The USP Performance Test Dissolution Systems Suitability Studies - PowerPoint PPT Presentation

1 / 23
About This Presentation
Title:

The USP Performance Test Dissolution Systems Suitability Studies

Description:

Mechanical Calibration. Engineering Approaches. Terminology. Apparatus. One vessel ... Dissolution test apparatus and assembly are mechanical systems ... – PowerPoint PPT presentation

Number of Views:1417
Avg rating:3.0/5.0
Slides: 24
Provided by: william533
Category:

less

Transcript and Presenter's Notes

Title: The USP Performance Test Dissolution Systems Suitability Studies


1
The USP Performance Test Dissolution Systems
Suitability Studies
  • Walter W. Hauck, Ph.D.
  • USP Consultant
  • Presentation to Advisory Committee for
    Pharmaceutical Sciences
  • October 25, 2005

2
Topics
  • Systems Suitability Studies for Dissolution
  • Collaborative Study Design and Analysis
  • Variances Correction to May 2005 Discussion
  • Next Steps

3
System Suitability
  • USP lt1225gt and ICH Q2b to ensure that the
    validity of the analytical method is maintained
    whenever used. Typical variations are the
    stability of the analytical solutions, equipment,
    and analysts. System suitability tests are
    based on the concept that the equipment,
    electronics, analytic operations, and samples to
    be analyzed constitute an integral system that
    can be evaluated as such.

4
Types of Dissolution Systems Suitability Studies
  • Chemical Calibration
  • The term is a misnomerit is a periodic Systems
    Suitability or system verification study
  • USP calibrator tablets support this type of study
  • The approach evaluates the system as a whole
    (apparatus plus operator plus procedures)
  • Acceptance ranges for dissolution values of the
    calibrator tablets determined by collaborative
    studies for each new lot (e.g., new Prednisone
    Lot P)
  • Other Approaches
  • Mechanical Calibration
  • Engineering Approaches

5
Terminology
  • Apparatus
  • One vessel
  • One stirring element
  • One position in the assembly
  • Assembly
  • One motor
  • Same temperature controller
  • Multiple apparatuses (6 or more)
  • Also called bath
  • Experiment
  • Single units tested in all apparatuses in the
    assembly (in the collaborative study, n 6)

6
Systems Suitability Chemical Calibration
  • Calibration procedure
  • For assembly under study, use standard
    dissolution procedure test all apparatuses in
    the assembly
  • Assembly/apparatus must be mechanically
    calibrated
  • Acceptance applied to each position (apparatus)
  • If one position result is unacceptable then
    entire assembly is also unacceptable
  • Increased risk of failing due to multiple testing
  • Evaluates the integrated function of the system
    to prepare a dissolution sample

7
Systems Suitability Mechanical Calibration
  • If a mechanical system is properly aligned and
    conforms to tolerances established for all such
    systems, then the operation of the system should
    be suitable
  • Dissolution test apparatus and assembly are
    mechanical systems
  • If mechanical calibration is the only control, no
    further check on the system is performed and
    system is assumed to be suitable

8
Systems Suitability Mechanical Calibration
  • Deficiency of purely mechanical calibration
  • Assumes that the integrated system is no more
    than the sum of its parts
  • Does not evaluate that standard flow of medium is
    achieved
  • Limited understanding of the appropriate range
    and measurement mode for vibration
  • Difficulty of determining that vessel is suitable
    (surface irregularity, deviation from hemisphere
    and cylinder)

List of suggested improved mechanical tolerances
in Dissolution Calibration Recommendations for
Reduced Chemical Testing and Enhanced Mechanical
Calibration, PhRMA Dissolution Calibration
Subcommittee, PF 26(4) July-Aug. 2000
9
Topics
  • Systems Suitability Studies for Dissolution
  • Collaborative Study Design and Analysis
  • Variances Correction to May 2005 Discussion
  • Next Steps

10
Collaborative Study Design
  • Standard product (tablet)
  • Standard procedures employed
  • Determine range of results from
    mechanically-calibrated assemblies
  • Determine bounds for acceptable results
  • 25-35 collaborating laboratories, international

11
Design Evolution
  • Originally, each lab performed one experiment per
    condition (e.g., basket at 50 rpm)
  • Beginning with 2003, each lab performs two
    experiments per condition with separate operator
    and equipment
  • Allows separation of inter-laboratory and
    intra-laboratory (intermediate) precision

12
Analysis Considerations
  • Analysis is preceded by a control chart analysis
    to select labs/experiments for inclusion in
    determination of acceptance ranges
  • Follows what had been PhRMA procedure
  • Statistical analysis determines three variances
  • apparatus/tablet/assay
  • between-experiment, with-in laboratory
  • between-laboratory

13
Sample Xbar and S Control Charts for Apparatus 2,
30 Minutes
14
Representative CVs ()
15
Comments
  • Variability due to the combination of apparatus,
    tablet, and assay is low, particularly for the
    basket (Apparatus 1)
  • Assay variability around 2 CV, so much of the
    contribution from apparatus/tablet/assay is due
    to the assay

16
Determination of Acceptance Ranges
  • Acceptance range based on sum of the three
    variances
  • Intended to represent range of values to be
    expected from a random tablet tested at a random
    laboratory using good practice

17
Issue for Acceptance Ranges
  • Ranges determined for a single tablet, but
    applied to sets of 6 and must pass all 6
  • Possible solutions
  • Standard statistical multiple testing adjustment
    i.e., widen the intervals to 99 from 95
  • Allow retesting
  • Set acceptance ranges on sample mean and standard
    deviation as done in collaborative study i.e.,
    would the data qualify for inclusion in the
    determination of the acceptance range

18
Topics
  • Systems Suitability Studies for Dissolution
  • Collaborative Study Design and Analysis
  • Variances Correction to May 2005 Discussion
  • Next Steps

19
Calibrator to Control Product Variability
Argument against USP RS in dissolution system
suitability delivered at various venues
  • Total product variability, s2total s2product
    s2measurement
  • s2total for calibrator s2 calibrator s2
    Cmeasurement
  • Since s2 calibrator is not known use s2total for
    calibrator
  • s2total for calibrator is given in the acceptable
    ranges
  • s2total s2product calibrator range

20
Restating Concerns About Calibrator Variability
  • This argument does not hold maybe s2total lt
    s2product calibrator range
  • Calibrator variability does not contribute to
    product variability

21
Topics
  • Systems Suitability Studies for Dissolution
  • Collaborative Study Design and Analysis
  • Variances Correction to May 2005 Discussion
  • Next Steps

22
Next Steps
  • Continued work to improve chemical calibration,
    e.g., approach to avoid multiple testing
  • Acknowledge that dissolution is case by
    caseresolved with flexible monographdifferenc
    es acceptable if bioequivalent
  • Assess various approaches to Systems Suitability
    for the dissolution procedure e.g., develop
    methodology to measure fluid flow that is
    applicable in QC environment (engineering
    approach)
  • Setting acceptance criteria for marketed dosage
    forms (not calibrators) tolerance interval
    approach based on clinical trial batches (see
    Hauck et al., Pharmaceutical Research 2005)
  • USP willing to work with all stakeholders on
    complex science/technical issues for Performance
    test Quality by Design approaches should be
    highly valuable.

23
(No Transcript)
Write a Comment
User Comments (0)
About PowerShow.com