This course is concerned with instrumental methods of analysis - PowerPoint PPT Presentation

1 / 79
About This Presentation
Title:

This course is concerned with instrumental methods of analysis

Description:

... measurement of voltage, current, or resistance mass spectrometry ... Line generated by least squares minimizes sum of the squares of residuals for all points ... – PowerPoint PPT presentation

Number of Views:514
Avg rating:3.0/5.0
Slides: 80
Provided by: UNC9151
Category:

less

Transcript and Presenter's Notes

Title: This course is concerned with instrumental methods of analysis


1
Chapter 1
  • This course is concerned with instrumental
    methods of analysis
  • These methods use various chemical and physical
    properties of analytes that can be measured
    instrumentally to determine qualitative (what is
    there) and quantitative (how much is there)
    information about these analytes

2
first 6 are spectroscopic-involve interaction of
light and matter next 4 are electrochemical-measur
ement of voltage, current, or resistance mass
spectrometry several other types
3
  • Chromatography and electrophoresis are considered
    instrumental techniques because they use many of
    the techniques mentioned above in separating
    (electrophoresis) and detecting (both) analytes
    after separation
  • As can be seen from topics list and laboratory
    schedule, we will cover many of the spectroscopic
    techniques, some electrochemistry,
    chromatography, and mass spectrometry

4
Course Objectives
  • basic theory behind each technique
  • instrument design and components
  • types of analytes than can be analyzed by a
    particular technique
  • qualitative and/or quantitative uses of a
    technique
  • limitations of each technique, including
  • precision - agreement among repeat measurements
  • accuracy - agreement of measurements with true
    value
  • comparisons of techniques so intelligent choices
    can be made between them

5
General instrument examples
6
transducer changes signal from nonelectrical to
electrical domains (and vice versa) detector is
the more general term, i.e. all transducers are
detectors but not all detectors are transducers
signal processing include amplification,
attenuation (decreasing signal), filtering,
integration, differentiation, etc. No matter
what type of instrument or method, quantitative
analyses require some sort of calibration curve
in which signal (A, V, i, conductivity, emission,
etc.) is plotted on y axis against known
concentrations of analyte on x axis
7
  • errors, absolute and relative, must also be
    considered
  • determinate errors - in principle can be
    accounted for, always high or always low (method,
    operator, instrument)
  • indeterminate - random and uncontrollable
  • Random means error will be high and low ()
  • Are preferred type of error because
  • they can be treated statistically
  • means analyst is getting the most out of the
    instruments and methods

8
Precision
  • precision agreement among repeat measurements
    absolute error is the standard deviation of the
    determination.
  • It determines number of significant digits in
    result.
  • example report the following number and its
    standard deviation to the correct number of
    significant figures
  • 45.6879 0.07895 mM

9
Precision
  • relative error in precision coefficient of
    variation (CV)
  • (absolute error/value) x 100
  • Why important?
  • compare relative errors in the following
  • 3.495 0.034 and 10.678 0.034

10
Accuracy
  • accuracy absolute error is difference between
    determined value and true value
  • relative error in accuracy (actual value
    true value)/ true value x 100
  • ex. determine absolute and relative errors in
    accuracy between determined value 4.58 mM and its
    accepted value of 4.62 mM

11
Method Selection
To select a method one must 1. define problem
(section 1E-1) 2. determine other performance
criteria (table 1-4)
12
  • 1. What accuracy and precision are required?
  • 2. How much sample is available?
  • ex. at least 5 mL needed for a flame AA
    measurement
  • 3. What is the concentration range of the
    analyte?
  • ex. ppm flame AA, ppb graphite furnace AA
  • 4. What sample components will cause
    interference?
  • ex. PO43- in AA analysis of calcium
  • 5. What are chemical and physical properties of
    sample matrix?
  • 6. How many samples are to be analyzed?

13
Matrix
rigorous definition of matrix all constituents
of a sample including analyte (other
constituents are called concomitants) if
analyte concentration is very low (so it does
not measurably affect other constituents) then
matrix is often defined as everything else in
sample except analyte(s)
14
Table 1-4 (modified)
Speed Ease and convenience Skill required of
operator Cost of instrumentation (to buy and
use) Availability of instrumentation Per-sample
cost
15
Accuracy Determination
Also must consider ways of determining
accuracy and other performance characteristics
in table 1-3 (p.11)
16
Accuracy Determination
  • Accuracy determined in one of three ways
  • analysis of standard reference material
  • intermethod comparison
  • spike recovery

17
analysis of standard reference material
  • reference material containing analyte in matrix
    essentially identical to that of unknown sample
    is analyzed.
  • If stated amount of analyte in reference
    standard is obtained within experimental error,
    method is accurate

18
intermethod comparison
  • Identical samples are analyzed by method in
    question and a method known to be accurate and
    independent of method being tested.
  • Results are then plotted against each other.
  • If slope of graph is equal to one (within
    experimental error) then method is accurate.
  • This can be used for unstable analytes.

19
(No Transcript)
20
spike recovery
  • Known amount of the analyte is added to a known
    amount of sample and analyzed.
  • If 100 of this spike is recovered (within
    experimental error) then method is considered
    accurate.
  • Is not recommended for unstable analytes.

21
(No Transcript)
22
This s is for multiple measurements of same number
23
This is not the way s is calculated when
concentration is determined from a calibration
curve. We will discuss this when calibration
curves are covered below.
24
Sensitivity
  • ability to discriminate between small differences
    in analyte concentration.
  • Limited by two factors
  • 1) slope of calibration curve
  • 2) precision of method

25
signal
concentration
26
calibration sensitivity
  • slope of calibration curve S mc Sbl
  • independent of concentration
  • does not take precision into account

27
analytical sensitivity (g)
  • g m/sS
  • 1) relatively insensitive to amplification
  • 2) independent of measurement units

often concentration dependent because sS often
varies with concentration

28
Detection Limit (LOD)
  • minimum concentration or mass that can be
    detected at a known confidence level
  • depends primarily on ratio of analytical signal
    size to size of standard deviation of blank
    signal
  • also is a function of slope of calibration curve

29
Detection Limit
  • Sm Sbl ksbl
  • Sm Minimum Signal
  • Sbl mean blank signal measured 20 to 30 times
    over extended period of time
  • sbl standard deviation of blank k 3


30
Detection Limit
  • use m from calibration curve of analyte
  • S mc Sbl
  • to convert Sm to cm (note that Sbl and Sbl
    are different)
  • cm (Sm - Sbl )/m
  • (Sbl ksbl - Sbl) / m
  • 3sbl/ m





31
Detection Limit
  • use m from calibration curve of analyte
  • S mc Sbl
  • to convert Sm to cm (note that Sbl and Sbl
    are different)
  • cm (Sm - Sbl )/m
  • (Sbl ksbl - Sbl) / m
  • 3sbl/ m





32
Detection Limit
  • use m from calibration curve of analyte
  • S mc Sbl
  • to convert Sm to cm (note that Sbl and Sbl
    are different)
  • cm (Sm - Sbl )/m
  • (Sbl ksbl - Sbl) / m
  • 3sbl/ m LOD





33
Detection Limit example
  • A method requires reaction of sample with
    enzyme and then DNPH reagent. 20 blanks were
    analyzed with one batch of enzyme. Another 20
    blanks were analyzed with another batch of enzyme
    one month later. Standard deviations of sets were
    same so results pooled to determine Sbl, sbl, Sm,
    and (with slope of calibration curve) LOD
  • Sbl 24.5 mAUs sbl 0.534 mAUs m
    52.9 mAUs/mM



34
Concentration range
  • Dynamic range range from lowest concentration
    at which quantitative measurements can be made
    (LOQ) to point where curve begins to deviate from
    linearity (LOL)

35
Figure 1-7
10sbl
36
Bias
  • measure of determinate error in method
  • Bias m xt
  • where m is population mean of analyte
    concentration in a standard reference sample
    (usually determined from minimum of 20
    measurements) and xt is true value.

37
Selectivity
  • degree to which method is free from interference
    by other species in matrix
  • interference can add to or subtract from signal.
  • ex. ion selective electrodes

38
Calibration Curves
  • All instrumental methods (except gravimetry
  • and coulometry) require calibration curve to
  • relate measured signal to concentration
  • (amount) of analyte

39
  • Titrations do not require calibration curves They
    are not mentioned in your text because when used
    to determine amounts or concentrations,
    titrations can be done without instrumentation
    (just burette, flask, and indicator needed)

40
Calibration Curves
  • Calibration curve which is usually linear.
  • Why?
  • vs
  • Ease of error analysis
  • Constant sensitivity
  • Fewer data points needed

41
Calibration Curves
  • Three types of calibration curves
  • External standard
  • Standard Addition
  • Internal Standard

42
External standard
  • constructed by measuring signals of several known
    concentrations of analyte and then plotting
    signal vs. concentration
  • known analyte solutions must have a similar
    matrix to that of sample
  • signal magnitude should not be a function of
    small sample volumes that may be difficult to
    reproduce Ex. GC
  • this procedure works well for simple matrices
  • error in result (concentration of unknown) is
    determined by equation from appendix 1

43
External Standard
  • Calibration curve prepared from known
    concentrations
  • Unknown solution measured

44
Concentration of unknown 1.71 ppm
45
Standard Addition
  • known amounts of analyte are added directly to
    sample to try to account for matrix effects
  • two main types of standard addition use either a
    constant total volume or a changing total volume.

46
Standard Addition Constant Total V
  • Analysis Conditions
  • constant total volume
  • constant volume of unknown analyte
  • varying volume of added analyte of known
    concentration to each sample
  • linear response of instrument to concentration y
    mx b

47
Standard Addition Constant Total V
  • S signal k proportionality constant
  • Vs volume of standard cs concentration of
    standard
  • Vt total volume
  • Vx volume of unknown cx unknown
    concentration

48
Standard Addition Constant Total V
  • all variables are constants except S and Vs
    therefore this is a straight line equation where
  • m and b

49
Standard Addition Constant Total V
  • Plot S vs Vs for several different Vs
  • Determine m and b
  • find cx using ratio of b to m
  • After rearranging -

50
Standard Addition Example
Six samples. Each contains 5.00 mL of unknown
copper solution, x mL of added 50.0 ppm copper
solution, and each sample is diluted to a final
volume of 10.00 mL.
51
2.000 ppm
52
Concentration of unknown 1.71 ppm
53
Standard addition vs external standard
  • How to determine if standard addition is needed?
  • Have to do analysis both ways and see if external
    standard analysis gives numbers that are always
    too high or too low

54
Standard addition vs external standard
  • Why isnt standard addition always used instead
    of external calibration?
  • Compare time involved for the preceding example
    which was an AA analysis

55
Standard addition vs external standard
  • External standard calibration curve analysis
  • One calibration curve (usually 4 to 6 points)
  • Single measurement for each sample
  • Standard Addition analysis
  • At least two measurements for each sample
  • Based on above information after 4 to 6 samples
    external standard analysis is faster

56
Standard addition vs external standard
  • External standard is more efficient if many
    samples and fast analysis time for each sample
  • Standard addition only used under these
    conditions if external standards do not give
    correct answers

57
Standard addition vs external standard
  • Voltammetric analyses almost always use standard
    addition because
  • each measurement that requires emptying and
    refilling cell takes several minutes
  • Addition of known to solution that has already
    been deoxygenated allows for quick measurement
  • Electrode surface is very sensitive to changes in
    solutions

58
Two point standard addition, constant total V
  • Can be done with only two points using the
    unknown and one addition to sample
  • Unknown concentration is determined from
  • where S1 is signal of diluted unknown and S2 is
    signal of diluted unknown plus standard

59
Two point standard addition, changing total V
  • Signal from known V of unknown concentration is
    measured (S1)
  • Signal after known V of standard added to this
    unknown sample (S2)
  • Unknown concentration determined from

60
Internal Standard
  • Corrects for random instrumental and method
    fluctuations
  • Ex. Methods in which signal depends on the sample
    volume used such as GC
  • Methods in which T variations can affect signal
    such as Atomic Emission (AE) spectroscopy

61
Internal Standard
  • A constant known amount of a substance that is
    similar to, but not the same as, analyte is added
    to all standards and unknown samples
  • Signals for calibration curve generated from
    ratio of analyte signal to internal standard
    signal
  • Ratio of unknown analyte signal to internal
    standard signal is used to determine analyte
    concentration from calibration curve.

62
Internal Standard Example
  • GC-MS analysis of octane using nonane as internal
    standard
  • 5 octane standards from 1.00 to 10.00 mg/mL each
    containing 3.00 mg/mL nonane
  • Octane analyte containing 3.00 mg/mL nonane

63
Internal standard example data
64
Octane concentration 1.81 mg/mL
65
Octane concentration 2.04 mg/mL
66
Error analysis for these methods
  • How to find error in concentration of an unknown
    determined by external standard calibration
    curve?
  • Review least squares
  • y mx b
  • Simple least squares assumes a constant error in
    y value (measurement signal) and no error in x
    axis values (concentration)

67
  • Residual is vertical deviation of each data point
    from line
  • Line generated by least squares minimizes sum of
    the squares of residuals for all points

68
Least Squares Equations
  • To determine slope, intercept, and errors these
    three quantities are very useful

69
  • sr standard deviation about regression
    (standard error of the estimate)
  • N number of points in calibration curve

70
  • sc standard deviation in results
    (concentrations) from curve
  • M number of replicate analyses of an unknown
  • N number of points in calibration curve
  • mean of signals from analysis of
    unknown
  • mean value of y signals used in
    calibration curve

71
External Standard
  • Calibration curve prepared from known
    concentrations
  • Unknown solution measured

72
Concentration of unknown 1.71 ppm
73
Excel version
Block out 2x5 matrix of cells linest(ys,xs,true
,true) hold down control and shift buttons and
then hit enter
74
Error Analysis for standard addition
  • Review propagation of error for random errors

75
Error Analysis for standard addition
cx is not determined by reading signal from
straight line
unknown signals are actually part of calibration
curve
determine sx in cx by assuming no error in cs or
Vx and that all error comes from b and m
Use propagation of error equation from Table a1-6
to come up with
76
Excel version
77
Internal standard example data
78
Octane concentration 1.81 mg/mL
Error analysis
79
Octane concentration 2.04 mg/mL
error analysis
Write a Comment
User Comments (0)
About PowerShow.com