Title: The USP Performance Test Dissolution Systems Suitability Studies
1The USP Performance Test Dissolution Systems
Suitability Studies
- Walter W. Hauck, Ph.D.
- USP Consultant
- Presentation to Advisory Committee for
Pharmaceutical Sciences - October 25, 2005
2Topics
- Systems Suitability Studies for Dissolution
- Collaborative Study Design and Analysis
- Variances Correction to May 2005 Discussion
- Next Steps
3System Suitability
- USP lt1225gt and ICH Q2b to ensure that the
validity of the analytical method is maintained
whenever used. Typical variations are the
stability of the analytical solutions, equipment,
and analysts. System suitability tests are
based on the concept that the equipment,
electronics, analytic operations, and samples to
be analyzed constitute an integral system that
can be evaluated as such.
4Types of Dissolution Systems Suitability Studies
- Chemical Calibration
- The term is a misnomerit is a periodic Systems
Suitability or system verification study - USP calibrator tablets support this type of study
- The approach evaluates the system as a whole
(apparatus plus operator plus procedures) - Acceptance ranges for dissolution values of the
calibrator tablets determined by collaborative
studies for each new lot (e.g., new Prednisone
Lot P) - Other Approaches
- Mechanical Calibration
- Engineering Approaches
5Terminology
- Apparatus
- One vessel
- One stirring element
- One position in the assembly
- Assembly
- One motor
- Same temperature controller
- Multiple apparatuses (6 or more)
- Also called bath
- Experiment
- Single units tested in all apparatuses in the
assembly (in the collaborative study, n 6)
6Systems Suitability Chemical Calibration
- Calibration procedure
- For assembly under study, use standard
dissolution procedure test all apparatuses in
the assembly - Assembly/apparatus must be mechanically
calibrated - Acceptance applied to each position (apparatus)
- If one position result is unacceptable then
entire assembly is also unacceptable - Increased risk of failing due to multiple testing
- Evaluates the integrated function of the system
to prepare a dissolution sample
7Systems Suitability Mechanical Calibration
- If a mechanical system is properly aligned and
conforms to tolerances established for all such
systems, then the operation of the system should
be suitable - Dissolution test apparatus and assembly are
mechanical systems - If mechanical calibration is the only control, no
further check on the system is performed and
system is assumed to be suitable
8Systems Suitability Mechanical Calibration
- Deficiency of purely mechanical calibration
- Assumes that the integrated system is no more
than the sum of its parts - Does not evaluate that standard flow of medium is
achieved - Limited understanding of the appropriate range
and measurement mode for vibration - Difficulty of determining that vessel is suitable
(surface irregularity, deviation from hemisphere
and cylinder)
List of suggested improved mechanical tolerances
in Dissolution Calibration Recommendations for
Reduced Chemical Testing and Enhanced Mechanical
Calibration, PhRMA Dissolution Calibration
Subcommittee, PF 26(4) July-Aug. 2000
9Topics
- Systems Suitability Studies for Dissolution
- Collaborative Study Design and Analysis
- Variances Correction to May 2005 Discussion
- Next Steps
10Collaborative Study Design
- Standard product (tablet)
- Standard procedures employed
- Determine range of results from
mechanically-calibrated assemblies - Determine bounds for acceptable results
- 25-35 collaborating laboratories, international
11Design Evolution
- Originally, each lab performed one experiment per
condition (e.g., basket at 50 rpm) - Beginning with 2003, each lab performs two
experiments per condition with separate operator
and equipment - Allows separation of inter-laboratory and
intra-laboratory (intermediate) precision
12Analysis Considerations
- Analysis is preceded by a control chart analysis
to select labs/experiments for inclusion in
determination of acceptance ranges - Follows what had been PhRMA procedure
- Statistical analysis determines three variances
- apparatus/tablet/assay
- between-experiment, with-in laboratory
- between-laboratory
13Sample Xbar and S Control Charts for Apparatus 2,
30 Minutes
14Representative CVs ()
15Comments
- Variability due to the combination of apparatus,
tablet, and assay is low, particularly for the
basket (Apparatus 1) - Assay variability around 2 CV, so much of the
contribution from apparatus/tablet/assay is due
to the assay
16Determination of Acceptance Ranges
- Acceptance range based on sum of the three
variances - Intended to represent range of values to be
expected from a random tablet tested at a random
laboratory using good practice
17Issue for Acceptance Ranges
- Ranges determined for a single tablet, but
applied to sets of 6 and must pass all 6 - Possible solutions
- Standard statistical multiple testing adjustment
i.e., widen the intervals to 99 from 95 - Allow retesting
- Set acceptance ranges on sample mean and standard
deviation as done in collaborative study i.e.,
would the data qualify for inclusion in the
determination of the acceptance range
18Topics
- Systems Suitability Studies for Dissolution
- Collaborative Study Design and Analysis
- Variances Correction to May 2005 Discussion
- Next Steps
19Calibrator to Control Product Variability
Argument against USP RS in dissolution system
suitability delivered at various venues
- Total product variability, s2total s2product
s2measurement - s2total for calibrator s2 calibrator s2
Cmeasurement - Since s2 calibrator is not known use s2total for
calibrator - s2total for calibrator is given in the acceptable
ranges - s2total s2product calibrator range
20Restating Concerns About Calibrator Variability
- This argument does not hold maybe s2total lt
s2product calibrator range - Calibrator variability does not contribute to
product variability
21Topics
- Systems Suitability Studies for Dissolution
- Collaborative Study Design and Analysis
- Variances Correction to May 2005 Discussion
- Next Steps
22Next Steps
- Continued work to improve chemical calibration,
e.g., approach to avoid multiple testing - Acknowledge that dissolution is case by
caseresolved with flexible monographdifferenc
es acceptable if bioequivalent - Assess various approaches to Systems Suitability
for the dissolution procedure e.g., develop
methodology to measure fluid flow that is
applicable in QC environment (engineering
approach) - Setting acceptance criteria for marketed dosage
forms (not calibrators) tolerance interval
approach based on clinical trial batches (see
Hauck et al., Pharmaceutical Research 2005) - USP willing to work with all stakeholders on
complex science/technical issues for Performance
test Quality by Design approaches should be
highly valuable.
23(No Transcript)