Title: This course is concerned with instrumental methods of analysis
1Chapter 1
- This course is concerned with instrumental
methods of analysis - These methods use various chemical and physical
properties of analytes that can be measured
instrumentally to determine qualitative (what is
there) and quantitative (how much is there)
information about these analytes
2first 6 are spectroscopic-involve interaction of
light and matter next 4 are electrochemical-measur
ement of voltage, current, or resistance mass
spectrometry several other types
3- Chromatography and electrophoresis are considered
instrumental techniques because they use many of
the techniques mentioned above in separating
(electrophoresis) and detecting (both) analytes
after separation - As can be seen from topics list and laboratory
schedule, we will cover many of the spectroscopic
techniques, some electrochemistry,
chromatography, and mass spectrometry
4 Course Objectives
- basic theory behind each technique
- instrument design and components
- types of analytes than can be analyzed by a
particular technique - qualitative and/or quantitative uses of a
technique - limitations of each technique, including
- precision - agreement among repeat measurements
- accuracy - agreement of measurements with true
value - comparisons of techniques so intelligent choices
can be made between them
5General instrument examples
6transducer changes signal from nonelectrical to
electrical domains (and vice versa) detector is
the more general term, i.e. all transducers are
detectors but not all detectors are transducers
signal processing include amplification,
attenuation (decreasing signal), filtering,
integration, differentiation, etc. No matter
what type of instrument or method, quantitative
analyses require some sort of calibration curve
in which signal (A, V, i, conductivity, emission,
etc.) is plotted on y axis against known
concentrations of analyte on x axis
7- errors, absolute and relative, must also be
considered - determinate errors - in principle can be
accounted for, always high or always low (method,
operator, instrument) - indeterminate - random and uncontrollable
- Random means error will be high and low ()
- Are preferred type of error because
- they can be treated statistically
- means analyst is getting the most out of the
instruments and methods
8Precision
- precision agreement among repeat measurements
absolute error is the standard deviation of the
determination. - It determines number of significant digits in
result. - example report the following number and its
standard deviation to the correct number of
significant figures - 45.6879 0.07895 mM
9Precision
- relative error in precision coefficient of
variation (CV) - (absolute error/value) x 100
- Why important?
- compare relative errors in the following
- 3.495 0.034 and 10.678 0.034
10Accuracy
- accuracy absolute error is difference between
determined value and true value - relative error in accuracy (actual value
true value)/ true value x 100 - ex. determine absolute and relative errors in
accuracy between determined value 4.58 mM and its
accepted value of 4.62 mM
11Method Selection
To select a method one must 1. define problem
(section 1E-1) 2. determine other performance
criteria (table 1-4)
12- 1. What accuracy and precision are required?
- 2. How much sample is available?
- ex. at least 5 mL needed for a flame AA
measurement - 3. What is the concentration range of the
analyte? - ex. ppm flame AA, ppb graphite furnace AA
-
- 4. What sample components will cause
interference? - ex. PO43- in AA analysis of calcium
- 5. What are chemical and physical properties of
sample matrix? - 6. How many samples are to be analyzed?
13Matrix
rigorous definition of matrix all constituents
of a sample including analyte (other
constituents are called concomitants) if
analyte concentration is very low (so it does
not measurably affect other constituents) then
matrix is often defined as everything else in
sample except analyte(s)
14Table 1-4 (modified)
Speed Ease and convenience Skill required of
operator Cost of instrumentation (to buy and
use) Availability of instrumentation Per-sample
cost
15Accuracy Determination
Also must consider ways of determining
accuracy and other performance characteristics
in table 1-3 (p.11)
16Accuracy Determination
- Accuracy determined in one of three ways
- analysis of standard reference material
- intermethod comparison
- spike recovery
17analysis of standard reference material
- reference material containing analyte in matrix
essentially identical to that of unknown sample
is analyzed. - If stated amount of analyte in reference
standard is obtained within experimental error,
method is accurate
18intermethod comparison
- Identical samples are analyzed by method in
question and a method known to be accurate and
independent of method being tested. - Results are then plotted against each other.
- If slope of graph is equal to one (within
experimental error) then method is accurate. - This can be used for unstable analytes.
19(No Transcript)
20spike recovery
- Known amount of the analyte is added to a known
amount of sample and analyzed. - If 100 of this spike is recovered (within
experimental error) then method is considered
accurate. - Is not recommended for unstable analytes.
21(No Transcript)
22This s is for multiple measurements of same number
23This is not the way s is calculated when
concentration is determined from a calibration
curve. We will discuss this when calibration
curves are covered below.
24Sensitivity
- ability to discriminate between small differences
in analyte concentration. - Limited by two factors
- 1) slope of calibration curve
- 2) precision of method
25signal
concentration
26calibration sensitivity
- slope of calibration curve S mc Sbl
- independent of concentration
- does not take precision into account
27analytical sensitivity (g)
- g m/sS
- 1) relatively insensitive to amplification
- 2) independent of measurement units
often concentration dependent because sS often
varies with concentration
28Detection Limit (LOD)
- minimum concentration or mass that can be
detected at a known confidence level - depends primarily on ratio of analytical signal
size to size of standard deviation of blank
signal - also is a function of slope of calibration curve
29Detection Limit
- Sm Sbl ksbl
- Sm Minimum Signal
- Sbl mean blank signal measured 20 to 30 times
over extended period of time - sbl standard deviation of blank k 3
30Detection Limit
- use m from calibration curve of analyte
- S mc Sbl
- to convert Sm to cm (note that Sbl and Sbl
are different) -
- cm (Sm - Sbl )/m
- (Sbl ksbl - Sbl) / m
- 3sbl/ m
31Detection Limit
- use m from calibration curve of analyte
- S mc Sbl
- to convert Sm to cm (note that Sbl and Sbl
are different) -
- cm (Sm - Sbl )/m
- (Sbl ksbl - Sbl) / m
- 3sbl/ m
32Detection Limit
- use m from calibration curve of analyte
- S mc Sbl
- to convert Sm to cm (note that Sbl and Sbl
are different) -
- cm (Sm - Sbl )/m
- (Sbl ksbl - Sbl) / m
- 3sbl/ m LOD
33Detection Limit example
- A method requires reaction of sample with
enzyme and then DNPH reagent. 20 blanks were
analyzed with one batch of enzyme. Another 20
blanks were analyzed with another batch of enzyme
one month later. Standard deviations of sets were
same so results pooled to determine Sbl, sbl, Sm,
and (with slope of calibration curve) LOD - Sbl 24.5 mAUs sbl 0.534 mAUs m
52.9 mAUs/mM
34Concentration range
- Dynamic range range from lowest concentration
at which quantitative measurements can be made
(LOQ) to point where curve begins to deviate from
linearity (LOL)
35Figure 1-7
10sbl
36Bias
- measure of determinate error in method
- Bias m xt
- where m is population mean of analyte
concentration in a standard reference sample
(usually determined from minimum of 20
measurements) and xt is true value.
37Selectivity
- degree to which method is free from interference
by other species in matrix - interference can add to or subtract from signal.
- ex. ion selective electrodes
38Calibration Curves
- All instrumental methods (except gravimetry
- and coulometry) require calibration curve to
- relate measured signal to concentration
- (amount) of analyte
39- Titrations do not require calibration curves They
are not mentioned in your text because when used
to determine amounts or concentrations,
titrations can be done without instrumentation
(just burette, flask, and indicator needed)
40Calibration Curves
- Calibration curve which is usually linear.
- Why?
- vs
- Ease of error analysis
- Constant sensitivity
- Fewer data points needed
41Calibration Curves
- Three types of calibration curves
- External standard
- Standard Addition
- Internal Standard
42External standard
- constructed by measuring signals of several known
concentrations of analyte and then plotting
signal vs. concentration - known analyte solutions must have a similar
matrix to that of sample - signal magnitude should not be a function of
small sample volumes that may be difficult to
reproduce Ex. GC - this procedure works well for simple matrices
- error in result (concentration of unknown) is
determined by equation from appendix 1
43External Standard
- Calibration curve prepared from known
concentrations - Unknown solution measured
44Concentration of unknown 1.71 ppm
45Standard Addition
- known amounts of analyte are added directly to
sample to try to account for matrix effects - two main types of standard addition use either a
constant total volume or a changing total volume.
46Standard Addition Constant Total V
- Analysis Conditions
- constant total volume
- constant volume of unknown analyte
- varying volume of added analyte of known
concentration to each sample - linear response of instrument to concentration y
mx b
47Standard Addition Constant Total V
- S signal k proportionality constant
- Vs volume of standard cs concentration of
standard - Vt total volume
- Vx volume of unknown cx unknown
concentration
48Standard Addition Constant Total V
- all variables are constants except S and Vs
therefore this is a straight line equation where
- m and b
49Standard Addition Constant Total V
- Plot S vs Vs for several different Vs
- Determine m and b
- find cx using ratio of b to m
- After rearranging -
50Standard Addition Example
Six samples. Each contains 5.00 mL of unknown
copper solution, x mL of added 50.0 ppm copper
solution, and each sample is diluted to a final
volume of 10.00 mL.
51 2.000 ppm
52Concentration of unknown 1.71 ppm
53Standard addition vs external standard
- How to determine if standard addition is needed?
- Have to do analysis both ways and see if external
standard analysis gives numbers that are always
too high or too low
54Standard addition vs external standard
- Why isnt standard addition always used instead
of external calibration? - Compare time involved for the preceding example
which was an AA analysis
55Standard addition vs external standard
- External standard calibration curve analysis
- One calibration curve (usually 4 to 6 points)
- Single measurement for each sample
- Standard Addition analysis
- At least two measurements for each sample
- Based on above information after 4 to 6 samples
external standard analysis is faster
56Standard addition vs external standard
- External standard is more efficient if many
samples and fast analysis time for each sample - Standard addition only used under these
conditions if external standards do not give
correct answers
57Standard addition vs external standard
- Voltammetric analyses almost always use standard
addition because - each measurement that requires emptying and
refilling cell takes several minutes - Addition of known to solution that has already
been deoxygenated allows for quick measurement - Electrode surface is very sensitive to changes in
solutions
58Two point standard addition, constant total V
- Can be done with only two points using the
unknown and one addition to sample - Unknown concentration is determined from
- where S1 is signal of diluted unknown and S2 is
signal of diluted unknown plus standard
59Two point standard addition, changing total V
- Signal from known V of unknown concentration is
measured (S1) - Signal after known V of standard added to this
unknown sample (S2) - Unknown concentration determined from
60Internal Standard
- Corrects for random instrumental and method
fluctuations - Ex. Methods in which signal depends on the sample
volume used such as GC - Methods in which T variations can affect signal
such as Atomic Emission (AE) spectroscopy
61Internal Standard
- A constant known amount of a substance that is
similar to, but not the same as, analyte is added
to all standards and unknown samples - Signals for calibration curve generated from
ratio of analyte signal to internal standard
signal - Ratio of unknown analyte signal to internal
standard signal is used to determine analyte
concentration from calibration curve.
62Internal Standard Example
- GC-MS analysis of octane using nonane as internal
standard - 5 octane standards from 1.00 to 10.00 mg/mL each
containing 3.00 mg/mL nonane - Octane analyte containing 3.00 mg/mL nonane
63Internal standard example data
64Octane concentration 1.81 mg/mL
65Octane concentration 2.04 mg/mL
66Error analysis for these methods
- How to find error in concentration of an unknown
determined by external standard calibration
curve? - Review least squares
- y mx b
- Simple least squares assumes a constant error in
y value (measurement signal) and no error in x
axis values (concentration)
67- Residual is vertical deviation of each data point
from line - Line generated by least squares minimizes sum of
the squares of residuals for all points
68Least Squares Equations
- To determine slope, intercept, and errors these
three quantities are very useful
69- sr standard deviation about regression
(standard error of the estimate) - N number of points in calibration curve
70- sc standard deviation in results
(concentrations) from curve - M number of replicate analyses of an unknown
- N number of points in calibration curve
- mean of signals from analysis of
unknown - mean value of y signals used in
calibration curve
71External Standard
- Calibration curve prepared from known
concentrations - Unknown solution measured
72Concentration of unknown 1.71 ppm
73Excel version
Block out 2x5 matrix of cells linest(ys,xs,true
,true) hold down control and shift buttons and
then hit enter
74Error Analysis for standard addition
- Review propagation of error for random errors
75Error Analysis for standard addition
cx is not determined by reading signal from
straight line
unknown signals are actually part of calibration
curve
determine sx in cx by assuming no error in cs or
Vx and that all error comes from b and m
Use propagation of error equation from Table a1-6
to come up with
76Excel version
77Internal standard example data
78Octane concentration 1.81 mg/mL
Error analysis
79Octane concentration 2.04 mg/mL
error analysis