    source by bombardment with electrons emitted from
    a heated filament e.g., rhenium wire, to which a
    voltage is applied - 150 V. The electron energy,
    Ee, can be varied (10 120 eV) and the molecular
    ion stability is inversely proportional to Ee .
    In an EI source the electrons are collected (go
    to earth). A high accelerating voltage is
    employed for EI in EB or BE MS and a low voltage
    is used for quadrupole MS. In the CI source the
    electrons are not collected.
  • Mass analysers
  • Magnetic sector/electric sector analysers MS is
    fundamentally based on magnetic sector analysers,
    such instruments are single focussing (SF) with
    respect to mass through use of a magnet only.
    Double focusing (DF) sector instruments either
    focus ions first through the electric sector
    (ESA), energy focusing and then the magnet (B),
    mass or momentum focussing, or vice versa in
    reversed geometry, i.e., EB or BE. The DF
    systems are capable of high mass resolution
    (0.001 u or 1 ppm), scanning metastables and
    linked scan functions (based on E, B and possibly
    Vacc together).

7
Diagram 1. EI Ion Source
8
  • Mass analysers continued
  • Although the equation describing the relationship
    between m/z (m/e), B2, Vacc , r, the field radius
    is based on the SF magnetic sector instrument
    that for the DF sector MS is more appropriate
    which describes the relationship between m/z, B2,
    the electric sector voltage, E, the distance
    between the ESA plates, d, the field radius, r ,
    and the radius of curvature of the ESA, R.
  • m/z B2 r2 d/ ER .
  • It can be seen that m/z is independent of Vacc .
    Scanning is conducted with respect to E, B (or V)
    or linked E/B or B2/E , etc. The accelerating
    voltage is 2 8 kV. The magnetic field flux is
    normally scanned, exponentially downwards, where
    m mo. exp(-2bt) with scan rates being described
    in terms of seconds per decade of mass. Until
    the 1980s the magnet scanning was hampered by
    long reset times imposed by hysteresis of the
    magnet the magnetic cycle involves scan-up,
    scan down and reset phases. Thus the scanning of
    a quadrupole was much faster at that stage in
    time. The introduction of low hysteresis magnets
    (laminated) with rapid reset times allowed
    improved scan speeds (lt 0.1 sec per decade)
    compatible with quadrupole systems.
  • See schematic diagram -2 and 3 of DF magnetic
    sector instrument

9
Diagram 2B ESA of a Double Focusing MS
Diagram 2C Normal Geometry DF MS energy and
momentum focusing to increase mass resolution
10
Diagram 3 Normal Geometry DF MS showing ion
optics
11
  • Mass analysers continued
  • Quadrupole mass analyser The basic quadrupole
    mass analyser is an electric sector device
    depending only on generation of electric fields.
    The analyser comprises circular (hyperbolic) four
    rods making up the quadrupole in which
    diametrically opposed rods (x and y directions)
    are electrically coupled to a combination of dc
    and rf amplitude voltages, U and V (see schematic
    diagram - 4).
  • To operate the analyser to enable mass analysis a
    combination of U and V must be applied to the x
    and y rods such that the ratio of U/V is
    constant. Varying the voltages over a set range
    allows ions of a selected m/z to undergo stable
    z-direction oscillations in the electric field
    generated at the appropriate U/V (the quadrupole
    is termed a mass filter for this reason). The
    Mathieu equations describing the trajectory in
    the combined field in x and y directions are not
    discussed here. However, it is useful to note
    that the application of U and V voltages is
    governed as follows
  • x-axis rods U Vcos(wt) y-axis rods -U
    Vcos(wt)
  • The coswt relationship implies that the rf
    amplitude voltage,V, is 180o out of phase with
    the dc voltage.The stability diagram (triangle)
    describing the region in which ions of a selected
    m/z is stable is obtained from solutions of the
    Mathieu equations (terms a and q) and provides
    information about the conditions under which mass
    analysis can be conducted. Essentially, only
    ions of a selected m/z can be analysed at a
    particular U/V and mass selection is maximised
    when ions of different m/z attain the same
    velocity (imposed by varying the accelerating
    voltage up to 20 V applied to the ions exiting
    from the ion source, coincidently with U/V) (see
    schematic diagrams 5 and 6).
  • Transmission of ions is lost towards the apex of
    the triangle where the resolution (mass
    selection) is greatest and a quadrupole mass
    analyser is essentially capable only of low mass
    resolution (unit mass accuracy) where as a DF
    magnetic/electrostatic sector analyser system is
    capable of low and high mass resolution.)

12
Diagram 4 Quadrupole MS showing a. X () and Y
(-) rods coupled to dc and rf voltages forming
the mass analyser and b. ion trajectory
13
  • Additional Notes
  • Parameters, a and q, for the Stability Diagram
  • Quadrupole and Ion Trap Analysers
  • The primary concern in the utilisation of a
    quadrupole device are the criteria that govern
    the stability of an ion trajectory within the
    field, i.e., the experimental conditions that
    determine whether the ion remains within the
    device or is lost to the environment. The
    trajectory of a charged particle under the
    influence of a two dimensional field is described
    as an oscillating system in which the restoring
    force is periodic in time. The system can be
    described by two parameters, a and q,
    formularised in terms of dc (U), rf (Vamp), ion
    mass (m), radiofrequency (?), charge (e) and the
    distance between the rods (ro)
  • a 8neU/m (ro ?)2 and q 4neV/m (ro ?)2
  • Being periodic in time, a and q could have a
    range of values but only discrete values allow
    the amplitude of the ion oscillation to remain
    bounded. Hence, it is customary to construct a
    stability diagram that reveals the range of a/q
    values for which the amplitude remains finite
    (stable) and other values for which oscillations
    become infinite (unstable).
  • The a/q diagram show regions of stability that
    may be satisfied by many combinations of U, V and
    m with the assumption that ro and ? are constant.
    The mass scan lines consist of ion values having
    a constant a/q ratio. Transmission of an ion
    occurs when U and V are such that the ion has
    particular values of a and q lying within the
    stable region. A diagram in (U,V) space,
    however, is specific to particular mass to charge
    ratio, and each ion will possess a different U,V
    diagram. Thus, the mass scan will successively
    pass through the apices of a family of diagrams
    (see stability diagrams 5 and 6).

14
Diagram 5 Stability Curve for ions, m1 , m2
and m3 only m2 within the region of stability,
is transmitted and detected.
Diagram 6 Stability curves for ions m1 , m2
and m3
15
  • Mass analysers continued
  • Quadrupole analysers are said to discriminate
    across the mass range as m/z increases, i.e., the
    transmission is reduced whereas magnetic
    analysers do not exhibit this effect.
  • The triple quadrupole is capable of MS-MS, see
    later.
  • Mass scanning is conducted linearly upwards with
    respect to the voltage applied.
  • Ion trap analyser The IT analyser can be
    described as a 3D quadrupole with three
    hyperbolic electrodes comprising a ring and two
    end caps that form the core of the instrument.
    Similarly to the quadrupole only an electric
    field is generated in the ion trap and ions
    generated are held in the ion trap when the dc
    and rf amplitude voltage are applied to the ring
    electrode. Ion ejection is initiated by selective
    ramping of the fundamental rf amplitude voltage
    that sequentially makes ion trajectories unstable
    allowing mass analysis and a mass spectrum to be
    generated by sequential ejection of fragment ions
    from low to high m/z.. Ions are ejected through
    holes in the endcap electrode and detected by an
    electron multiplier. (see schematic diagram - 7)
  • There are two scan modes of operation
  • mass selective stability whereby dc and rf
    voltage applied to the ring electrodes are ramped
    at a constant rf/dc ratio to allow stability and
    storage of ions of a single m/z for mass analysis
  • mass selective instability whereby all ions
    created over a given time period are trapped and
    then sequentially ejected from the IT to the
    detector (EM). Thus ions of differing m/z are
    stored while mass analysis is performed. Unlike
    the mode above this mode coupled with the
    presence of helium gas (1 mtorr) in the trapping
    volume greatly improves mass resolution by
    contracting the ion trajectories to the centre of
    the trap and reducing the kinetic energy of the
    ions allowing ions of a given m/z to form a
    packet. These advances allowed the successful
    development of the commercial ITMS.

16
  • Ion Trapping Ions created by EI, ESI or MALDI
    are focused using an electrostatic lensing system
    into the ion trap. An electrostatic ion gate
    pulses open (-V) and closed (V) to inject ions
    into the ion trap. The pulsing of the gate
    differentiates ion traps from beam instruments
    where ions continually enter the mass analyser.
    The time during which the ions are allowed into
    the trap, termed the ionisation period, is set to
    maximise the signal while minimising the
    space-charge effects. Space-charge results from
    too many ions in the trap that cause a distortion
    of the electrical fields leading to an overall
    reduction in performance. The ion trap is filled
    typically with helium (1 mtorr). Collisions with
    helium molecules dampens the kinetic energy of
    the ions and serve to quickly contract ion
    trajectories toward the centre of the trap,
    enabling trapping of injected ions. Trapped ions
    are further focused toward the centre of the trap
    the use of an oscillating potential, referred to
    as the fundamental (rf), applied to the ring
    electrode. An ion will be stably trapped
    depending on the values of m/z of the ion, the
    size of the ion trap (r), the oscillating
    frequency (?), and Vamp . The dependence of ion
    motion on these parameters is described by the
    dimensionless parameter qz,
  • qz 4eV/mr2w2
  • (see schematic diagram 7 diagram - 8 ).

17
Ring Electrode rf ? Vamp
End Cap Electrode tickle (ac) voltage
Diagram 7 Ion trap schematic
Diagram 8 Stability Diagram (SD) for ions in an
ion trap showing the region of overlap of axial
and radial stability. Depending on the Vamp an
ion of given m/z will have a qz value that falls
within the boundaries of the SD and the ion will
be trapped, If the qz at Vamp falls outside the
boundaries SD, the ion will hit the electrodes
and be lost.
18
  • Ion trap Glossary
  • ac voltage- a voltage placed on the endcap
    electrodes used to induce resonance excitation
    and resonance ejection.
  • bath gas, damping gas, target gas He- in the
    trapping volume at 1 mtorr serves to cool the
    kinetic energies and focus ion trajectories into
    a tight packet in the centre of the trap.
  • fundamental rf- typically 1.1 MHz potential
    applied to the ring electrode, the amplitude of
    which determines the range of m/z values that can
    be trapped and is ramped to eject ions.
  • high resolution- the mass scan of the instrument
    is reduced resulting in an increase in the
    density of data points per unit m/z thus
    increasing mass resolution.
  • qz- dimensionless parameter that determines
    stability of ion trajectories depending upon
    their m/z ratio, the size of the trap, and the
    amplitude and frequency of the rf.
  • Resonance- an ac voltage is applied to the endcap
    and the qz value of an ion of interest is changed
    until the secular frequency of the ion matches
    the frequency of the applied voltage. When
    resonance occurs, the amplitudes of the ion
    trajectories increase linearly with time. A high
    amplitude ac voltage will cause resonance
    ejection, while a low ac voltage will cause
    resonance excitation.
  • secular frequency- the frequency dependent upon
    the qz value, with which an ion oscillates in the
    trap.
  • space-charge effects- too many ions in the trap
    distort the electric fields, leading to
    significantly impaired performance.
  • tickle voltage- an ac voltage applied to the
    endcaps during the excitiation period. The
    amplitude is generally small so as to enable
    fragmentation of the ions by collision with the
    helium damping gas rather than ejection.

19
  • Ion excitation and ejection Ions oscillate with
    a frequency, known as the secular frequency, that
    is determined by the values of az and qz and rf,
    ?. Resonance conditions are induced by matching
    the frequency of a supplementary potential by the
    application of a low amplitude ac resonance
    signal across the end cap electrodes causing the
    ion kinetic energies to increase and leads to ion
    dissociation due to many collisions with the
    helium damping gas. This process causes random
    fragmentation of the ion analogous to that
    obtained in TSQ MS. A mass spectrum is generated
    by sequentially ejecting fragment ions from low
    to high m/z by choosing amplitudes of the
    fundamental rf potential that sequentially make
    ion trajectories unstable. Ions are ejected
    through holes in the encap electrodes and
    detected using an electron multiplier.

20
  • The mass scanning is conducted in a manner
    similar to that for a quadrupole (linearly for
    m/z with respect to voltage).
  • The generation of the electric field is governed
    by complex Mathieu equations that do not need to
    be discussed here.
  • The nominal mass range has been extended from m/z
    650 to 70,000 greater than a quadrupole MS but
    less than a TOF-MS. Worth noting is that
    multi-stage MS-MS (or MSn where n 1 12) can
    be carried out in an ion trap directly by
    introduction of a collision gas (argon) causing
    collisionally-induced decomposition (CID) of the
    initial ions trapped. This feature is beyond the
    capability of of TSQ or TOF-MS and greatly
    increases the the structural information
    obtainable for a given molecule.
  • The ITMS is extremely sensitive being able to
    detect as few as 1.5 million peptide molecules
    and this feature together with others has created
    an interest in applications of ITMS to biological
    molecules. Easy access to different modes of
    scanning for different experiments is provided
    without the need for extensive manual set-up and
    tuning and in-depth understanding of theory thus
    extending the power to biologist and biochemists.

21
  • Mass analysers continued
  • Time of flight mass analyser The TOF analyser
    discriminates between ions of differing m/z by
    the time taken to traverse a flight tube, drift
    to the detector A pulse of ions is accelerated
    out of the source by application of a reasonably
    high voltage. Ions of all m/z have the same
    energy . Thus the velocity v decreases with
    increasing m/z and the time taken to reach the
    detector increases with m/z. The separation of
    ions of differing m/z depends on the time of
    traverse. A time signal is initiated when ions
    are pulsed and also when ions arrive at the
    detector so that the time of flight for each m/z
    can be measured.
  • See schematic diagram - 9
  • The velocity, v, of an ion is determined by the
    accelerating voltage. The kinetic energy is given
    by
  • m2v/2 zV
  • where m mass z charge on ion (electronic
    charge) and V accelerating voltage
  • The time of flight of an ion, t, is given by t
    l/v where l is the length of the flight. The mass
    of ion can be calculated if the time of flight,
    the length of flight, the charge on the electron
    and the accelerating voltage are known, as
    follows
  • m/z 2V2t2 / l2
  • The mass range of a TOF analyser is almost
    infinite and so for this reason with MALD it has
    found application recently in the analysis of
    high mass polymers for biomolecular (protein) and
    synthetic polymer research and applications. The
    resolution of a TOF-MS depends on the time spread
    of the ion beam and was limited to ca. 2000 u.
    Recent advances employ reflectron lenses
    (electrostatic repeller field ion mirror) and
    the resulting delayed extraction allows improved
    resolution by minimising small differences in ion
    energies and resolutions of up to 12000 u
  • . Higher energy ions penetrate deeper into the
    ion mirror than lower energy ions of the same m/z
    and will be turned around and arrive at the
    detector at the same time.

22
Diagram 9 Reflectron TOF MS
Reflectron TOF
23
  • Ionisation methods
  • Ionisation of a compound to produce positive ions
    has been exemplified so far by the
    well-established conventional EI (FD) in an ion
    source under high vacuum conditions. Other
    ionisation methods are available leading to the
    stabilisation of pseudo molecular ions, negative
    ions and adduct ions. Many methods employ
    energetic electrons to initiate the ionisation
    primary ionisation. However, if a reagent
    (reactant) gas is introduced a reaction of the
    primary ion occurs and chemical ionisation
    results as a secondary process leading to the
    production of a secondary ion from molecules of
    the sample to be analysed. Ionisation can be
    produced electrically (ESI) Atoms and ions can
    also be used to ionise molecules under soft
    ionisation conditions (FAB). Lasers can be used
    to desorb and ionise molecules (MALDI)
  • Chemical ionisation positive ion CI Chemical
    ionisation occurs when a reagent gas such CH4,
    NH3 or isopropane is introduced into the ion
    source at low pressure (lt 1 torr) ionisation
    occurs in the reagent gas to produce positive
    ions characteristic of the reagent gas
  • CH4 e- ? CH4 2e- CH4 CH4 ? CH3.
    CH5 M CH5 ? MH CH4
  • The positive ion produced because of its high
    acidity will protonate a neutral molecule of the
    sample introduced to generate a pseudo-molecular
    ion (M H) characteristic of the compound
    analysed. If this ion is produced in an
    energetic state, excess internal energy imparted
    during protonation, as is the case with methane,
    it will decompose leading to lower m/z fragment
    ions. However, if a softer chemical ionisation
    is employed as with NH3 or isopropane , the
    pseudo molecular ion will be more stable. NH3
    leads to the formation of adduct ions at m/z, (M
    18) and (M 35) but only with molecules
    containing heteroatoms with lone pairs, such as O
    and N.

24
  • Ionisation methods continued
  • Chemical ionisation Negative ion CI NICI can
    be achieved in several ways. When a reagent gas
    is introduced the electrons emitted from the
    filament ionise this gas producing more electrons
    which have lower energies brought about through
    inelastic collisions with neutral molecules.
    Thermal electrons are subsequently produced when
    the electron energy is reduced to that of the
    surroundings and these electrons undergo
    resonance capture reaction with molecules of a
    compound to be analysed having a finite electron
    affinity. The process of electron capture is
    simply as follows
  • M e-t ? M.- ? A- B.
  • The radical anion produced is sometimes unstable
    and undergoes dissociative electron capture
    resulting in the formation of stabilised A-
    characteristic of the molecule. Little
    fragmentation often occurs in NICI.
  • NICI can also be conducted by using a reactant
    gas, X, from which reactant gas ions, X-, are
    generated on interaction with electrons produced
    by ionisation of the reactant gas. The reactant
    gas ion undergoes charge transfer with a molecule
    of a compound introduced frequently by proton
    transfer X- M ? XH (M H)- .
  • Fast atom bombardment and secondary ion MS The
    molecular ion generated through electron
    bombardment is very frequently unstable because
    the molecule is sensitive to breakdown. A much
    softer ionisation must be employed to produce
    stable molecular ions in this instance. FAB
    employs fast atoms or ions to sputter ions from
    molecules of the analyte in a liquid matrix, e.g.
    glycerol in which the solid compound is
    dispersed. The process is essentially one of
    desorption ionisation. Fast atoms are prepared by
    accelerating xenon (or argon) ions to an energy
    in the range of 6 9 KeV and then neutralising
    these ions by charge exchange (electron transfer)
    as they pass almost stationary xenon atoms at low
    pressures, the excess kinetic energy is
    transferred to the xenon atoms formed.

25
? Xe Xe ? Xe Xe
  • When the fast atoms impart energy into the
    solution of the sample in the matrix molecules of
    the compound are desorbed, often as ions, by
    momentum transfer. FAB has the capability to
    analyse large polar molecules and it is
    beneficial to couple a FAB source to magnetic
    sector instrument having a mass range of up to
    10, 000 12, 000 daltons at full accelerating
    voltage. An alternative matrix is
    thioglycerol/diglyerol (1/1). It is important
    that the sample dissolves in the matrix and is
    marginally more hydrophobic than the matrix thus
    occupying the matrix/vacuum interface. Cs ions
    are also employed in FAB analysis (see schematic
    diagram - 10).
  • Positive and negative ion FAB mass
    spectra can be recorded, MH and (M-H)- the
    molecular ion species are usually the most
    abundant ions and since large molecules can be
    analysed RMM determination is extremely easy by
    these techniques.
  • Plasma desorption The principle of plasma
    desorption employing 252Cf PD source is
    illustrated in the diagram (see schematic diagram
    - 11). The sample is deposited on a thin metal
    foil, usually nickel. Spontaneous fission of
    252Cf occurs and each fission event results in
    the formation of two fragments travelling
    diametrically at 180o (momentum conservation). A
    typical pair of fragments is 142Ba18 and
    106Tc22 with kinetic energies of roughly 79 and
    104 Mev, respectively. When a high-energy
    fission fragment passes through the sample foil,
    extremely localised heating occurs, producing a
    temperature in the region of 10,000 K. The
    molecules in this plasma zone are desorbed with
    the production of both positive and negative
    ions. The ions formed may be accelerated out
    into the analyser system.
  • 252Cf PD is more able to produce
    molecular ion signals than FAB or SIMS in the RMM
    range 10,000 20,000 daltons, where the
    precision of mass determination may be /- 10
    20 mass units.

26
5. MALDI-MS Matrix-Assisted Laser Desorption
Ionisation TOF.MS is an alternative to PD-MS, see
Diagram 12. For this technique a matrix is
required in which the target compound is
dissolved usually a viscous organic liquid
glycerol, cinnamic acid, nitrophenol derivative
Since lasers only generate coherent light at
specific wavelengths not all compounds undergo
MALDI. The firing of the laser is synchronised
with the TOF to enable accurate flight times and
m/z to be recorded.
27
Diagram 10 FAB mechanism
28
Diagram 11 Plasma Desorption MS
29
Diagram 12 MALDI MS
30
  • Ionisation methods and interfaces for
    chromatography
  • Ionisation devices have been specially designed
    for applications with chromatography in addition
    to the universally used EI and CI mode (stand
    alone MS). The development has been rapid and GC
    was first coupled to mass spectrometry via a jet
    separator for packed column analysis and by
    direct insert for capillary column mode.
    Whereas, for LC applications the advances were
    more recent and have been slower, progressing
    through more design stages Moving belt,
    Particle Beam, (FAB also stand alone),
    Thermospray, PD, Atmospheric Pressure Chemical
    Ionisation and API/Electrospray (originally
    developed in 1930s). Because the mobile phase is
    a liquid and a flow rate of 1 ml/min entering the
    MS on vaporisation, would create a volume of
    vapour which the mass spectrometer pumping system
    could not cope with. Therefore, there is a need
    in the design of the interface to reduce the
    volume of vapour entering.
  • The directly coupled interface for GC requires an
    interface oven heated to 300oC between the GC and
    MS through which the capillary column passes
    before entry into the ion source. The column exit
    will be at low pressure because of the vacuum
    effect of MS manifold reducing the column
    performance and shortening retention times. To
    overcome this effect the column can be connected
    via a split coupling to allow ambient pressure at
    the column exit by introducing helium gas via an
    auxiliary line (and cutting of the
    analytes/solvent eluting from the column if
    required). This device enables matching of
    retention data from GC/MS analysis to that
    obtained from conventional GC.
  • Interfaces for LC The interfaces developed
    specifically for LC include MBI, PBI, TSP, APCI
    and ESI. The discussion will focus mainly on APCI
    and API/ESI. The interface can be divided into
    two types, ionisation interfaces, TSP and ESI and
    pre-ionisation interfaces, APCI (MBI and PBI)
    whereby ionisation occurs in the ion source
    region.

31
  • Electrospray ionisation ESI is a technique used
    to produce gaseous ionised species from a liquid
    solution generated by creating a fine spray of
    charged molecules in the presence of a strong
    electric field. The charged molecules are
    sprayed from the tip of metal nozzle held at
    approximately 3 4 kV and then electrostatically
    directed towards the mass spectrometer inlet.
    Either gas, heat or a combination of both are
    applied to the droplets to induce vaporisation
    prior to entering the MS. As the solvent
    evaporates, the electric field density increases,
    hence the mutual Coulombic repulsion between
    individual species becomes so great that it
    exceeds the surface tension and at the Rayleigh
    limit the droplet explodes forming a series of
    smaller, lower charged droplets. Ions leave the
    droplet through what is known as a Taylor Cone.
    The process of shrinking is repeated until
    individually charged analyte ions are formed.
    Increasing the solvent evaporation, by
    introducing a drying gas flow counter current to
    the sprayed ions increases multiple charging.
  • A Schematic diagram - 13 of an ESI Source and
  • B diagram 14 demonstrating formation of a
    Taylor cone and ion ejection
  • Atmospheric Pressure Ionisation can be carried
    out in two ways
  • AP MS employing ESI which is termed, ionisation
    in the interface. To aid solvent evaporation and
    reduction in droplet size a dryng gas N2 is
    introduced prior to discharge also later as a
    curtain gas to reduce clustering of ions which is
    a feature of ESI.
  • AP MS using chemical ionisation whereby a fine
    spray from a nebuliser is directed into the ion
    source to which an ionisation device is fitted,
    such as discharge needle (Corona) at 7 8 kV or
    plasma generated by 252Cf. Here also N2 is
    introduced to aid evaporation.
  • The AP assists in reducing the flow rate of
    liquid eluting from the HPLC column from 0.5
    ml/min to 10 100 ?l/min so that the vacuum
    system of the MS can handle the inflow of the
    vapour generated from the liquid (NB. 1 ml of
    liquid, usually water, produces over 1000 ml
    vapour).

32
Diagram 13 Electrospray ion source
33
Diagram 14 ESI and formation of Taylor Cone
(coulombic explosion)
34
  • Tandem (Hybrid) Mass Spectrometry for Multiple
    Modes of Scanning (MS-MS)
  • T(H)MS is an instrumental technique in which
    multiple MS-MS modes of scanning can be
    conducted, is divided into three categories
    comprising, for the first two, combinations of
    analysers
  • one type of mass spectrometer only EB-EB, BE-EB
    or Q-q-Q both pure tandem MS-MS (the lower case q
    denotes a transmission only quadrupole)
  • two or more types of analyser in a hybrid
    configuration EBqQ Q-TOF EB-Q-TOF
  • OR a single MS system
  • Ion trap MS in which MSn can be conducted where n
    1 - 12.
  • The tandem MS-MS involving E an B sector
    analysers has been largely superseded by 1. TMS
    involving Q analysers, 2. hybrid MS and 3. ion
    trap MS.
  • The advantages of TMHS are several higher
    specificity for characterising a target compound
    in a complex mixture and higher sensitivity,
    particularly at higher mass resolution for the
    compound of interest (reduced ion background
    interference, higher S/N ratio).
  • The TMS configuration, QqQ, for example, allows
    primary or reactant ions to be selected in the
    first Q, to undergo collision-induced
    deactivation (CID) in second Q (collision gas
    argon transmission only) and the product ions
    resulting to be analysed by dynamic mass scanning
    of the third Q (see type (i) below). For a full
    demo of the operation for MS-MS see Applied
    BioSystems HPLC TSQ MSppt notes.
  • The scanning can be of three types
  • Static mass scanning of the first Q and dynamic
    mass or selected mass scanning of the third Q
    (reactant ions fragmenting to the product
    ions
  • Dynamic mass scanning of the first Q and static
    mass scanning of the third Q (product ions
    having the same reactant ion)
  • Static mass scanning of the first Q and static
    offset mass scanning of the third Q to monitor
    constant neutral loss (mass difference between
    reactant ion 1 and product ion 2).

35
  • Detectors
  • There are several types of detector suitable for
    use with MS
  • Electron and Photomultipliers The EM relies on
    a conversion electrode (dynode) at high voltage
    ( 1 to 3 kV range), at which a positive ion is
    transformed into an electron, an electron gain
    multiplication factor, 106, pre-amplification of
    the electron current, conversion of the electron
    current into an analogue voltage prior to
    digitisation via a fast ADC interfaced to the
    Data System. The modern EM is of a continuous
    (trumpet) dynode type. channeltron c oate on
    the inside with an electron emissive material).
  • see Schematic Diagram - 15
  • The detection in the PM is brought about by
    transformation of a positive ion at a conversion
    dynode into an electron that subsequently
    impinges on a phosphor screen producing a light
    photon (h?), followed by multiplication,
    pre-amplification and current-voltage conversion
    as indicated above.
  • The detection and conversion of a negative ion
    is complicated by the fact that the first dynode
    of the EM cannot be used to directly detect the
    negative ion. A further conversion dynode (at
    high positive kV) is added to transform the
    negative ion into a positive ion which in turn is
    converted by a dynode at -kV into an electron
    prior to multiplication, pre-amplification and
    current to voltage conversion.
  • see Schematic Diagram - 16
  • Microchannel plate This detector consists of an
    array of glass capillaries (10 25 ?m i.d.) that
    area coated on the inside with an electron
    emissive material. The capillaries are biased at
    a high voltage, and like the channeltron, an ion
    that strikes the inner wall of one of the
    capillaries creates a cascade of secondary
    electrons. This cascading effect creates a gain
    of 104 and produces a current at the output.

36
Schematic Diagram - 15
37
Schematic Diagram - 16
Schematic Diagram - 16
38
  • Data Systems
  • Data system electronically interfaced to MS is
    vitally important for
  • Control of MS modes of scanning, voltages,
    currents, etc., tuning and mass calibration
  • data acquisition the digital voltage signal
    response for each ion scanned transmitted from
    the fast ADC
  • data processing to obtain mass spectra, RA
    abundance lists, elemental composition, selected
    mass chromatograms, library search NBS/Wiley,
    quantitation (total ion chromatogram, TIC)
  • Following tuning to optimise the response mass
    peak shape/intensity using a calibrant gas
    pftba (quadrupole) or pfk (EB or BE) and (PEG
    oligomers and Cs ion clusters for high mass) and
    mass calibration, acquisition of data can be
    carried out in conjunction with chromatography
    (GC or LC) or solid probe with ionisation modes
    - EI, FD/FI, CI, FAB, APCI/ESI, laser - LD/I,
    252Cf - PD
  • Centroiding of the mass peak The digitised
    voltage and intensity of the signal (Im/z) of
    the ion peak detected (m/z) is proportional to
    the current carried by that ion and related to
    the time of appearance (tm/z) of that ion in the
    scan (a typical total scan time, span, is 0 - 1
    s). The calibration table stored on the DS is
    used to relate the tm/z to the mass (m/z) and
    thus the intensity v. m/z relationship can
    obtained to build up the mass spectrum from the
    molecular ion and all the fragment ions (m/z)
    detected for the compound being analysed. For
    accurately obtaining the mass of each ion
    detected a process of mass centroiding is
    require. An adequate sampling (data points per
    unit time) of the mass peak is required so that
    the centre of mass can be determined from the
    mass peak profile. The width of the mass peak
    decreases with resolution and thus the ADC
    sampling rate must be extremely fast (MHz range)
    for EB instruments.

see Schematic Diagrams 17 18
39
Schematic Diagram - 17
40
Schematic Diagram - 18
Schematic Diagram - 18
41
  • Scan modes
  • Full Scan For acquisition of the complete mass
    spectrum of a component in a complex mixture (GC
    or LC) or single compound on a probe the mode of
    scanning is full scan over the required mass
    range and 40 650 u is adequate for most organic
    compounds that are volatile (gas
    chromatographable). The mass range for a
    quadrupole is 1000 3000 u while that for a
    magnetic sector is dependent on the accelerating
    voltage.
  • Selected ion monitoring/recording (SIM or SIR)
    For detection of low levels of a particular known
    target compound in a complex mixture, monitoring
    of a single ion, or selected ions characteristic
    of the compound is required. Accurate mass
    tuning and optimisation is required prior to SIM.
    The enhancement is proportional to the time of
    residence on the ion selected which is typically
    0.1 sec for a mass span of /- 0.02 u as compared
    to full scan, 1 - 2 msec for a mass span of 1 u.
    The employment of internal standards (13C or 2H
    labelled analogues) assists in the quantitation
    of target compounds. For EB or BE instruments a
    device to lock the mass selected for the standard
    to that of the target is required and accurate
    mass is carried by out by reference to the mass
    calibration gas (pfk)
  • Multiple or selected reaction monitoring (MRM)
    This mode of scanning is possible only for tandem
    MS and similar to above the MRM can be general
    used or specifically to detect low levels of a
    target compound in a complex mixture by selecting
    the reactant ion and product ions generated, in
    one or several targets to be analysed, such that
    the CID is operated in a unique fashion.

42
Examples of Data Processing
  • Total ion chromatogram (TIC), sum of all the ions
    detected during a chromatographic run in each GC
    peak and the background (chemical noise). Diagram
    19 20
  • Mass Spectrum, raw and refined, RA of each m/z
    with respect to the base ion m/z (100). Diagram
    21
  • Relative Abundance Mass Lists, raw refined, RA
    of an ion, m/z, relative to the base m/z (100).
  • Mass Chromatogram / fragmentogram processing for
    location of target compounds in the TIC. using
    selected characteristic ions, m/z. Diagram 22
  • Library search and match factors for raw and
    refined mass spectra to determine absolute,
    probable, possible and tentative ID.

Examples of Scan function operation
  • Full Scan involves a selected mass range, 45
    450 u, depending on maximum m/z expected
    generating a Total Ion Chromatogram (TIC).
  • Selected or Single Ion Monitoring (SIM/SIR)
    requires mass calibration and full scan
    chromatogram operation (TIC) to obtain the
    accurate masses of the ions to be monitored in
    the GC peak of the target compounds, usually the
    100 m/z and two diagnostic ions, prior to SIM
    chromatogram generation.
  • Selected or Multiple Reaction Monitoring
    (SRM/MRM) in MS-MS involves scanning of
    characteristic precursor ions in Q1, CID in Q2
    and scanning of the product ions in Q3.
    Precursor ions for different compounds undergo
    different fragmentation reactions that may be
    compound specific. SIM can also be conducted
    using MS-MS.

43
Diagram 19 Typical MRM chromatogram
Desethylatrazine 188/146 2ng/mL
Atrazine 216/174 1ng/mL
Propazine 230/146 1ng/mL
Simazine 202/68 1ng/mL
Desisopropylatrazine 174/68 10ng/mL
API 2000 LC/MS/MS system with TurboIonSpray
source
44
Diagram 20 Single quadrupole vs. triple
quadrupole sensitivity/selectivity
analyte
analyte
MS
MS/MS
matrix
Injection of 8.0 ng/ml Acyclovir
much better detection limits in MS/MS due to
higher selectivity in MS-MS
45
Diagram 21 Typical Mass Spectra using MS/MS
Base Mass (ion) Peak, 100
Possible Molecular Mass (ion)
Intensity, or relative
Product Ion Scan 35V Haloperidol
100x more sensitivity in full scan
Mass/Charge Ratio, m/z
46
Diagram 22. Mass Chromatograms (Fragmentograms)
generated through Data Processing from the Total
Ion Chromatogram (m/z 191 and 218 selected),
Write a Comment
User Comments (0)
About PowerShow.com