Title: Monitoring Respiratory Status
1Monitoring Respiratory Status
2Pulse Oximetry
- Measures oxygen saturation
- Documents peripheral oxygen availability
3Oximetry History
- Became standard of care in the 1980s
- 1935 Carl Matthes
- first oximeter
- 1930s J.R. Squires
- self calibrating oximeter
4Oximetry History (Contd)
- 1940s Glen Milliken
- aviation oximeter
- 1951
- First reported use of oximetry in the operating
room
5Has pulse oximetry improved the outcome of
patients receiving anesthesia?
6Improved Safety
- Research is not ethical
- Closed claim studies
- capnography
- pulse oximetry
- Insurance rates have dropped
7What information can be obtained from the pulse
oximeter?
8Physics Engineering
- Lambert-Beer law
- concentration of a liquid is related to the
amount of light that will pass through it - Oxygenated hemoglobin absorbs a different
wavelength of light than does deoxygenated blood
9Cyanosis
- Is cyanosis a reliable clinical sign?
- What is required for cyanosis to occur?
- At what saturation will the average patient
become cyanotic?
10Light Absorption
- Tissues
- Nonpulsatile vessels
- Pulsatile arterial blood
11Pulse Oximeter Wavelengths
- Red (660 nm)
- absorbed by unoxygenated hemoglobin
- Near infrared (940 nm)
- absorbed by oxygenated hemoglobin
12Cause of False Readings
- Accurate only with saturation 80
- most accurate 95
- shows trends at low saturation
- Abnormal hemoglobin
13False Readings (Contd)
- Intravenous dyes
- Diminished pulse
- Movement of finger
- Ambient light
- Nail polish
14Abnormal Hemoglobin
- Carboxyhemoglobin
- false high reading in carbon monoxide patients
- Methemaglobin
- reads 85 regardless of actual saturation
- Fetal hemoglobin
- little effect on pulse oximetry
15Dyes
- Methylene blue
- Indocyanine green
- Indigo carmine
- Acrylic finger nails
- nail polish
16Capnometry vsCapnography
17Monitoring Expired CO2
- Capnometry gives numerical value
- Capnography gives wave form
18Monitoring ETCO2
- Confirms the movement of air in and out of the
lungs - Assumed to reflect alveolar CO2
- Assumed to indicate adequacy of ventilation
19Monitoring ETCO2 (cont)
- Better indicator of ventilation
- Measures high point of the expiratory plateau
- Normally less than the PaCO2
- Normal gradient about 5-8
20Sampling Sites
- Main stream
- measuring device is in line
- measures only CO2
- patient must be intubated
21Sampling Systems
22Sampling Sites (Contd)
- Side stream
- takes sample to processing machine
- removes gas from circuit
23CO2 Increases with
- Hypoventilation
- Malignant hyperthermia
- Sepsis
- Rebreathing
- Bicarbonate administration
- Insufflation of CO2
24CO2 Decreases with
- Hyperventilation
- Hypothermia
- Low cardiac output
- pulmonary embolism
- Circuit disconnect
- Cardiac arrest
25Describe Wave Forms representing the following
- Normal wave form
- COPD
- Inadequate neuromuscular relaxation
- Unequal lung emptying
26Describe Wave Forms (Contd)
- Restrictive lung disease
- Esophageal intubation
- Malignant hyperthermia
- Cardiac arrest
- Pulmonary embolism
27Clinical Uses of Capnography
- Detection of untoward events
- Maintenance of normocarbia
- Weaning from mechanical ventilation
- Evaluating effectiveness of CPR
28Monitoring Anesthetic Gases
- What types of gases are present?
- What are their concentrations?
- partial pressure
- volume percent
29Anesthetic Gas Monitoring
- What types of gases are present?
- What are their concentrations?
- partial pressure
- volume percent
30Sampling Systems
31Mass Spectrometry
- Gas enters high vacuum area
- Bombarded by electron beam
- Charged particles passed over strong magnet
32Mass Spectrometry (Contd)
- Different components are deflected according to
their chemical composition - Specific collectors measure composition
33Mass Spectrometry Display
- Assumption that the sum of the gases 100
- Calibrated to ambient pressure minus 47 mm/Hg
- greater chance for error with mm/Hg
- percentages remain accurate
34Infrared Analyzers
- Measures energy absorbed from narrow band of
wavelengths of infrared light passing through a
gas sample
35Infrared Analyzers (Contd)
- Molecules that absorb energy
- carbon dioxide
- nitrous oxide
- water vapor
- volatile anesthetics
36Infrared Analyzers (Contd)
- Molecules that do not absorb energy
- oxygen
- argon
- nitrogen
- helium
- xenon
37Ramon Spectrometer
- When light strikes gas molecules, most of the
scattered energy is absorbed and re-emitted in
the same direction
38Ramon Spectrometer (Contd)
- Small amount of light is scattered and detected
by optical detection system - Both polar non-polar molecules are detected
39Ramon Spectrometer (Contd)
- Monoatomic gases lacking intermolecular bonds are
not detected - helium/xenon/argon
40Neurological Monitoring
41Common Parameters Monitored
- Cerebral blood flow
- EEG
- SSEP
- EMG
- Wake-up test
42What to look for?
- Blood flow
- Too much
- Too little
- Location of nerve
- Integrity of nerve
43Monitoring Cerebral Blood Flow
- Level of consciousness
- Blood pressure
- Intracranial pressure
- Jugular venous oxygen content
- Radioactive venous washout
- Transcranial doppler
44Essentials of Neurological Monitoring
- Must assess function of area at risk
- Operator must understand pathways assessed
- Monitoring should be continuous
- Minimal interference
- Strict quality control
45Preoperative Assessment
- Essential to understand baseline status
- Patient teaching
46Wake-up Test
- Test neurologic function following reversible
surgical manipulation - Movement must not cause damage
- Patient is allowed to awaken
- Amnesia must be maintained
47Wake-up Test
- After awakening, patient follows verbal commands
- Evaluates corticospinal tracts (thoracic)
- Response to painful stimuli
- Lumbar cord function
48Disadvantage of Wake-up
- Test is intermittent
- not assessed as distraction is applied
- gray matter damaged quickly
- white matter damaged slowly
- Surgery is interrupted
49Wake-up disadvantage
- Risk of patient injury
- Danger of recall
- In spite of disadvantages, wake-up test is still
used successfully
50Awake Clinical Observation
- Carotid endarterectomy
- Resection of seizure focus
- Resection of brain tumor
- Non-neurological surgery on patient with head or
neck trauma - Patient selection is critical
51Sedation / Analgesia
- May mask neurological changes
- Propofol infusion
- Rapid on and off
- Impairs continuous monitoring
- Risk loss of airway
52CBF Monitoring
- Brain has high metabolic rate
- Oxygen supply is critical
- Even brief interruption of CBF can cause
perminent brain damage
53Cerebral Blood Flow
- Autoregulation attempts to maintain constant CBF
- CBF constant with CPP in the range of 50-150
mm/hg - At extremes, CBF related to MAP
54Cerebral Perfusion Pressure
- CPP MAP - ICP or CVP (whichever is higher
- Unless ICP is measured, CPP can only be estimated
- If BP is constant, CPP falls as ICP increases
55Measuring ICP
- Ventricular catheter
- Subdural bolt
- Lumbar CSF catheter
- Scanning techniques
56Ventricular Catheter
- Small burr hole
- Catheter into lateral ventricle
- Connected to transducer
- note do not flush
- This is the most accurate technique
- Fluid can be removed to release pressure
57Subdural Bolt
- Small drill hole
- Dura open
- Hollow fluid filled bolt rests directly against
the brain - Less Accurate than ventricular cath.
- Can not remove fluid
58Lumbar CSF Catheter
- Place epidural type catheter
- Attach to transducer
- Correlates with ICP if zeroed at level of the
foramen of monroe - CSF blockage will alter accuracy
59Scanning Techniques
- CT or MRI may show edema or distended ventricle
- Does not provide quantative number
- Does not substitute for ICP monitor
60Measuring Cerebral Blood Flow
- Direct
- Radioactive scan
- Indirect
- Transcranial doppler
61CBF Scan
- Inject radioactive xenon into carotid artery
- Measure radioactivity and washout
- Not continuous
- Not used in the operating room
62Transcranial Doppler
- Direct, continuous, non-invasive
- Ultrasound waves to basal artery
- Assumed that blood velocity is related to CBF
63Uses of Transcranial Doppler
- Carotic endarterectomy
- Cardiopulmonary bypass
- Detection of vasospasm
- Confirmation of brain death
64EEG
- 10-20 electrodes placed on scalp
- Each channel is electrical activity between 1
pair of electrodes - Requires continuous monitoring by trained
technician - Even 1-2 channels may be helpful
65Types of EEG waves
66Alpha Waves
- Occipital area of the brain
- Patient is alert, relaxed with eyes open
- May be seen in light plane of anesthesia
67Beta Waves
- Mental concentration in the awake patient
- Seen with low doses of sedatives or hypnotic
drugs - Abolished with deep anesthesia or ischemia
68Delta Waves
- Deep sleep or deep anesthesia
- Ischemia
- Drug overdose
- Severe mental derangements
69Theta Waves
- Commonly seen during general anesthesia
- May be seen in same pathological states as delta
waves
70Using the EEG
- Equipment must be properly applied
- Set acceptable parameters
- Learn anticipated changes
- If possible, have EEG technician monitor the
patient
71Uses of the EEG
- Determine cerebral ischemia
- need for shunt placement
- hypotension
- barbiturate suppression of CMRO2
72Drug effects on the EEG
- Low dose of GA with nitrous oxide
- active EEG with alpha and beta
- Deep anesthesia with volatile agent
- Similar pattern to ischemia
- Very deep anesthesia
- flat EEG
73Drug Effects
- Drug bolus may resemble ischemia
- Steady state anesthesia pruduces stable EEG
74What are somatosensory Evoked Potentials?
75The Somesthetic Nervous System
- Extends throughout peripheral and central nervous
system - Peripheral nerves
- Spinal cord
- Subcortical structures
- Cortical structures
76Somesthetic Sensory System
- Carries sensory information
- Vibration
- Proprioception
- Light touch
77SSEPs
- Peripheral nerve stimulated
- Response recorded proximal to stimulation and at
cerebral cortex - Should monitor both sides of the body even though
surgery is only on one side - Each stimulus should reach brain
78SSEP Helpful For
- Spinal surgery
- Thoracic aneurysm
- Cerebral blood flow
- cerebral aneurysm
- carotid endarterectomy
79Anesthesia may alter SSEP
- Hypnotics and volatile agents increase latency
and decrease amplitude - Muscle relaxants have no effect
- Narcotics have minimal effects if given in
moderate doses
80Other Factors Affecting SSEP
- Hypothermia
- Cold irrigation
- Hypoxemia
- Sudden changes in PaCO2 or anesthetic agent
81Minimizing the effects of anesthesia on SSEPs
- Do not change anesthetic concentration during
critical times - Monitor area of brain not at risk
- Use cervical response rather than cortical
response when possible - Use favorable anesthetic technique
82Motor Evoked Potentials
- Similar to SSEP
- Motor cortex in brain is stimulated
- Descending motor pathways conduct the impulse
- Response recorded by EMG
83Clinical Application of MEP
- Not commonly used
- Problems
- pathways yet to be determined
- experience limited
- May not be better than SSEP
- profound effects by anesthetic agents
84Brainstem Auditory Evoked Potentials
85BAEP
- Monitors the 8th cranial nerve
- Auditory stimulus in the ear
- Scalp electrodes record response
86Clinical use of BAEP
- Monitors 8th cranial nerve function during
surgery of the posterior fossa - Used during acoustic neuroma surgery
87Advantage of BAEP
- Specific function of 8th nerve
- Less affected by hypothermia or anesthetic agents
than other EPs
88Visual Evoked Potentials
89Visual Evoked Potentials
- Not commonly used
- Assesses function of the optic nerve
- Goggles emit bright light through closed eyes.
- Stimulus detected by the brain
90Use of VEP
- Monitors optic nerve function in operations near
the optic nerve or the chiasm - pituitary surgery
- meningiomas that compress the optic nerve
91Facial Nerve Monitoring
92Facial Nerve Observation
- Simple method of assessment
- Surgeon observes for muscle twitch when nerve is
touched - Twitch may not be readily visible
- Muscle relaxants abolish the twitch
93Facial Nerve Monitoring
- Needle placement
- Orbicularis Oculi muscle
- Orbicularis Oris muscle
- EMG activity is recorded
- Muscles twitch if facial nerve is touched
94Advantage of Facial Nerve Monitoring
- Audible signal when nerve is touched
- More reliable than visual observation
- Note The patient must have a twitch for this
technique to work